Literature DB >> 31243814

In vivo demonstration of microscopic anisotropy in the human kidney using multidimensional diffusion MRI.

Fabio Nery1, Filip Szczepankiewicz2,3,4, Leevi Kerkelä1, Matt G Hall1,5, Enrico Kaden6, Isky Gordon1, David L Thomas7,8, Chris A Clark1.   

Abstract

PURPOSE: To demonstrate the feasibility of multidimensional diffusion MRI to probe and quantify microscopic fractional anisotropy (µFA) in human kidneys in vivo.
METHODS: Linear tensor encoded (LTE) and spherical tensor encoded (STE) renal diffusion MRI scans were performed in 10 healthy volunteers. Respiratory triggering and image registration were used to minimize motion artefacts during the acquisition. Kidney cortex-medulla were semi-automatically segmented based on fractional anisotropy (FA) values. A model-free analysis of LTE and STE signal dependence on b-value in the renal cortex and medulla was performed. Subsequently, µFA was estimated using a single-shell approach. Finally, a comparison of conventional FA and µFA is shown.
RESULTS: The hallmark effect of µFA (divergence of LTE and STE signal with increasing b-value) was observed in all subjects. A statistically significant difference between LTE and STE signal was found in the cortex and medulla, starting from b = 750 s/mm2 and b = 500 s/mm2 , respectively. This difference was maximal at the highest b-value sampled (b = 1000 s/mm2 ) which suggests that relatively high b-values are required for µFA mapping in the kidney compared to conventional FA. Cortical and medullary µFA were, respectively, 0.53 ± 0.09 and 0.65 ± 0.05, both respectively higher than conventional FA (0.19 ± 0.02 and 0.40 ± 0.02).
CONCLUSION: The feasibility of combining LTE and STE diffusion MRI to probe and quantify µFA in human kidneys is demonstrated for the first time. By doing so, we show that novel microstructure information-not accessible by conventional diffusion encoding-can be probed by multidimensional diffusion MRI. We also identify relevant technical limitations that warrant further development of the technique for body MRI.
© 2019 The Authors Magnetic Resonance in Medicine published by Wiley Periodicals, Inc. on behalf of International Society for Magnetic Resonance in Medicine.

Entities:  

Keywords:  diffusion; fractional anisotropy; kidney; microscopic anisotropy; spherical tensor encoding; tensor-valued diffusion encoding

Mesh:

Year:  2019        PMID: 31243814      PMCID: PMC6988820          DOI: 10.1002/mrm.27869

Source DB:  PubMed          Journal:  Magn Reson Med        ISSN: 0740-3194            Impact factor:   4.668


