| Literature DB >> 33898035 |
Aida Kafai Golahmadi1,2, Danyal Z Khan3,4, George P Mylonas2, Hani J Marcus3,4.
Abstract
BACKGROUND: Excessive tool-tissue interaction forces often result in tissue damage and intraoperative complications, while insufficient forces prevent the completion of the task. This review sought to explore the tool-tissue interaction forces exerted by instruments during surgery across different specialities, tissues, manoeuvres and experience levels. MATERIALS &Entities:
Keywords: Force feedback; Force sensors; Haptic feedback; Operative forces; Robotic surgery; Robotics; Smart instruments; Surgical forces; Tool-tissue interaction
Year: 2021 PMID: 33898035 PMCID: PMC8058906 DOI: 10.1016/j.amsu.2021.102268
Source DB: PubMed Journal: Ann Med Surg (Lond) ISSN: 2049-0801
Fig. 1PRISMA flow chart of paper identification, screening and eventual inclusion.
Summary forces averages across speciality, tissue type surgical tasks.
| Category | Mean of average (N) | Mean of maximum (N) | Number of studies |
|---|---|---|---|
| Ophthalmology | 0.04 | 0.05 | 5 |
| Vascular | 0.07 | 0.65 | 3 |
| Neurosurgery | 0.68 | 1.48 | 13 |
| Cardiothoracic | 1.47 | 1.76 | 3 |
| General surgery | 4.67 | 11.4 | 14 |
| Otorhinolaryngology | 8.49 | 15.6 | 2 |
| Obstetrics & gynaecology | 8.69 | 10.1 | 2 |
| Urology | 9.79 | 15.6 | 1 |
| Orthopaedics | 210 | 1708 | 2 |
| Nervous | 0.4 | 1.7 | 17 |
| Epithelial | 3.8 | 9.7 | 18 |
| Muscle | 4.1 | 6.7 | 3 |
| Connective | 45.8 | 347.9 | 10 |
| Sharp dissection | 0.03 | NA | 1 |
| Clamping | 0.05 | 0.10 | 2 |
| Suction | 0.13 | 0.21 | 1 |
| Coagulation | 0.29 | 0.90 | 4 |
| Retraction without grasping | 0.95 | 2.51 | 3 |
| Blunt penetration | 1.21 | 3.58 | 4 |
| Blunt dissection | 2.08 | 3.69 | 7 |
| Cutting | 2.28 | 4.48 | 5 |
| Sharp penetration | 2.34 | 4.67 | 11 |
| Retraction with grasping | 3.65 | 10.25 | 13 |
| Drilling | 14.09 | 25.99 | 2 |
Fig. 2Selected speciality-specific tasks with overview of forces.
Overview of impact of experience levels on force exertion.
| Study | Task | Speciality | % Difference Novice Vs Intermediate | % Difference Novice Vs Expert | % Difference Intermediate Vs Expert | |||
|---|---|---|---|---|---|---|---|---|
| Horeman 2012 | Sharp penetration | General surgery (laparoscopic) | 4.5 | 2.7 | 40% more | |||
| Kobler 2015 | Drilling | Otolaryngology | 13.85 | 12.29 | 11% more | |||
| Wottawa 2016 | Retraction with Grasping | General surgery (laparoscopic) | 3.65 | 3.15 | 14% more | |||
| Gonenc 2017 | Sharp penetration | Ophthalmology (robotic) | 0.05 | 0.07 | 0.05 | 29% | 0% more | 29% more |
| Rafii-Tari 2017 | Sharp penetration | Vascular surgery | 0.32 | 0.12 | 62% more | |||
| Talasaz 2017 | Sharp penetration | Cardiothoracic surgery (robotic) | 2.54 | 2.49 | 2% more | |||
| Ebrahimi 2019 | Sharp penetration | Ophthalmology (robotic) | 0.10 | 0.14 | 0.07 | 29% | 30% more | 50% more |
| Kwong 2019 | Retraction with grasping | Orthopaedic | 470.7 | 941.4 | 50% less |
Overview of force feedback mechanism reported.
| Study | Feedback groups | Task | Speciality | Mean of average forces pre-feedback | Mean of average forces post-feedback | Absolute Difference | % Difference |
|---|---|---|---|---|---|---|---|
| Gonenc 2012 | Audio (vs. no feedback) | Sharp penetration | Ophthalmology (robotic) | 0.1 | 0.07 | 0.03 | 30% |
| Horeman 2013 | Visual force feedback (vs. visual time feedback) | Retraction with Grasping | General surgery (laparoscopic) | 0.79 (Visual time feedback) | 0.51 (Visual force feedback) | 0.28 | 35.4% |
| Alleblas 2017 | Haptic (vs. no feedback) | Retraction with Grasping | General surgery (laparoscopic) | 4.6 | 1.7 | 2.9 | 63% |
| Diez 2018 | Haptic (vs. no feedback) | Multiple (laparoscopic tasks) | Obstetrics and Gynaecology (laparoscopic) | 0.81 | 0.48 | 0.33 | 40.7% |
| Ebrahimi 2018 | Haptic (vs. no feedback) | Sharp penetration | Ophthalmology (robotic) | 0.12 | 0.09 | 0.03 | 25% |
| Ebrahimi 2018 | Audio (vs. no feedback) | Sharp penetration | Ophthalmology (robotic) | 0.12 | 0.08 | 0.04 | 33.3% |
| Wottawa 2016 | Tactile (vs. no feedback) | Retraction with Grasping | General surgery (laparoscopic) | 3.4 | 2.3 | 1.1 | 32.4% |
| Horeman 2012 | Visual (vs. no feedback) | Sharp penetration | General surgery (laparoscopic) | 2.6 | 1.3 | 1.3 | 50% |