AIM: The objective of this study was to evaluate skin dose based on retrofit prototype multileaf collimators (MLCs), designed for cobalt-60 teletherapy machine. Since patient's skin is sensitive to radiation, evaluation of skin dose is of utmost importance for investigating the risk of late effects.Entities:
Keywords: Cobalt-60 teletherapy machine; multileaf collimators; radiochromic films; surface dose
Year: 2021 PMID: 33953496 PMCID: PMC8074722 DOI: 10.4103/jmp.JMP_25_20
Source DB: PubMed Journal: J Med Phys ISSN: 0971-6203
Figure 1Leaf cross-sectional diagram of (a) prototype-1 (b) prototype-2 multileaf collimators system
Figure 2A typical experimental setup with prototype multileaf collimators attached to the telecobalt machine
Figure 3Surface dose (as a percentage of the dmax dose) from open cobalt fields as a function of the size of the edge of the equivalent square field
Figure 4Surface dose (as a percentage of the dmax dose) from prototype-1 multileaf collimator fields as a function of the size of the edge of the equivalent square field
Skin dose from cobalt-60 beam influenced by the presence of add-on prototype-1 multileaf collimator for field sizes ranging from 2×2 cm² to 14×14 cm² defined by the multileaf collimators
| Percentage surface dose | |||||||
|---|---|---|---|---|---|---|---|
| Field size (cm²) | SSD 60 cm | SSD 65cm | SSD 70 cm | SSD 75 cm | SSD 80 cm | SSD 85 cm | SSD 90 cm |
| 2 | 24 | 22 | 20 | 20 | 22 | 15 | 13 |
| 4 | 32 | 26 | 28 | 25 | 27 | 22 | 24 |
| 6 | 46 | 35 | 31 | 32 | 29 | 29 | 29 |
| 8 | 54 | 43 | 37 | 34 | 36 | 34 | 35 |
| 10 | 48 | 48 | 41 | 37 | 39 | 38 | 36 |
| 12 | 66 | 49 | 46 | 38 | 40 | 39 | 41 |
| 14 | 61 | 49 | 46 | 41 | 42 | 40 | 44 |
SSD: Source-to-surface distance
Figure 5Surface dose (as a percentage of the dmax dose) from prototype-2 multileaf collimator fields as a function of the size of the edge of the equivalent square field
Skin dose from cobalt-60 beam influenced by the presence of add-on prototype-2 multileaf collimators for field sizes ranging from 2×2 cm² to 14×14 cm² defined by multileaf collimators
| Percentage surface dose | |||||||
|---|---|---|---|---|---|---|---|
| Field size (cm²) | SSD 60 cm | SSD 65cm | SSD 70 cm | SSD 75 cm | SSD 80 cm | SSD 85 cm | SSD 90 cm |
| 2 | 29 | 24 | 23 | 21 | 20 | 22 | 13 |
| 4 | 37 | 30 | 27 | 25 | 28 | 23 | 23 |
| 6 | 49 | 34 | 31 | 32 | 30 | 30 | 31 |
| 8 | 42 | 41 | 35 | 33 | 31 | 34 | 29 |
| 10 | 62 | 49 | 40 | 37 | 37 | 35 | 37 |
| 12 | 65 | 50 | 42 | 39 | 39 | 41 | 38 |
| 14 | 66 | 52 | 48 | 41 | 43 | 41 | 43 |
SSD: Source-to-surface distance