Instrumented implants are a promising approach to further improve the clinical outcome of total hip arthroplasties. For the integrated sensors or active functions, an electrical power supply is required. Energy harvesting concepts can provide autonomous power with unlimited lifetime and are independent from external equipment. However, those systems occupy space within the mechanically loaded total hip replacement and can decrease the life span due to fatigue failure in the altered implant. We previously presented a piezoelectric energy harvesting system for an energy-autonomous instrumented total hip stem that notably changes the original implant geometry. The aim of this study was to investigate the remaining structural fatigue failure strength of the metallic femoral implant component in a worst-case scenario. Therefore, the modified hip stem was tested under load conditions based on ISO 7206-4:2010. The required five million cycles were completed twice by all samples (n = 3). Additionally applied cycles with incrementally increased load levels up to 4.7 kN did not induce implant failure. In total, 18 million cycles were endured, outperforming the requirements of the ISO standard. Supplementary finite element analysis was conducted to determine stress distribution within the implant. A high stress concentration was found in the region of modification. The stress level showed an increase compared to the previously evaluated physiological loading situation and was close to the fatigue data from the literature. The stress concentration factor compared to the original geometry amounted to 2.56. The assessed stress level in accordance with the experimental fatigue testing can serve as a maximum reference value for further implant design modifications and optimisations.
Instrumented implants are a promising approach to further improve the clinical outcome of total hip arthroplasties. For the integrated sensors or active functions, an electrical power supply is required. Energy harvesting concepts can provide autonomous power with unlimited lifetime and are independent from external equipment. However, those systems occupy space within the mechanically loaded total hip replacement and can decrease the life span due to fatigue failure in the altered implant. We previously presented a piezoelectric energy harvesting system for an energy-autonomous instrumented total hip stem that notably changes the original implant geometry. The aim of this study was to investigate the remaining structural fatigue failure strength of the metallic femoral implant component in a worst-case scenario. Therefore, the modified hip stem was tested under load conditions based on ISO 7206-4:2010. The required five million cycles were completed twice by all samples (n = 3). Additionally applied cycles with incrementally increased load levels up to 4.7 kN did not induce implant failure. In total, 18 million cycles were endured, outperforming the requirements of the ISO standard. Supplementary finite element analysis was conducted to determine stress distribution within the implant. A high stress concentration was found in the region of modification. The stress level showed an increase compared to the previously evaluated physiological loading situation and was close to the fatigue data from the literature. The stress concentration factor compared to the original geometry amounted to 2.56. The assessed stress level in accordance with the experimental fatigue testing can serve as a maximum reference value for further implant design modifications and optimisations.
Entities:
Keywords:
Total hip replacement; energy harvesting; fatigue; finite element analysis; testing
Total hip arthroplasty is a common treatment for hip joint-related diseases and the
standard therapy when conservative measures are exhausted.
Demographic changes in the growing world population as well as the increased
use of implants in younger patients lead to a rising number of total hip
arthroplasties.[2,3]
Despite the actual low revision burden the absolute number of failed implants will
increase proportionally to the number of primary arthroplasties.
Thus, a longer service life of implants is required as revision surgeries
come with the risk of additional complications and a decrease in the patient’s
quality of life[4,5]
as well as an increase of the economic burden.[6-8] This highlights the necessity
of further developments with regard to continuously improving the clinical outcome.
Research in the field of instrumented implants is a promising approach to reduce
deficits and increase the options of treatment. Sensors for diagnostic functions can
monitor implant stability, implant loosening or wear to support clinical decisions
for possible counteractive measures. Recording the patients’ activity and implant
loading allows for an individual retrospective analysis in case of implant failure.