INTRODUCTION

Diffusion tensor imaging (DTI) is a widely used diffusion-weighted imaging approach that has been successful because of its sensitivity to changes in tissue microstructure. However, tissue that is heterogeneous on a sub-voxel scale cannot be fully captured by the single diffusion tensor in DTI, meaning that the information on microscopic anisotropy and orientation dispersion is lost. The problem is not solely because of poor signal representation or tissue modelling but depends also on the diffusion encoding technique itself. If an acquisition is limited to a single diffusion encoding b-tensor shape, such as conventional linear tensor encoding (LTE), situations arise where vastly different tissue structures may yield virtually identical signal behavior.[1-3] In this setting, fractional anisotropy (FA) estimates are confounded by microscopic anisotropy and orientation dispersion, making their interpretation ambiguous.[4,5] Rotation invariant diffusion data (whose signal attenuation does not vary on rotations) can be acquired using isotropic diffusion encoding[6,7] (also referred to as spherical tensor encoding [STE][8]). Recent advances in the design of multidimensional diffusion MRI (MD-dMRI) acquisition schemes,[9] including the use of optimized gradient waveforms,[10] have enabled efficient spherical tensor encoding (STE) on clinical systems.[3] A joint analysis of LTE and STE provides more specific information on the underlying distribution of diffusion tensors compared to what is possible using LTE alone, allowing the effects of microscopic anisotropy and orientation dispersion to be disentangled and therefore to measure microscopic diffusion anisotropy (μFA) independently of orientation dispersion.[2,11-13] More generally, μFA can be probed in acquisitions where diffusion encoding is performed using measurement b-tensors with different shapes.[8,9] In addition to STE, double-diffusion encoding methods,[14-19] which encode diffusion along 2 directions before readout, provide planar b-tensors that also allow quantification of μFA.[20] Contrast-agent free quantitative biomarkers are actively being sought-after in the field of renal MRI.[21-25] In diffusion tensor imaging (DTI),[26] FA has been widely used in the kidneys as a non-invasive probe of tubular integrity[22,27] and the geometric arrangement of microscopic structure. Several studies consistently demonstrate a greater degree of anisotropy in the kidney medulla compared to the cortex (Kataoka et al,[28] Kido et al,[29] and Chan et al,[30] among others). A significant decrease in FA has been found in chronic kidney disease patients compared to healthy controls.[31] Furthermore, DTI has been used to assess renal allograft function early following transplantation[32] and has shown reduced medullary FA and FA-based cortico–medullary differentiation in renal transplant recipients with impaired allograft function compared to those with good or moderate function.[33] However, FA remains unable to differentiate between different pathophysiological mechanisms underlying renal dysfunction.[22] A recent study on pediatric renal allografts has shown a moderate correlation of medullary FA with several Banff histopathology scores (particularly at higher b-values) but not cortical FA.[34] Neither cortical nor medullary FA correlated with the glomerulitis (g) Banff score. This suggests that more specific biomarkers of renal microstructure are desirable. However, at the time of writing, and to the best of our knowledge, no studies have been reported investigating microscopic anisotropy in the renal parenchyma.[35,36] The aim of this work is to investigate, for the first time, the feasibility of using STE in combination with conventional LTE in the human kidney to probe and quantify tissue microscopic anisotropy in vivo. A detailed analysis of the LTE and STE diffusion-weighted signal in healthy subjects has been performed, followed by quantification of cortical and medullary μFA using a single-shell approach, along with a comparison to conventional FA, and an overview of current challenges in MD-dMRI of the kidneys.

