Jessica Hanae Zafra-Tanaka1, David Beran2, Beatrice Vetter3, Rangarajan Sampath3, Antonio Bernabe-Ortiz1. 1. CRONICAS Centre of Excellence in Chronic Diseases, Universidad Peruana Cayetano Heredia, Lima, Peru. 2. Division of Tropical and Humanitarian Medicine, University of Geneva and Geneva University Hospitals, Geneva, Switzerland. 3. Foundation for Innovative New Diagnostics, Geneva, Switzerland.
Abstract
BACKGROUND: Self-management is an important pillar for diabetes control and to achieve it, glucose self-monitoring devices are needed. Currently, there exist several different devices in the market and many others are being developed. However, whether these devices are suitable to be used in resource constrained settings is yet to be evaluated. AIMS: To assess existing glucose monitoring tools and also those in development against the REASSURED which have been previously used to evaluate diagnostic tools for communicable diseases. METHODS: We conducted a scoping review by searching PubMed for peer-review articles published in either English, Spanish or Portuguese in the last 5 years. We selected papers including information about devices used for self-monitoring and tested on humans with diabetes; then, the REASSURED criteria were used to assess them. RESULTS: We found a total of 7 continuous glucose monitoring device groups, 6 non-continuous, and 6 devices in development. Accuracy varied between devices and most of them were either invasive or minimally invasive. Little to no evidence is published around robustness, affordability and delivery to those in need. However, when reviewing publicly available prices, none of the devices would be affordable for people living in low- and middle-income countries. CONCLUSIONS: Available devices cannot be considered adapted for use in self-monitoring in resource constraints settings. Further studies should aim to develop less-invasive devices that do not require a large set of components. Additionally, we suggest some improvement in the REASSURED criteria such as the inclusion of patient-important outcomes to increase its appropriateness to assess non-communicable diseases devices.
BACKGROUND: Self-management is an important pillar for diabetes control and to achieve it, glucose self-monitoring devices are needed. Currently, there exist several different devices in the market and many others are being developed. However, whether these devices are suitable to be used in resource constrained settings is yet to be evaluated. AIMS: To assess existing glucose monitoring tools and also those in development against the REASSURED which have been previously used to evaluate diagnostic tools for communicable diseases. METHODS: We conducted a scoping review by searching PubMed for peer-review articles published in either English, Spanish or Portuguese in the last 5 years. We selected papers including information about devices used for self-monitoring and tested on humans with diabetes; then, the REASSURED criteria were used to assess them. RESULTS: We found a total of 7 continuous glucose monitoring device groups, 6 non-continuous, and 6 devices in development. Accuracy varied between devices and most of them were either invasive or minimally invasive. Little to no evidence is published around robustness, affordability and delivery to those in need. However, when reviewing publicly available prices, none of the devices would be affordable for people living in low- and middle-income countries. CONCLUSIONS: Available devices cannot be considered adapted for use in self-monitoring in resource constraints settings. Further studies should aim to develop less-invasive devices that do not require a large set of components. Additionally, we suggest some improvement in the REASSURED criteria such as the inclusion of patient-important outcomes to increase its appropriateness to assess non-communicable diseases devices.
Globally, around 463 million people were estimated to have diabetes mellitus in
2019 and this number is expected to increase to 700 million in 2045.
Moreover, diabetes accounted for 1.6 million deaths worldwide in 2015.
Due to these reasons, diabetes has been prioritized by the World
Health Organization (WHO) in its Global Action Plan (GAP) 2013 to 2020,
together with other noncommunicable diseases (NCDs) such as cardiovascular
disease (CVD), cancer, and chronic respiratory diseases (CRD).[3,4]The GAP advocates for the strengthening of primary health care facilities to
improve prevention, early detection, treatment and sustained management of
NCDs.[3,4] Moreover, the Package of Essential
Noncommunicable Disease Interventions for primary care in low-resource
settings (WHO PEN package) identifies the essential interventions that need
to be implemented regarding diabetes management which include early
detection and glycaemic management[5,6] and highlights the
importance of implementing technologies to measure blood glucose (glucose
meters and blood glucose test strips) and protein urine test strips in
primary health care facilities.[3,7] In addition, WHO
self-care recommendations state that patients with diabetes should be
offered self-monitoring of blood glucose based on individual clinical need.