It may also provide relevant daily life load data for implant development. Beside
passive sensors, active functions can be used for direct therapeutic measures such
as electrical stimulation to promote bone growth and implant stability.[9-12] For most sensors and
actuators, the question of an optimal power source arises. In the beginning, energy
and data transmission was realised by percutaneous wires,[13-15] which came with the obvious
risk of infection. The use of telemetry and batteries allowed wireless
measurements.[16,17] Drawbacks of batteries (finite lifetime and risk of leakage)
where overcome by external power transmission via inductive coupling.[12,18,19] However, the
requirement of external equipment for this method limits is use to a temporary
application in a clinical or laboratory environment. Hence, to address these
shortcomings energy harvesting represents a promising solution. Instrumented
implants would benefit from an autonomous energy supply, rendering replacement of a
power source unnecessary and allowing for continuous operation during daily
life.We previously presented an energy harvesting concept for a load-bearing total hip replacement.
A schematic of the new proposed concept is shown in Figure 1. The loads acting on the implant
are transmitted through the hip stem and the ultra-high-molecular-weight
polyethylene (UHMW-PE) housing to a multilayer piezoelectric element. Thereby
physical activity generates voltage as previously described by us in more detail.
The piezoelectric element functions as a power source for any
instrumentation; however, as a self-powered sensor the output voltage can also
directly be related to the implant loading.
Figure 1.
Energy harvesting concept, integrated in the hip stem (with detail of
modified implant geometry). UHMW-PE housing in transparent view.
Energy harvesting concept, integrated in the hip stem (with detail of
modified implant geometry). UHMW-PE housing in transparent view.Like any other instrumentation system (sensors, circuitry, transmission units, etc.),
the energy harvesting system requires space. Orthopaedic load-bearing implants that
have been used in humans for in vivo joint force measurements were modified in
relatively unloaded regions, that is, primary along neutral axes, to maintain the
implant safety.[19,21-23] In contrast,
it is inherent in our design concept that the piezoelectric element is placed in a
mechanically loaded area to transmit the necessary forces for generation of power.
Therefore, a cavity to house the energy harvesting system was introduced. However,
this notably changes the cross-sectional area in the hip stem and alters the flow of
forces. It is likely to impair the structural fatigue failure strength of the
implant. Previous numerical simulations of a physiologically based loading situation
revealed that, compared to the unmodified geometry, stress concentrated at the
cavity base.
In this study, a worst-case loading situation, simulating the proximal
implant loosening, was used to evaluate whether the safety of the mechanically
loaded metallic hip stem component was sufficient.
The endurance testing conducted may prove the applicability of the new
concept and is based on the standardised procedure of ISO 7206-4:2010.
In order to study the implant loading in more detail, the fatigue experiments
were accompanied by a finite element analysis (FEA) to evaluate the maximum stress
and strain levels and the load distribution. The finite element model was validated
by a quasi-static experiment.
Materials and methods
Quasi-static and fatigue testing
The modified geometry of the hip stem, based on our previous work,
was milled from an original hip stem (Exeter V40, size 37.5 mm N°3,
Stryker, Howmedica Osteonics Corp, Mahwah, New Jersey, USA) by an external
manufacturer (Figure
2). A V40 cobalt-chrome head of 32 mm diameter with the maximal allowed
offset of 8 mm was selected. For validation of the FEA, a linear strain gauge
(DMS 1.5/120 LY15, Hottinger Baldwin Messtechnik GmbH, Darmstadt, Germany) was
applied to the lateral hip stem region at the cavity level. For the placement of
the strain gauge in a reproducible way and in accordance with the FE model, an
additive manufactured template based on the CAD geometry was used to transfer
the centre and measuring direction axis of the sensor. The position was based on
preliminary simulation results, reconciling the requirement of a preferably
homogenously loaded area with a location near the region of interest. The cavity
base itself with the expected stress concentration is not suitable for strain
gauge application due to the high curvature.
Figure 2.
Modified implant geometry with detail of the milled cavity. The new
surface showed a rougher structure from milling and polishing than the
original mirror finish.
