Subsecond and Submillimeter Resolution Positional Verification for Stereotactic Irradiation of Spinal Lesions.

Colien Hazelaar1, Max Dahele2, Hassan Mostafavi3, Lineke van der Weide2, Ben J Slotman2, Wilko F A R Verbakel2.   

Abstract

PURPOSE: Spine stereotactic body radiation therapy (SBRT) requires highly accurate positioning. We report our experience with markerless template matching and triangulation of kilovoltage images routinely acquired during spine SBRT, to determine spine position. METHODS AND MATERIALS: Kilovoltage images, continuously acquired at 7, 11 or 15 frames/s during volumetric modulated spine SBRT of 18 patients, consisting of 93 fluoroscopy datasets (1 dataset/arc), were analyzed off-line. Four patients were immobilized in a head/neck mask, 14 had no immobilization. Two-dimensional (2D) templates were created for each gantry angle from planning computed tomography data and registered to prefiltered kilovoltage images to determine 2D shifts between actual and planned spine position. Registrations were considered valid if the normalized cross correlation score was ≥0.15. Multiple registrations were triangulated to determine 3D position. For each spine position dataset, average positional offset and standard deviation were calculated. To verify the accuracy and precision of the technique, mean positional offset and standard deviation for twenty stationary phantom datasets with different baseline shifts were measured.
RESULTS: For the phantom, average standard deviations were 0.18 mm for left-right (LR), 0.17 mm for superior-inferior (SI), and 0.23 mm for the anterior-posterior (AP) direction. Maximum difference in average detected and applied shift was 0.09 mm. For the 93 clinical datasets, the percentage of valid matched frames was, on average, 90.7% (range: 49.9-96.1%) per dataset. Average standard deviations for all datasets were 0.28, 0.19, and 0.28 mm for LR, SI, and AP, respectively. Spine position offsets were, on average, -0.05 (range: -1.58 to 2.18), -0.04 (range: -3.56 to 0.82), and -0.03 mm (range: -1.16 to 1.51), respectively. Average positional deviation was <1 mm in all directions in 92% of the arcs.
CONCLUSIONS: Template matching and triangulation using kilovoltage images acquired during irradiation allows spine position detection with submillimeter accuracy at subsecond intervals. Although the majority of patients were not immobilized, most vertebrae were stable at the sub-mm level during spine SBRT delivery.
Copyright © 2016 Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2016        PMID: 27026317     DOI: 10.1016/j.ijrobp.2016.01.006

Source DB:  PubMed          Journal:  Int J Radiat Oncol Biol Phys        ISSN: 0360-3016            Impact factor:   7.038


  11 in total

1.  Treatment planning for spinal radiosurgery : A competitive multiplatform benchmark challenge.

Authors:  Christos Moustakis; Mark K H Chan; Jinkoo Kim; Joakim Nilsson; Alanah Bergman; Tewfik J Bichay; Isabel Palazon Cano; Savino Cilla; Francesco Deodato; Raffaela Doro; Jürgen Dunst; Hans Theodor Eich; Pierre Fau; Ming Fong; Uwe Haverkamp; Simon Heinze; Guido Hildebrandt; Detlef Imhoff; Erik de Klerck; Janett Köhn; Ulrike Lambrecht; Britta Loutfi-Krauss; Fatemeh Ebrahimi; Laura Masi; Alan H Mayville; Ante Mestrovic; Maaike Milder; Alessio G Morganti; Dirk Rades; Ulla Ramm; Claus Rödel; Frank-Andre Siebert; Wilhelm den Toom; Lei Wang; Stefan Wurster; Achim Schweikard; Scott G Soltys; Samuel Ryu; Oliver Blanck
Journal:  Strahlenther Onkol       Date:  2018-05-25       Impact factor: 3.621

Review 2.  Review of Real-Time 3-Dimensional Image Guided Radiation Therapy on Standard-Equipped Cancer Radiation Therapy Systems: Are We at the Tipping Point for the Era of Real-Time Radiation Therapy?