Moreover, self-monitoring and self-adjustment of insulin dosage are
recommended for patients with type 1 diabetes
and thus require a tool for these individuals to monitor their blood
glucose levels.Given the large number of devices to monitor diabetes such as blood glucose
meters or continuous glucose monitoring systems (CGMs) that exist,
having clear criteria to select which one to use is important for
health care systems. Thus, the ASSURED (Affordable, Sensitive, Specific,
User-friendly, Rapid and robust, Equipment-free and Deliverable to
end-users) criteria have been suggested to systematically assess the
appropriateness of diagnostic tests for resource-constrained settings.
These criteria were updated to include real-time connectivity and
ease of specimen collection and are now called REASSURED criteria.[3,10]
One previous study assessed monitoring technologies and found some pitfalls;
variable accuracy, low affordability and availability at both the health
care facilities and as self-management tools at the patients’ homes.
However, other important criteria proposed in the REASSURED have not
been systematically assessed which prevents a complete assessment of the
appropriateness of the technologies for resource-constrained settings and
individuals’ needs.Our study aimed to assess existing glucose monitoring tools, including those in
development, against the REASSURED criteria which have been previously used
to evaluate diagnostic tools for communicable diseases. Besides, this
exercise allowed the evaluation of the appropriateness of the REASSURED
criteria for tools used in NCDs versus communicable diseases. Our study
systematically assessed available devices and helped identify gaps to tailor
future developments.
Methods
Study Design
We decided to conduct a scoping review as they are useful for
synthesizing research evidence and are often used to map existing
literature in a given field in terms of its nature, features, and
volume. Moreover, scoping reviews allow to summarize and disseminate
research findings, to identify research gaps, and to make
recommendations for future research. Our study followed the Preferred
Reporting Items for Systematic Reviews and Meta-analysis Protocols (PRISMA-P)
and the Guidance for conducting systematic scoping reviews from
the Joanna Briggs Institute.
The reporting of the results followed the PRISMA Extension for
Scoping Reviews (PRISMA-ScR).
Eligibility Criteria
To identify and assess glucose monitoring technologies, we included all
devices to monitor glucose (test systems for use at health care
establishments or at home to measure the amount of blood glucose
): (1) that could be used for self-monitoring of diabetes; (2)
that had been validated in humans with diabetes; (3) whose validation
results were published as scientific articles in peer-review journals
in the last 5 years; and (4) whose full-text articles were available
in English, Spanish or Portuguese. We decided to include studies that
had been validated in humans with diabetes given that our objective
was to assess monitoring devices in terms of accuracy and ease of
use.
Information Sources and Search
To identify potentially relevant documents, we searched PubMed from
August 2014 to May 2020. We choose to include documents published in
the last 5 years as we are studying a rapidly evolving field and we
targeted devices that are currently in the market and more recent
developments that have an opportunity to result in a marketable
product. The search strategies were drafted by the research team and
included terms related to glucose monitoring, diabetes, equipment or
devices. The final search strategy can be found in Supplemental material 1. The search was conducted by
one researcher who uploaded the results to Rayyan, an open software
used for study selection.
The search was last updated on June 9, 2020 and included papers
published up to that date.
Selection of Sources of Evidence
To identify and assess glucose monitoring technologies, 2 independent
reviewers screened titles, abstracts using Rayyan, and checked
disagreements to reach a consensus. The studies that passed this first
phase were downloaded for full-text review. We created a list of all
the documents using Microsoft Excel to keep track of the selection
process. Disagreements on study selection were resolved by
consensus.
