Exposure and dose in panoramic radiology.

A Goldstein1.   

Abstract

Panoramic radiology entrance skin exposure (ESE) was measured for a Siemens Orthopantomograph 10 E using a thimble chamber and TLDs. Thimble chamber measurements are shown to be inaccurate for the moving, narrow beam panoramic scan geometry due to partial volume and partial charged-particle equilibrium effects. For 75 KVp, 8 mA panoramic scans the TLD measured beam central ray ESE on the scan central plane (patient midsagittal plane) was an approximately constant 20 mR for the first 7.5 cm from the entrance slit. A theoretical model of central plane ESE for a fixed rotation center agrees very well with the TLD measurements and is consistent with all of the known ESE (and dose) properties near rotation centers. Head phantom skin dose measurements demonstrate the dependence of skin dose on exact equipment mode of use and patient anatomy. Central plane, beam central ray ESE is judged to be a convenient, universal measure of panoramic radiology radiation output.

Entities:  

Mesh:

Year:  1998        PMID: 9650194     DOI: 10.1118/1.598285

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  4 in total

1.  Influence of lead apron shielding on absorbed doses from panoramic radiography.

Authors:  D Rottke; L Grossekettler; K Sawada; P Poxleitner; D Schulze
Journal:  Dentomaxillofac Radiol       Date:  2013-10-30       Impact factor: 2.419

2.  Dosimetry analysis of panoramic-imaging devices in different-sized phantoms.

Authors:  Muizz A Wahid; Ella Choi; David S MacDonald; Nancy L Ford
Journal:  J Appl Clin Med Phys       Date:  2017-02-25       Impact factor: 2.102

3.  Assessment of Regional Pediatric Diagnostic Reference Levels for Panoramic Radiography Using Dose Area Product.

Authors:  Amal Jose; A Saravana Kumar; K N Govindarajan; P Manimaran
Journal:  J Med Phys       Date:  2020-10-13

4.  GafChromic XR-QA film in testing panoramic dental radiography.

Authors:  Robert Y Chu; Tin Lam; Yongguang Liang
Journal:  J Appl Clin Med Phys       Date:  2007-04-19       Impact factor: 2.102

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.