| Literature DB >> 34471413 |
Jing Dong1, Yaqin Zhang2, Xiaogang Wang2.
Abstract
AIM: To investigate the interdevice agreement for differences in toric power calculated using data on anterior corneal astigmatism obtained with corneal topography/ray-tracing aberrometry (iTrace), partial coherence interferometry (IOLMaster 500), and Scheimpflug imaging (Pentacam).Entities:
Year: 2021 PMID: 34471413 PMCID: PMC8405298 DOI: 10.1155/2021/7712345
Source DB: PubMed Journal: J Trop Med ISSN: 1687-9686
Patient demographics.
| Characteristic | Patients ( |
|---|---|
| Eyes (% right eyes) | 55 (54.5%) |
| Age, y (mean ± SD) | 66 ± 10 |
| Sex (% male) | 56 (55.4%) |
| Axial length, mm (mean ± SD) | 23.37 ± 0.91 |
| Anterior chamber depth, mm (mean ± SD) | 3.10 ± 0.37 |
| With-the-rule astigmatism (% eyes) | 29 (28.7%) |
| Against-the-rule astigmatism (% eyes) | 65 (64.4%) |
| Oblique astigmatism (% eyes) | 7 (6.9%) |
SD: standard deviation; corneal astigmatism measured with IOLMaster 500.
Mean anterior corneal keratometry values obtained using iTrace, IOLMaster, and Pentacam.
| iTrace at 3 mm | IOLMaster at 2.3 mm | Pentacam at 3 mm | |
|---|---|---|---|
| ( | ( | (Centered on the corneal apex) ( | |
| Astigmatism magnitude (D) | 1.86 ± 0.77 | 1.95 ± 0.87 | 1.83 ± 0.79 |
| Range: 0.61–4.23 | Range: 0.34–4.50 | Range: 0.40–3.80 | |
| Astigmatism axis (degree) | 97 ± 68 | 99 ± 64 | 94 ± 47 |
| Range: 0–179 | Range: 1–179 | Range: 0–179 | |
| Flat keratometry (D) | 43.69 ± 1.58 | 44.04 ± 1.61 | 44.01 ± 1.58 |
| Range: 39.58–47.74 | Range: 39.57–47.74 | Range: 39.40–47.60 | |
| Steep keratometry (D) | 45.55 ± 1.62 | 45.99 ± 1.57 | 45.84 ± 1.60 |
| Range: 41.02–48.77 | Range: 42.13–49.41 | Range: 41.10–49.10 | |
| Mean keratometry (D) | 44.62 ± 1.55 | 45.02 ± 1.53 | 44.92 ± 1.54 |
| Range: 40.30–48.26 | Range: 40.85–48.51 | Range: 40.25–48.25 |
D: diopter.
Mean difference between anterior corneal keratometry values for each pair of devices.
| iTrace-IOLMaster | iTrace-Pentacam | IOLMaster-Pentacam | |
|---|---|---|---|
| Astigmatism magnitude (D) | −0.10 ± 0.33 | 0.03 ± 0.27 | 0.12 ± 0.38 |
| Flat keratometry (D) | −0.35 ± 0.27 | −0.31 ± 0.33 | 0.04 ± 0.36 |
| Steep keratometry (D) | −0.45 ± 0.26 | −0.29 ± 0.30 | 0.16 ± 0.24 |
| Mean keratometry (D) | −0.40 ± 0.20 | −0.30 ± 0.29 | 0.10 ± 0.24 |
D: diopter.
Figure 1Double-angle plots of differences in anterior corneal astigmatism measured by iTrace (a), IOLMaster 500 (b), and Pentacam (c). Note: centroid (solid square); 95% confidence of the ellipse of the centroid (red color); 95% confidence ellipse of the dataset (blue color); each ring = 1.0 D.
Intraocular lens power and toricity calculated using the online Barrett toric calculator for each device.
| iTrace | IOLMaster | Pentacam | |
|---|---|---|---|
| IOL power (D) | 20.8 ± 2.8 | 20.3 ± 2.9 | 20.4 ± 2.8 |
| Range: 12.5–27.0 | Range: 12.5–27.0 | Range: 12.5–27.0 | |
| Toricity (T) | 5 ± 2 | 5 ± 2 | 5 ± 2 |
| Range: 2–9 | Range: 2–9 | Range: 2–9 | |
| Axis (degree) | 62 ± 66 | 59 ± 64 | 64 ± 65 |
| Range: 1–180 | Range: 0–179 | Range: 1–179 |
D: diopter.
Figure 2Participants whose intraocular lens power difference changed from −1.0 to 1.5 D as calculated using data obtained by iTrace, IOLMaster, and Pentacam.
Figure 3Participants whose toricity difference changed by ≤2 as calculated using astigmatism data obtained by iTrace, IOLMaster, and Pentacam.