Data Extraction
A data-charting form was developed on an Excel-based sheet using the
REASSURED criteria. Two independent researchers extracted the
following characteristics from the devices: name of the device,
manufacturer/developer name, version or generation of the device, and
the REASSURED criteria items. Discrepancies on data extraction were
discussed and articles were reviewed to check the information.The REASSURED criteria include the following
: real-time connectivity (the possibility to connect the device
to a reader or mobile phone to provide the required data to decide
about clinical management), ease of specimen collection (designed for
non-invasive collection of specimens), affordability (affordable to
end-users and the health systems), sensitivity, specificity,
user-friendliness (the procedure is simple or performed in a few steps
that require minimal training), rapidity (results are available within
5 minutes to 2 h), robustness (eg, devices do not require special
conditions related to temperature), equipment-free (do not require
special equipment or can be operated in a very simple device that uses
solar or battery power), and delivered to those who need it (available
to those in need of the test).As part of the ease of specimen collection, we defined invasiveness as
follows: invasive (devices that are implanted in the patient’s body or
that invade the body to access a blood sample), minimally invasive
(devices that painlessly invade a very small part of the patient’s
body, such as skin to collect a minimal sample, or extract some other
form of body fluid, such as sweat, tears, and saliva), and
non-invasive devices (devices that do not invade the patient’s body)
according to the literature.[17,18]The REASSURED criteria include sensitivity and specificity. However, they
can only be quantified for categorical variables (eg, presence versus
absence of disease) and not for numerical variables such as blood
glucose. For this reason, instead of assessing sensitivity and
specificity, we assessed measured of accuracy.
To assess the accuracy of blood glucose measurements, three
indicators were used: (1) the mean-average-relative-difference (MARD)
which is the average of all the absolute errors between the measured
points and those set as the reference (in other cases, the median
average relative difference measurement (MedARD) was included); (2)
the error grids; two-dimension grids divided in ‘risk zones’, based on
the agreement between the glucose measuring device and the reference
method (eg, Clark error grid or similar)
; and (3) others like bias (eg, Bland-Altman test) or
correlations (eg, Pearson test or similar). In our study, we are
reporting MARD (or MedARD) and error grids as these are the most
commonly used approaches. We considered a value of less than 10% for
MARD as accurate.Affordability is a complex concept given that the definition of what is
“affordable” varies according to the context and according to who is
the payer (health care systems versus patients). Thus, we decided to
evaluate affordability by searching for unitary costs or
cost-effectiveness assessments reported in selected articles. If not
available, a search was conducted on web pages, however as this
information was not published in peer-review journals, we did not
consider them in the REASSURED assessment but discussed in the results
section. The same methodology was followed to assess the “delivered to
those in need” criteria, for which we sought information regarding
distribution zones. We did not consider prices as they are already
being considered within the affordability criteria.For each of the categories, we present a description of the device and to
summarize the findings we used color coding: green whether the device
fulfils the criteria, red if it does not, orange where results were
inconclusive or criteria were only partially met, and grey if no
information was found in the reviewed manuscripts.As part of the charting process, we classified devices in groups
according to how they are used (continuous and non-continuous glucose
monitoring) and other characteristics such as how long they last,
calibration needs, application to the body and manufacturer.
Results
We found a total of 1945 documents in the database search and we included 56
documents that provided information for a total of 13 device groups whose
use had been validated on patients with diabetes (see PRISMA flowchart,
Figure 1). For
the proof-of-concept devices refer to Supplemental material 2.
Figure 1.
PRISMA flowchart.
PRISMA flowchart.
Characteristics of the Devices
Based on the continuity of measurement of glucose, we classified the
devices in non-continuous glucose and continuous monitoring systems.
With regards to continuous systems, we considered those monitoring
devices that provided information about glucose levels in a continuous
and automated fashion; while non-continuous monitoring systems were
those that required patients to perform any action to obtain a glucose
measurement.Within continuous glucose monitoring devices, we defined a total of 7
groups based on the similarity of the devices and the manufacturer:
Medtronic devices,[21-24] Eversense (Senseonics),[21,25-28] Dexcom devices,[21,23,29-36] Glucotrack (Integrity Applications),[37,38] an infrared
sensor for subcutaneous microdialysis (Fraunhofer ICT-IMM),
subcutaneous monitoring systems (Biovotion),
and NovioSense glucose sensor (NovioSense).
All of the mentioned devices measured glucose levels on
interstitial tissue or tears (see Figure 2 and Supplemental Table 1).
