Markus Nilsson1, Greta Eklund2, Filip Szczepankiewicz3, Mikael Skorpil4,5, Karin Bryskhe2, Carl-Fredrik Westin6, Claes Lindh7, Lennart Blomqvist4,8, Fredrik Jäderling4,8,9. 1. Clinical Sciences Lund, Radiology, Lund University, Lund, Sweden. 2. Random Walk Imaging AB, Lund, Sweden. 3. Clinical Sciences Lund, Medical Radiation Physics, Lund University, Lund, Sweden. 4. Department of Molecular Medicine and Surgery, Karolinska Institutet, Stockholm, Sweden. 5. Department of Neuroradiology, Karolinska University Hospital, Solna, Stockholm, Sweden. 6. Brigham and Women's Hospital, Harvard Medical School, Boston, Massachusetts, USA. 7. Department of Oncology-Pathology, Karolinska Institutet, Stockholm, Sweden. 8. Department of Diagnostic Radiology, Karolinska University Hospital, Solna, Sweden. 9. Department of Radiology, Capio S:t Görans Hospital, Stockholm, Sweden.
Abstract
PURPOSE: Tensor-valued diffusion encoding provides more specific information than conventional diffusion-weighted imaging (DWI), but has mainly been applied in neuroimaging studies. This study aimed to assess its potential for the imaging of prostate cancer (PCa). METHODS: Seventeen patients with histologically proven PCa were enrolled. DWI of the prostate was performed with linear and spherical tensor encoding using a maximal b-value of 1.5 ms/µm2 and a voxel size of 3 × 3 × 4 mm3 . The gamma-distribution model was used to estimate the mean diffusivity (MD), the isotropic kurtosis (MKI ), and the anisotropic kurtosis (MKA ). Regions of interest were placed in MR-defined cancerous tissues, as well as in apparently healthy tissues in the peripheral and transitional zones (PZs and TZs). RESULTS: DWI with linear and spherical encoding yielded different image contrasts at high b-values, which enabled the estimation of MKA and MKI . Compared with healthy tissue (PZs and TZs combined) the cancers displayed a significantly lower MD (P < .05), higher MKI (P < 10-5 ), and lower MKA (P < .05). Compared with the TZ, tissue in the PZ showed lower MD (P < 10-3 ) and higher MKA (P < 10-3 ). No significant differences were found between cancers of different Gleason scores, possibly because of the limited sample size. CONCLUSION: Tensor-valued diffusion encoding enabled mapping of MKA and MKI in the prostate. The elevated MKI in PCa compared with normal tissues suggests an elevated heterogeneity in the cancers. Increased in-plane resolution could improve tumor delineation in future studies.
PURPOSE: Tensor-valued diffusion encoding provides more specific information than conventional diffusion-weighted imaging (DWI), but has mainly been applied in neuroimaging studies. This study aimed to assess its potential for the imaging of prostate cancer (PCa). METHODS: Seventeen patients with histologically proven PCa were enrolled. DWI of the prostate was performed with linear and spherical tensor encoding using a maximal b-value of 1.5 ms/µm2 and a voxel size of 3 × 3 × 4 mm3 . The gamma-distribution model was used to estimate the mean diffusivity (MD), the isotropic kurtosis (MKI ), and the anisotropic kurtosis (MKA ). Regions of interest were placed in MR-defined cancerous tissues, as well as in apparently healthy tissues in the peripheral and transitional zones (PZs and TZs). RESULTS: DWI with linear and spherical encoding yielded different image contrasts at high b-values, which enabled the estimation of MKA and MKI . Compared with healthy tissue (PZs and TZs combined) the cancers displayed a significantly lower MD (P < .05), higher MKI (P < 10-5 ), and lower MKA (P < .05). Compared with the TZ, tissue in the PZ showed lower MD (P < 10-3 ) and higher MKA (P < 10-3 ). No significant differences were found between cancers of different Gleason scores, possibly because of the limited sample size. CONCLUSION: Tensor-valued diffusion encoding enabled mapping of MKA and MKI in the prostate. The elevated MKI in PCa compared with normal tissues suggests an elevated heterogeneity in the cancers. Increased in-plane resolution could improve tumor delineation in future studies.