Authors:  Paul J Keall; Doan Trang Nguyen; Ricky O'Brien; Pengpeng Zhang; Laura Happersett; Jenny Bertholet; Per R Poulsen
Journal:  Int J Radiat Oncol Biol Phys       Date:  2018-04-14       Impact factor: 7.038

3.  Less Time Is Less Motion: Analysis of Practical Efficiencies Gained With a Modified Workflow Integrating Planar kV Midimaging With CBCT for Spine Stereotactic Body Radiation Therapy.

Authors:  David Y Hu; Yiwen Xu; Yu-Hui Chen; Marjan Khosravi; Yulia Lyatskaya; Jeremy S Bredfeldt; Fred L Hacker; Tracy A Balboni; Alexander Spektor; Daniel Cagney; Raymond Mak; Mai Anh Huynh
Journal:  Adv Radiat Oncol       Date:  2022-04-08

4.  Decompose kV projection using neural network for improved motion tracking in paraspinal SBRT.

Authors:  Xiuxiu He; Weixing Cai; Feifei Li; Qiyong Fan; Pengpeng Zhang; John J Cuaron; Laura I Cerviño; Xiang Li; Tianfang Li
Journal:  Med Phys       Date:  2021-10-28       Impact factor: 4.506

5.  Technical Note: Synthetic treatment beam imaging for motion monitoring during spine SBRT treatments - a phantom study.

Authors:  Tianfang Li; Feifei Li; Weixing Cai; Pengpeng Zhang; Xiang Li
Journal:  Med Phys       Date:  2020-12-07       Impact factor: 4.071

6.  Residual intra-fraction error in robotic spinal stereotactic body radiotherapy without immobilization devices.

Authors:  Eleonora Rossi; Claudio Fiorino; Andrei Fodor; Chiara Deantoni; Paola Mangili; Nadia Gisella Di Muzio; Antonella Del Vecchio; Sara Broggi
Journal:  Phys Imaging Radiat Oncol       Date:  2020-10-03

7.  Evaluation of inter- and intra-fraction 6D motion for stereotactic body radiation therapy of spinal metastases: influence of treatment time.

Authors:  Ahmed Hadj Henni; David Gensanne; Maximilien Roge; Chantal Hanzen; Guillaume Bulot; Elyse Colard; Sebastien Thureau
Journal:  Radiat Oncol       Date:  2021-08-30       Impact factor: 3.481

8.  Feasibility, Method and Early Outcome of Image-Guided Volumetric Modulated Arc Radiosurgery Followed by Resection for AJCC Stage IIA-IIIB High-Risk Large Intraocular Melanoma.

Authors:  Maja Guberina; Ekaterina Sokolenko; Nika Guberina; Sami Dalbah; Christoph Pöttgen; Wolfgang Lübcke; Frank Indenkämpen; Manfred Lachmuth; Dirk Flühs; Ying Chen; Christian Hoffmann; Cornelius Deuschl; Leyla Jabbarli; Miltiadis Fiorentzis; Andreas Foerster; Philipp Rating; Melanie Ebenau; Tobias Grunewald; Nikolaos Bechrakis; Martin Stuschke
Journal:  Cancers (Basel)       Date:  2022-09-28       Impact factor: 6.575

9.  Markerless Real-Time 3-Dimensional kV Tracking of Lung Tumors During Free Breathing Stereotactic Radiation Therapy.

Authors:  Kimmie de Bruin; Max Dahele; Hassan Mostafavi; Berend J Slotman; Wilko F A R Verbakel
Journal:  Adv Radiat Oncol       Date:  2021-04-20

10.  Kilovoltage projection streaming-based tracking application (KiPSTA): First clinical implementation during spine stereotactic radiation surgery.

Authors:  Jihun Kim; Yang-Kyun Park; David Edmunds; Kevin Oh; Gregory C Sharp; Brian Winey
Journal:  Adv Radiat Oncol       Date:  2018-06-07
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.