Figure 2.
RE-ASSURED criteria for continuous glucose monitoring
devices.
Coding: green if it fulfils the criteria, red if it does not,
orange were results were inclusive or partially met, and
grey if no information was found in the revised
articles.
RE-ASSURED criteria for continuous glucose monitoring
devices.Coding: green if it fulfils the criteria, red if it does not,
orange were results were inclusive or partially met, and
grey if no information was found in the revised
articles.In the case of non-continuous monitoring devices, we grouped the devices
into 6 categories: glucose meters,[42-49] FLASH (Abbott),[23,29,30,50-52] TensorTip
Combo Glucometer (CNOGA Medical),
analysis software that works with a smartphone (iXensor Co),
GASA (Guaiacol diazo derivative, 4[(4-Hydroxy-3-methoxyphenyl)
azo]-benzenesulfonic acid),
and Breathotron for diabetes screening, and hypoglycemia
monitoring.[55,56] These devices measured glucose levels on
capillary blood, breath, saliva and interstitial tissue (see Figure 3 and
Supplemental Table 2).
Figure 3.
RE-ASSURED criteria for non-continuous glucose monitoring
devices.
Coding: green if it fulfils the criteria, red if it does not,
orange were results were inclusive or partially met, and
grey if no information was found in the revised
articles.
RE-ASSURED criteria for non-continuous glucose monitoring
devices.Coding: green if it fulfils the criteria, red if it does not,
orange were results were inclusive or partially met, and
grey if no information was found in the revised
articles.
REASSURED Criteria
When using the REASSURED approach, none of the assessed devices met all
of the criteria. For continuous glucose monitoring devices where
information was available for all device groups, the “rapid” and
“user-friendly” criteria were met by all but 2 devices. However, their
accuracy varied widely: As an example, for Medtronic devices, MARD had
a mean value of 13.6% (SD: 11.0%), with variation by glycaemic range,
and it was greater when used at home (MARD: 19.9% (SD:
20.5%)).[21-24] As for Dexcom, MARD varied according to the
device: Dexcom G6 (7.7% to 10.0%), Dexcom G5 (9.0% to 16.3%), and
Dexcom G4 (10.8 to 19%).[21,23,30-36]Regarding real-time connectivity, we only found information for devices
such as Medtronic devices,[21-24] Eversense (Senseonics)[21,25-28] and Dexcom[21,23,29-36] which could connect to smartphones through
Bluetooth and/or download the information in a USB. On the other hand,
none of the devices was equipment free (ie, they need strips) and
information related to robustness was not available for any of the
devices. Little information regarding affordability and delivery for
those in need was found (see Figure 2 and Supplemental Table 1 for details).As for non-continuous glucose monitoring devices, most of them were
user-friendly and rapid, but accuracy varied. In the case of glucose
meters, MARD ranged from 2.3% to 21.2%[42,44-49] and FLASH
from 11.2% to 22%.[23,29,30,50-52] None of the
devices was equipment free. No information about affordability and
deliver to those in need was found for most of the devices. No
information related to robustness was found for any of the devices
(see Figure 3
and Supplemental Table 2 for details).
Discussion
Main Findings
We have found several devices that can be used as self-monitoring tools
for glucose levels at the patients’ home. However, currently available
devices cannot be considered as adapted to be used for self-monitoring
in resource constraints settings. Most of these devices are either
invasive or minimally invasive and rapid even when using different
methods to measure glucose; but all devices were dependent on more
than one piece of equipment and not all of them provided information
regarding connectivity to other devices. Accuracy was assessed for all
devices and the results varied widely, and little to no information
was found concerning robustness, deliver to those in need, and
affordability.Information published on monitoring tools for diabetes tends to focus on
accuracy, an important measurement to assess the usefulness of these
tools. Accuracy studies are relevant as they allow quality assessments
of medical devices, comparisons between devices and are a requirement
for approval from stringent regulatory agencies such as FDA.
Currently, there are different methods to measure accuracy, but
all of them have limitations.[20,57] In this study, we opted to use MARD, an
indicator that evaluates variability between measurements, with a
stringent cutoff point of 10% and we found that the only device that
met the criteria were TensorTip Combo Glucometer
and the Opto-fluidic near-infrared (NIR) continuous glucose monitor.