Prostate cancer (PCa) is the second most frequent cancer diagnosis in men, and the fifth leading cause of death world-wide.[1] In the Nordic countries, the lifetime risk for a man to be diagnosed with PCa is approximately 1 in 6, and the risk of dying from the disease is 1 in 30.[2] There is a large variation in cancer aggressiveness and prognosis, which poses a challenge for diagnosis and treatment planning. The introduction of MRI in combination with targeted biopsies in the diagnostic workflow has been a paradigm shift,[3,4] leading to a reduction of the number of performed biopsies. Compared with systematic biopsies alone, the use of MRI and targeted biopsies increase detection of significant cancers and reduce the overdetection of indolent low-grade cancers.[5,6]A cornerstone for the detection of significant cancers with MRI is diffusion-weighted imaging (DWI). Apart from providing images for radiological assessment, it also provides maps of potential quantitative imaging biomarkers. A simple but powerful example is the apparent diffusion coefficient (ADC), which correlates with cellularity in many different types of cancers.[7] Another is the fractional anisotropy (FA) obtained from diffusion tensor imaging (DTI).[8] This parameter is sensitive to the presence of elongated cell structures and has been proposed to improve the detection and grading of PCa.[9,10] However, previous studies found contradicting results, with FA being either higher or lower in PCas compared with normal glandular tissue.[10-12] This variation could be caused by multiple factors. One is that prostatic FA tends to decrease with echo time and increase with diffusion time,[13,14] and these imaging parameters can differ between studies. Another factor is the limited spatial resolution of MRI, which causes each imaging voxel to include cell structures with different orientations. This in itself reduces the FA[15,16] and leads to a voxel-size dependence of the FA in organs with high orientation dispersion such as the prostate.[17] To resolve this problem, anisotropy must be detected on the micrometer level, rather than on the millimeter level as is done in DTI.Tensor-valued diffusion encoding has recently been proposed as a viable approach for the estimation of microscopic diffusion anisotropy.[18-20] The approach is based on two pillars. First, it uses high b-value DWI to reveal the effects of non-Gaussian diffusion,[21] which in this context refers to the presence of multiple components undergoing approximately Gaussian diffusion. Second, it uses two different shapes of the diffusion encoding tensor to separate two contributions to the multi-Gaussian diffusion: microscopic diffusion anisotropy and isotropic heterogeneity. These contributions are quantified in terms of the anisotropic kurtosis (MKA) and isotropic kurtosis (MKI), respectively.[16,19] In brain tumors, these parameters have a link to features from histology slides[22]: between MKA and the density of elongated cell structures and between MKI and the intravoxel variation in cell density. Tensor-valued diffusion encoding has been applied to study microscopic anisotropy in a wide range of conditions, including schizophrenia,[19] brain tumors,[22,23] epilepsy,[24] and kidney variations.[25] It has also been used to learn about cerebellar structure[26] and in conjunction with motion-compensated encoding to improve DWI of the heart.[27,28] However, tensor-valued encoding has so far not been applied in the context of PCa.The purpose of this study was to test the potential of tensor-valued diffusion encoding for imaging of the prostate in patients with PCa, and to provide an initial survey of the microscopic diffusion anisotropy and isotropic heterogeneity in this organ and its tumors. Our hypothesis was that tissues of the prostate with more elongated cell structures would exhibit higher microscopic diffusion anisotropy (MKA), and that cancers that often exhibit a relatively chaotic microenvironment would exhibit higher isotropic heterogeneity (MKI) than normal-appearing tissue.
METHODS
Patient inclusion and preparation
Seventeen patients with histologically proven PCas were included in the study, after providing written informed consent to participation. The study was approved by the regional ethical review board in Stockholm (Dnr. 2013/1655-31/2). Patients underwent a prostate MRI before surgery on a Siemens 3T scanner (Siemens Magnetom, Skyra). Patients were advised to refrain from sexual activity with ejaculation 3 days before the examination to increase fluid content in the peripheral zone (PZ) and seminal vesicles. On the day of the examination, patients were instructed to fasten and refrain from nicotine-containing substances 1 h before examination to reduce bowel movements. All patients had a bowel preparation 2 h before examination using a small enema Microlax (McNeil Sweden AB). In some patients, a small catheter was introduced to evacuate gas in the rectum to minimize image artifacts. One milligram of Glucagon (Novo Nordisk Scandinavia AB) was administered intramuscularly just before the examination to reduce bowel motility. The body coil was placed over the small pelvis (Body 18; Siemens) and run together with a spine coil (Spine Coil 32; Siemens). After MRI, all patients underwent robot-assisted laparoscopic radical prostatectomy using a three-armed DaVinci system (Intuitive Surgical).
