Literature DB >> 33487931

Assessment of Regional Pediatric Diagnostic Reference Levels for Panoramic Radiography Using Dose Area Product.

Amal Jose1, A Saravana Kumar1, K N Govindarajan1, P Manimaran1.   

Abstract

AIM: The current work aims to calculate dose area product (DAP) and to determine regional diagnostic reference level (DRL) for pediatric panoramic radiography in Tamil Nadu.
MATERIALS AND METHODS: In this study, DAP was calculated after finding the product of air kerma on the detector side of scanner with the corresponding exposed area. The obtained DAP values were further analyzed, and DRL was calculated using Microsoft Excel. The study was carried out with routine pediatric exposure parameters.
RESULTS: The obtained mean, range, and third quartile values for pediatric panoramic radiography are found to be 65 mGycm2, 11-148 mGycm2, and 82 mGycm2, respectively. The proposed DRL is comparable with the other countries' DRL.
CONCLUSION: Based on the results of the present study, it was observed that there exists a wide difference in mean doses among the panoramic scanners. The variation in radiation doses between the clinics/hospitals and similar scanners suggests a large potential for optimization of panoramic procedures. Copyright:
© 2020 Journal of Medical Physics.

Entities:  

Keywords:  Diagnostic reference level; dose area product; panoramic radiography; radiation protection

Year:  2020        PMID: 33487931      PMCID: PMC7810145          DOI: 10.4103/jmp.JMP_106_19

Source DB:  PubMed          Journal:  J Med Phys        ISSN: 0971-6203


INTRODUCTION

Globally, 480 million diagnostic dental examinations are performed yearly.[1] Although the radiation dose associated with dental radiography is comparatively less, the patient may undergo a dental radiography diagnosis process many times during their life span. While considering the health effects associated with dental radiography, the cumulative doses have to be estimated.[2] Diagnostic reference level (DRL) is a tool used to optimize the radiation exposure to a level appropriate for the medical imaging task. DRLs can be used for all modes of diagnostic radiology. Several methods have been used for assessing the panoramic reference level.[3456789101112] Dose width product method used by Napier[3] can be related to dose area product (DAP) by multiplying it with the beam height. Lee et al.[13] and Doyle et al.[14] used a solid-state dosimeter for dose measurements in panoramic radiography. Establishing reference levels for dental radiography will ensure a safer diagnosis procedure from the patient's viewpoint. Many radiographers and dentists knew that optimization of the exposure parameter is a key tool for dose reduction. Thus, a reference dose level is required for safe practice. It is an effective method for the significant reduction in collective dose, especially for repeated procedures and in the procedures containing more radiosensitive patients, like children. The present work, initiated by PSG Institute of Medical Sciences and Research in consultation with and grant from the Atomic Energy Regulatory Board (AERB), is a logical extension of the previous study,[15] conducted for the adult panoramic procedures in Tamil Nadu as a part of the establishment of national DRL for dental radiography for the country. Thus, the objective of our study was to calculate DAP for pediatric panoramic radiography to propose DRL and compare it with other country DRL.

MATERIALS AND METHODS

To establish a regional DRL, it is necessary to sample as many dental hospitals/clinics as possible in the region. The hospitals included should be evenly spread across all geopolitical zones of the region. These should encompass all government-owned X-ray centers and major private practitioners since they admit the highest number of patients. The choice of hospitals/clinics for the study was based on convenience and their willingness as opposed to a purely random selection of hospitals. This constraint was accepted because of practical constraints on the time and resources available for the study. The 75 panoramic scanners that have been used for the current study are spread over 18 major cities in Tamil Nadu, as shown in Figure 1. Out of the total, 60 panoramic units used in this study were equipped with a direct digital (digital) image capturing system (charge-coupled devices [CCDs]) and the remaining 15 units were used indirect digital image systems that use computed radiography (CR) cassette (phosphor storage plate) as image receptors. CCD type of image capturing systems helps to get a digital image directly after exposure is made.
Figure 1

Regional distribution of the panoramic units used in the present study

Regional distribution of the panoramic units used in the present study A calibrated solid-state dosimeter (Piranha 557, RTI Electronics, Sweden) was used to measure the air kerma from the detector side of the panoramic unit. A questionnaire was prepared, and data about the exposure parameters routinely used for the pediatric procedures, radiation safety status, periodic quality assurance (QA) status, and the detector type used in the panoramic scanners were collected. Table 1 summarizes the panoramic scanners, its beam area, and mean routine pediatric exposure parameters used for the present study.
Table 1

