A method to evaluate the dose increase in CT with iodinated contrast medium.

Ernesto Amato1, Domenico Lizio, Nicola Settineri, Andrea Di Pasquale, Ignazio Salamone, Ignazio Pandolfo.   

Abstract

PURPOSE: The objective of this study is to develop a method to calculate the relative dose increase when a computerized tomography scan (CT) is carried out after administration of iodinated contrast medium, with respect to the same CT scan in absence of contrast medium.
METHODS: A Monte Carlo simulation in GEANT4 of anthropomorphic neck and abdomen phantoms exposed to a simplified model of CT scanner was set up in order to calculate the increase of dose to thyroid, liver, spleen, kidneys, and pancreas as a function of the quantity of iodine accumulated; a series of experimental measurements of Hounsfield unit (HU) increment for known concentrations of iodinated contrast medium was carried out on a Siemens Sensation 16 CT scanner in order to obtain a relationship between the increment in HU and the relative dose increase in the organs studied. The authors applied such a method to calculate the average dose increase in three patients who underwent standard CT protocols consisting of one native scan in absence of contrast, followed by a contrast-enhanced scan in venous phase.
RESULTS: The authors validated their GEANT4 Monte Carlo simulation by comparing the resulting dose increases for iodine solutions in water with the ones presented in literature and with their experimental data obtained through a Roentgen therapy unit. The relative dose increases as a function of the iodine mass fraction accumulated and as a function of the Hounsfield unit increment between the contrast-enhanced scan and the native scan are presented. The data shown for the three patients exhibit an average relative dose increase between 22% for liver and 74% for kidneys; also, spleen (34%), pancreas (28%), and thyroid (48%) show a remarkable average increase.
CONCLUSIONS: The method developed allows a simple evaluation of the dose increase when iodinated contrast medium is used in CT scans, basing on the increment in Hounsfield units observed on the patients' organs. Since many clinical protocols employ multiple scans at different circulatory phases after administration of contrast medium, such a method can be useful to evaluate the total dose to the patient, also in view of potential clinical protocol optimizations.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20879585     DOI: 10.1118/1.3460797

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  12 in total

1.  Optimization of beam quality for photon-counting spectral computed tomography in head imaging: simulation study.

Authors:  Han Chen; Cheng Xu; Mats Persson; Mats Danielsson
Journal:  J Med Imaging (Bellingham)       Date:  2015-11-06

2.  Determination of contrast media administration to achieve a targeted contrast enhancement in computed tomography.

Authors:  Pooyan Sahbaee; Paul P Segars; Daniele Marin; Rendon Nelson; Ehsan Samei
Journal:  J Med Imaging (Bellingham)       Date:  2016-01-20

3.  The effect of iodine uptake on radiation dose absorbed by patient tissues in contrast enhanced CT imaging: Implications for CT dosimetry.

Authors:  Kostas Perisinakis; Antonis Tzedakis; Kostas Spanakis; Antonios E Papadakis; Adam Hatzidakis; John Damilakis
Journal:  Eur Radiol       Date:  2017-07-14       Impact factor: 5.315

4.  Investigation of 90Y-avidin for prostate cancer brachytherapy: a dosimetric model for a phase I-II clinical study.

Authors:  Francesca Botta; Marta Cremonesi; Mahila E Ferrari; Ernesto Amato; Francesco Guerriero; Andrea Vavassori; Anna Sarnelli; Stefano Severi; Guido Pedroli; Giovanni Paganelli
Journal:  Eur J Nucl Med Mol Imaging       Date:  2013-05-03       Impact factor: 9.236

5.  Systematic survey of the dose enhancement in tissue-equivalent materials facing medium- and high-Z backscatterers exposed to X-rays with energies from 5 to 250 keV.

Authors:  M Seidenbusch; D Harder; D Regulla
Journal:  Radiat Environ Biophys       Date:  2014-03-15       Impact factor: 1.925

Review 6.  Enhanced radiation dose and DNA damage associated with iodinated contrast media in diagnostic X-ray imaging.

Authors:  Richard Harbron; Elizabeth A Ainsbury; Simon D Bouffler; Rick J Tanner; Jonathan S Eakins; Mark S Pearce
Journal:  Br J Radiol       Date:  2017-08-22       Impact factor: 3.039

7.  Personalized administration of contrast medium with high delivery rate in low tube voltage coronary computed tomography angiography.

Authors:  Sock Keow Tan; Kwan Hoong Ng; Chai Hong Yeong; Raja Rizal Azman Raja Aman; Fadhli Mohamed Sani; Yang Faridah Abdul Aziz; Zhonghua Sun
Journal:  Quant Imaging Med Surg       Date:  2019-04

Review 8.  Estimation of the effects of medical diagnostic radiation exposure based on DNA damage.

Authors:  Lin Shi; Satoshi Tashiro
Journal:  J Radiat Res       Date:  2018-04-01       Impact factor: 2.724

9.  Decreasing the radiation dose for contrast-enhanced abdominal spectral CT with a half contrast dose: a matched-pair comparison with a 120 kVp protocol.

Authors:  Daisuke Sakabe; Takeshi Nakaura; Seitaro Oda; Masafumi Kidoh; Daisuke Utsunomiya; R T Masahiro Hatemura; Yoshinori Funama
Journal:  BJR Open       Date:  2020-12-11

10.  Dosimetric effect of CT contrast agent in CyberKnife treatment plans.

Authors:  Hee Jung Kim; Ah Ram Chang; Yang-Kyun Park; Sung-Joon Ye
Journal:  Radiat Oncol       Date:  2013-10-18       Impact factor: 3.481

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.