MRI and statistical analysis
MRI was performed using a T2-weighted turbo-spin–echo sequence (voxel size = 0.5 × 0.5 mm2, slice thickness = 3 mm, TR = 4590 ms, TE = 109 ms), and DWI (voxel size = 1.5 × 1.5 mm2, slice thickness = 4.8 mm, TR = 6100 ms, TE = 63 ms). In addition, DWI was performed using a prototype spin-echo echo-planar imaging sequence, customized for tensor-valued diffusion encoding with user-defined free gradient waveforms.[27] Data were acquired with two different types of gradient waveforms. These were optimized to yield linear (LTE) and spherical (STE) diffusion-encoding tensors.[29] Each waveform was used to acquire five distinct b-values (0.2, 0.5, 0.8, 1.2, and 1.5 ms/μm2) in eight different directions. Figure 1 shows the waveforms used and how they support estimation of microscopic anisotropy and isotropic heterogeneity. In total, each data set comprised 80 image volumes acquired in approximately 6 min. All images were measured using a matrix size = 128 × 128, 17 slices, in-plane voxel size = 3 × 3 mm2, slice thickness = 4 mm, TR = 4000 ms, and TE = 101 ms. Image registration was applied to correct for motion and eddy current artifacts, using ElastiX.[30] To avoid errors introduced by the registration algorithm itself, only in-plane adjustments were allowed.
FIGURE 1
Method overview. The leftmost panels show gradient waveforms used for linear tensor encoding (LTE) and spherical tensor encoding (STE) at the top and bottom, respectively. For LTE, waveforms are applied in one direction only, whereas STE waveforms are applied in all three spatial dimensions (represented by red, dark-gray, and light-gray waveforms, respectively). In addition to the diffusion-encoding waveforms, the plots also show the radiofrequency pulses (90° and 180°) and the imaging readout block (echo-planar imaging). The middle and right panels show diffusion-weighted signal curves for a synthetic voxel comprised of environments in which the diffusion is described by anisotropic but randomly ordered diffusion tensors (middle) and isotropic diffusion tensors with variable diffusivity (right). Note that the diffusion is isotropic on the voxel level in both cases. For the case with microscopically anisotropic diffusion, the curves from LTE and STE deviate at higher b-values, whereas they overlap for the other case. This provides a mechanism to disentangle microscopic anisotropy (MKA) from isotropic heterogeneity (MKI)
The data acquired with tensor-valued diffusion encoding were analyzed to yield the two different types of diffusional kurtosis supported by this acquisition.[19,31] Prior to analysis, data were averaged across the diffusion encoding directions (referred to as powder averaging).[31,32] The following equation was then fitted to the data[31]:
where S0 is the nondiffusion-weighted signal intensity, MD is the mean diffusivity, , and MKA is the anisotropic kurtosis, MKI is the isotropic kurtosis, and bΔ is the shape of the b-tensor, which is zero for STE and unity for LTE. The open-source analysis software is available at https://github.com/markus-nilsson/md-dmri,[33] (SHA-1 hash 989ebab), with this method referred to as dtd_gamma.Regions of interest (ROIs) were manually drawn on the high-resolution DWI and T2-weighted (T2W) images by an experienced radiologist (FJ), after which they were transferred to the low-resolution space of the images acquired with LTE and STE, corresponding to tumor locations on histology slides. The transform was obtained from a rigid body registration from the high- to the low-resolution space using ElastiX. Three types of ROIs were defined. The first covered the tumor lesions (one per lesion, up to three different ROIs per patient). In total, 32 histologically proven PCas were defined (7 in the transitional zone (TZ) and 25 in the PZ). The second and third covered normal-appearing regions with no tumor at the corresponding location at histology of the TZs and PZs. From each ROI, values were extracted of the three diffusion parameters (MD, MKI, MKA). The values were then assigned to one of five groups: cancers with Gleason score of 3 + 3, 3 + 4, or 4 + 3, and TZ or PZ. Three types of comparisons were performed: between each cancer type, between cancer and healthy-appearing prostate tissue (3 + 3, 3 + 4, and 4 + 3 vs TZ and PZ), and between TZ and PZ. Tests for differences in medians were done using nonparametric Wilcoxson rank-sum tests (chosen because of limited sample sizes), with a significance threshold of 0.05.
