Damien Racine1, Nick Ryckx1, Alexandre Ba1, Fabio Becce2, Anais Viry1, Francis R Verdun1, Sabine Schmidt3.
Abstract
OBJECTIVE: We investigated the variability in diagnostic information inherent in computed tomography (CT) images acquired at 68 different CT units, with the selected acquisition protocols aiming to answer the same clinical question.Entities:
Keywords: Abdominal computed tomography; Image quality; Model observer; Standardisation; Task-based assessment
Mesh:
Year: 2018 PMID: 29858638 PMCID: PMC6223860 DOI: 10.1007/s00330-018-5518-8
Source DB: PubMed Journal: Eur Radiol ISSN: 0938-7994 Impact factor: 5.315
Fig. 1CT images of the anthropomorphic abdomen phantom. From left to right: small, medium and large phantom
The 68 CT scanners involved in this study
| CT scanner | Number | |
|---|---|---|
| GE | BrightSpeed S | 1 |
| Discovery CT 750 HD | 2 | |
| LightSpeed VCT | 7 | |
| LightSpeed16 | 1 | |
| Optima CT520 | 2 | |
| Optima CT580 | 1 | |
| Optima CT660 | 7 | |
| Revolution | 1 | |
| Revolution EVO | 2 | |
| Philips | Brilliance 40 | 1 |
| Brilliance 64 | 6 | |
| iCT 256 | 4 | |
| Ingenuity Core 128 | 4 | |
| Ingenuity CT | 4 | |
| Ingenuity Flex | 2 | |
| IQon - Spectral CT | 1 | |
| Siemens | Perspective | 1 |
| Sensation 64 | 2 | |
| SOMATOM Definition AS | 2 | |
| SOMATOM Force | 1 | |
| Canon Medical System | Activion16 | 4 |
| Aquilion | 6 | |
| Aquilion PRIME | 6 |
Fig. 2CTDIvol obtained for the three phantom sizes as a function of the abdominal clinical CT protocol settings. The black line in the middle of the coloured rectangles represents the median. The bottom edge of the rectangle corresponds to the first quartile and the top edge to the third quartile. The bottom line represents the fifth percentile and the top line the 95th percentile. The red dots outside these two lines are outliers
Fig. 3AUC as a function of CTDIvol for the small-sized phantom
Fig. 4AUC as a function of CTDIvol for the medium-sized phantom
Fig. 5AUC as a function of CTDIvol for the large-sized phantom
Correlation matrix based on the area under the ROC curve (AUC) obtained with automatic tube current modulation (ATCM) that maintained absolute noise levels close to target values
| Size | S | M | L |
|---|---|---|---|
| S | 1 | ||
| M | 0.39 | 1 | |
| L | 0.33 | 0.49 | 1 |
Correlation matrix based on the area under the ROC curve (AUC) obtained with automatic tube current modulation (ATCM) that maintained a constant level of overall diagnostic quality for all patient sizes relating to a reference image
| Size | S | M | L |
|---|---|---|---|
| S | 1 | ||
| M | 0.49 | 1 | |
| L | 0.40 | 0.58 | 1 |