Literature DB >> 19647951

Nonrandom intrafraction target motions and general strategy for correction of spine stereotactic body radiotherapy.

Lijun Ma1, Arjun Sahgal, Sabbir Hossain, Cynthia Chuang, Martina Descovich, Kim Huang, Alex Gottschalk, David A Larson.   

Abstract

PURPOSE: To characterize nonrandom intrafraction target motions for spine stereotactic body radiotherapy and to develop a method of correction via image guidance. The dependence of target motions, as well as the effectiveness of the correction strategy for lesions of different locations within the spine, was analyzed. METHODS AND MATERIALS: Intrafraction target motions for 64 targets in 64 patients treated with a total of 233 fractions were analyzed. Based on the target location, the cases were divided into three groups, i.e., cervical (n = 20 patients), thoracic (n = 20 patients), or lumbar-sacrum (n = 24 patients) lesions. For each case, time-lag autocorrelation analysis was performed for each degree of freedom of motion that included both translations (x, y, and z shifts) and rotations (roll, yaw, and pitch). A general correction strategy based on periodic interventions was derived to determine the time interval required between two adjacent interventions, to overcome the patient-specific target motions.
RESULTS: Nonrandom target motions were detected for 100% of cases regardless of target locations. Cervical spine targets were found to possess the highest incidence of nonrandom target motion compared with thoracic and lumbar-sacral lesions (p < 0.001). The average time needed to maintain the target motion to within 1 mm of translation or 1 degrees of rotational deviation was 5.5 min, 5.9 min, and 7.1 min for cervical, thoracic, and lumbar-sacrum locations, respectively (at 95% confidence level).
CONCLUSIONS: A high incidence of nonrandom intrafraction target motions was found for spine stereotactic body radiotherapy treatments. Periodic interventions at approximately every 5 minutes or less were needed to overcome such motions.

Entities:  

Mesh:

Year:  2009        PMID: 19647951     DOI: 10.1016/j.ijrobp.2009.04.027

Source DB:  PubMed          Journal:  Int J Radiat Oncol Biol Phys        ISSN: 0360-3016            Impact factor:   7.038


  14 in total

1.  Safety and utility of kyphoplasty prior to spine stereotactic radiosurgery for metastatic tumors: a clinical and dosimetric analysis.

Authors:  Ori Barzilai; Natalie DiStefano; Eric Lis; Yoshiya Yamada; D Michael Lovelock; Andrew N Fontanella; Mark H Bilsky; Ilya Laufer
Journal:  J Neurosurg Spine       Date:  2017-11-03

2.  Is helical tomotherapy accurate and safe enough for spine stereotactic body radiotherapy?

Authors:  Yoonsun Chung; Hong In Yoon; Joo Ho Kim; Ki Chang Nam; Woong Sub Koom
Journal:  J Cancer Res Clin Oncol       Date:  2012-10-02       Impact factor: 4.553

3.  Optimizing the flattening filter free beam selection in RapidArc®-based stereotactic body radiotherapy for Stage I lung cancer.

Authors:  J-Y Lu; Z Lin; P-X Lin; B-T Huang
Journal:  Br J Radiol       Date:  2015-07-02       Impact factor: 3.039

Review 4.  Radiation myelopathy following stereotactic body radiation therapy for spine metastases.

Authors:  Wee Loon Ong; Shun Wong; Hany Soliman; Sten Myrehaug; Chia-Lin Tseng; Jay Detsky; Zain Husain; Pejman Maralani; Lijun Ma; Simon S Lo; Arjun Sahgal
Journal:  J Neurooncol       Date:  2022-06-23       Impact factor: 4.506

Review 5.  Stereotactic body radiotherapy: a new paradigm in the management of spinal metastases.

Authors:  Zain A Husain; Isabelle Thibault; Daniel Letourneau; Lijun Ma; Harald Keller; John Suh; Veronica Chiang; Eric L Chang; Raja K Rampersaud; James Perry; David A Larson; Arjun Sahgal
Journal:  CNS Oncol       Date:  2013-05

6.  Prospective evaluation of target and spinal cord motion and dosimetric changes with respiration in spinal stereotactic body radiation therapy utilizing 4-D CT.

Authors:  Xin Wang; Amol J Ghia; Zhongxiang Zhao; Jinzhong Yang; Dershan Luo; Tina M Briere; Ramiro Pino; Jing Li; Mary F McAleer; David C Weksberg; Eric L Chang; Paul D Brown; James N Yang
Journal:  J Radiosurg SBRT       Date:  2016

7.  Correlation between small-volume spinal cord doses for spine stereotactic body radiotherapy (SBRT).

Authors:  Lijun Ma; Lei Wang; Young Lee; Chia-Lin Tseng; Scott Soltys; Steve Braunstein; Arjun Sahgal
Journal:  J Radiosurg SBRT       Date:  2018

8.  Comparison of Two RapidArc Delivery Strategies in Stereotactic Body Radiotherapy of Peripheral Lung Cancer with Flattening Filter Free Beams.

Authors:  Bao-Tian Huang; Jia-Yang Lu; Pei-Xian Lin; Jian-Zhou Chen; Yu Kuang; Chuang-Zhen Chen
Journal:  PLoS One       Date:  2015-07-01       Impact factor: 3.240

9.  Study of Spinal Cord Substructure Expansion Margin in Esophageal Cancer.

Authors:  Dingjie Li; Shengtao Wei; Tian Li; Yang Liu; Jing Cai; Hong Ge
Journal:  Technol Cancer Res Treat       Date:  2021 Jan-Dec

10.  Monitor unit optimization in stereotactic body radiotherapy for small peripheral non-small cell lung cancer patients.

Authors:  Bao-Tian Huang; Zhu Lin; Pei-Xian Lin; Jia-Yang Lu; Chuang-Zhen Chen
Journal:  Sci Rep       Date:  2015-12-18       Impact factor: 4.379

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.