Abstract
AIM: The aim of this study was to calculate dose area product (DAP) and to determine diagnostic reference level (DRL) for adult panoramic procedures in Tamil Nadu.Entities:
Keywords: Dental radiography; diagnostic reference level; dose area product; radiation protection
Year: 2019 PMID: 31908390 PMCID: PMC6936199 DOI: 10.4103/jmp.JMP_77_19
Source DB: PubMed Journal: J Med Phys ISSN: 0971-6203
Average exposure parameters used for different scanners and beam area
| Number of scanners | Name of the scanner | Beam area (cm2) | kVp | mA | Time (s) |
|---|---|---|---|---|---|
| 1 | Newtom Go 3D | 0.60*14.6 | 78 | 8 | 12.7 |
| 1 | Rotograph Evo D | 0.60*14.60 | 72 | 6 | 14.4 |
| 1 | Villa Sistemi, Rotograph plus | 0.60*14.3 | 78 | 10 | 17.0 |
| 1 | HYPERION X-7 | 0.60*14.60 | 67 | 5 | 9.1 |
| 1 | Xmind Novus | 0.61*14.75 | 70 | 7 | 9.0 |
| 1 | Gendex Dental Systems | 0.60*15.10 | 70 | 10 | 16.0 |
| 1 | X-MIND PANO D+ | 0.61*14.75 | 73 | 10 | 17.6 |
| 1 | CS 9300 | 0.50*14.9 | 75 | 12 | 14.3 |
| 1 | GME Pantograph 10 | 0.61*14.75 | 75 | 10 | 17.0 |
| 1 | Xtropan 2000 | 0.60*15.00 | 70 | 12 | 17.0 |
| 2 | ADVAPEX - DENTOPAN-10 | 0.69*15.10 | 70 | 10 | 14.0 |
| 2 | Planmeca ProOne | 0.60*14.6 | 69 | 7 | 9.8 |
| 2 | Soredex | 0.61*14.75 | 73 | 10 | 9.0 |
| 3 | Planmeca proline -EC | 0.50*13.6 | 66 | 11 | 16.0 |
| 3 | ORTHOPHOS XG 3 | 0.325*13 | 67 | 13 | 14.1 |
| 3 | Genoray Co. Ltd, EXTOR-P | 0.64*15.00 | 72 | 11 | 12.0 |
| 4 | CS 8100 | 0.50*13.12 | 73 | 11 | 10.7 |
| 5 | ASAHI IIIECM roentgen | 0.69*15.10 | 73 | 10 | 12.0 |
| 6 | Vatech Pax-i | 0.60*15.04 | 72 | 9 | 13.2 |
| 6 | Planmeca Promax | 0.60*14.70 | 67 | 9 | 15.8 |
| 6 | ORTHOPHOS XG 5 | 0.325*13 | 66 | 12 | 14.1 |
| 15 | Kodak 8000 | 0.40*12.80 | 71 | 10 | 13.7 |
Figure 1Graphical representation shown in Piranha Ocean software after inaccurate positioning (a) and accurate positioning (b) of dosimeter
Figure 2Panoramic Piranha holder with Piranha dosimeter
Figure 3Pictorial representation of experimental procedure
Figure 4Screenshot of the reading obtained in Ocean 2014 Connect software for exposure parameters of 70 kVp, 10 mA and 13.9 s
Figure 5Comparison of machine dose area product and calculated dose area product
Figure 6Proposed adult diagnostic reference level and calculated dose area product values of different panoramic units
Mean exposure parameters along with range, mean and 3rd quartile of dose area product values for digital and analogue types of units
| Type of unit | Exposure mean (kV) | Mean (mA) | Mean exposure time (s) | DAP range (mGycm2) | Mean DAP (mGycm2) | DRL (mGycm2) |
|---|---|---|---|---|---|---|
| Analogue | 71.2 | 10.3 | 14.4 | 65-165 | 105.9 | 125.5 |
| Digital | 70.1 | 10.0 | 13.2 | 41-148 | 90.7 | 112.4 |
| Total | 70.3 | 10.0 | 13.5 | 41-165 | 93.6 | 114.3 |
DRL: Diagnostic reference level, DAP: Dose area product
Figure 7Comparison of other country panoramic diagnostic reference level and the present study diagnostic reference level
Mean exposure parameters for scanners which were operated by professionally and not professionally qualified radiographers
| Type | kVp | mA | Time (s) | Average DAP |
|---|---|---|---|---|
| Scanners operated by professionally qualified and experienced radiographers | 67 | 8 | 13 | 74 |
| Scanners NOT operated by professionally qualified and experienced radiographers | 71 | 11 | 14 | 102 |
DAP: Dose area product