METHODS

Ten healthy volunteers (age 31 ± 6 y, 5 male) were scanned on a 3T Prisma MR system (Siemens Healthineers, Erlangen, Germany) using a prototype spin-echo sequence with EPI readout that facilitates diffusion encoding with variable b-tensor shapes (Figure 1). Data was acquired with LTE and STE using FOV = 288 × 288 mm2, voxel size = 3 × 3 × (4–4.6) mm3, 11 coronal oblique slices, TE = 87 ms; 3/4 partial-Fourier and parallel imaging in-plane acceleration R = 2 (GRAPPA). TR was 3000 ms (following a consensus recommendation)[37] to ensure significant T1 recovery between volumes; the data acquisition window was limited to 1500 ms to reduce motion artefacts. Furthermore, setting the minimum TR to 3000 ms also minimizes the likelihood of triggering events not related to respiration (i.e., no more than 1 triggering event per respiratory cycle is allowed).[38] Diffusion data was acquired at 4 encoding strengths (b-values of 250, 500, 750, 1000 s/mm2), each with 2 signal averages, in addition to a non-diffusion-weighted reference scan (b = 0 s/mm2) repeated 4 times. The number of encoding directions was chosen such that a rotation invariant powder signal could be obtained with the acquisition protocol. The minimum number of directions necessary to fulfil this requirement was estimated by following a previously proposed simulation framework.[39] Most renal DTI studies have found FA in the renal medulla to be (1) lower than 0.5, and (2) higher than cortical FA (see above). Therefore, assuming a maximum FA of 0.5 (see Results for details of our FA estimates in this data set), ~11 directions yield a rotation invariant powder signal for b × MD < 3,[39] where MD = mean diffusivity. As such, considering the highest b-value used in this study (1000 s/mm2), ~11 directions yield a rotation invariant powder signal for a MD of 3 × 10−3 mm2/s (diffusion coefficient of water at body temperature),[40] which is above the typically observed MD in the renal cortex and medulla[41-43] (also verified in our data, see Results). The number of encoding directions was therefore set to 12 for all LTE and STE acquisitions. Encoding waveforms were optimized numerically[10] assuming a maximum gradient amplitude of 75 mT/m and slew rate of 100 T/m/s, heat dissipation factor η = 0.6, and using the Euclidean norm to obtain rotatable waveforms[39] (Figure 1). The LTE waveform was generated from the STE waveform such that the magnitude of the q-vector remained constant.[44,45] Respiratory triggering (at end-expiration) was used. Experience with respiratory-triggered acquisitions on our system suggests that using a short trigger delay increases the likelihood of the initial portion of the image readout occurring during the quiescent part of the respiratory cycle. This is consistent with a previous study.[28] The respiratory triggering parameters used in this work were threshold 20% and trigger delay 200 ms. The nominal scan time (i.e., without accounting for the delays associated with respiratory triggering) was 11 min 00 s (5 min 30 s for each of the encoding schemes). The time penalty associated with using respiratory triggering compared to non-triggered acquisitions was investigated (see Results). Retrospective motion correction (image registration) of individual diffusion-weighted volumes to the reference b = 0 s/mm2 data before powder averaging was performed separately for each kidney using elastix.[46] The whole kidney parenchyma (excluding the renal hilum) was manually segmented in the mean b = 0 s/mm2 image. To reduce operator bias, subsequent cortical-medullary regions of interest (ROIs) were segmented using an automatic algorithm developed in-house. Briefly, it uses Gaussian mixture modeling to determine an optimal subject-wise threshold to separate cortex from medulla based on conventional FA voxel-wise estimates (Figure 2), followed by a 2D morphological opening operation to remove spurious voxels in the resulting ROIs. Multi-shell diffusion tensor fitting was performed with FSL[47] to obtain voxel-wise conventional FA and MD estimates using all b-values in the LTE data set. Quantification of μFA was performed using a single-shell approach, using the highest sampled b-value (1000 s/mm2) with the following equation (see Lasič et al[2]) where , the difference in scaled variance of apparent diffusion coefficients, is given by where ELTE and ESTE are, respectively, the powder-averaged LTE and STE signal normalized to the b = 0 s/mm2 reference scan, is the mean diffusivity and b is the encoding strength. First, the feasibility of voxel-wise μFA mapping was assessed by determining the proportion of voxels for which μFA could be not quantified (i.e., because of a negative LTE-STE difference). Subsequently, to maximize the SNR of the LTE-STE difference and therefore the reliability of the μFA estimation, a ROI-based approach was used. Here, LTE and STE powder averaged signal was separately averaged within the cortical and medullary ROIs before μFA quantification in each tissue type and in all subjects.
FIGURE 1

Schematic of the spin-echo EPI sequence with optimized gradient waveforms to yield spherical and linear b-tensor encoding (STE and LTE).[10] Note that STE and LTE measurements are performed separately (i.e., require separate RF excitation pulses)

FIGURE 2

Fractional anisotropy (FA)-based automatic segmentation of cortex and medulla on a single representative subject (after manual segmentation of kidney parenchyma to exclude the hilum). A Gaussian mixture model (2 components) is fitted to the histogram of the FA values of the kidney parenchyma (obtained from the LTE acquisition) and a FA threshold to separate cortex from medulla is obtained from the intersection of the 2 resulting Gaussian distributions (green marker and dashed line). Note that the value of the FA threshold indicated in this figure is specific for this particular subject. Even though a single central slice is shown, 3 central slices are used to obtain the FA histogram and subsequent regions of interest (ROIs). (A) Non-diffusion weighted (b = 0 s/mm2) image. (B) Cropped renal parenchyma on the b = 0 s/mm2 image. (C) FA map. (D) FA intensity histogram and result of Gaussian mixture model fitting and subsequent estimation of the FA threshold to segment cortex–medulla. (E) Cortical ROIs. (F) Masked cortex in the FA map. (G) Medulla ROI. (H) Masked medulla in the FA map

