Evaluation of a metal artefact reduction tool on different positions of a metal object in the FOV.

Polyane M Queiroz1, Gustavo M Santaella1, Thais D J da Paz1, Deborah Q Freitas1.   

Abstract

OBJECTIVES: To evaluate the action of a metal artefact reduction (MAR) tool when artefact-generator metal object is at different positions in the field of view (FOV).
METHODS: A cylindrical utility wax phantom, with a metal alloy sample inside, was made. The phantom was positioned centrally and peripherally in the FOV for image acquisition, with and without the MAR tool activation. The standard deviation values (image noise levels) from areas around the metal sample and the control area were obtained. The numbers were compared by Student's t-test (α = 0.05).
RESULTS: When the tool was activated, a significant difference of image noise was observed for central and peripheral positioning, for both control area (p = 0.0012) and metal area (p = 0.03), and a smaller level of noise was observed for images with phantoms in central positioning. A decrease in image noise with the tool activated was found only in phantoms with the metal object positioned centrally in the FOV.
CONCLUSIONS: For the MAR tool to be effective, the artefact-generator object needs to be in the central region of the FOV.

Entities:  

Keywords:  CBCT; artefacts

Mesh:

Substances:

Year:  2017        PMID: 27910700      PMCID: PMC5606272          DOI: 10.1259/dmfr.20160366

Source DB:  PubMed          Journal:  Dentomaxillofac Radiol        ISSN: 0250-832X            Impact factor:   2.419


  18 in total

1.  Effects of image artifacts on gray-value density in limited-volume cone-beam computerized tomography.

Authors:  Akitoshi Katsumata; Akiko Hirukawa; Shinji Okumura; Munetaka Naitoh; Masami Fujishita; Eiichiro Ariji; Robert P Langlais
Journal:  Oral Surg Oral Med Oral Pathol Oral Radiol Endod       Date:  2007-04-20

2.  Location dependency of the spatial resolution of cone beam computed tomography for dental use.

Authors:  Yoshihiro Ozaki; Hiroshi Watanabe; Yoshikazu Nomura; Eiichi Honda; Yasunori Sumi; Tohru Kurabayashi
Journal:  Oral Surg Oral Med Oral Pathol Oral Radiol       Date:  2013-11

3.  What is cone-beam CT and how does it work?

Authors:  William C Scarfe; Allan G Farman
Journal:  Dent Clin North Am       Date:  2008-10

4.  A new volumetric CT machine for dental imaging based on the cone-beam technique: preliminary results.

Authors:  P Mozzo; C Procacci; A Tacconi; P T Martini; I A Andreis
Journal:  Eur Radiol       Date:  1998       Impact factor: 5.315

5.  Analysis of intensity variability in multislice and cone beam computed tomography.

Authors:  Olivia Nackaerts; Frederik Maes; Hua Yan; Paulo Couto Souza; Ruben Pauwels; Reinhilde Jacobs
Journal:  Clin Oral Implants Res       Date:  2011-01-18       Impact factor: 5.977

6.  Contrast-to-noise ratio with different large volumes in a cone-beam computerized tomography machine: an in vitro study.

Authors:  Boulos Bechara; C Alex McMahan; William S Moore; Marcel Noujeim; Hassem Geha
Journal:  Oral Surg Oral Med Oral Pathol Oral Radiol       Date:  2012-11

7.  Cone beam CT scans with and without artefact reduction in root fracture detection of endodontically treated teeth.

Authors:  B Bechara; C Alex McMahan; W S Moore; M Noujeim; F B Teixeira; H Geha
Journal:  Dentomaxillofac Radiol       Date:  2013-03-21       Impact factor: 2.419

8.  Assessment of metal artefact reduction around dental titanium implants in cone beam CT.

Authors:  A Parsa; N Ibrahim; B Hassan; K Syriopoulos; P van der Stelt
Journal:  Dentomaxillofac Radiol       Date:  2014-08-19       Impact factor: 2.419

9.  Artifacts in CT: recognition and avoidance.

Authors:  Julia F Barrett; Nicholas Keat
Journal:  Radiographics       Date:  2004 Nov-Dec       Impact factor: 5.333

10.  Reduction of scatter-induced image noise in cone beam computed tomography: effect of field of view size and position.

Authors:  Ruben Pauwels; Reinhilde Jacobs; Ria Bogaerts; Hilde Bosmans; Soontra Panmekiate
Journal:  Oral Surg Oral Med Oral Pathol Oral Radiol       Date:  2015-10-20
View more
  15 in total

1.  Do the number of basis images and metal artifact reduction affect the production of artifacts near and far from zirconium dental implants in CBCT?

