BISS: concept and biomechanical investigations of a new screw system for electromagnetically induced internal osteostimulation.

W Mittelmeier1, S Lehner, W Kraus, H P Matter, L Gerdesmeyer, E Steinhauser.   

Abstract

INTRODUCTION: The aim of an intraosseous application of electromagnetic alternating fields is to speed up both the regeneration of osteonecroses and bone regeneration. In clinical studies, the efficiency of the technique could be successfully proven by using a transducer coil. The advantage of the traditional technique was the variety of its applications in connection with various osteosynthesis systems; the disadvantage was a possible failure of the contacting leads and the resulting functional breakdown.
MATERIALS AND METHODS: A newly developed BISS screw (bipolar induction screw system) with integrated coil and electrodes was compared to a standard cannulated screw used in the traditional technique. The strength of BISS screws ( n=6) and of cannulated screws ( n=6) was evaluated in comparative biomechanical tests. Examinations consisted of torsional and static and dynamic cantilever tests. All screws were made of the same material (TiAl(6)V(4)) and had identical outer dimensions.
RESULTS: No significantly lower strengths could be observed when we compared BISS screws with cannulated screws. The BISS screws even showed significantly higher mechanical values due to a reinforcing effect by the attached electrode.
CONCLUSION: In the modified concept of the new BISS screw, both coil and electrodes are housed in only one cannulated screw. No negative effects concerning mechanical strength and durability were associated with the new screw concept. This provides for a simpler implantation and makes removal easier, while the risk of a cable tear is avoided.

Entities:  

Mesh:

Substances:

Year:  2003        PMID: 14593486     DOI: 10.1007/s00402-003-0594-9

Source DB:  PubMed          Journal:  Arch Orthop Trauma Surg        ISSN: 0936-8051            Impact factor:   3.067


  8 in total

1.  [Arthroscopically assisted therapy of avascular necrosis of the femoral head].

Authors:  M Ellenrieder; T Tischer; P C Kreuz; S Fröhlich; A Fritsche; W Mittelmeier
Journal:  Oper Orthop Traumatol       Date:  2013-02       Impact factor: 1.154

2.  Numerical study on the effect of capacitively coupled electrical stimulation on biological cells considering model uncertainties.

Authors:  Julius Zimmermann; Richard Altenkirch; Ursula van Rienen
Journal:  Sci Rep       Date:  2022-03-18       Impact factor: 4.996

3.  Long-term stimulation with alternating electric fields modulates the differentiation and mineralization of human pre-osteoblasts.

Authors:  Franziska Sahm; Vivica Freiin Grote; Julius Zimmermann; Fiete Haack; Adelinde M Uhrmacher; Ursula van Rienen; Rainer Bader; Rainer Detsch; Anika Jonitz-Heincke
Journal:  Front Physiol       Date:  2022-09-30       Impact factor: 4.755

4.  Relationship between mechanical properties and bone mineral density of human femoral bone retrieved from patients with osteoarthritis.

Authors:  Yvonne Haba; Tobias Lindner; Andreas Fritsche; Ann-Kristin Schiebenhöfer; Robert Souffrant; Daniel Kluess; Ralf Skripitz; Wolfram Mittelmeier; Rainer Bader
Journal:  Open Orthop J       Date:  2012-10-19

5.  Magnetically induced electrostimulation of human osteoblasts results in enhanced cell viability and osteogenic differentiation.

Authors:  Bettina Hiemer; Josefin Ziebart; Anika Jonitz-Heincke; Philip Christian Grunert; Yukun Su; Doris Hansmann; Rainer Bader
Journal:  Int J Mol Med       Date:  2016-05-16       Impact factor: 4.101

6.  Continuous Electrical Stimulation Affects Initial Growth and Proliferation of Adipose-Derived Stem Cells.

Authors:  Peer W Kämmerer; Vivien Engel; Franz Plocksties; Anika Jonitz-Heincke; Dirk Timmermann; Nadja Engel; Bernhard Frerich; Rainer Bader; Daniel G E Thiem; Anna Skorska; Robert David; Bilal Al-Nawas; Michael Dau
Journal:  Biomedicines       Date:  2020-11-08

7.  Performance of a Piezoelectric Energy Harvesting System for an Energy-Autonomous Instrumented Total Hip Replacement: Experimental and Numerical Evaluation.

Authors:  Hans-E Lange; Nils Arbeiter; Rainer Bader; Daniel Kluess
Journal:  Materials (Basel)       Date:  2021-09-08       Impact factor: 3.623

8.  Endurance testing and finite element simulation of a modified hip stem for integration of an energy harvesting system.

Authors:  Hans-E Lange; Rainer Bader; Daniel Kluess
Journal:  Proc Inst Mech Eng H       Date:  2021-06-17       Impact factor: 1.617

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.