Comparison of low-contrast detectability between two CT reconstruction algorithms using voxel-based 3D printed textured phantoms.

Justin Solomon1, Alexandre Ba2, François Bochud2, Ehsan Samei3.   

Abstract

PURPOSE: To use novel voxel-based 3D printed textured phantoms in order to compare low-contrast detectability between two reconstruction algorithms, FBP (filtered-backprojection) and SAFIRE (sinogram affirmed iterative reconstruction) and determine what impact background texture (i.e., anatomical noise) has on estimating the dose reduction potential of SAFIRE.
METHODS: Liver volumes were segmented from 23 abdominal CT cases. The volumes were characterized in terms of texture features from gray-level co-occurrence and run-length matrices. Using a 3D clustered lumpy background (CLB) model, a fitting technique based on a genetic optimization algorithm was used to find CLB textures that were reflective of the liver textures, accounting for CT system factors of spatial blurring and noise. With the modeled background texture as a guide, four cylindrical phantoms (Textures A-C and uniform, 165 mm in diameter, and 30 mm height) were designed, each containing 20 low-contrast spherical signals (6 mm diameter at nominal contrast levels of ∼3.2, 5.2, 7.2, 10, and 14 HU with four repeats per signal). The phantoms were voxelized and input into a commercial multimaterial 3D printer (Object Connex 350), with custom software for voxel-based printing (using principles of digital dithering). Images of the textured phantoms and a corresponding uniform phantom were acquired at six radiation dose levels (SOMATOM Flash, Siemens Healthcare) and observer model detection performance (detectability index of a multislice channelized Hotelling observer) was estimated for each condition (5 contrasts × 6 doses × 2 reconstructions × 4 backgrounds = 240 total conditions). A multivariate generalized regression analysis was performed (linear terms, no interactions, random error term, log link function) to assess whether dose, reconstruction algorithm, signal contrast, and background type have statistically significant effects on detectability. Also, fitted curves of detectability (averaged across contrast levels) as a function of dose were constructed for each reconstruction algorithm and background texture. FBP and SAFIRE were compared for each background type to determine the improvement in detectability at a given dose, and the reduced dose at which SAFIRE had equivalent performance compared to FBP at 100% dose.
RESULTS: Detectability increased with increasing radiation dose (P = 2.7 × 10-59) and contrast level (P = 2.2 × 10-86) and was higher in the uniform phantom compared to the textured phantoms (P = 6.9 × 10-51). Overall, SAFIRE had higher d' compared to FBP (P = 0.02). The estimated dose reduction potential of SAFIRE was found to be 8%, 10%, 27%, and 8% for Texture-A, Texture-B, Texture-C and uniform phantoms.
CONCLUSIONS: In all background types, detectability was higher with SAFIRE compared to FBP. However, the relative improvement observed from SAFIRE was highly dependent on the complexity of the background texture. Iterative algorithms such as SAFIRE should be assessed in the most realistic context possible.

Mesh:

Year:  2016        PMID: 27908164     DOI: 10.1118/1.4967478

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  22 in total

1.  Channelized Hotelling observer correlation with human observers for low-contrast detection in liver CT images.

Authors:  Alexandre Ba; Craig K Abbey; Damien Racine; Anaïs Viry; Francis R Verdun; Sabine Schmidt; François O Bochud
Journal:  J Med Imaging (Bellingham)       Date:  2019-05-20

2.  Design and fabrication of heterogeneous lung nodule phantoms for assessing the accuracy and variability of measured texture radiomics features in CT.

Authors:  Ehsan Samei; Jocelyn Hoye; Yuese Zheng; Justin B Solomon; Daniele Marin
Journal:  J Med Imaging (Bellingham)       Date:  2019-06-21

3.  Systematic analysis of bias and variability of texture measurements in computed tomography.

Authors:  Marthony Robins; Justin Solomon; Jocelyn Hoye; Ehsan Abadi; Daniele Marin; Ehsan Samei
Journal:  J Med Imaging (Bellingham)       Date:  2019-07-12

4.  Paper-based 3D printing of anthropomorphic CT phantoms: Feasibility of two construction techniques.

Authors:  Paul Jahnke; Stephan Schwarz; Marco Ziegert; Felix Benjamin Schwarz; Bernd Hamm; Michael Scheel
Journal:  Eur Radiol       Date:  2018-08-16       Impact factor: 5.315

5.  Use of a channelized Hotelling observer to assess CT image quality and optimize dose reduction for iteratively reconstructed images.

Authors:  Christopher P Favazza; Andrea Ferrero; Lifeng Yu; Shuai Leng; Kyle L McMillan; Cynthia H McCollough
Journal:  J Med Imaging (Bellingham)       Date:  2017-10-03

6.  Accounting for reconstruction kernel-induced variability in CT radiomic features using noise power spectra.

Authors:  Muhammad Shafiq-Ul-Hassan; Geoffrey G Zhang; Dylan C Hunt; Kujtim Latifi; Ghanim Ullah; Robert J Gillies; Eduardo G Moros
Journal:  J Med Imaging (Bellingham)       Date:  2017-12-14

7.  Localization of liver lesions in abdominal CT imaging: II. Mathematical model observer performance correlates with human observer performance for localization of liver lesions in abdominal CT imaging.

Authors:  Samantha K N Dilger; Shuai Leng; Baiyu Chen; Rickey E Carter; Chris P Favazza; Joel G Fletcher; Cynthia H McCollough; Lifeng Yu
Journal:  Phys Med Biol       Date:  2019-05-10       Impact factor: 3.609

8.  Three-dimensionally-printed anthropomorphic physical phantom for mammography and digital breast tomosynthesis with custom materials, lesions, and uniform quality control region.

Authors:  Andrea H Rossman; Matthew Catenacci; Christine Zhao; Dhiraj Sikaria; John E Knudsen; Danielle Dawes; Michael E Gehm; Ehsan Samei; Benjamin J Wiley; Joseph Y Lo
Journal:  J Med Imaging (Bellingham)       Date:  2019-03-22

9.  Harmonization of in-plane resolution in CT using multiple reconstructions from single acquisitions.

Authors:  Gonzalo Vegas-Sánchez-Ferrero; Gabriel Ramos-Llordén; Raúl San José Estépar
Journal:  Med Phys       Date:  2021-09-14       Impact factor: 4.071

10.  Virtual monoenergetic images from dual-energy CT: systematic assessment of task-based image quality performance.

Authors:  Davide Cester; Matthias Eberhard; Hatem Alkadhi; André Euler
Journal:  Quant Imaging Med Surg       Date:  2022-01
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.