Postoperative pathology
All prostatic tissue was fixed in 10% buffered formaldehyde. The specimens were subjected to the whole-mount, step-section according to the Stanford procedure and sliced horizontally, perpendicular to the dorsal aspect of the prostate in 3- to 4-mm slices. The slicing corresponded to the angle of the transversal images on MRI, augmenting the correlation of tumor location at the base, mid, and apex. The slices were subsequently stained with hematoxylin-eosin for tissue microstructure visualization.
RESULTS
A T2-weighted image and a set of DWIs acquired at different b-values with LTE and STE are shown in Figure 2. The appearance of the LTE and STE images is similar at low b-values, as expected, whereas subtle contrast differences appear at higher b-values. The LTE and STE data were jointly analyzed to obtain the parameter maps that are shown in Figure 3. All parameters are unitless, except for MD, which is given in units of μm2/ms. Three of the maps (S0, MD, and MKI) are most intense in the PZ, whereas one (MKA) is most intense in the TZ and in the anterior fibromuscular stroma.
FIGURE 2
Overview of acquired images. The left image shows a T2-weighted transversal slice across the pelvis from the turbo-spin–echo sequence. The right image shows cropped versions of the diffusion-weighted images, where the columns show different b-values and the rows different shapes of the b-tensor. The contrast obtained with linear tensor encoding (top row) and spherical tensor encoding (bottom row) is similar at low b-values, but starts to deviate at higher b-values. The b-values are reported in units of ms/μm2. LTE, linear tensor encoding; STE, spherical tensor encoding; T2w, T2-weighted
FIGURE 3
Overview of parameter maps. Columns show fitted T2-weighted signal intensity (S0) with the tumor region of interest marked in yellow, mean diffusivity (MD in units of μm2/ms), isotropic kurtosis (MKI), and anisotropic kurtosis (MKA). High MKA indicating high microscopic diffusion anisotropy is found symmetrically around the midline. The high values of MKA in the top part of the image are probably caused by a low signal-to-noise ratio because it coincides with the low values of S0
Values of MD, MKI, and MKA for prostate cancer tissue (characterized by the Gleason score) and in healthy prostate tissue within both the PZ and the TZ are shown in Figure 4. The distributions of MKI and MKA are summarized by their mean and variance. Cancer tissues showed a significantly lower MD (P < .05), higher MKI (P < 10−5), and higher MKA (P < .05) compared with normal tissue (TZ and PZ combined). Moreover, compared with healthy tissue in the TZ, we found that the PZ showed lower MD (P < 10−3) and higher MKA (P < 10−3). No significant differences were found between tumors of different Gleason scores.
FIGURE 4
Quantitative analysis of prostatic tumors versus healthy tissue. A statistically significant difference was found between the cancers (with Gleason scores of 3 + 3, 3 + 4, and 4 + 3) and healthy tissue (peripheral zone [PZ] and transitional zone [TZ]) for mean diffusivity (MD; P < .05), isotropic kurtosis (MKI; P < 10−5), and anisotropic kurtosis (MKA; P < .05). Significant differences were found between PZ and TZ for MD (P < 10−3) and for MKA (P < 10−3). No significant differences were found between cancers with different Gleason scores
Figure 5 shows hematoxylin-eosin–stained histopathological data from cancers with values of MKI and MKA in the lowest and highest percentiles of all cases. The observed stroma surrounding the ducts in the high MKI case appears to add to the heterogeneity in diffusion coefficients compared with the low MKI case where the tissue microstructure is more uniform. In the high MKA case, the tissue appears to have a higher prevalence of elongated cellular structures compared with the case with low MKA.