Modified implant geometry with detail of the milled cavity. The new
surface showed a rougher structure from milling and polishing than the
original mirror finish.The testing was conducted based on the ISO standard 7206-4:2010.
The hip stem specimen was oriented (α = 9°, β = 10°) within a specimen
holder with the help of an align fixture. Subsequently, a filled casting resin
(RenCast® FC 52/53 Isocyanate/FC 53 Polyol, filler DT 082;
Young’s modulus of 2.4 GPa; Huntsman Advanced Materials GmbH, Basel,
Switzerland) was filled up to the embedding level for potting (80 mm distance to
the head centre). The specimen was placed under an electrodynamic uniaxial
testing machine (LTM 5, ZwickRoell GmbH & Co. KG, Ulm, Germany equipped with
a 5 kN load cell). A ball ring, held in position by soft springs, was integrated
between the head and the actuator to prevent the transmission of lateral
forces.For validation purposes, the maximum force of 2300 N was applied stepwise within
10 increments in a quasi-static experiment while measuring the strains with the
strain gauge. At each load level, the force was held for 5 s. This experiment
was repeated three times. The strains were averaged over the holding time of
each increment and over the three specimens. The following fatigue testing was
conducted with a sinusoidal load of maximum 2300 N with R = 0.1
at a frequency of 10 Hz for 5 million cycles. After successful completion of the
test, a repetition of another five million cycles was allowed, followed by an
incremental increase of the loading level (+300 N for one million cycles per
increment, R = 0.1) to induce fracture. During the fatigue
testing, the specimen was unloaded every 5000 cycles to allow self-recentring of
the ball ring by the soft springs. Figure 3 shows a schematic of the
testing procedure.
Figure 3.
Schematic of the testing procedure. The quasi-static testing for
validation purpose was followed by the fatigue testing based on ISO
7206-4:2010 (twice, if no fracture for the first run) and then continued
with successively increased load levels to force failure until
4.7 kN (below load cell limit).
Schematic of the testing procedure. The quasi-static testing for
validation purpose was followed by the fatigue testing based on ISO
7206-4:2010 (twice, if no fracture for the first run) and then continued
with successively increased load levels to force failure until
4.7 kN (below load cell limit).The whole experiment was conducted for three specimens. Figure 4 displays the overall test
set-up.
Figure 4.
Test set-up and detail of the hip stem with modification and the applied
strain gauge (arrow).
Test set-up and detail of the hip stem with modification and the applied
strain gauge (arrow).
Finite element analysis
The finite element model reproduced the quasi-static testing procedure. The
geometry was generated in SolidWorks 2018 (Dassault Systèmes,
Vélizy-Villacoublay, France) and imported to ANSYS V18.2 (Ansys Inc, Canonsburg,
Pennsylvania, USA). Only the modified hip stem and the embedding material were
considered. For the representation of the strain gauge measure grid, a surface
patch was constructed on the lateral stem area (see Figure 5, detail) together with a
centred, local coordinate system with the x-axis pointing in
the measuring direction.
Figure 5.
Boundary conditions and loading of the FE model (femoral head centre
pt. C with the force distributed on the taper’s
outer surface, and fixed support in blue), detail of the strain gauge
measuring grid with local coordinate system.
Boundary conditions and loading of the FE model (femoral head centre
pt. C with the force distributed on the taper’s
outer surface, and fixed support in blue), detail of the strain gauge
measuring grid with local coordinate system.The force was applied to a remote point, representing the head centre point C
(pt. C) and was distributed to the taper’s outer surface.
The loading was in direction of the cylinder axis of the embedding material. The
outer surface of the embedding material was fully constrained, apart from the
top area. Figure 5
shows the loading and boundary conditions.A symmetric frictional contact (µ = 0.3) was defined between the hip stem and the
embedding material. Both components were meshed with quadratic solid tetrahedral
elements. The strain gauge was approximated by a single linear shell element.
