Reliability of noninvasive oximetry in black subjects during exercise and hypoxia.

R J Zeballos1, I M Weisman.   

Abstract

The effect of skin pigmentation on the reliability of noninvasive oximetry, especially during exercise and hypoxia, has not been thoroughly investigated. This is the first study, to our knowledge, that specifically addresses this question. Thirty-three young black men performed multistage, steady-state cycle ergometry, breathing gas mixtures simulating different altitudes: 33 breathed gas simulating sea level (PIO2 = 146 mm Hg), 11 breathed gas simulating 2,300 m (PIO2 = 110 mm Hg), and 22 breathed gas simulating 4,000 m (PIO2 = 85 mm Hg). Co-oximeter SaO2 determinations were performed in arterial blood samples obtained concurrently with ear oximetry that was measured using Hewlett-Packard 47201A (HP) and Blox IIA oximeters. The mean error or bias for the [HP - SaO2] and for [Biox IIA - SaO2] +/- 95% CI were: at simulated sea level (SaO2 greater than 96%): -0.4 +/- 0.3% and 2.1 +/- 0.3%; at simulated 2,300 m (range of SaO2 means, 89 to 94%): -0.8 +/- 0.5% and 3.5 +/- 0.9%; for simulated 4,000 m (range of SaO2 means, 75 to 84%): -4.8 +/- 1.6% and 9.8 +/- 1.8%, respectively. A better coefficient correlation was observed for all the pairs between SaO2 versus HP (r = 0.94, p less than 0.001, n = 279) than for the SaO2 versus Biox IIA (r = 0.80, p less than 0.001, n = 242). In conclusion, the HP oximeter appears to estimate SaO2 more accurately than the Biox IIA oximeter. The previously described overestimation for the Biox IIA ear oximeter and the underestimation for the HP ear oximeter at low SaO2 values in whites is exaggerated in blacks.(ABSTRACT TRUNCATED AT 250 WORDS)

Entities:  

Mesh:

Year:  1991        PMID: 1741533     DOI: 10.1164/ajrccm/144.6.1240

Source DB:  PubMed          Journal:  Am Rev Respir Dis        ISSN: 0003-0805


  11 in total

1.  Arterial blood gas and pulse oximetry in initial management of patients with community-acquired pneumonia.

Authors:  K P Levin; B H Hanusa; A Rotondi; D E Singer; C M Coley; T J Marrie; W N Kapoor; M J Fine
Journal:  J Gen Intern Med       Date:  2001-09       Impact factor: 5.128

2.  Effect of newborn screening for critical CHD on healthcare utilisation.

Authors:  Rie Sakai-Bizmark; Hiraku Kumamaru; Eliza J Webber; Dennys Estevez; Laurie A Mena; Emily H Marr; Ruey-Kang R Chang
Journal:  Cardiol Young       Date:  2020-07-02       Impact factor: 1.093

3.  Racial and Ethnic Discrepancy in Pulse Oximetry and Delayed Identification of Treatment Eligibility Among Patients With COVID-19.

Authors:  Ashraf Fawzy; Tianshi David Wu; Kunbo Wang; Matthew L Robinson; Jad Farha; Amanda Bradke; Sherita H Golden; Yanxun Xu; Brian T Garibaldi
Journal:  JAMA Intern Med       Date:  2022-07-01       Impact factor: 44.409

4.  Influence of Sex, BMI, and Skin Color on the Accuracy of Non-Invasive Cuffless Photoplethysmography-Based Blood Pressure Measurements.

Authors:  Dean Nachman; Arik Eisenkraft; Nir Goldstein; Arik Ben-Ishay; Meir Fons; Roei Merin; Yftach Gepner
Journal:  Front Physiol       Date:  2022-06-29       Impact factor: 4.755

Review 5.  Skin Pigmentation Influence on Pulse Oximetry Accuracy: A Systematic Review and Bibliometric Analysis.

Authors:  Ana M Cabanas; Macarena Fuentes-Guajardo; Katina Latorre; Dayneri León; Pilar Martín-Escudero
Journal:  Sensors (Basel)       Date:  2022-04-29       Impact factor: 3.847

6.  Pulse oximetry in sickle cell disease.

Authors:  P Pianosi; T D Charge; D W Esseltine; A L Coates
Journal:  Arch Dis Child       Date:  1993-06       Impact factor: 3.791

7.  The Big Consequences of Small Discrepancies: Why Racial Differences in Pulse Oximetry Errors Matter.

Authors:  Andre L Holder; An-Kwok Ian Wong
Journal:  Crit Care Med       Date:  2022-02-01       Impact factor: 7.598

8.  Pulse oximeter accuracy and precision at five different sensor locations in infants and children with cyanotic heart disease.

Authors:  Jyotirmoy Das; Amit Aggarwal; Naresh Kumar Aggarwal
Journal:  Indian J Anaesth       Date:  2010-11

9.  Pulse oximetry.

Authors: 
Journal:  Crit Care       Date:  1999       Impact factor: 9.097

10.  Accuracy and precision of pulse oximeter at different sensor locations in patients with heart failure.

Authors:  Alaa Thabet Hassan; Soher Mostafa Ahmed; Azza Salah AbdelHaffeez; Sherif A A Mohamed
Journal:  Multidiscip Respir Med       Date:  2021-07-06
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.