FIGURE 5
Examples of histology slices. The slices were obtained from selected cases with lowest and highest values of isotropic kurtosis (MKI) and anisotropic kurtosis (MKA). Top row shows examples of cancers having low MKI to the left (Gleason score = 4 + 4, peripheral zone [PZ]) and high MKI (Gleason score = 3 + 4, transitional zone [TZ]) to the right. The higher MKI seems to be related to a higher amount of stroma surrounding the ducts, which appears to add to the heterogeneity in diffusion coefficients. Bottom row shows examples of cancers having low MKA to the left (Gleason score = 4 + 4, TZ) and high MKA to the right (Gleason score = 3 + 4, TZ). High MKA appears to have a higher prevalence of elongated cellular structures. Scale bars = 100 μm
DISCUSSION
This study aimed to test the feasibility of tensor-valued diffusion encoding in the prostate to map tissue heterogeneity and microscopic anisotropy by the isotropic and anisotropic kurtosis (MKI and MKA). Results showed plausible maps (Figure 3), where the PZ showed high values of MKI, indicating a heterogeneous microenvironment with compartments having a wide variation of diffusivities within the voxel.[22] On average, the PZ was found to have significantly higher MKA than the TZ (Figure 3), but as we can see in Figure 3 the transitional zone showed a mix of high and low values of MKA. Higher values of MKA are expected where there is a larger presence of elongated cell structures.[22] In the prostate, regions with stromal smooth muscle have high diffusion anisotropy on the microscopic level.[17] However, this only leads to high values of the FA obtained with DTI if the resolution is high enough—otherwise, the high dispersion in orientation between muscle fibers in the voxel leads to low values of FA.[17] Using tensor-valued encoding, we can detect and quantify the level of microscopic diffusion anisotropy even in the presence of orientation dispersion.[16,31] Thus, we interpret regions of the TZ with high MKA as regions with high amounts of smooth muscle. In the cancers, histology shows how discrete well-formed glands are replaced with poorly formed and fused glands as the cancers progress from Gleason pattern 3 to pattern 4.[34] This can be seen as a progression towards a more disorganized and heterogeneous tissue, which is aligned with the finding of increased MKI in the cancers. However, predicting MRI parameters based on histology is no easy task because tissue components contribute not by their volume fractions alone but rather by their T2-weighted signal fractions.[13] Future studies could combine tensor-valued encoding with acquisitions featuring multiple echo times to further increase specificity,[35] which could be particularly useful because diffusion-relaxation correlation experiments can improve prostate characterization.[36,37]The question of diagnostic relevance is whether the cancers exhibited characteristics that were different from those of the healthy tissue. Indeed, we found lower MD, higher MKI and higher MKA in the cancers, with the strongest effect found for MKI (Figure 4). Some of the cancers had values of MKI that were considerably higher than any of the values in PZ and TZ. Future studies with larger sample sizes could assess the receiver-operating characteristic curve for using MKI to distinguish cancers. Previous studies have shown that using diffusional kurtosis imaging—a method similar to the one used herein but that can only assess the sum of the two kurtosis components MKA and MKI—may provide increased value compared with standard DWI in PCa assessment.[38,39] Another study did, however, not find reasons to replace DWI as the clinical standard to characterize PCa.[40]Note that we report MD values higher than expected from previous studies.[41] There are at least four different factors that could explain this: the data analysis model, the image resolution, the effective diffusion time, and the relaxation weighting. First, the data analysis was based on the so-called gamma model, which tends to overestimate the MD—in particular in the presence of high MKI.[42] A separate analysis of the low b-value part of our data using a conventional monoexponential ADC analysis corroborates this, as it yielded values on average 25% lower than with the gamma model. In an analysis of prostate tissue, Hector et al similarly found that DKI analysis—comparable to the gamma model—yielded approximately 25% higher values of MD (referred to as ADC) than a monoexponential analysis of low b-value data.[43] Second, the image resolution was low in this preliminary study (3 × 3 × 4 mm3). This limited the accuracy by which tumors can be delineated and increased the level of partial volume artifacts. As PC has low MD, partial inclusion of healthy tissue in the ROI increases the observed MD of the cancers. Third, the MD depends on the diffusion time of the gradient waveforms used for diffusion encoding. Shorter diffusion times are associated with higher MD values.[14] The gradient waveforms used in b-tensor encoding have no well-defined diffusion time,[20] however, as they use free gradient waveforms (Figure 1). Analysis of time-dependent diffusion can nevertheless be made using so-called spectral theory.