Pulse oximeter probes. A comparison between finger, nose, ear and forehead probes under conditions of poor perfusion.

D G Clayton1, R K Webb, A C Ralston, D Duthie, W B Runciman.   

Abstract

The performances of 10 pulse oximeters using finger probes were compared with the same pulse oximeters using alternative probes (eight finger probes, two nose probes and a forehead probe) in poorly perfused patients. All readings were then compared with directly measured arterial blood oxygen saturations. The mean difference (bias, 'accuracy'), standard deviation (precision) and 'drop out' rate for each pulse oximeter combination was determined. An overall ranking of performance of each pulse oximeter was calculated using five criteria (accuracy, precision, number of readings within 3% of standard, percentage of readings given within 3% of standard, expected overread limit in 95% of cases). Nose and forehead probes performed poorly. Some ear probes performed well compared to some finger probes, but the overall performance of probes in other sites compared to finger probes was worse, (p = 0.05). Two of eight ear probes and no nose or forehead probes would be expected to be within 4% of the reference value in 95% of readings. The use of finger probes rather than probes in other sites is recommended in the patient with poor peripheral perfusion.

Entities:  

Mesh:

Year:  1991        PMID: 2024741     DOI: 10.1111/j.1365-2044.1991.tb11492.x

Source DB:  PubMed          Journal:  Anaesthesia        ISSN: 0003-2409            Impact factor:   6.955


  12 in total

1.  Arterial line for monitoring SpO2 in patients with ischemic peripheries.

Authors:  Goneppanavar Umesh; Kaur Jasvinder; Shetty Nanda; Shetty Ranjan
Journal:  J Clin Monit Comput       Date:  2010-07-16       Impact factor: 2.502

2.  Uses and abuses of pulse oximetry.

Authors:  D G Kruchek
Journal:  Arch Dis Child       Date:  1996-08       Impact factor: 3.791

3.  The peripheral pulse wave: information overlooked.

Authors:  W B Murray; P A Foster
Journal:  J Clin Monit       Date:  1996-09

4.  A comparison of finger and forehead pulse oximeters in heart failure patients during maximal exercise.

Authors:  Kevin L Kelly; Alex R Carlson; Thomas G Allison; Bruce D Johnson
Journal:  Heart Lung       Date:  2019-11-14       Impact factor: 2.210

5.  Arterial blood gas analysis or oxygen saturation in the assessment of acute asthma?

Authors:  D M Carruthers; B D Harrison
Journal:  Thorax       Date:  1995-02       Impact factor: 9.139

6.  Assessment of sensory sensitivity through critical flicker fusion frequency thresholds after a maximum voluntary apnoea.

Authors:  Francisco de Asís Fernández; Fernando González-Mohino; José M González-Ravé
Journal:  Diving Hyperb Med       Date:  2019-09-30       Impact factor: 0.887

7.  Pulse oximeter accuracy and precision at five different sensor locations in infants and children with cyanotic heart disease.

Authors:  Jyotirmoy Das; Amit Aggarwal; Naresh Kumar Aggarwal
Journal:  Indian J Anaesth       Date:  2010-11

8.  Pulse oximetry.

Authors: 
Journal:  Crit Care       Date:  1999       Impact factor: 9.097

Review 9.  Pulse oximetry.

Authors:  Amal Jubran
Journal:  Crit Care       Date:  2015-07-16       Impact factor: 9.097

10.  In vivo investigation of ear canal pulse oximetry during hypothermia.

Authors:  K Budidha; P A Kyriacou
Journal:  J Clin Monit Comput       Date:  2017-01-27       Impact factor: 2.502

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.