Reference doses in dental radiodiagnostic facilities.

L González1, E Vañó, R Fernández.   

Abstract

The present work describes an experiment undertaken using thermoluminescent dosemeters to obtain a local reference dose level in orthopantomography, based on patient measurements in 11 dentistry installations. In addition, a critical audit of the dose recommended by the International Atomic Energy Agency for intraoral X-ray diagnostics is performed, based on data gathered from over 300 intraoral X-ray facilities. The provisional local reference levels proposed are 0.7 mGy entrance surface dose at the occipital region for orthopantomography and 3.5 mGy entrance surface dose for intraoral radiology.

Entities:  

Mesh:

Year:  2001        PMID: 11718387     DOI: 10.1259/bjr.74.878.740153

Source DB:  PubMed          Journal:  Br J Radiol        ISSN: 0007-1285            Impact factor:   3.039


  3 in total

1.  Radiation dose saving through the use of cone-beam CT in hearing-impaired patients.

Authors:  N Faccioli; M Barillari; S Guariglia; E Zivelonghi; A Rizzotti; R Cerini; R Pozzi Mucelli
Journal:  Radiol Med       Date:  2009-09-30       Impact factor: 3.469

2.  Assessment of Regional Pediatric Diagnostic Reference Levels for Panoramic Radiography Using Dose Area Product.

Authors:  Amal Jose; A Saravana Kumar; K N Govindarajan; P Manimaran
Journal:  J Med Phys       Date:  2020-10-13

3.  Estimation of Surface Radiation Dosage to Thyroid Gland and Lower Abdomen While Using Intraoral Periapical Radiography: A Phantom Study.

Authors:  Krishnamachari Janani; T Malarkodi; Sathasivasubramanian Sankarapandian
Journal:  Cureus       Date:  2021-11-18
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.