W M vanNiekerk1, T Lazeroms1, S J Rogers1, N Lomax1, R A Hälg1, V Gajdos2, I Özden3, E Kessler1, O Riesterer1.
Abstract
Background and purpose: This study evaluated translational and rotational intra-fractional patient movement during spinal stereotactic body radiotherapy (SBRT) using 6D positioning based on 3D cone beam computerized tomography (CBCT) and stereoscopic kilovoltage imaging (ExacTrac). The aim was to determine whether additional intra-fractional image verification reduced intra-fractional motion without significantly prolonging treatment time, whilst maintaining acceptable imaging related dose. Materials and methods: A retrospective analysis of 38 patients with 41 primary tumour volumes treated with SBRT between September 2018 and May 2021 was performed. Three different image-guided radiotherapy (IGRT) workflows were assessed. The translational and rotational positioning errors for the different imaging workflows, 3D translational vectors and estimates of imaging dose delivered for the different imaging workflows were evaluated.Entities:
Keywords: 3D cone beam computerized tomography (CBCT); Dose; Intra-fractional motion; Radiotherapy; Six degrees of freedom (6 DoF); Spinal metastasis; Stereotactic body radiosurgery (SBRS); Stereotactic body radiotherapy (SBRT); Treatment time; image guided radiotherapy (IGRT); stereoscopic kilovoltage imaging (ExacTrac)
Year: 2022 PMID: 36203742 PMCID: PMC9530958 DOI: 10.1016/j.tipsro.2022.09.007
Source DB: PubMed Journal: Tech Innov Patient Support Radiat Oncol ISSN: 2405-6324
Characteristics of patients and spinal SBRT treatments for all three IGRT workflows.
| Characteristics of patients and treatment | IGRT workflow 1 | IGRT workflow 2 | IGRT workflow 3 | Total |
|---|---|---|---|---|
| Number of patients | 12 | 12 | 14 | 38 |
| Number of volumes | 13 | 13 | 15 | 41 |
| Cervical spine | 2 | 0 | 0 | 2 |
| Thoracic spine | 8 | 8 | 11 | 27 |
| Lumbar spine | 3 | 5 | 4 | 12 |
| Number of fractions | 105 | 85 | 93 | 283 |
| 5 × 7 Gy | 1 | 0 | 0 | 1 |
| 5 × 8 Gy | 5 | 9 | 10 | 24 |
| 10 × 3 Gy | 0 | 0 | 1 | 1 |
| 3 × 4.85 Gy | 0 | 0 | 1 | 1 |
| 10 × 4.85 Gy | 6 | 4 | 3 | 13 |
| 15 × 2.6 Gy | 1 | 0 | 0 | 1 |
| 2 Arcs | 13 | 11 | 8 | 32 |
| 3 Arcs | 0 | 1 | 7 | 8 |
| 4 Arcs | 0 | 1 | 0 | 1 |
| Wingboard | 11 | 13 | 12 | 36 |
| Treatment mask | 2 | 0 | 3 | 5 |
Fig. 1Standard positioning utilized during patient setup for thoracic and lumbar spine stereotactic body radiation therapy using BrainLab Infrared (IR) Reflective Reference Star and TrueBeam STx with Novalis Radiosurgery platform.
Fig. 2General IGRT workflow procedure.
Fig. 3Treatment procedure of IGRT workflow 1.
Fig. 4Treatment procedure of IGRT workflow 2.
Fig. 5Treatment procedure of IGRT workflow 3.
Total ExacTrac (EXT) and CBCT images for all three IGRT workflows.
| Total images | Workflow 1 | Workflow 2 | Workflow 3 | Total |
|---|---|---|---|---|
| Varian CBCTs | 210 | 169 | 185 | 564 |
| ExacTrac images | 103 | 170 | 454 | 727 |
| Total CBCTs & ExacTrac | ||||
| CBCTs suitable | 204 | 165 | 184 | 553 |
| ExacTrac suitable | 103 | 170 | 454 | 727 |
Fig. 6Graph showing the percentage of verifications that are within tolerance at various stages from the first ExacTrac verification following the corrections based on the pre-treatment CBCT, through the mid-treatment ExacTrac verifications to the final CBCT verification, for the three different IGRT workflows. *ExacTrac (EXT), Radiation Treatment (RT).
Maximum “out of tolerance” values in mm/° for all three IGRT workflows.
| EXT pre RT Arc 1 | EXT mid RT Arc 1 | EXT pre RT Arc 2 | EXT mid RT Arc 2 | EXT pre RT Arc 3 | EXT mid RT Arc 3 | CBCT post RT | ||
|---|---|---|---|---|---|---|---|---|
| Workflow 1 | lateral | 2.9 mm | N/A | N/A | N/A | N/A | N/A | 1.8 mm |
| longitudinal | 2.2 mm | N/A | N/A | N/A | N/A | N/A | 3.3 mm | |
| vertical | 1.9 mm | N/A | N/A | N/A | N/A | N/A | 1.7 mm | |
| pitch | – | N/A | N/A | N/A | N/A | N/A | – | |
| roll | 1.4° | N/A | N/A | N/A | N/A | N/A | 1.4° | |
| rotation | 1.5° | N/A | N/A | N/A | N/A | N/A | 1.4° | |
| Workflow 2 | lateral | 2.9 mm | N/A | 1.9 mm | N/A | N/A | N/A | 2.2 mm |
| longitudinal | 2.6 mm | N/A | 1.3 mm | N/A | N/A | N/A | 1.3 mm | |
| vertical | 2.6 mm | N/A | 1.7 mm | N/A | N/A | N/A | 1.3 mm | |
| pitch | 2.3° | N/A | 1.2° | N/A | N/A | N/A | 1.1° | |
| roll | 1.3° | N/A | 1.3° | N/A | N/A | N/A | 1.6° | |
| rotation | 1.2° | N/A | 1.3° | N/A | N/A | N/A | – | |
| Workflow 3 | lateral | 2.1 mm | 1.2 mm | 1.9 mm | 1.2 mm | 1.2 mm | 1.0 mm | 1.3 mm |
| longitudinal | 1.6 mm | 1.5 mm | 1.0 mm | – | – | – | 1.3 mm | |
| vertical | 2.5 mm | 1.4 mm | – | 1.3 mm | 1.4 mm | 1.1 mm | 1.4 mm | |
| pitch | 1.1° | – | 1.1° | 1.0 mm | 1.8° | – | – | |
| roll | 1.9° | 1.7° | 1.3° | – | – | – | – | |
| rotation | 1.4° | 1.1° | – | – | – | – | – | |
*ExacTrac (EXT), Radiation Treatment (RT).
Fig. 7Comparison of the positional, and rotational errors recorded by the initial ExacTrac verification (left) combined for all workflows and the final post-treatment CBCT (right) for the three different workflows. *Lateral (Lat), Longitudinal (Long), Vertical (Vert).
Effective whole-body dose in mSv for the ExacTrac imaging in the different IGRT workflows.
| Average number of ExacTrac image pairs per treatment | Effective dose per treatment, mSv | |
|---|---|---|
| IGRT workflow 1 | 12 | 0.6 mSv |
| IGRT workflow 2 | 19 | 0.95 mSv |
| IGRT workflow 3 | 36 | 1.8 mSv |
*millisievert (mSv).