Authors:  Eduarda Helena Leandro Nascimento; Hugo Gaêta-Araujo; Rocharles Cavalcante Fontenele; Nicolly Oliveira-Santos; Christiano Oliveira-Santos; Deborah Queiroz Freitas
Journal:  Clin Oral Investig       Date:  2021-02-24       Impact factor: 3.573

2.  Influence of acquisition parameters on the magnitude of cone beam computed tomography artifacts.

Authors:  Deborah Queiroz Freitas; Rocharles Cavalcante Fontenele; Eduarda Helena Leandro Nascimento; Taruska Ventorini Vasconcelos; Marcel Noujeim
Journal:  Dentomaxillofac Radiol       Date:  2018-06-27       Impact factor: 2.419

3.  Effect of anatomical region on the formation of metal artefacts produced by dental implants in cone beam computed tomographic images.

Authors:  Alessiana Helena Machado; Karolina Aparecida Castilho Fardim; Camila Furtado de Souza; Bruno Salles Sotto-Maior; Neuza Maria Souza Picorelli Assis; Karina Lopes Devito
Journal:  Dentomaxillofac Radiol       Date:  2018-01-10       Impact factor: 2.419

4.  Are metal artefact reduction algorithms effective to correct cone beam CT artefacts arising from the exomass?

Authors:  Amanda Pelegrin Candemil; Benjamin Salmon; Deborah Queiroz Freitas; Glaucia Maria Bovi Ambrosano; Francisco Haiter-Neto; Matheus Lima Oliveira
Journal:  Dentomaxillofac Radiol       Date:  2019-01-28       Impact factor: 2.419

5.  Quantitative assessment of variation in CBCT image technical parameters related to CBCT detector lateral-offset position.

Authors:  Gustavo Machado Santaella; Pedro Luiz Rosalen; Polyane Mazucatto Queiroz; Francisco Haiter-Neto; Ann Wenzel; Rubens Spin-Neto
Journal:  Dentomaxillofac Radiol       Date:  2019-09-06       Impact factor: 2.419

6.  Influence of scan mode (partial/full rotations) and FOV size in the formation of artefacts in cone beam CT.

Authors:  Eliana Dantas da Costa; Polyane Mazucatto Queiroz; Gustavo Machado Santaella; Ana Lúcia Alvares Capelozza; Glaucia Maria Bovi Ambrosano; Deborah Queiroz Freitas
Journal:  Dentomaxillofac Radiol       Date:  2019-01-11       Impact factor: 2.419

7.  CBCT image artefacts generated by implants located inside the field of view or in the exomass.

Authors:  Husniye Demirturk Kocasarac; Lisa J Koenig; Gulbahar Ustaoglu; Matheus Lima Oliveira; Deborah Queiroz Freitas
Journal:  Dentomaxillofac Radiol       Date:  2021-07-29       Impact factor: 2.419

8.  Artefacts at different distances from titanium and zirconia implants in cone-beam computed tomography: effect of tube current and metal artefact reduction.

Authors:  Arthur Xavier Maseti Mancini; Matheus Urias Cruz Santos; Hugo Gaêta-Araujo; Camila Tirapelli; Ruben Pauwels; Christiano Oliveira-Santos
Journal:  Clin Oral Investig       Date:  2021-02-05       Impact factor: 3.573

9.  Influence of size of field of view (FOV), position within the FOV, and scanning mode on the detection of root fracture and observer's perception of artifacts in CBCT images.

Authors:  Martina Gerlane de Oliveira Pinto; Saulo L Sousa Melo; Fernanda Clotilde Mariz Suassuna; Luiz Eduardo Marinho; José Bruno da Silva Leite; Andre Ulisses Dantas Batista; Patrícia Meira Bento; Daniela Pita Melo
Journal:  Dentomaxillofac Radiol       Date:  2021-02-10       Impact factor: 3.525

10.  The Efficacy of Metal Artifact Reduction Mode in Cone-Beam Computed Tomography Images on Diagnostic Accuracy of Root Fractures in Teeth with Intracanal Posts.

Authors:  Zahra Dalili Kajan; Mehran Taramsari; Negar Khosravi Fard; Farnoosh Khaksari; Forough Moghasem Hamidi
Journal:  Iran Endod J       Date:  2018
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.