Ahmed Hadj Henni1, David Gensanne2, Maximilien Roge2, Chantal Hanzen2, Guillaume Bulot2, Elyse Colard2, Sebastien Thureau2.
Abstract
BACKGROUND: The objective of this study was to analyze the amplitude of translational and rotational movements occurring during stereotactic body radiotherapy (SBRT) of spinal metastases in two different positioning devices. The relevance of intra-fractional imaging and the influence of treatment time were evaluated.Entities:
Keywords: CBCT; Inter-fraction; Intra-fraction; SBRT; Spinal; Treatment time
Mesh:
Year: 2021 PMID: 34461953 PMCID: PMC8404277 DOI: 10.1186/s13014-021-01892-5
Source DB: PubMed Journal: Radiat Oncol ISSN: 1748-717X Impact factor: 3.481
Fig. 1Patient positioning systems: a vacuum cushion, arms raised and resting on a clegecel and b AIO Orfit™ system and vacuum cushion
Patients and treatment characteristics
| Patients (n = 20) | |
| Sex | |
| Men | 35% (n = 7) |
| Women | 65% (n = 13) |
| Median age (years), (range) | 66 (48–76) |
| Primary cancer | |
| Breast | 50% (n = 10) |
| Prostate | 25% (n = 5) |
| Lung | 20% (n = 4) |
| Small bowel | 5% (n = 1) |
| Histological types | |
| Adenocarcinoma | 55% (n = 11) |
| Infiltrating ductal carcinoma | 35% (n = 7) |
| Infiltrating lobular carcinoma | 5% (n = 1) |
| Neuro-endocrine tumor | 5% (n = 1) |
| Vertebral lesions | 22 |
| Vertebrae treated in post-operative situation | 1 |
| Location of spinal lesions | |
| Thoracic spine | 55% (n = 12) |
| Lumbar spine | 41% (n = 9) |
| Sacrum | 4% (n = 1) |
| Treatment | |
| Mean GTV (cm3), (range) | 9.96 (0.40–42.30) |
| Mean CTV (cm3), (range) | 32.35 (8.74–99.50) |
| Prescribed dose | |
| 5 × 7 Gy | 100% (n = 20) |
Summary of translational and rotational deviation of each positioning system for each type of CBCT
| CBCT | Abosolute translational shifts (mm) | Absolute rotational shifts (°) | ||||
|---|---|---|---|---|---|---|
| X (Lat.) | Y (Vert.) | Z (Long.) | Rx (Pitch) | Ry (Yaw) | Rz (Roll) | |
| Pos_Old (179) | ||||||
| CBCTpTT (60) | 4.0 (3.7, 15.9) | 2.4 (2.2, 9.5) | 5.0 (5.5, 26.4) | 0.8 (0.6, 2.2) | 1.0 (0.7, 2.7) | 0.9 (0.7, 2.7) |
| CBCT12 + 23 (119) | 0.9 (0.9,5.2) | 0.6 (0.7, 4.1) | 0.9 (1.5, 9.6) | 0.2 (0.3, 2.5) | 0.2 (0.3, 1.8) | 0.2 (0.3, 2.2) |
| Pos_New (150) | ||||||
| CBCTpTT (50) | 3.1 (3.7, 22.0) | 1.6 (1.3, 6.2) | 3.6 (3.1, 10.9) | 0.7 (0.6, 2.3) | 0.9 (0.7, 2.0) | 0.9 (0.7, 2.4) |
| CBCT12 + 23 (100) | 0.5 (0.5, 2.7) | 0.6 (0.7, 4.7) | 0.5 (0.5, 2.9) | 0.2 (0.3, 1.3) | 0.2 (0.3, 1.9) | 0.2 (0.3, 1.9) |
3D vector (mm) and max angle (°) calculated for each positioning system for each CBCT type
| CBCT | 3D vect (mm) | Max angle (°) |
|---|---|---|
| 3D vect = | Max angle = Max(RX; Ry; Rz) | |
| Pos_Old (179) | ||
| CBCTpTT (60) | 7.8 (5.9, 28.1) | 1.5 (0.6, 2.7) |
| CBCT12 + 23 (119) | 1.7 (1.7, 9.7) | 0.4 (0.4, 2.5) |
| Pos_New (150) | ||
| CBCTpTT (50) | 5.9 (3.8, 2.2) | 1.4 (0.5, 2.4) |
| CBCT12 + 23 (100) | 1.1 (0.8, 5.7) | 0.4 (0.3, 1.9) |
Fig. 2Proportion (%) of treatment fractions setup errors within given tolerance, translational (3D Vect on top) and rotational (MaxAngle at the bottom), measured on pre-treatment (CBCTpTT) and intra fractions (CBCT12 + 23) CBCTs. The thresholds denoted be a red line represent a 3D vector = 3.5 mm corresponding to a setup errors of 2 mm in the 3 directions and a maximum rotation angle of 2°
Fig. 3Relationship between intra-fraction 3D Vector (CBCT12 + 23) motion (mm) for Pos_Old (solid blue dots) and Pos_New (empty red dots) as a function of time (min) measured either between CBCTpTT start and CBCT12 start or between CBCT12 start and CBCT23 start. The linear fits are shown with their corresponding equations
Comparison of standard deviations (SD) in mm of intra-fraction positioning errors found in the literature as well as the Pos_Old and Pos_New positioning systems
| Authors | Positioning systems | SD (mm) for intra-fraction CBCT |
|---|---|---|
| Hyde et al | Elekta BodyFIX system | LR = 0.6; CC = 0.5 and AP = 0.5 |
| Our center | Pos_New | LR = 0.5; CC = 0.5 and AP = 0.7 |
| Dahele et al | Posirest | LR = 0.9; CC = 0.6 and AP = 0.7 |
| Li et al | Elekta BodyFIX system | LR = 0.9, CC = 0.7 and AP = 0.9 |
| Li et al | Evacuated cushion | LR = 1.3; CC = 1.2 and AP = 1.0 |
| Li et al | Thermoplastic S-frame | LR = 1.3, CC = 0.9, and AP = 1.1 |
| Foster et al | Evacuated cushion | LR = 1.05; CC = 1.23 and AP = 1.04 |
| Our center | Pos_Old | LR = 0,9; CC = 1,5 and AP = 0,7 |