RESULTS

Based on visual inspection of SNR and image artefacts, all LTE and STE data had sufficiently high quality for analysis, and the 3 central slices from each subject were used for all further analyses. The true scan time (both encoding schemes), including delays because of respiratory triggering and subject-specific respiratory rates was 17 ± 4 min (average ± SD). Respiratory triggering significantly increased scan time (approximately doubling it in 1 subject) compared to a non-triggered acquisition. However, respiratory triggering has been shown to improve data quality in DTI of native kidneys even when image registration is used[48] which motivated its use in this study. The conventional FA threshold calculated via Gaussian mixture modeling to differentiate cortex from medulla was 0.28 ± 0.02 (mean ± SD across all subjects). All cortical and medullary ROIs obtained by the proposed segmentation approach were reviewed by 2 authors with 5 and 35 y of experience in renal imaging (the latter being a radiologist) and were deemed anatomically accurate (see Figure 2E-H for an example on a representative subject). This supports the initial assumption that FA could be used to separate cortex and medulla (at least in healthy kidneys). This is consistent with most renal DTI publications that have demonstrated medullary FA to be significantly higher than cortical FA in healthy volunteers.[22] The FA estimates using the resulting ROIs were 0.19 ± 0.02 and 0.40 ± 0.03, respectively, for kidney cortex and medulla. Mean cortical and medullary MD were, respectively, 1.84 × 10−3 ± 0.08 × 10−3 mm2/s and 1.77 × 10−3 ± 0.09 × 10−3 mm2/s (2-tailed paired t-test, P < 5 × 10−4), also reflecting the expected higher diffusivity in the renal cortex compared to the medulla in healthy subjects.[22] Figure 3A and B show the dependence of the LTE and STE powder average signal, and the normalized signal difference (i.e., (Signal(LTE) – Signal(STE))/Signal(LTE)), on b-value for medulla and cortical ROIs averaged across the 10 subjects. The hallmark of microscopic anisotropy (μFA > 0) is divergence of the STE and LTE signals with increasing b-value.[2] Figure 3B highlights the clear trend of increasing relative difference between the LTE and STE signals with increasing b-value in both kidney parenchyma regions (cortex and medulla). This effect was observed individually in each of the subjects (see Supporting Information Table S1). Statistically significant LTE-STE signal differences (two-tailed paired t-test, P < 0.05) were found from b = 500 s/mm2 in the medulla and b = 750 s/mm2 in the cortex. The relative LTE-STE difference was greatest at b = 1000 s/mm2 (the highest b-value sampled in this study) for both medulla (22 ± 3%; P < 10−5) and cortex (15 ± 5%; P < 10−4) (see Supporting Information Table S1). This suggests that relatively high b-values (compared to most renal DTI studies) are required to capture microscopic anisotropy information in the kidneys, particularly in the cortex. Figure 4 shows a comparison of conventional FA and LTE-STE relative difference maps with increasing b-value on a single representative subject. Excellent cortico–medullary differentiation is seen in the FA maps, whereas the LTE-STE relative difference is shown with a more homogenous intensity distribution throughout the renal parenchyma. In our voxel-wise analysis, a lack of divergence of LTE and STE signal consistent with the hallmark of μFA (i.e., a positive LTE-STE difference) resulted in a failure to estimate μFA using the single-shell method in a non-trivial proportion of voxels (5 ± 9% in the cortex and 2 ± 4% in the medulla [median ± interquartile range]). Across all subjects, ROI-based cortical and medullary μFA were, respectively, 0.53 ± 0.09 and 0.65 ± 0.05 (two-tailed paired t-test, P < 10−4) (individual results for each subject shown in Supporting Information Table S2). Furthermore, μFA was shown to be significantly larger than conventional FA in both the cortex and medulla (2-tailed paired t-tests, P < 10−5). Figure 5 shows a comparison of conventional FA and μFA maps in a central slice of 2 subjects. These are, respectively, examples of worst- and best-case scenarios in terms of image quality of the μFA maps (chosen on the basis of the highest and lowest proportion of μFA calculation fails in the cortex). The lower cortico-medullary differentiation in the μFA map compared to the standard FA in Figure 5 (mostly because of higher μFA vs. standard FA in the cortex) suggests that orientation dispersion in the cortex enhances the cortico–medullary differentiation seen in conventional DTI FA measures.
FIGURE 3

