Safety of using real-time sensor glucose values for treatment decisions in adolescents with poorly controlled type 1 diabetes mellitus: a pilot study.

Larry A Fox1, Emilie Balkman1, Kim Englert1, Jobayer Hossain2, Nelly Mauras1.   

Abstract

BACKGROUND: This study explored the safety of using real-time sensor glucose (SG) data for treatment decisions in adolescents with poorly controlled type 1 diabetes.
METHODS: Ten adolescents with type 1 diabetes, HbA1c ≥9% on insulin pumps were admitted to the clinical research center and a continuous glucose sensor was inserted. Plasma glucose was measured at least hourly using Yellow Springs Instrument's (YSI) glucose analyzer. Starting at dinner, SG rather than YSI was used for treatment decisions unless YSI was <70 mg/dL (<3.9 mmol/L) or specific criteria indicating SG and YSI were very discordant were met. Participants were discharged after lunch the next day.
RESULTS: Ten participants (seven males; 15.2-17.8 year old) completed the study. The range of differences between high glucose correction doses using SG vs YSI for calculations was -2 (SG < YSI dose) to +1 (SG > YSI dose); this difference was two units in only 2 of 23 correction doses given (all SG < YSI dose). There were five episodes of mild hypoglycemia in two patients, two of which occurred after using SG for dose calculations. There was no severe hypoglycemia and no YSI glucose >350 mg/dL (19.4 mmol/L). Mean (±SE) pre- and postmeal YSI glucose were 163 ± 11 and 183 ± 12 mg/dL (9.1 ± 0.6 and 10.2 ± 0.7 mmol/L), respectively.
CONCLUSION: Use of real-time continuous glucose monitoring for treatment decisions was safe and did not result in significant over- or undertreatment. Use of SG for treatment decisions under supervised inpatient conditions is a suitable alternative to repeated fingerstick glucose monitoring. Outpatient studies using SG in real-time are needed.
© 2016 John Wiley & Sons A/S. Published by John Wiley & Sons Ltd.

Entities:  

Keywords:  continuous glucose monitoring; pediatrics; type 1 diabetes

Mesh:

Substances:

Year:  2016        PMID: 27435145      PMCID: PMC5250611          DOI: 10.1111/pedi.12404

Source DB:  PubMed          Journal:  Pediatr Diabetes        ISSN: 1399-543X            Impact factor:   4.866


  24 in total

1.  Improved glycemic control in poorly controlled patients with type 1 diabetes using real-time continuous glucose monitoring.

Authors:  Dorothee Deiss; Jan Bolinder; Jean-Pierre Riveline; Tadej Battelino; Emanuele Bosi; Nadia Tubiana-Rufi; David Kerr; Moshe Phillip
Journal:  Diabetes Care       Date:  2006-12       Impact factor: 19.112

2.  Day and Night Closed-Loop Control Using the Integrated Medtronic Hybrid Closed-Loop System in Type 1 Diabetes at Diabetes Camp.

Authors:  Trang T Ly; Anirban Roy; Benyamin Grosman; John Shin; Alex Campbell; Salman Monirabbasi; Bradley Liang; Rie von Eyben; Satya Shanmugham; Paula Clinton; Bruce A Buckingham
Journal:  Diabetes Care       Date:  2015-06-06       Impact factor: 19.112

3.  Current state of type 1 diabetes treatment in the U.S.: updated data from the T1D Exchange clinic registry.

Authors:  Kellee M Miller; Nicole C Foster; Roy W Beck; Richard M Bergenstal; Stephanie N DuBose; Linda A DiMeglio; David M Maahs; William V Tamborlane
Journal:  Diabetes Care       Date:  2015-06       Impact factor: 19.112

4.  Accuracy of newer-generation home blood glucose meters in a Diabetes Research in Children Network (DirecNet) inpatient exercise study.

Authors:  Stuart A Weinzimer; Roy W Beck; H Peter Chase; Larry A Fox; Bruce A Buckingham; William V Tamborlane; Craig Kollman; Julie Coffey; Dongyuan Xing; Katrina J Ruedy
Journal:  Diabetes Technol Ther       Date:  2005-10       Impact factor: 6.118

5.  Experiences of continuous subcutaneous insulin infusion in pregnant women with type 1 diabetes during delivery from four Italian centers: a retrospective observational study.

Authors:  Raffaella Fresa; Natalia Visalli; Vincenzo Di Blasi; Vincenzo Cavallaro; Egle Ansaldi; Oria Trifoglio; Santina Abbruzzese; Marzia Bongiovanni; Mariano Agrusta; Angela Napoli
Journal:  Diabetes Technol Ther       Date:  2013-03-28       Impact factor: 6.118

6.  Microcirculation and its relation to continuous subcutaneous glucose sensor accuracy in cardiac surgery patients in the intensive care unit.

Authors:  Sarah E Siegelaar; Temo Barwari; Jeroen Hermanides; Peter H J van der Voort; Joost B L Hoekstra; J Hans DeVries
Journal:  J Thorac Cardiovasc Surg       Date:  2013-07-21       Impact factor: 5.209

7.  Randomized Evaluation of Glycemic Control in the Medical Intensive Care Unit Using Real-Time Continuous Glucose Monitoring (REGIMEN Trial).

Authors:  Christophe E M De Block; Jens Gios; Nina Verheyen; Begoña Manuel-y-Keenoy; Peter Rogiers; Philippe G Jorens; Cosimo Scuffi; Luc F Van Gaal
Journal:  Diabetes Technol Ther       Date:  2015-08-25       Impact factor: 6.118

8.  A randomized clinical trial to assess the efficacy and safety of real-time continuous glucose monitoring in the management of type 1 diabetes in young children aged 4 to <10 years.

Authors:  Nelly Mauras; Roy Beck; Dongyuan Xing; Katrina Ruedy; Bruce Buckingham; Michael Tansey; Neil H White; Stuart A Weinzimer; William Tamborlane; Craig Kollman
Journal:  Diabetes Care       Date:  2011-12-30       Impact factor: 19.112

9.  Real-time continuous glucose monitoring versus conventional glucose monitoring in critically ill patients: a systematic review study protocol.

Authors:  Weidong Zhu; Libing Jiang; Shouyin Jiang; Yuefeng Ma; Mao Zhang
Journal:  BMJ Open       Date:  2015-01-23       Impact factor: 2.692

10.  Accuracy of Continuous Glucose Monitoring Measurements in Normo-Glycemic Individuals.

Authors:  Abimbola A Akintola; Raymond Noordam; Steffy W Jansen; Anton J de Craen; Bart E Ballieux; Christa M Cobbaert; Simon P Mooijaart; Hanno Pijl; Rudi G Westendorp; Diana van Heemst
Journal:  PLoS One       Date:  2015-10-07       Impact factor: 3.240

View more
  2 in total

1.  Ambulatory glucose profile analysis of the juvenile diabetes research foundation continuous glucose monitoring dataset-Applications to the pediatric diabetes population.

Authors:  Gregory P Forlenza; Laura L Pyle; David M Maahs; Timothy C Dunn
Journal:  Pediatr Diabetes       Date:  2016-11-23       Impact factor: 4.866

Review 2.  Technologies for Diabetes Self-Monitoring: A Scoping Review and Assessment Using the REASSURED Criteria.

Authors:  Jessica Hanae Zafra-Tanaka; David Beran; Beatrice Vetter; Rangarajan Sampath; Antonio Bernabe-Ortiz
Journal:  J Diabetes Sci Technol       Date:  2021-03-09
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.