Average exposure parameters, beam area used for different scanners, and the corresponding calculated dose area product

Number of scannersMake (model) of the scannersType of detectorsBeam area (cm2)Mean exposure parameterCalculated median DAP (mGycm2)

kVpmATime (s)
1Villa Sistemi Medicali, Buccinasco (Rotograph Evo D)Digital0.60×14.606661445
1Myray (Hyperion X-7)Digital0.60×14.60704726
1Soredex (Xmind Novus)Digital0.61×14.75607932
1Gendex Dental Systems (GXDP 700)Digital0.60×15.106651436
1Soredex (X-Mind Pano D+)Digital0.61×14.7573101498
1M/s. Villa Sistemi Medicali, Buccinasco (Rotograph Plus)CR0.60×14.3701017138
1Trophy (CS 9300 C)Digital0.5×11.970101439
1GME (Pantograph Digi-10)CR0.61×14.7570101791
1Xtronics Imaging Systems (Xtropan 2000)CR0.60×15.00651217128
2Soredex (Cranex Novus e)Digital0.61×14.756610940
2Cefla Dental Group Italy (Newtom Go 3D)Digital0.6×14.676812115
2The Samon Imaging Systems (Dentopan-10)CR0.69×15.1065101478
2Planmeca OY (Planmeca ProOne)Digital0.6×14.6667989
3Genoray Co. Ltd. (Papaya)Digital0.64×15.006251556
3Sirona Dental Systems (Orthophos XG 3)Digital0.32×136681462
4Trophy (CS 8100)Digital0.5×126881071
4Planmeca OY (Planmeca Proline EC)CR0.5×13.66791792
6Vatech Global Co Ltd (Pax-I PCH-2500)Digital0.60×15.046881262
6Asahi Roentgen Ind. Co. Limited (Auto IIIECM)CR0.69×15.1069101279
7Planmeca OY (Planmeca Promax)Digital0.60×14.706161572
8Sirona Dental Systems (Orthophos XG 5)Digital0.32×136381447
17Trophy (Kodak 8000)Digital0.4×12.86881352

DAP: Dose area product, CR: Computed radiography, GME: General medical equipments

Average exposure parameters, beam area used for different scanners, and the corresponding calculated dose area product DAP: Dose area product, CR: Computed radiography, GME: General medical equipments After the data collection, the dosimeter was connected to the computer using “OCEAN CONNECT 2014 software (RTI Electronics, Sweden)“ by Bluetooth or wired connection. The dosimeter was placed on the detector side of the scanner parallel to the slit and positioned visually with the slit [Figure 2].
Figure 2

Piranha dosimeter is attached at the detector side of the panoramic unit

Piranha dosimeter is attached at the detector side of the panoramic unit After the proper positioning of the dosimeter, QA test was performed for each machine, and the results were compared with the AERB-specified QA tolerance limit to ensure the proper working of the scanner as well as to ensure the accurate position of the dosimeter.[15] The experimental methods were based on Lee et al.[13] study and the National Radiological Protection Board assessment of panoramic X-ray sets proposed by Napier.[3] The air kerma was measured for a routine pediatric exposure cycle in the absence of the subject. However, the measured air kerma dose is integrated over the movement of the X-ray beam across the dosimeter. The total air kerma obtained from the Piranha dosimeter is multiplied with the corresponding beam area to calculate DAP with the respective scanners. Most of the digital units were displayed DAP either in the console monitor or in the panoramic scanner itself after every exposure. To verify the consistency of DAP with and without the patient, the machine DAP was analyzed for both the manner and no significant differences were observed. As per ICRP 135 recommendations, the median value was calculated from the obtained range of DAP for proposing the DRL.[16] The third quartile value was calculated by Microsoft Excel software (Microsoft Corporation, Washington, USA) using the formula “PERCENTILE (array, k),“ where the array is the list of median calculated DAP values and k is 0.75 in the present study.

