Enamel nail polish does not interfere with pulse oximetry among normoxic volunteers.

Thomas M Brand1, Mary E Brand, Gregory D Jay.   

Abstract

OBJECTIVE: To determine if enamel nail polish interferes with pulse oximetry.
DESIGN: Laboratory investigation. PARTICIPANTS: 12 healthy nonsmoking volunteers.
METHODS: Spectrophotometry was performed on polystyrene cuvettes painted with 3 uniform layers of nail polish enamel in triplicate. Absorbances were compared at 660 and 940 nm for 10 colors across the visible spectrum and documented by the Institute of Electrical and Electronic Engineers color code against Pantone matches on a color scanner. Colors were then selected that had the greatest A660-A940 difference and pulse oximetry (Nellcor N-209A. Pleasanton , CA) was performed on nails painted with these colors on 12 subjects using an unpainted digit as a control.
RESULTS: When tested by the oximeter, there were no SpO2 differences detected between nail polish colors, nor were there any differences between the painted and the unpainted control digits across subjects (F = 0.51, ANOVA; p = 0.67). Blue (B), green (G), and lime green (G') possessed the largest A660-A940 difference and potentially could interfere with pulse oximetry but did not do so in the clinical model.
CONCLUSION: Enamel finger nail polish pigments do not interfere with pulse oximetry as previously reported, the A660-A940 difference must be greater than 1.88 +/- 0.23 SD AU in order to affect pulse oximetry.

Entities:  

Mesh:

Year:  2002        PMID: 12212998     DOI: 10.1023/a:1016385222568

Source DB:  PubMed          Journal:  J Clin Monit Comput        ISSN: 1387-1307            Impact factor:   2.502


  11 in total

Review 1.  Pulse oximetry. Uses and abuses.

Authors:  L M Schnapp; N H Cohen
Journal:  Chest       Date:  1990-11       Impact factor: 9.410

2.  Pulse oximeter desaturation due to methemoglobinemia.

Authors:  J B Eisenkraft
Journal:  Anesthesiology       Date:  1988-02       Impact factor: 7.892

3.  Effects of methemoglobinemia on pulse oximetry and mixed venous oximetry.

Authors:  S J Barker; K K Tremper; J Hyatt
Journal:  Anesthesiology       Date:  1989-01       Impact factor: 7.892

4.  Nail polish and oximetry.

Authors:  P F White; W A Boyle
Journal:  Anesth Analg       Date:  1989-04       Impact factor: 5.108

5.  Ambient light affects pulse oximeters.

Authors:  L Hanowell; J H Eisele; D Downs
Journal:  Anesthesiology       Date:  1987-11       Impact factor: 7.892

6.  Nail polish does not affect pulse oximeter saturation.

Authors:  B K Kataria; R Lampkins
Journal:  Anesth Analg       Date:  1986-07       Impact factor: 5.108

7.  The effect of nail polish on pulse oximetry.

Authors:  C J Coté; E A Goldstein; W H Fuchsman; D C Hoaglin
Journal:  Anesth Analg       Date:  1988-07       Impact factor: 5.108

8.  Pulse oximetry.

Authors:  W New
Journal:  J Clin Monit       Date:  1985-04

9.  The effect of acrylic nails on the measurement of oxygen saturation as determined by pulse oximetry.

Authors:  S M Peters
Journal:  AANA J       Date:  1997-08

10.  Pulse oximetry is accurate in acute anemia from hemorrhage.

Authors:  G D Jay; L Hughes; F P Renzi
Journal:  Ann Emerg Med       Date:  1994-07       Impact factor: 5.721

View more
  5 in total

Review 1.  Pulse oximetry in supportive and palliative care.

Authors:  Vandana A Vora; Sam H Ahmedzai
Journal:  Support Care Cancer       Date:  2004-11       Impact factor: 3.603

2.  Effects of passive smoking on students at College of Applied Medical Sciences, King Saud Bin Abdulaziz University for Health Sciences, Riyadh.

Authors:  Abdullah Alanazi; Farhan Al Enezi; Mohammd Mesfer Alqahtani; Turki Faleh Alshammari; Mumtaz Ahmed Ansari; Saleh Al-Oraibi; Shoeb Qureshi
Journal:  J Nat Sci Biol Med       Date:  2015 Jan-Jun

3.  The Predictive Value of Pulse Oximeters for Pulse Improvement after Angiography in Infants and Children.

Authors:  Mohammad-Reza Alipour; Mazyar Rastegar; Mehdi Ghaderian; Seyedeh-Mahdieh Namayandeh; Reza Faraji; Zohreh Pezeshkpour
Journal:  Iran J Pediatr       Date:  2016-07-10       Impact factor: 0.364

4.  Pulse oximetry in bronchiolitis: is it needed?

Authors:  Mohamed A Hendaus; Fatima A Jomha; Ahmed H Alhammadi
Journal:  Ther Clin Risk Manag       Date:  2015-10-12       Impact factor: 2.423

5.  Accuracy and precision of pulse oximeter at different sensor locations in patients with heart failure.

Authors:  Alaa Thabet Hassan; Soher Mostafa Ahmed; Azza Salah AbdelHaffeez; Sherif A A Mohamed
Journal:  Multidiscip Respir Med       Date:  2021-07-06
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.