| Literature DB >> 33783569 |
Mahta Mazloumi1, Gert Van Gompel2, Veerle Kersemans3, Johan de Mey2, Nico Buls2.
Abstract
OBJECTIVES: Routine dosimetry calculations do not account for the presence of iodine in organs and tissues during CT acquisition. This study aims to investigate the impact of contrast agent (CA) on radiation dose.Entities:
Keywords: Humans; Iodine; Phantoms, Imaging; Radiation dosage; Tomography, X-ray Computed
Mesh:
Substances:
Year: 2021 PMID: 33783569 PMCID: PMC8452580 DOI: 10.1007/s00330-021-07763-7
Source DB: PubMed Journal: Eur Radiol ISSN: 0938-7994 Impact factor: 5.315
Fig. 1a CT image of the experimental phantom, and (b) the reconstructed virtual phantom
Fig. 2a CTDIvol normalized dose versus iodine concentration in the ROI in syringes for 80, 120, 140 kVp, and DECT. b Dose in the ROI in the iodine syringes versus iodine concentration for virtual model and experimental phantom for DECT. Horizontal bars indicate uncertainty in the solutions and vertical bars indicate the uncertainty in the simulated dose (6.3%)
Fig. 3DSSDE in the presence and in the absence of CA in (a) liver, (b) liver parenchyma, (c) aorta, (d) right kidney, (e) left kidney, and (f) spleen in the 20 patients
Fig. 4a A Pearson correlation test shows a linear agreement between %DSSDE increase and iodine content in all the organs of the patient study. b DSSDE in the segmented organs and iodine solutions with respect to their iodine content
Fig. 5Parametric dose maps of three patients in the absence (left column) and presence (right column) of CA. a 23-year-old female administered to 370 mg I/ml (Iopromide), (b) 35-year-old female administered to 350 mg I/ml (Iomeprol), (c) 76-year-old male administered to 350 mg I/ml (Iomeprol), Abbreviations: Iadm, administered iodine
The mean HU values ± SD, the mean iodine concentrations ± SD, and the DSSDE ± SD in the segmented organs and tissues of the 20 patients
| Parameter | Liver | Liver parenchyma | Right kidney | Left kidney | Aortic blood | Spleen |
|---|---|---|---|---|---|---|
| Mean CT values (HU) ± SD in VUE images | 50 ± 12 | 49 ± 14 | 34 ± 5 | 34 ± 4 | 50 ± 5 | 44 ± 4 |
| Mean CT values (HU) ± SD in CE images | 120 ± 32 | 116 ± 33 | 190 ± 42 | 191 ± 39 | 131 ± 19 | 173 ± 30 |
| Mean iodine concentration ± SD (mg I/ml) | 3.00 ± 0.91 | 2.85 ± 0.86 | 6.38 ± 1.60 | 6.43 ± 1.50 | 5.32 ± 1.10 | 3.48 ± 0.71 |
| Mean dose increase ± SD (mGy) | 21 ± 5% | 20 ± 5% | 37 ± 7% | 39 ± 7% | 34 ± 6% | 26 ± 4% |
Abbreviations: HU Hounsfield units, SD standard deviation, VUE virtual-unenhanced, CE contrast-enhanced, SSDE size-specific dose estimate
The administered iodine, CTDIvol, SSDE, mean absolute dose ± SD in the absence of contrast, mean absolute dose ± SD in the presence of contrast, and the dose enhancement factor ± SD (DEF) in the organs and tissues (excluding aorta) for each patient
| Patient | Administered iodine (g) | CTDIvol (mGy) | SSDE (mGy) | DEF = | ||
|---|---|---|---|---|---|---|
| PA1 | 38.5 | 9.42 | 11.59 | 10.16 ± 0.42 | 12.70 ± 0.98 | 1.25 ± 0.11 |
| PA2 | 42.0 | 10.10 | 12.42 | 10.88 ± 0.31 | 13.32 ± 0.92 | 1.22 ± 0.09 |
| PA3 | 38.5 | 9.03 | 11.92 | 10.83 ± 0.59 | 13.53 ± 1.30 | 1.25 ± 0.14 |
| PA4 | 42.0 | 11.18 | 15.32 | 13.84 ± 0.42 | 18.04 ± 1.56 | 1.30 ± 0.12 |
| PA5 | 36.7 | 7.95 | 12.16 | 10.88 ± 0.65 | 13.64 ± 1.43 | 1.25 ± 0.15 |
| PA6 | 42.0 | 9.03 | 11.92 | 10.44 ± 0.37 | 13.99 ± 0.50 | 1.34 ± 0.07 |
| PA7 | 42.0 | 5.80 | 8.58 | 8.11 ± 0.26 | 11.21 ± 1.45 | 1.38 ± 0.18 |
| PA8 | 44.4 | 9.03 | 12.37 | 9.74 ± 0.69 | 12.97 ± 1.39 | 1.33 ± 0.17 |
| PA9 | 42.0 | 7.95 | 10.18 | 8.65 ± 0.64 | 10.78 ± 1.19 | 1.25 ± 0.17 |
| PA10 | 42.0 | 10.10 | 12.42 | 11.55 ± 0.80 | 15.28 ± 2.09 | 1.32 ± 0.20 |
| PA11 | 44.4 | 12.25 | 14.58 | 13.08 ± 0.43 | 17.43 ± 1.07 | 1.33 ± 0.09 |
| PA12 | 42.0 | 7.95 | 11.77 | 11.45 ± 0.27 | 14.57 ± 1.25 | 1.27 ± 0.11 |
| PA13 | 40.7 | 9.03 | 11.56 | 10.70 ± 0.53 | 12.92 ± 1.76 | 1.21 ± 0.17 |
| PA14 | 42.0 | 7.93 | 9.04 | 7.25 ± 0.60 | 9.41 ± 1.02 | 1.30 ± 0.18 |
| PA15 | 38.8 | 10.10 | 12.02 | 10.45 ± 0.81 | 13.04 ± 1.26 | 1.25 ± 0.15 |
| PA16 | 42.0 | 7.95 | 11.37 | 9.96 ± 0.35 | 12.50 ± 1.07 | 1.26 ± 0.12 |
| PA17 | 30.4 | 11.18 | 15.32 | 13.29 ± 1.51 | 16.62 ± 2.42 | 1.25 ± 0.23 |
| PA18 | 44.4 | 9.03 | 11.92 | 11.13 ± 1.15 | 14.64 ± 1.70 | 1.32 ± 0.20 |
| PA19 | 35.2 | 7.73 | 11.83 | 10.53 ± 0.67 | 14.28 ± 1.82 | 1.36 ± 0.19 |
| PA20 | 42.0 | 12.25 | 11.27 | 9.40 ± 1.19 | 11.90 ± 1.43 | 1.27 ± 0.22 |
Abbreviations: g gram, CTDI volume computed tomography dose index, mGy milli-gray, SSDE size-specific dose estimate, SD standard deviation, mean absolute dose in the absence of contrast agent excluding aorta, mean absolute dose in the presence of contrast agent excluding aorta, DEF dose enhancement factor