Measurement of relative output factors for the 8 and 4 mm collimators of Leksell Gamma Knife Perfexion by film dosimetry.

Josef Novotny1, Jagdish P Bhatnagar, Mubina A Quader, Greg Bednarz, L Dade Lunsford, M Saiful Huq.   

Abstract

Three types of films, Kodak EDR2, Gafchromic EBT, and Gafchromic MD-V2-55, were used to measure relative output factors of 4 and 8 mm collimators of the Leksell Gamma Knife Perfexion. The optical density to dose calibration curve for each of the film types was obtained by exposing the films to a range of known doses. Ten data points were acquired for each of the calibration curves in the dose ranges from 0 to 4 Gy, 0 to 8 Gy, and 0 to 80 Gy for Kodak EDR2, Gafchromic EBT, and Gafchromic MD-V2-55 films, respectively. For the measurement of relative output factors, five films of each film type were exposed to a known dose. All films were scanned using EPSON EXPRESSION 10000 XL scanner with 200 dpi resolution in 16 bit gray scale for EDR2 film and 48 bit color scale for Gafchromic films. The scanned images were imported in the red channel for both Gafchromic films. The background corrections from an unexposed film were applied to all films. The output factors obtained from film measurements were in a close agreement both with the Monte Carlo calculated values of 0.924 and 0.805 for 8 and 4 mm collimators, respectively. These values are provided by the vendor and used as default values in the vendor's treatment planning system. The largest differences were noted for the Kodak EDR 2 films (-2.1% and -4.5% for 8 and 4 mm collimators, respectively). The best agreement observed was for EBT Gafchromic film (-0.8% and +0.6% differences for 8 and 4 mm collimators, respectively). Based on the present values, no changes in the default relative output factor values were made in the treatment planning system.

Entities:  

Mesh:

Year:  2009        PMID: 19544795     DOI: 10.1118/1.3113904

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  12 in total

Review 1.  Stereotactic radiosurgery alone for multiple brain metastases? A review of clinical and technical issues.

Authors:  Arjun Sahgal; Mark Ruschin; Lijun Ma; Wilko Verbakel; David Larson; Paul D Brown
Journal:  Neuro Oncol       Date:  2017-04-01       Impact factor: 12.300

2.  Replacing gamma knife beam-profiles on film with point-detector scans.

Authors:  Benedikt Rudek; Kenneth Bernstein; Sunshine Osterman; Tanxia Qu
Journal:  J Appl Clin Med Phys       Date:  2022-01-09       Impact factor: 2.102

3.  Target and peripheral dose from radiation sector motions accompanying couch repositioning of patient coordinates with the Gamma Knife(®) Perfexion(™).

Authors:  Tuan-Anh Tran; Vincent Wu; Harish Malhotra; James P Steinman; Dheerendra Prasad; Matthew B Podgorsak
Journal:  Radiol Oncol       Date:  2011-04-23       Impact factor: 2.991

4.  Is It Possible to Publish a Calibration Function for Radiochromic Film?

Authors:  Maria F Chan; David Lewis; Xiang Yu
Journal:  Int J Med Phys Clin Eng Radiat Oncol       Date:  2014-01-29

5.  Clinical commissioning and use of the Novalis Tx linear accelerator for SRS and SBRT.

Authors:  Jinkoo Kim; Ning Wen; Jian-Yue Jin; Nicole Walls; Sangroh Kim; Haisen Li; Lei Ren; Yimei Huang; Anthony Doemer; Kathleen Faber; Tina Kunkel; Ahssan Balawi; Kimberly Garbarino; Kenneth Levin; Samir Patel; Munther Ajlouni; Brett Miller; Teamor Nurushev; Calvin Huntzinger; Raymond Schulz; Indrin J Chetty; Benjamin Movsas; Samuel Ryu
Journal:  J Appl Clin Med Phys       Date:  2012-05-10       Impact factor: 2.102

6.  Gamma Knife radiosurgery for vestibular schwannomas: Evaluation of planning using the sphericity degree of the target volume.

Authors:  Crystian Wilian Chagas Saraiva; Simone Coutinho Cardoso; Daniela Piai Groppo; Antônio Afonso Ferreira De Salles; Luiz Fernando de Ávila; Luiz Antonio Ribeiro da Rosa
Journal:  PLoS One       Date:  2020-01-10       Impact factor: 3.240

7.  Estimation of Skin doses for Retrofit Prototype Multileaf Collimators Designed for Telecobalt Therapy Machine.

Authors:  Akula Roopa Rani; Komanduri Ayyangar; A R Reddy; Ayyalasomayajula Anil Kumar; Pal Reddy Yadagiri Reddy
Journal:  J Med Phys       Date:  2021-02-02

8.  Calculation of excess dose to the eye phantom due to a distanced shielding for electron therapy in head and neck cancers.

Authors:  Keyvan Jabbari; Mahnaz Roayaei; Hosein Saberi
Journal:  J Med Signals Sens       Date:  2012-07

9.  Dosimetry for small fields in stereotactic radiosurgery using gafchromic MD-V2-55 film, TLD-100 and alanine dosimeters.

Authors:  Guerda Massillon-J L; Diego Cueva-Prócel; Porfirio Díaz-Aguirre; Miguel Rodríguez-Ponce; Flor Herrera-Martínez
Journal:  PLoS One       Date:  2013-05-09       Impact factor: 3.240

10.  A novel and effective method for validation and measurement of output factors for Leksell Gamma Knife® Icon™ using TRS 483 protocol.

Authors:  Swapna Lilly Cyriac; Jian Liu; Emel Calugaru; Jenghwa Chang
Journal:  J Appl Clin Med Phys       Date:  2020-09-06       Impact factor: 2.102

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.