The final mesh density was based on a mesh independence study with converging
results for all reported output parameters. Linear-elastic material behaviour
was assumed, with Poisson’s ratio of ν = 0.3. All material
properties are listed in Table 1.
Material properties.For calculation of a stress concentration factor, we accordingly performed a
simulation of the original design without cavity.A sensitivity analysis was performed to identify the influence of the input
parameters on the stress and strain results. The Young’s moduli of the implant
and embedding material were altered by ±2.5%. With regard to the potting
process, different embedding levels (±1 mm) and deviations in the alignments
(±3° for α and β) were considered. Additionally, the head centre pt.
C was moved ±0.5 mm along the neck axis. Main imprecision is
hypothesised to result from the strain gauge application. To account for
geometric deviations between simulation and experiment, anterior, posterior,
proximal and distal displacement of ±0.5 mm from the original position were
simulated, as well as a rotation of the measuring direction about the local
coordinate system by ±3°.
Results
Quasi-static testing and FE-model validation
The average measured strain value at the maximum force level amounted to 614 µm/m
(standard deviation of 19 µm/m, equals 3% of the maximum). The strain rose
nearly linearly by around 61 µm/m for each load level increment.The percentage deviation of the simulated strain gauge was ≤13.1% at each load
level compared to the average strain values for the three specimens. The maximum
difference occurred at 2300 N (92 µm/m). Figure 6 shows the linear regression
between the experiments and the FE simulation with a very high coefficient of
determination (R2 = 0.997). The slope of 0.8686
reflects the above-mentioned percentage deviation.
Figure 6.
Linear regression results between FE data and experimental strain gauge
values. To each numerical result, three experimental strain values were
assigned according to the number of tested specimens.
Linear regression results between FE data and experimental strain gauge
values. To each numerical result, three experimental strain values were
assigned according to the number of tested specimens.
FE loading distribution and sensitivity analysis
For the hip stem, several local load concentrations were found (since stress and
strain strongly correlated for the linear-elastic material behaviour, only the
latter is shown in Figure
7 for convenience).
Figure 7.
Von Mises strain distribution (µm/m) for the hip stem with load
concentration at the cavity ground.
Von Mises strain distribution (µm/m) for the hip stem with load
concentration at the cavity ground.Local load concentrations occurred at the anterior and lateral hip stem region
near the embedding level. Due to the change of contact situation, the local
singularity led to non-converging stress and strain results. Further load
concentrations were located at the upper and lower neck at the transition to the
taper. For the lower position, it amounted to 1797 µm/m (respectively von Mises
stress of 349 MPa). At the cavity base, the maximum strain value was 2137 µm/m
(respectively 417 MPa). It was slightly shifted to the lateral side in the
region of the highest overall loading gradient. In contrast, the opposed side
where the strain gauge is located showed notably more uniform strain
distribution. For the original design, the maximum von Mises stress in the
region where the cavity would be located amounted to 163 MPa. Thus, the changed
geometry resulted in a stress concentration factor of 2.56.For the sensitivity analysis, the loading in the cavity base was reported as von
Mises stress in order to compare to fatigue data for metallic implant materials
from the literature, which is usually reported as stress. The strain gauge
position and orientation had no influence and are therefore not shown in Figure 8. No parameter
notably changed the stress in the cavity base. The percentage deviation for all
was<0.2%, except for the position of the head centre pt. C.
The latter had an influence of >0.8%, which is still small. For the largest
distance between pt. C and the taper’s frontal surface, the
maximum stress amounted to 420 MPa. All modifications that contributed to an
increased stress value sum up to an influence of 1.3% (5 MPa).
Figure 8.