Nevertheless, each of the devices had only 1 validation study,
suggesting a need to conduct more studies to compare and contrast the
results as the MARD is usually very variable.In contrast to the abundance of articles published that assess accuracy,
there is little to no evidence with regards to robustness, delivery to
those in need and affordability. Even when manufacturers do determine
temperature and humidity ranges for their devices, studies that assess
the robustness of these devices in terms of different climatological
conditions (heat, humidity, etc.) and settings need to be
published.Delivery to those in need and affordability are not always disclosed in
scientific publications which can be justified by different factors.
One study pointed out that, for European countries, coverage for
glucose monitoring devices (either glucose meter plus test strips or
continuous glucose monitoring devices) has increased in the last decade.
However, these tools are not always accessed by all persons in need.
We believe that collecting and sharing information regarding
delivery to those in need could help take actions to improve access to
monitoring tools.In the case of affordability, there are only a limited number of
cost-effectiveness studies and health technology assessments that have
been conducted mostly in high-income countries.[60-63] Given that
prices and the cutoff point to call a tool “cost-effective” vary
between countries we cannot use the available information to assess
the REASSURED criteria when thinking about resource constrains
settings. One suggestion would be to have an open repository to
publicly publish costs of the devices in different countries to be
able to estimate the cost-effectiveness of each device per country.
This would also increase transparency regarding the prices offered to
different countries.To simulate what a patient without cover for these devices would pay
out-of-pocket, we searched the official webpages of the devices that
were in the market for prices and found that: FLASH cost was around
50 USD (the reader) and 50 USD the sensor which has to be changed
every 14 days
; DEXCOM G6 cost was around 50 USD each sensor (change after
30 days), 210 USD the transmitter (change after 3 months) and 300 USD
the receiver
; Medtronic’s Enlite cost was around 80 USD the sensor (change
every 5 days) and 40 USD the One-press Serter,
and Eversense costs were not available on their webpage.
Our objective was not to conduct a costs study, however, when
looking at public prices we can estimate the expenditure of at least
150 USD per month; therefore, the devices would not be affordable
(less than one day of work) for patients living in countries where the
mean monthly income is below 4500 USD.
In the case of glucose meters, the global average manufacturer
selling price per strip is 0.6 USD.
If patients measured glucose 6 times per day, as suggested by ADA,
the costs will reach a total of 108 USD per month and would not
be affordable for patients with incomes below 3240 USD. Thus, both
continuous monitoring systems and glucose meters would not be
affordable.
Limitations of the REASSURED Criteria
The REASSURED criteria have some limitations given that this approach was
created to assess tests for infectious diseases placed at healthcare
facilities. They do not contemplate other criteria that might be
relevant for NCD monitoring devices which are supposed to be designed
for patients as end-users instead of healthcare professionals. A
systematic review found that some patient-reported outcomes that
impacted on the usage of monitoring tools were: (a) satisfaction, (b)
quality of life, (c) emotional distress, and (d) self-efficacy.
Moreover, other important criteria that should be taken into
account when assessing a tool for NCDs’ monitoring should be: (a)
information that leads to action (timely information), (b) minimally
invasive and also minimally disruptive with the patient’s life, and
(c) affordability in the long term (considering that a patient would
need a piece of equipment, batteries, test strips, and others for a
lifetime). This highlights the need to create devices not only
accurate, affordable and easy to use, but also tools that have value
for the end-user. With this in mind, we suggest that the REASSURED
criteria include at least one of the patients reported outcomes found
to be linked to the usage of these devices.
Limitations of the Review
Among our study limitations, we can consider that not all the information
needed to assess the suitability of monitoring devices using the
REASSURED criteria can be found in scientific publications, we
searched only 1 database and the inclusion criteria were rather
strict. Moreover, we only included articles published in English,
Spanish, or Portuguese. However, we believe this was the best approach
to find available published evidence regarding the devices that have
been previously tested in patients with diabetes.