(A) Signal versus b-value averaged across 10 subjects within cortical and medullary regions of interest (ROIs). (B) Highlights the trend of increasing relative difference between the LTE and the STE signal with increasing b-value. The P-value corresponding to the statistically significant differences between the LTE and the STE signal is shown in blue (2-tailed paired t-test)

FIGURE 4

Non-diffusion-weighted (top), fractional anisotropy (FA) and LTE-STE relative difference images in 3 kidney slices of 1 subject (slice 2: anterior; slice 4: central; slice 6: posterior). The relative difference images are shown with a constant intensity scale across all b-values (ranging from 0–40%)

FIGURE 5

Central slice b = 0 s/mm2 reference image, conventional FA map and μFA map (left to right) for 2 subjects (rows, respectively the worst and best-case scenarios as judged by the proportion of cortical μFA calculation failures across the 10 subjects). Both conventional FA and μFA range from [0, 1]. However, for figure displaying purposes the intensity range of the conventional FA map is set to [0, 0.7]. Note that the b = 0 s/mm2 image shows the whole kidneys whereas for both quantitative maps the hilum was removed to avoid biasing any FA or μFA estimates. An artificial boundary (cyan) depicting the boundary of the kidneys (excluding hilum) was added to the μFA map for easy visualization of voxels where μFA calculation was not possible (shown as dark regions in the grayscale color map). Note that even though voxel-wise calculations of μFA were necessary to generate the maps in this figure, the reported μFA values (text) were obtained using the ROI-based approach. See results for all subjects in Supporting Information Table S2

DISCUSSION

This work provides evidence that MD-dMRI methods are capable of probing microscopic anisotropy in the human kidneys in vivo and provides information on the lower bounds of the range of b-values required to map it. Similarly to previous studies,[48,49] the use of retrospective image registration improved data quality by reducing motion artefacts and enhancing the cortico–medullary differentiation. The hallmark effect of microscopic anisotropy (divergence between LTE and STE signal with increasing b-value) is demonstrated in the renal cortex and more predominantly in the medulla without the need for any model assumptions or fitting. Mapping μFA in the kidneys in a voxel-wise manner was found challenging in a subset of subjects, particularly in patches of cortical tissue (as shown in Figure 5). This is consistent with previous work outside the kidneys that has shown estimation of μFA to be problematic in tissue with low intrinsic μFA.[2,45] This was a minor issue in the medulla with <5% of μFA calculation failures in the majority of subjects (see Supporting Information Table S2). Nevertheless, to overcome this, we resorted to a ROI-based approach where quantification of μFA in all subjects in both cortex and medulla was successful. One might hypothesize use of μFA to help elucidate the pathophysiological mechanisms that decrease diffusion anisotropy in the kidneys.[33,50] Indeed, where orientation dispersion is a plausible feature of tissue microstructure (e.g., convoluted tubules), μFA may provide a more accurate assessment of microstructural integrity compared to conventional FA. This hypothesis will require future clinical studies, whereas the present study focused on establishing the feasibility of measuring μFA and obtaining baseline data from healthy volunteers. Several limitations must be addressed in future research. First, the waveforms used here are not compensated with respect to concomitant fields that may cause a hyper-attenuation of the STE and potentially yield overestimation of the LTE-STE difference signal as recently reported.[51] Second, the contribution of flow on anisotropy measures has not been investigated in this work. This would be required to disentangle fast pseudo-diffusion effects due to microscopic capillary and/or tubular flow from passive diffusion effects from which tissue microstructure properties can be estimated. This is especially relevant in the kidneys and in the cortex in particular where one might expect the convoluted tubule (through which flow is significant) to be a candidate for locally anisotropic diffusion. Using flow-compensation techniques may provide the additional benefit of compensating for spin dephasing (and therefore signal loss) due to bulk motion effects which may improve image quality.[52] Furthermore, a potential limitation is that LTE and STE waveforms were not matched with respect to the diffusion time, which may cause a parameter bias in systems with diffusion time dependencies.[53] An investigation of time-dependency of our diffusion measurements was beyond the scope of this work and is frequently overlooked in the renal diffusion MR literature. Our estimate for mean diffusivity (~1.8 × 10−3 mm2/s) suggests mean displacements of ~11–16 μm for diffusion times on the order of 35–70 ms (note the time elapsed between the start and end of the diffusion encoding gradients is ~70 ms). The tubular diameters within the human kidney nephron components range in sizes from 10–70 μm (tubular diameter).[54,55] Therefore, an overlap between the sizes of the diffusion-restricting structures and the diffusion propagator is expected to a considerable extent. This suggests that a diffusion-time dependency (i.e., mean diffusivity not being independent from the gradient waveform) is likely to exist and this should be investigated in future studies. Finally, for clinical applications where scan time is limited, optimization of parsimonious time-efficient protocols is warranted.