Results of the sensitivity analysis (absolute values in blue and
percentage deviation in red). Deviations were relative to the original
value of the reference model (dashed line). For convenience, the
absolute values of the percentages are shown: (a) von Mises stress
maximum at the implant’s cavity base (MPa) and (b) simulated strains for
the strain gauge (µm/m). Configuration names according to the changed
parameters: Young’s modulus implant material –‘E Imp.’, Young’s modulus
embedding material –‘E Emb.’, Embedding level –‘Emb. level’, Angles
specified by ISO 7206-4:2010 – ‘Alpha’ resp. ‘Beta’, Displacement of the
head centre along the neck axis –‘Position pt. C’, Proximal, distal,
anterior or posterior displacement of strain gauge –‘SG prox.’, ‘SG
dist.’, ‘SG ant.’ or ‘SG post.’, Rotation of the strain gauge’s
measuring direction about the local coordinate system –‘SG rot.’
Results of the sensitivity analysis (absolute values in blue and
percentage deviation in red). Deviations were relative to the original
value of the reference model (dashed line). For convenience, the
absolute values of the percentages are shown: (a) von Mises stress
maximum at the implant’s cavity base (MPa) and (b) simulated strains for
the strain gauge (µm/m). Configuration names according to the changed
parameters: Young’s modulus implant material –‘E Imp.’, Young’s modulus
embedding material –‘E Emb.’, Embedding level –‘Emb. level’, Angles
specified by ISO 7206-4:2010 – ‘Alpha’ resp. ‘Beta’, Displacement of the
head centre along the neck axis –‘Position pt. C’, Proximal, distal,
anterior or posterior displacement of strain gauge –‘SG prox.’, ‘SG
dist.’, ‘SG ant.’ or ‘SG post.’, Rotation of the strain gauge’s
measuring direction about the local coordinate system –‘SG rot.’With regard to the model validation, the strain in the strain gauge was
predominantly influenced by the Young’s modulus of the implant material, see
Figure 8(b). A
change of 2.5% altered the strain by the same value; for a weaker material the
maximum strain was 725 µm/m (18 µm/m higher than for the reference model). The
change of position of the head centre pt. C still had an
influence of around 1%. All other geometrical adaptations as well as the Young’s
modulus of the embedding material had an influence below 0.05%. In contrast,
displacement and rotation of the simulated strain gauge notably changed the
measured strain in the range of 1.1%–2.2%. The deviation for summarising all
single parameters contributing to a change in the strain is ±9% (respectively
64 µm/m).
Fatigue-testing
All three specimens passed five million cycles of the testing procedure without
fracture or notable plastic deformation. Even the optional repetition and the
increase in the magnitude of cyclic loading in eight steps with testing for one
million cycles at each level were endured; hence, 18 million cycles in total.
Figure 9 shows the
measured data for an exemplary specimen. The specified force levels were always
attained with no relevant deviations. Small fluctuations are shown in the
displacement curve.
Figure 9.
Complete testing for an exemplary specimen (periodically measured maximum
values of force in red and displacements of the femoral head in blue).
Since the testing was force controlled, the displacement curve showed
small fluctuations.
Complete testing for an exemplary specimen (periodically measured maximum
values of force in red and displacements of the femoral head in blue).
Since the testing was force controlled, the displacement curve showed
small fluctuations.Beyond the testing protocol, we additionally loaded the first specimen in a
quasi-static experiment using a uniaxial testing machine (Z050, ZwickRoell GmbH
& Co. KG, Ulm, Germany) with an increased test load since no fracture could
be effected from the dynamic testing at the load cell force limit (5 kN) at a
cycle number of 18 million. The maximum force of 10.1 kN was reached after
4.74 mm of vertical displacement of the prosthetic head. Subsequently, the force
decreased. At 17.43 mm of vertical displacement, the experiment was stopped due
to large dislocation of the ball ring under the actuator. The hip stem showed
strong plastic deformation at the embedding level (Figure 10), however, no fracture
occurred.
Figure 10.
Plastically deformed specimen after quasi-static testing (maximum
vertical displacement 17.43 mm, max. force
10.1 kN at 4.74 mm) and after removal from
the embedding medium. Note that the deformations occurred in the region
of the embedding level (in contrast, the cavity was located on the
opposite side of the still attached strain gauge).