RESULTS AND DISCUSSION

The accuracy of calculated DAPs was analyzed by comparing it with the displayed DAP. From Figure 3, it can be observed that the difference between the calculated and measured DAPs has complied well within ±18%. The observed variation may be attributed because of the difference in the type of dosimeter, experimental technique, time slab of measurement, beam quality, and tube filtration.
Figure 3

Comparison of machine dose area product and calculated dose area product

Comparison of machine dose area product and calculated dose area product Figure 4 shows the calculated DAP values and the proposed DRL assessed from 75 dental panoramic X-ray scanners. The horizontal line illustrates the proposed DRL of the present study. The study proposes 82 mGycm2 as DRL for pediatric panoramic dental radiography in the Tamil Nadu region. Further, a wide range of DAP discloses the difference in the techniques used at various clinics/hospitals. It may be ascribed to the variation in facial sizes of pediatric patients, X-ray operator experience, preset exposure parameters, beam quality, inherent filters used, manufacturing period, beam area, and type of detectors.
Figure 4

Calculated dose area product values and proposed diagnostic reference level for different panoramic units

Calculated dose area product values and proposed diagnostic reference level for different panoramic units Table 2 shows the number of readings, mean exposure parameters, DAP range and DRL values of CR, and digital and total panoramic scanners obtained from the present study. The number of readings is the product of the total number of units and total air kerma measurements performed for each unit (20 air kerma measurements for each unit). It was noticed that digital DAP values varied from 11 mGycm2 to 148 mGycm2. Almost 14-fold difference was observed between the minimum and maximum digital DAP values, whereas there is only a 2.5-fold difference between the minimum and maximum values of CR units. The minimum DAP for a digital unit was observed from a scanner that was operating at 60.0 kVp, 2.0 mA, and 8.8 s, whereas comparatively higher exposure parameters (78.0 kVp, 8.0 mA, and 12.7 s) were noted from the digital unit with maximum DAP. Compared to digital units, the difference in exposure parameters between the minimum DAP (60.0 kVp, 10.0 mA, and 12.0 s) and maximum DAP (85.0 kVp, 10.0 mA, and 12.0 s) observed from CR type of detectors was comparatively less. Few digital units were used to make a change in the shape and size of the focal trough automatically while selecting the pediatric mode. All panoramic scanners included in the present study feature preset exposure parameters (separately for adult and pediatric) that can modify manually to fit various patient sizes. However, some of the clinics/hospitals were found using adult exposure parameters or altered adult exposure parameters for pediatric cases. This trend was mainly observed in scanners equipped with digital detectors which were operated by the radiographers who were not professionally qualified.
Table 2

Mean exposure parameters, dose area product range, and third quartile values obtained from the present study

Type of UnitNumber of readingsMean kV (SD)Mean mA (SD)Mean exposure time (s) (SD)DAP range (mGycm2)Mean DAP (mGycm2)Calculated DRL (mGycm2)
CR300 (15×20)68 (6)10 (1)14 (2)55-13990106
Digital1200 (60×20)66 (4)8 (2)13 (2)11-1485968
Total1500 (75*20)67 (4)8 (2)13 (2)11-1486582

SD: Standard deviation, DAP: Dose area product, DRL: Diagnostic reference level, SD: Standard deviation, CR: Computed radiography

Mean exposure parameters, dose area product range, and third quartile values obtained from the present study SD: Standard deviation, DAP: Dose area product, DRL: Diagnostic reference level, SD: Standard deviation, CR: Computed radiography Tierris et al.[7] and Poppe et al.[8] have assessed pediatric DRL using DAP meter after attaching it on the X-ray tube head at the exit slit of the beam in the presence and absence of patients, respectively. Tierris et al.[7] determined the mean DAP as 68 mGycm2 and proposed 77 mGycm2 as DRL, and Poppe et al.[8] calculated the mean DAP as 59.3 mGycm2 and found 75.4 mGycm2 as DRL. The proposed DRL value in Great Britain (Hart et al.[9]), the UK (HPA[10]), Korea (Kim et al.[11]), and Kosovo (Hodolli et al.[12]) were 82 mGycm2, 67 mGycm2, 95.9 mGycm2, and 73 mGycm2, respectively. The DAP proposed in the present study is similar to Hart et al.,[9] slightly higher than Tierris et al.,[7] Poppe et al.,[8] G. Hodolli et al.,[12] and HPA report[10] and lower than Kim et al.[11] reported values. The slight variation was observed because of the difference in the type of dosimeter used for the measurement, manufacturing period, and time slab of the measurement used by other investigators referred to here. The summary of the comparison between the other country DRL and the present study DRL is shown in Table 3.
Table 3

Comparison of the present study diagnostic reference level and other study diagnostic reference level

AuthorsYear of studyCountryDRL mGycm2 (third quartile)
Tierris et al.[7]2004Greece77
Poppe et al.[8]2007Germany75.4
Hart et al.[9]2009Great Britain82
HPA[10]2010UK67
Kim et al.[11]2014Korea95.9
Hodolli et al.[12]2019Kosovo73
Present study2019India82