CONCLUSION

This work demonstrates the technical feasibility of tensor-valued diffusion encoding for renal imaging and provides pilot data to demonstrate the minimum b-values required to probe microscopic fractional anisotropy. We then harness this information to provide the first estimation for microscopic anisotropy in human kidney in vivo. Finally, we highlight current limitations, motivating further investigation of the microstructural information offered by multidimensional diffusion MRI for renal applications. This approach may ultimately enable a more specific in vivo characterization of the microstructure of human kidney in healthy and diseased subjects, compared to methods based on conventional diffusion encoding. TABLE S1 Relative signal difference (%) between LTE and STE signal in cortical and medullary ROIs, for all b-values and subjects, calculated as ((SLTE – SSTE)/SLTE) × 100, where SLTE and SSTE are, respectively, the mean LTE and STE signal in the ROIs. The last row shows the P-value row corresponding to the differences between the LTE and the STE signal (2-tailed paired t-test). Significant differences (P < 0.05) are underlined TABLE S2 Percentage of voxels for which μFA calculation was not possible (μFA “fails”). ROI-based μFA and conventional FA results shown individually for all subjects, for cortex and medulla
  47 in total

1.  Noninvasive bipolar double-pulsed-field-gradient NMR reveals signatures for pore size and shape in polydisperse, randomly oriented, inhomogeneous porous media.

Authors:  Noam Shemesh; Evren Ozarslan; Tal Adiri; Peter J Basser; Yoram Cohen
Journal:  J Chem Phys       Date:  2010-07-28       Impact factor: 3.488

2.  Comparison of physiological triggering schemes for diffusion-weighted magnetic resonance imaging in kidneys.

Authors:  Tobias Binser; Harriet C Thoeny; Ute Eisenberger; Alto Stemmer; Chris Boesch; Peter Vermathen
Journal:  J Magn Reson Imaging       Date:  2010-05       Impact factor: 4.813

3.  elastix: a toolbox for intensity-based medical image registration.

Authors:  Stefan Klein; Marius Staring; Keelin Murphy; Max A Viergever; Josien P W Pluim
Journal:  IEEE Trans Med Imaging       Date:  2009-11-17       Impact factor: 10.048

4.  Multiple wave-vector extensions of the NMR pulsed-field-gradient spin-echo diffusion measurement.

Authors: 
Journal:  Phys Rev B Condens Matter       Date:  1995-06-01

5.  Comparison of Turbo Spin Echo and Echo Planar Imaging for intravoxel incoherent motion and diffusion tensor imaging of the kidney at 3Tesla.

Authors:  Fabian Hilbert; Tobias Wech; Henning Neubauer; Simon Veldhoen; Thorsten Alexander Bley; Herbert Köstler
Journal:  Z Med Phys       Date:  2017-04-12       Impact factor: 4.820

6.  Arterial spin labelling MRI to measure renal perfusion: a systematic review and statement paper.

Authors:  Aghogho Odudu; Fabio Nery; Anita A Harteveld; Roger G Evans; Douglas Pendse; Charlotte E Buchanan; Susan T Francis; María A Fernández-Seara
Journal:  Nephrol Dial Transplant       Date:  2018-09-01       Impact factor: 5.992

7.  Magnetic resonance diffusion tensor imaging for evaluation of histopathological changes in a rat model of diabetic nephropathy.

Authors:  Katja Hueper; Dagmar Hartung; Marcel Gutberlet; Faikah Gueler; Holger Sann; Bettina Husen; Frank Wacker; Dania Reiche
Journal:  Invest Radiol       Date:  2012-07       Impact factor: 6.016

Review 8.  From single-pulsed field gradient to double-pulsed field gradient MR: gleaning new microstructural information and developing new forms of contrast in MRI.