Plastically deformed specimen after quasi-static testing (maximum
vertical displacement 17.43 mm, max. force
10.1 kN at 4.74 mm) and after removal from
the embedding medium. Note that the deformations occurred in the region
of the embedding level (in contrast, the cavity was located on the
opposite side of the still attached strain gauge).
Discussion
A linear strain increase for the strain gauge values, visible both in the
experimental and the numerical data, was observed, since no stress hot spot exceeded
the yield strength Rp0.2 of 436 MPa
and thereby only linear-elastic material behaviour occurred and for the
numerical simulation, only linear elasticity was defined. Furthermore, the detected
deformations of the implant were negligible, that is, relevant structural stiffness
changes may not occur. A coefficient of determination of nearly 1 (Figure 6) confirmed the
correlation between the experimental and numerical data, thus supporting the
validity of the numerical model. However, the pronounced deviation between the
maximum strain values must be considered. With regard to the summarised deviation of
64 µm/m, when all parameters from the sensitivity analysis are taken into account,
only around two-thirds of the deviation from the experiment (92 µm/m) can be
explained. The variations for the implant material’s Young’s modulus and the
position of the head centre pt. C were rather low, however stronger
effects are possible when assuming higher changes. It was correctly hypothesised
that the maximum influence on the measured strain results from the strain gauge
positioning. Nevertheless, further effects with influence on the simulated strain
may exist that could not be addressed within the sensitivity analysis. Particularly
the strain gauge application has to be mentioned since gluing on the polished hip
stem surface is delicate. Incomplete moistening with glue or thick adhesive layers
will alter the measured strain. The slight curvature of the hip stem additionally
complicates the gluing process. This curvature was also neglected in the FE
modelling of the strain gauge by a single shell. In the end, the origin of the 13.1%
divergence between the numerical and experimental results could not be resolved
exclusively. Since the small standard deviation confirms the reliable experimental
results, the mentioned percentage deviation should be considered when discussing
fatigue data on basis of the simulation results. However, the values of the FE
analysis were higher than the measured strains and thereby a rather conservative
estimation.The simulated strain and stress distribution within the implant revealed several
locations prone to fatigue damage and therefore enhanced general understanding of
the loading situation. This demonstrates that supplementary finite element analysis
can deliver additional knowledge compared to ISO 7206-4:2010 experimental fatigue
testing which determines only whether the specified test is passed.
The particular computed values can be compared to literature data and might
serve as a reference in design optimisation processes. For this study, hot spots at
the embedding level and for the upper and lower neck were noted, with tensile and
compression areas, respectively, at the opposed sides. These findings are in
accordance with our previous work for a similar set-up.
The stress concentration at the cavity base was expected since the increased
deformation is necessary for the functioning of the energy harvesting by
transmitting the forces to the piezoelectric element. The high gradient indicates
the inexpediency of strain gauge application and illustrates the advantage of FEA.
The stress maximum of 417 MPa increases the stress locally by 254 MPa compared to
the original design from the manufacturer. The stress maxima exceed the maxima of
our previous numerical investigations for a physiologically based loading situation,
and raised the stress maximum at the cavity base and at the lower neck by nearly 150
and 60 MPa, respectively.
The reasons are, predominantly, the increased force acting in the prosthetic
head centre, the altered embedding situation and also the absent housing and
piezoelectric element which usually would support the structure.The influence of the maximum stress level by model variations is negligible, as the
sensitivity analysis showed. Since the changes related to the strain gauge had only
numerical effects (small variations of mesh or time increments), they may be
disregarded. The high sensitivity to the position of the head centre pt.
