Amal Jose1, A Saravana Kumar1, K N Govindarajan1, P Manimaran1.
Abstract
AIM: The current work aims to calculate dose area product (DAP) and to determine regional diagnostic reference level (DRL) for pediatric panoramic radiography in Tamil Nadu.Entities:
Keywords: Diagnostic reference level; dose area product; panoramic radiography; radiation protection
Year: 2020 PMID: 33487931 PMCID: PMC7810145 DOI: 10.4103/jmp.JMP_106_19
Source DB: PubMed Journal: J Med Phys ISSN: 0971-6203
Figure 1Regional distribution of the panoramic units used in the present study
Average exposure parameters, beam area used for different scanners, and the corresponding calculated dose area product
| Number of scanners | Make (model) of the scanners | Type of detectors | Beam area (cm2) | Mean exposure parameter | Calculated median DAP (mGycm2) | ||
|---|---|---|---|---|---|---|---|
| kVp | mA | Time (s) | |||||
| 1 | Villa Sistemi Medicali, Buccinasco (Rotograph Evo D) | Digital | 0.60×14.60 | 66 | 6 | 14 | 45 |
| 1 | Myray (Hyperion X-7) | Digital | 0.60×14.60 | 70 | 4 | 7 | 26 |
| 1 | Soredex (Xmind Novus) | Digital | 0.61×14.75 | 60 | 7 | 9 | 32 |
| 1 | Gendex Dental Systems (GXDP 700) | Digital | 0.60×15.10 | 66 | 5 | 14 | 36 |
| 1 | Soredex (X-Mind Pano D+) | Digital | 0.61×14.75 | 73 | 10 | 14 | 98 |
| 1 | M/s. Villa Sistemi Medicali, Buccinasco (Rotograph Plus) | CR | 0.60×14.3 | 70 | 10 | 17 | 138 |
| 1 | Trophy (CS 9300 C) | Digital | 0.5×11.9 | 70 | 10 | 14 | 39 |
| 1 | GME (Pantograph Digi-10) | CR | 0.61×14.75 | 70 | 10 | 17 | 91 |
| 1 | Xtronics Imaging Systems (Xtropan 2000) | CR | 0.60×15.00 | 65 | 12 | 17 | 128 |
| 2 | Soredex (Cranex Novus e) | Digital | 0.61×14.75 | 66 | 10 | 9 | 40 |
| 2 | Cefla Dental Group Italy (Newtom Go 3D) | Digital | 0.6×14.6 | 76 | 8 | 12 | 115 |
| 2 | The Samon Imaging Systems (Dentopan-10) | CR | 0.69×15.10 | 65 | 10 | 14 | 78 |
| 2 | Planmeca OY (Planmeca ProOne) | Digital | 0.6×14.6 | 66 | 7 | 9 | 89 |
| 3 | Genoray Co. Ltd. (Papaya) | Digital | 0.64×15.00 | 62 | 5 | 15 | 56 |
| 3 | Sirona Dental Systems (Orthophos XG 3) | Digital | 0.32×13 | 66 | 8 | 14 | 62 |
| 4 | Trophy (CS 8100) | Digital | 0.5×12 | 68 | 8 | 10 | 71 |
| 4 | Planmeca OY (Planmeca Proline EC) | CR | 0.5×13.6 | 67 | 9 | 17 | 92 |
| 6 | Vatech Global Co Ltd (Pax-I PCH-2500) | Digital | 0.60×15.04 | 68 | 8 | 12 | 62 |
| 6 | Asahi Roentgen Ind. Co. Limited (Auto IIIECM) | CR | 0.69×15.10 | 69 | 10 | 12 | 79 |
| 7 | Planmeca OY (Planmeca Promax) | Digital | 0.60×14.70 | 61 | 6 | 15 | 72 |
| 8 | Sirona Dental Systems (Orthophos XG 5) | Digital | 0.32×13 | 63 | 8 | 14 | 47 |
| 17 | Trophy (Kodak 8000) | Digital | 0.4×12.8 | 68 | 8 | 13 | 52 |
DAP: Dose area product, CR: Computed radiography, GME: General medical equipments
Figure 2Piranha dosimeter is attached at the detector side of the panoramic unit
Figure 3Comparison of machine dose area product and calculated dose area product
Figure 4Calculated dose area product values and proposed diagnostic reference level for different panoramic units
Mean exposure parameters, dose area product range, and third quartile values obtained from the present study
| Type of Unit | Number of readings | Mean kV (SD) | Mean mA (SD) | Mean exposure time (s) (SD) | DAP range (mGycm2) | Mean DAP (mGycm2) | Calculated DRL (mGycm2) |
|---|---|---|---|---|---|---|---|
| CR | 300 (15×20) | 68 (6) | 10 (1) | 14 (2) | 55-139 | 90 | 106 |
| Digital | 1200 (60×20) | 66 (4) | 8 (2) | 13 (2) | 11-148 | 59 | 68 |
| Total | 1500 (75*20) | 67 (4) | 8 (2) | 13 (2) | 11-148 | 65 | 82 |
SD: Standard deviation, DAP: Dose area product, DRL: Diagnostic reference level, SD: Standard deviation, CR: Computed radiography
Comparison of the present study diagnostic reference level and other study diagnostic reference level
| Authors | Year of study | Country | DRL mGycm2 (third quartile) |
|---|---|---|---|
| Tierris | 2004 | Greece | 77 |
| Poppe | 2007 | Germany | 75.4 |
| Hart | 2009 | Great Britain | 82 |
| HPA[ | 2010 | UK | 67 |
| Kim | 2014 | Korea | 95.9 |
| Hodolli | 2019 | Kosovo | 73 |
| Present study | 2019 | India | 82 |
DRL: Diagnostic reference level