Authors:  Noam Shemesh; Evren Ozarslan; Michal E Komlosh; Peter J Basser; Yoram Cohen
Journal:  NMR Biomed       Date:  2010-08       Impact factor: 4.044

9.  The link between diffusion MRI and tumor heterogeneity: Mapping cell eccentricity and density by diffusional variance decomposition (DIVIDE).

Authors:  Filip Szczepankiewicz; Danielle van Westen; Elisabet Englund; Carl-Fredrik Westin; Freddy Ståhlberg; Jimmy Lätt; Pia C Sundgren; Markus Nilsson
Journal:  Neuroimage       Date:  2016-07-20       Impact factor: 6.556

Review 10.  Diffusion-weighted imaging outside the brain: Consensus statement from an ISMRM-sponsored workshop.

Authors:  Bachir Taouli; Ambros J Beer; Thomas Chenevert; David Collins; Constance Lehman; Celso Matos; Anwar R Padhani; Andrew B Rosenkrantz; Amita Shukla-Dave; Eric Sigmund; Lawrence Tanenbaum; Harriet Thoeny; Isabelle Thomassin-Naggara; Sebastiano Barbieri; Idoia Corcuera-Solano; Matthew Orton; Savannah C Partridge; Dow-Mu Koh
Journal:  J Magn Reson Imaging       Date:  2016-02-19       Impact factor: 4.813

View more
  7 in total

1.  Nonparametric 5D D-R2 distribution imaging with single-shot EPI at 21.1 T: Initial results for in vivo rat brain.

Authors:  Jens T Rosenberg; Samuel C Grant; Daniel Topgaard
Journal:  J Magn Reson       Date:  2022-06-15       Impact factor: 2.734

2.  Mapping kidney tubule diameter ex vivo by diffusion MRI.

Authors:  Darya Morozov; Neda Parvin; Jennifer R Charlton; Kevin M Bennett
Journal:  Am J Physiol Renal Physiol       Date:  2021-03-15

3.  Mapping prostatic microscopic anisotropy using linear and spherical b-tensor encoding: A preliminary study.

Authors:  Markus Nilsson; Greta Eklund; Filip Szczepankiewicz; Mikael Skorpil; Karin Bryskhe; Carl-Fredrik Westin; Claes Lindh; Lennart Blomqvist; Fredrik Jäderling
Journal:  Magn Reson Med       Date:  2021-05-31       Impact factor: 3.737

4.  Motion-compensated b-tensor encoding for in vivo cardiac diffusion-weighted imaging.

Authors:  Samo Lasič; Filip Szczepankiewicz; Erica Dall'Armellina; Arka Das; Christopher Kelly; Sven Plein; Jürgen E Schneider; Markus Nilsson; Irvin Teh
Journal:  NMR Biomed       Date:  2019-11-25       Impact factor: 4.044

5.  Motion-compensated gradient waveforms for tensor-valued diffusion encoding by constrained numerical optimization.

Authors:  Filip Szczepankiewicz; Jens Sjölund; Erica Dall'Armellina; Sven Plein; Jürgen E Schneider; Irvin Teh; Carl-Fredrik Westin
Journal:  Magn Reson Med       Date:  2020-10-13       Impact factor: 4.668

6.  MR Fingerprinting with b-Tensor Encoding for Simultaneous Quantification of Relaxation and Diffusion in a Single Scan.

Authors:  Maryam Afzali; Lars Mueller; Ken Sakaie; Siyuan Hu; Yong Chen; Filip Szczepankiewicz; Mark A Griswold; Derek K Jones; Dan Ma
Journal:  Magn Reson Med       Date:  2022-06-17       Impact factor: 3.737

7.  Evidence for microscopic kurtosis in neural tissue revealed by correlation tensor MRI.

Authors:  Rafael Neto Henriques; Sune N Jespersen; Noam Shemesh
Journal:  Magn Reson Med       Date:  2021-07-30       Impact factor: 3.737

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.