[44] The waveforms we use tend to oscillate as the q-vector builds the b-tensor (Figure 1),[19] which means more energy at high frequencies. This corresponds to shorter diffusion times. How to efficiently design gradient waveforms to yield a given b-tensor while featuring specific diffusion times and high efficiency is still an open question,[20,45] but will be essential to consider when planning future b-tensor encoding studies. Fourth—and last but certainly not least—we need to consider the TE. The prostate consists of both cellular and luminal tissues, which have different T2 relaxation times. At our TE of 101 ms, the luminal component is much less attenuated than the cellular component because of differences in T2 relaxation times (~300 ms vs 50 ms).[13] As the T2 and the ADC of the luminal component are both higher than in the cellular component, the voxel-averaged ADC increases with TE.[46] We can similarly expect a dependence of TE in MKA and MKI, but this remains to be investigated. In this context we must also note that b-tensor encoding is less efficient than conventional encoding (lower b-values per unit time), which necessitates longer TEs. This leads to lower signal-to-noise ratio, which can be compensated by reducing the resolution at the expense of accurate tumor delineation. Longer echo times also affect the magnitude by which tissue components contribute to the signal. The impact of the tradeoffs between maximal b-value, resolution, and echo times in the context of b-tensor encoding in the prostate needs to be carefully analyzed, ideally using an MRI system with high-performance gradients which can acquire data at short echo times with high b-values.[47]This study had four main limitations. First, it included only 17 subjects. This limits the statistical power for detection of differences between cancers of different Gleason scores. Future studies with larger sample sizes are needed to assess the diagnostic potential of b-tensor encoding in PCa. Second, the present protocol featured voxel sizes of 3 × 3 × 4 mm3, but higher resolution is warranted for more precise delineation of the lesions. A preliminary study suggests that 2 × 2 × 4 mm3 is achievable at a system with 80 mT/m gradients.[48] Higher resolution gives less partial volume effects and thus reduce parameter bias. Third, the gradient waveforms applied here were not Maxwell compensated.[49] This can lead to a negative bias in MKI in locations far from the isocenter where the concomitant fields are larger. This probably had at most a mild impact on the present study because the prostate was positioned close to the isocenter. Fourth, zoomed-in sections of the histology sections shown in Figure 5 were obtained in regions manually selected to match the position of the MRI ROIs, so we lack voxel-to-voxel histology that is needed for careful assessment of what microstructure features that explain variations in diffusion MRI parameters.[50] However, the histology results illustrated in Figure 5 provide an initial hypothesis about what causes variation in MKA and MKI.
CONCLUSION
We found that it was feasible to perform tensor-valued diffusion encoding in the prostate for the mapping of MKA and MKI. We observed an elevated MKI in PCa compared with normal tissues in the PZs and TZs. This suggests a potential of this parameter to support a PCa diagnosis; however, this hypothesis needs to be validated in a study with a larger sample size. Future studies would benefit from the use of an increased in-plane resolution for increased accuracy in the tumor delineation.
Authors: Stefan Klein; Marius Staring; Keelin Murphy; Max A Viergever; Josien P W Pluim Journal: IEEE Trans Med Imaging Date: 2009-11-17 Impact factor: 10.048
Authors: Aritrick Chatterjee; Roger M Bourne; Shiyang Wang; Ajit Devaraj; Alexander J Gallan; Tatjana Antic; Gregory S Karczmar; Aytekin Oto Journal: Radiology Date: 2018-02-02 Impact factor: 11.105
Authors: Filip Szczepankiewicz; Danielle van Westen; Elisabet Englund; Carl-Fredrik Westin; Freddy Ståhlberg; Jimmy Lätt; Pia C Sundgren; Markus Nilsson Journal: Neuroimage Date: 2016-07-20 Impact factor: 6.556
Authors: Björn Lampinen; Filip Szczepankiewicz; Johan Mårtensson; Danielle van Westen; Oskar Hansson; Carl-Fredrik Westin; Markus Nilsson Journal: Magn Reson Med Date: 2020-03-06 Impact factor: 4.668
Authors: Fabio Nery; Filip Szczepankiewicz; Leevi Kerkelä; Matt G Hall; Enrico Kaden; Isky Gordon; David L Thomas; Chris A Clark Journal: Magn Reson Med Date: 2019-06-26 Impact factor: 4.668
Authors: Stamatios N Sotiropoulos; Saad Jbabdi; Junqian Xu; Jesper L Andersson; Steen Moeller; Edward J Auerbach; Matthew F Glasser; Moises Hernandez; Guillermo Sapiro; Mark Jenkinson; David A Feinberg; Essa Yacoub; Christophe Lenglet; David C Van Essen; Kamil Ugurbil; Timothy E J Behrens Journal: Neuroimage Date: 2013-05-20 Impact factor: 6.556
Authors: Maryam Afzali; Lars Mueller; Ken Sakaie; Siyuan Hu; Yong Chen; Filip Szczepankiewicz; Mark A Griswold; Derek K Jones; Dan Ma Journal: Magn Reson Med Date: 2022-06-17 Impact factor: 3.737