C is based on altering the lever of the acting force. An estimated
overall increase for all parameters amounted to 422 MPa. This is below the fatigue
data for this material of around 470 MPa (high-nitrogen stainless steel
(Orthinox®), at 107 cycles, 10 Hz, in Eagle’s medium at
37°C) reported in the literature,
but with the same stress ratio of R = 0.1 which is typical
for implant loadings.[22,27] It is important to note that this value depends on various
parameters, for example, manufacturing, heat treatment, surface finish, size, and so
on and can, therefore, yield only approximate evaluations. Thus, the remaining
margin of around 50 MPa between the simulated maximum stress at the cavity base and
the fatigue value from literature is small also with respect to FE model deviations
from the quasi-static testing. This highlights the relevance of our experimental
investigations.The fatigue testing was based on the procedure of ISO 7206-4:2010. In particular, the
implant alignment, embedding, force specifications and cycle number were in
accordance. We applied the largest head offset allowed for the present hip stem
design, leading to the highest possible lever and thereby a worst-case loading.
However, the hip stem size was selected with regard to our previous study
but configurations with smaller bodies or longer necks are available. A
further limitation was the use of a reduced sample size of only three instead of six
samples.Advanced research will be required to evaluate the need of different energy
harvesting system geometries for different hip stem sizes and the influence on the
structural fatigue failure strength. All of our three samples completed the
prescribed number of load cycles and also a second run. Since no implant fracture
occurred for the additional load steps at higher forces, the required cycle number
was accomplished 3.6 times in total. Therefore, the design changes to the metallic
implant component can be considered as safe with regard to fatigue. Additional
quasi-static testing led to exceeding of the implant material’s yield point. The
plastic deformation occurred mainly at the embedding level and not in the cavity
region, as to assume for the original implant geometry. It indicates the more
critical loading at this latter position. We want to emphasise that no crack or
fracture propagated from the modified area even if the surface finish was impaired
in comparison with the original condition by the manufacturing process. In general,
a rougher surface decreases the structural fatigue failure strength.
Conclusion
Research for energy-autonomous instrumented implants mainly focusses on the
development of new concepts. Nevertheless, basic implant functionalities must be
guaranteed. Within this study, we analysed a modified hip stem for integration of a
piezoelectric energy harvesting system by experimentally investigating the
structural fatigue failure strength within a standardised testing procedure. By
means of finite element analysis we correlated the experimental results with a
numerically calculated stress level. This value may serve as reference for implant
modifications with regard to further instrumentation or design adaptation. These
should aim at increased load transmission to the piezoelectric element in order to
optimise for higher output of electrical energy.In future studies, we plan to investigate the energy conversion and fatigue behaviour
of a fully assembled total hip implant system.
Authors: W Mittelmeier; S Lehner; W Kraus; H P Matter; L Gerdesmeyer; E Steinhauser Journal: Arch Orthop Trauma Surg Date: 2003-10-31 Impact factor: 3.067
Authors: Michelle J Lespasio; Assem A Sultan; Nicolas S Piuzzi; Anton Khlopas; M Elaine Husni; George F Muschler; Michael A Mont Journal: Perm J Date: 2018
Authors: D T Davy; G M Kotzar; R H Brown; K G Heiple; V M Goldberg; K G Heiple; J Berilla; A H Burstein Journal: J Bone Joint Surg Am Date: 1988-01 Impact factor: 5.284
Authors: Sanjeev Patil; Donald S Garbuz; Nelson V Greidanus; Bassam A Masri; Clive P Duncan Journal: J Arthroplasty Date: 2007-10-23 Impact factor: 4.757
Authors: Joana Reis; Clara Frias; Carlos Canto e Castro; Maria Luísa Botelho; António Torres Marques; José António Oliveira Simões; Fernando Capela e Silva; José Potes Journal: J Biomed Biotechnol Date: 2012-05-31
Authors: Marco P Soares Dos Santos; Ana Marote; T Santos; João Torrão; A Ramos; José A O Simões; Odete A B da Cruz E Silva; Edward P Furlani; Sandra I Vieira; Jorge A F Ferreira Journal: Sci Rep Date: 2016-07-26 Impact factor: 4.379