When there is an absence of scattering material adjacent to the skin on the exit side of a megavoltage beam the dose to the skin is less than would be calculated using depth dose tables or isodose distributions measured in "semi-infinite" phantoms. Ionization measurements using a thin-window parallel-plate chamber show that the dose at 4 mg/cm2 from the exit surface is 14% to 16% less for cobalt-60 gamma rays and about 8% less for 25-MV x rays compared to the dose with full backscatter. As the angle of incidence increases the skin dose increases due to radiation scattered toward the surface. A method for the calculation of skin doses from tangential therapy beams is described.

Entities:  

Mesh:

Substances:

Year:  1979        PMID: 113655     DOI: 10.1118/1.594583

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  3 in total

1.  Estimation of Skin doses for Retrofit Prototype Multileaf Collimators Designed for Telecobalt Therapy Machine.

Authors:  Akula Roopa Rani; Komanduri Ayyangar; A R Reddy; Ayyalasomayajula Anil Kumar; Pal Reddy Yadagiri Reddy
Journal:  J Med Phys       Date:  2021-02-02

2.  Skin dose during radiotherapy: a summary and general estimation technique.

Authors:  Stephen F Kry; Susan A Smith; Rita Weathers; Marilyn Stovall
Journal:  J Appl Clin Med Phys       Date:  2012-05-10       Impact factor: 2.102

3.  A simple technique to improve calculated skin dose accuracy in a commercial treatment planning system.

Authors:  Lilie Wang; Anthony J Cmelak; George X Ding
Journal:  J Appl Clin Med Phys       Date:  2018-02-06       Impact factor: 2.102