Literature DB >> 25186395

Quantum noise properties of CT images with anatomical textured backgrounds across reconstruction algorithms: FBP and SAFIRE.

Justin Solomon1, Ehsan Samei2.   

Abstract

PURPOSE: Quantum noise properties of CT images are generally assessed using simple geometric phantoms with uniform backgrounds. Such phantoms may be inadequate when assessing nonlinear reconstruction or postprocessing algorithms. The purpose of this study was to design anatomically informed textured phantoms and use the phantoms to assess quantum noise properties across two clinically available reconstruction algorithms, filtered back projection (FBP) and sinogram affirmed iterative reconstruction (SAFIRE).
METHODS: Two phantoms were designed to represent lung and soft-tissue textures. The lung phantom included intricate vessel-like structures along with embedded nodules (spherical, lobulated, and spiculated). The soft tissue phantom was designed based on a three-dimensional clustered lumpy background with included low-contrast lesions (spherical and anthropomorphic). The phantoms were built using rapid prototyping (3D printing) technology and, along with a uniform phantom of similar size, were imaged on a Siemens SOMATOM Definition Flash CT scanner and reconstructed with FBP and SAFIRE. Fifty repeated acquisitions were acquired for each background type and noise was assessed by estimating pixel-value statistics, such as standard deviation (i.e., noise magnitude), autocorrelation, and noise power spectrum. Noise stationarity was also assessed by examining the spatial distribution of noise magnitude. The noise properties were compared across background types and between the two reconstruction algorithms.
RESULTS: In FBP and SAFIRE images, noise was globally nonstationary for all phantoms. In FBP images of all phantoms, and in SAFIRE images of the uniform phantom, noise appeared to be locally stationary (within a reasonably small region of interest). Noise was locally nonstationary in SAFIRE images of the textured phantoms with edge pixels showing higher noise magnitude compared to pixels in more homogenous regions. For pixels in uniform regions, noise magnitude was reduced by an average of 60% in SAFIRE images compared to FBP. However, for edge pixels, noise magnitude ranged from 20% higher to 40% lower in SAFIRE images compared to FBP. SAFIRE images of the lung phantom exhibited distinct regions with varying noise texture (i.e., noise autocorrelation/power spectra).
CONCLUSIONS: Quantum noise properties observed in uniform phantoms may not be representative of those in actual patients for nonlinear reconstruction algorithms. Anatomical texture should be considered when evaluating the performance of CT systems that use such nonlinear algorithms.

Entities:  

Mesh:

Year:  2014        PMID: 25186395     DOI: 10.1118/1.4893497

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  30 in total

1.  Anthropomorphic model observer performance in three-dimensional detection task for low-contrast computed tomography.

Authors:  Alexandre Ba; Miguel P Eckstein; Damien Racine; Julien G Ott; Francis Verdun; Sabine Kobbe-Schmidt; François O Bochud
Journal:  J Med Imaging (Bellingham)       Date:  2015-12-29

2.  Full Dose-Reduction Potential of Statistical Iterative Reconstruction for Head CT Protocols in a Predominantly Pediatric Population.

Authors:  A E Mirro; S L Brady; R A Kaufman
Journal:  AJNR Am J Neuroradiol       Date:  2016-04-07       Impact factor: 3.825

3.  DukeSim: A Realistic, Rapid, and Scanner-Specific Simulation Framework in Computed Tomography.

Authors:  Ehsan Abadi; Brian Harrawood; Shobhit Sharma; Anuj Kapadia; William P Segars; Ehsan Samei
Journal:  IEEE Trans Med Imaging       Date:  2018-12-12       Impact factor: 10.048

4.  Design and fabrication of heterogeneous lung nodule phantoms for assessing the accuracy and variability of measured texture radiomics features in CT.

Authors:  Ehsan Samei; Jocelyn Hoye; Yuese Zheng; Justin B Solomon; Daniele Marin
Journal:  J Med Imaging (Bellingham)       Date:  2019-06-21

5.  Correlation between human detection accuracy and observer model-based image quality metrics in computed tomography.

Authors:  Justin Solomon; Ehsan Samei
Journal:  J Med Imaging (Bellingham)       Date:  2016-09-22

6.  Paper-based 3D printing of anthropomorphic CT phantoms: Feasibility of two construction techniques.

Authors:  Paul Jahnke; Stephan Schwarz; Marco Ziegert; Felix Benjamin Schwarz; Bernd Hamm; Michael Scheel
Journal:  Eur Radiol       Date:  2018-08-16       Impact factor: 5.315

7.  Assessment of structural similarity in CT using filtered backprojection and iterative reconstruction: a phantom study with 3D printed lung vessels.

Authors:  Raoul M S Joemai; Jacob Geleijns
Journal:  Br J Radiol       Date:  2017-08-22       Impact factor: 3.039

8.  Characterization and validation of the thorax phantom Lungman for dose assessment in chest radiography optimization studies.

Authors:  Sunay Rodríguez Pérez; Nicholas William Marshall; Lara Struelens; Hilde Bosmans
Journal:  J Med Imaging (Bellingham)       Date:  2018-02-06

9.  Evaluation of low-contrast detectability for iterative reconstruction in pediatric abdominal computed tomography: a phantom study.

Authors:  Nicholas Rubert; Richard Southard; Susan M Hamman; Ryan Robison
Journal:  Pediatr Radiol       Date:  2019-11-09

10.  Statistical model based iterative reconstruction in clinical CT systems. Part III. Task-based kV/mAs optimization for radiation dose reduction.

Authors:  Ke Li; Daniel Gomez-Cardona; Jiang Hsieh; Meghan G Lubner; Perry J Pickhardt; Guang-Hong Chen
Journal:  Med Phys       Date:  2015-09       Impact factor: 4.071

View more