Alaa Thabet Hassan1, Soher Mostafa Ahmed2, Azza Salah AbdelHaffeez3, Sherif A A Mohamed1.
Abstract
BACKGROUND: Despite its wide use in clinical practice, few studies have assessed the role of pulse oximetry in patients with heart failure. We aimed to evaluate the accuracy and precision of the pulse oximeter in patients with heart failure and to determine this accuracy at three different sensor locations.Entities:
Keywords: Pulse oximeter; accuracy; evaluation; heart failure; location; myocardial infarction; sensor
Year: 2021 PMID: 34316367 PMCID: PMC8278778 DOI: 10.4081/mrm.2021.742
Source DB: PubMed Journal: Multidiscip Respir Med ISSN: 1828-695X
Demographic data of the study and control groups.
| Study group (n=235) | Control group (n=90) | p | |
|---|---|---|---|
| N (%) | N (%) | ||
| Age (years) | 0.815 | ||
| Mean ± SD | 56.01 ± 12.38 | 53.25 ± 12.16 | |
| Median (range) | 58 (22-76) | 55 (26-76) | |
| Gender | 0.804 | ||
| Female | 130 (55.3) | 48 (53.3) | |
| Male | 105 (44.7) | 42 (46.7) | |
| Comorbidities | |||
| | 70 (29.8) | 30(33.3) | 0.591 |
| HTN | 91 (38.7) | 28(31.1) | 0.247 |
| Hypothyroidism | 12 (5.1) | 4 (4.4) | 1.000 |
| Hyperlipidemia | 126 (53.6) | 35(38.9) | 0.019 |
| Hyperthyroidism | 8 (3.4) | 3(3.3) | 1.000 |
| Bronchial asthma | 45(19.1) | 15(16.7) | 0.749 |
SD, standard deviation; DM, diabetes mellitus; HTN, hypertension.
Oxygen saturation measured by pulse oximetry and ABGs and in different locations among different groups of HF patients.
| Control | Group A | Group B | Group C | p | |
|---|---|---|---|---|---|
| (EF>40%) | (EF<40%) | (Acute HF) | |||
| O2 sat in ABGs | 96.6±3.14 | 94.8±5.1 | 93.32±3.63 | 92.84±7.45 | 0.056 |
| O2 sat in finger | 96.2±4.2 | 93.2±5.4 | 88.43±8.2 | 90±6.3 | 0.029* |
| O2 sat in toe | 95.7±4,7 | 92.9± 2.8 | 89.3±6.1 | 89.3±10.6 | 0.049* |
| O2 sat in ear | 96.9±2.8 | 91.89±3.5 | 92.78±8.7 | 93.6±6.2 | 0.051 |
| p | 0.09 | 0.054 | 0.005* | 0.008* |
ABGs, arterial blood gases; HF, heart failure; EF, ejection fraction; O2 sat, oxygen saturation.
Figure 1.Mean O2 saturation measured by pulse oximetry and arterial blood gases (ABG) in different locations.
Mean bias in different oxygen saturations measured by pulse oximetry in comparison to arterial blood gases among different groups of HF patients.
| Group A | Group B | Group C | p | ||
|---|---|---|---|---|---|
| Mean bias±SD | Mean bias±SD | Mean bias±SD | |||
| O2 saturation in the finger | O2 sat <80% | -0.15±1.26 | -4.15±3.15 | -1.99±1.75 | 0.027* |
| O2 sat >80% - <90% | -0.9±1.01 | -3.03±2.49 | -1.38±1.01 | 0.041* | |
| O2 sat> 90% | -0.4±0.43 | -1.02±1.28 | -0.5±0.28 | 0.009* | |
| O2 saturation in the toe | O2 sat <80% | -0.29±0.22 | -5.39±4.25 | -2.06±1.83 | 0.03* |
| O2 sat >80% - <90% | -1.08±0.05 | -4.78±3.96 | -1.76±1.30 | 0.040* | |
| O2 sat> 90% | -0.87±0.77 | -1.09±0.65 | -1±0.02 | 0.09 | |
| O2 saturation in the ear | O2 sat <80% | 1.21±0.89 | 0.43±0.35 | 0.49±0.23 | 0.06 |
| O2 sat >80% - <90% | 1.64±1.23 | 0.75±0.54 | 0.64±0.45 | 0.075 | |
| O2 sat> 90% | 2.23±0.76 | 1.0±0.91 | 0.88±0.62 | 0.049* | |
| 0.009* | 0.000* | 0.01* |
SD, standard deviation; O2 sat, oxygen saturation; *p<0.05.
Figure 2.Pattern of mean bias in oxygen saturation measured with pulse oximeter in Group A.
Figure 3.Pattern of mean bias in oxygen saturation measured with pulse oximeter in Group B.
Figure 4.Pattern of mean bias in oxygen saturation measured with pulse oximeter in Group C.