DRL: Diagnostic reference level

Comparison of the present study diagnostic reference level and other study diagnostic reference level DRL: Diagnostic reference level A notable finding in the present study is the difference in minimum and maximum DAPs observed for the digital type of detectors. Many clinics and hospitals having CR types of detectors expose the patient with comparable or even lower doses than direct digital systems. During patient exposure, overexposure is difficult to observe in digital detectors by the operators as it would not create an argumentative impact on the image.[8] Image noise can also be controlled by increasing the exposure parameter. Thus, there is always an affinity to improve the image quality by increasing the dose.[8] This study agrees further with that inclination of increasing dose to achieve image quality. However, more consideration has to be paid toward dose optimization in digital radiography. Frequent training programs to radiographers and dentists are suggested regarding the dose optimization and radiation protection tools. It is also suggested to conduct dose surveys and QA at regular intervals and also after altering the equipment and methods that affect the patient dose levels. However, there is a limitation to the present study. Considering this as a preliminary study for proposing pediatric panoramic DRL in India, the collection of data from a large number of panoramic units installed at different dental facilities in Tamil Nadu over 2 years was limited. A longer timeframe would have allowed more data collection and supported this study.

CONCLUSION

In this study, the proposed DRL for pediatric panoramic dental radiography was 82 mGycm2. More than 80% of the clinics in the current study were working with digital detectors. The DRL obtained in this study is comparable with other countries' DRLs. However, DAP at many clinics/hospitals was comparatively higher. Considering the radiosensitivity of the child to ionizing radiation, the optimization of radiation dose is required at many clinics/hospitals. Although, after considering this as the first study in establishing pediatric panoramic dental DRL in India, many future studies are suggested at different states and regions for establishing criteria for taking quality images with optimal doses.

Financial support and sponsorship

This work was supported by the Atomic Energy Regulatory Board (AERB) of India and PSG Institute of Medical Sciences and Research, Coimbatore.

Conflicts of interest

There are no conflicts of interest.
  12 in total

1.  Reference doses for dental radiography.

Authors:  I D Napier
Journal:  Br Dent J       Date:  1999-04-24       Impact factor: 1.626

2.  Reference doses in dental radiodiagnostic facilities.

Authors:  L González; E Vañó; R Fernández
Journal:  Br J Radiol       Date:  2001-02       Impact factor: 3.039

3.  Reference dose levels for dental panoramic radiography in Gwangju, South Korea.

Authors:  Jae-Seo Lee; Young-Hee Kim; Suk-Ja Yoon; Byung-Cheol Kang
Journal:  Radiat Prot Dosimetry       Date:  2010-08-11       Impact factor: 0.972

4.  Techniques for measurement of dose width product in panoramic dental radiography.

Authors:  P Doyle; C J Martin; J Robertson
Journal:  Br J Radiol       Date:  2006-02       Impact factor: 3.039

5.  Exposure and dose in panoramic radiology.

Authors:  A Goldstein
Journal:  Med Phys       Date:  1998-06       Impact factor: 4.071

6.  Radiation exposure to children in intraoral dental radiology.

Authors:  H K Looe; A Pfaffenberger; N Chofor; F Eenboom; M Sering; A Rühmann; A Poplawski; K Willborn; B Poppe
Journal:  Radiat Prot Dosimetry       Date:  2006-06-17       Impact factor: 0.972

7.  Measurement of dose in panoramic dental radiology.

Authors:  J R Williams; A Montgomery
Journal:  Br J Radiol       Date:  2000-09       Impact factor: 3.039

8.  National reference doses for common radiographic, fluoroscopic and dental X-ray examinations in the UK.

Authors:  D Hart; M C Hillier; B F Wall
Journal:  Br J Radiol       Date:  2008-10-13       Impact factor: 3.039

9.  Dose area product reference levels in dental panoramic radiology.

Authors:  Christine E Tierris; Emmanuel N Yakoumakis; George N Bramis; Evangelos Georgiou
Journal:  Radiat Prot Dosimetry       Date:  2004-07-20       Impact factor: 0.972

10.  Assessment of Adult Diagnostic Reference Levels for Panoramic Radiography in Tamil Nadu Region.

Authors:  Amal Jose; A Saravana Kumar; K N Govindarajan; B Devanand; N Elango
Journal:  J Med Phys       Date:  2019-12-11
View more
  1 in total

1.  Assessment of Regional Diagnostic Reference Levels in Dental Radiography in Tamil Nadu.

Authors:  Amal Jose; A Saravana Kumar; K N Govindarajan; Sunil Dutt Sharma
Journal:  J Med Phys       Date:  2022-03-31
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.