| Literature DB >> 35010055 |
Helena Torné-Morató1,2, Paolo Donati1, Pier Paolo Pompa1.
Abstract
Nowadays, there is an increasing interest in Point-of-care (POC) devices for the noninvasive glucose assessment. Despite the recent progress in glucose self-monitoring, commercially available devices still use invasive samples such as blood or interstitial fluids, and they are not equipment-free and affordable for the whole population. Here, we report a fully integrated strip test for the semi-quantitative detection of glucose in whole saliva. The colorimetric mechanism consists of an enzyme-mediated reshaping of multibranched gold nanoparticles (MGNPs) into nanospheres with an associated plasmonic shift and consequent blue-to-red color change, clearly detectable in less than 10 min.Entities:
Keywords: colorimetric sensor; gold nanoparticles; hyperglycemia; plasmonics; point-of-care
Year: 2021 PMID: 35010055 PMCID: PMC8746375 DOI: 10.3390/nano12010105
Source DB: PubMed Journal: Nanomaterials (Basel) ISSN: 2079-4991 Impact factor: 5.076
Figure 1Device structural and functional scheme. (a) Representative photographs of the whole device and its comprising parts. The detection zone is constituted by a polyamide membrane functionalized with MGNPs and GOx. The loading strip is a Fusion5® membrane, where the phosphate buffer and sodium iodide have been dried. The reaction zone and loading strip are held together on an adhesive backing card. (b) Sensing scheme and colorimetric results. This device is designed to recognize a healthy level of glucose and two increasing levels of hyperglycemia. Depending on the glucose concentration in the salivary fluid, MGNPs retain their surface structure (blue color), are partially (purple) or are totally (red) and reshaped into spherical GNPs. The colorimetric results can be read 10 min after the sample loading.
Figure 2Detection mechanism characterization. Reflectance spectroscopy (diffuse mode) spectra of (a) a healthy glucose level saliva sample (1.5 mg/dL) and (b) a sample spiked with D-(+)-Glucose to simulate a hyperglycemic state (9.5 mg/dL), monitored for 18 min. The sample with healthy levels of glucose maintained its reflectance spectra stable over the time, while the hyperglycemic sample reflectance peak shifted from 670 to 560 nm. Bottom: HR-SEM micrographs of polyamide membranes casted with homogenously dispersed MGNPs before the test (c) and reshaped GNP spheres after the test (d) (scale bars: low magnification 1 µm; high magnification 50 nm).
Figure 3Device optimization and performance. (a) Buffering and pH condition optimization of the detection zone using four healthy donors’ saliva samples. The non-buffered condition (H2O) was prepared with deionized water, while buffered conditions were prepared with phosphate buffer (300 mM) at pH 6, 6.5, 7, 7.5. Considering a basal glucose concentration of 1.5 mg/dL, samples were spiked with 2.5 and 4.5 mg/dL. (b) Device performance was assessed with seven saliva samples from healthy volunteers. Samples were spiked with 1, 2.5 and 4.5 mg/dL. One-way ANOVA and Tukey’s multiple comparison test were used to determinate the statistical significance (**** p < 0.0001 and ** p < 0.01). The insert image is representative of the assay results of four different glucose concentrations (1, 2, 3 and 4 correspond to basal glucose concentration 1.5 mg/dL and spikes of 1, 2.5, 4.5 mg/dL, respectively). The RGB coordinates values were always measured from pictures at t = 0 and t = 10 min with ImageJ software. In both experiments, all conditions were tested in triplicates.