Relevance of Findings
After assessing the identified tools, we recognize that accuracy and ease
of specimen collection should be improved for currently available
tools. Moreover, we have identified gaps in the reporting of certain
criteria chiefly affordability, delivered to those in need, and
robustness. Finally, we identified the need to incorporate new
criteria related to patient-reported outcomes such as satisfaction or
quality of life. Thus, we suggest that the devices improvements focus
on the above-mentioned qualities, especially invasiveness and reducing
the need to have multiple components (sensor, transmitter, and
reader). Also, scientific papers should report all the needed criteria
to assess the appropriateness of devices including costs and
cost-effectiveness studies.
Conclusions
We conducted a scoping review of glucose monitoring devices and assessed them
against the REASSURED criteria. We found that currently available devices do
not meet these criteria and are not adapted be used for self-monitoring in
resource constraints settings.Click here for additional data file.Supplemental material, sj-docx-1-dst-10.1177_1932296821997909 for
Technologies for Diabetes Self-Monitoring: A Scoping Review and
Assessment Using the REASSURED Criteria by Jessica Hanae Zafra-Tanaka,
David Beran, Beatrice Vetter, Rangarajan Sampath and Antonio
Bernabe-Ortiz in Journal of Diabetes Science and TechnologyClick here for additional data file.Supplemental material, sj-docx-2-dst-10.1177_1932296821997909 for
Technologies for Diabetes Self-Monitoring: A Scoping Review and
Assessment Using the REASSURED Criteria by Jessica Hanae Zafra-Tanaka,
David Beran, Beatrice Vetter, Rangarajan Sampath and Antonio
Bernabe-Ortiz in Journal of Diabetes Science and TechnologyClick here for additional data file.Supplemental material, sj-docx-3-dst-10.1177_1932296821997909 for
Technologies for Diabetes Self-Monitoring: A Scoping Review and
Assessment Using the REASSURED Criteria by Jessica Hanae Zafra-Tanaka,
David Beran, Beatrice Vetter, Rangarajan Sampath and Antonio
Bernabe-Ortiz in Journal of Diabetes Science and TechnologyClick here for additional data file.Supplemental material, sj-docx-4-dst-10.1177_1932296821997909 for
Technologies for Diabetes Self-Monitoring: A Scoping Review and
Assessment Using the REASSURED Criteria by Jessica Hanae Zafra-Tanaka,
David Beran, Beatrice Vetter, Rangarajan Sampath and Antonio
Bernabe-Ortiz in Journal of Diabetes Science and Technology
Authors: Rabab Z Jafri; Courtney A Balliro; Firas El-Khatib; Michele M Maheno; Mallory A Hillard; Alexander O'Donovan; Rajendranath Selagamsetty; Hui Zheng; Edward R Damiano; Steven J Russell Journal: Diabetes Technol Ther Date: 2020-11 Impact factor: 6.118
Authors: Emma Louise Klatman; Alicia Josephine Jenkins; Muhammad Yakoob Ahmedani; Graham David Ogle Journal: Lancet Diabetes Endocrinol Date: 2018-07-30 Impact factor: 32.069
Authors: Andrea C Tricco; Erin Lillie; Wasifa Zarin; Kelly K O'Brien; Heather Colquhoun; Danielle Levac; David Moher; Micah D J Peters; Tanya Horsley; Laura Weeks; Susanne Hempel; Elie A Akl; Christine Chang; Jessie McGowan; Lesley Stewart; Lisa Hartling; Adrian Aldcroft; Michael G Wilson; Chantelle Garritty; Simon Lewin; Christina M Godfrey; Marilyn T Macdonald; Etienne V Langlois; Karla Soares-Weiser; Jo Moriarty; Tammy Clifford; Özge Tunçalp; Sharon E Straus Journal: Ann Intern Med Date: 2018-09-04 Impact factor: 25.391
Authors: J Kropff; D Bruttomesso; W Doll; A Farret; S Galasso; Y M Luijf; J K Mader; J Place; F Boscari; T R Pieber; E Renard; J H DeVries Journal: Diabetes Obes Metab Date: 2014-09-10 Impact factor: 6.577