Vicki T Taasti1, Christian Bäumer2, Christina V Dahlgren3, Amanda J Deisher4, Malte Ellerbrock5, Jeffrey Free6, Joanna Gora7, Anna Kozera8, Antony J Lomax9, Ludovic De Marzi10, Silvia Molinelli11, Boon-Keng Kevin Teo12, Patrick Wohlfahrt13,14, Jørgen B B Petersen1, Ludvig P Muren1, David C Hansen1, Christian Richter13,14.
Abstract
BACKGROUND ANDEntities:
Keywords: CT scan protocol; Dual-energy CT; Hounsfield look-up-table; Inter-centre comparison; Particle treatment planning; Stopping-power ratio prediction
Year: 2018 PMID: 33458385 PMCID: PMC7807627 DOI: 10.1016/j.phro.2018.04.006
Source DB: PubMed Journal: Phys Imaging Radiat Oncol ISSN: 2405-6316
Fig. 1Anatomical sites treated at the different particle therapy centres. The category “Other” includes cranio-spinal treatments and irradiations close to the spine.
CT scanner specification used at the twelve different institutions. Abbreviations: BHC: Beam-hardening correction; FBP: Filtered back-projection. 2 ×/4 ×: Two/four CT scanners of the same model.
| Institution number | Scanner model | Scan and reconstruction protocol | ||
|---|---|---|---|---|
| Tube potential (kVp) | Reconstruction | |||
| Type | Type of BHC | |||
| 1 | Siemens Sensation Open | 120, 140 | FBP | Bone-BHC for head |
| 2 | Philips Brilliance Big Bore | 120 | FBP | Water-BHC |
| 3 | Philips Brilliance Big Bore | 120, 140 | FBP | Water-BHC |
| 4 | Toshiba Aquilion LB Siemens Definition AS Siemens Sensation Open GE LightSpeed RT16 | 100, 120, 135 100, 120 120 120, 140 | FBP Iterative FBP FBP | Water-BHC Bone-BHC for head/Water-BHC for abdomen Bone-BHC for head/Water-BHC for abdomen Water-BHC |
| 5 | Siemens Definition AS | 120 | FBP | Water-BHC |
| 6 | Siemens Sensation Open Siemens Definition – part of PET/CT Biograph40 Siemens Confidence Siemens Sensation Open | 120 | FBP | Bone-BHC for head |
| 7 | Siemens Definition AS | 120 | FBP | Water-BHC |
| 8 | 4 × Siemens Definition AS | 120 | Iterative | Water-BHC |
| 9 | Siemens Sensation | 120 | FBP | Water-BHC |
| 10 | Siemens Sensation Open Philips TF Big Bore PET-CT | 120 | FBP | Water-BHC |
| 11 | 2 × Siemens Definition AS | SECT: 120 | Iterative | Bone-BHC |
| 12 | Siemens Definition AS | 120 for adults | Iterative | Not provided |
Specification of CT-to-SPR conversion (HLUT). Institution numbers are the same as in Table 1. Abbreviations: N/A: not applicable; WET: water-equivalent thickness. TPS: Treatment planning system.
| Institution number | Calibration procedure | Calibration phantom | Reference human tissue data | # of line segments | # of HLUT | Same HLUT for multiple scanners | Validation procedure |
|---|---|---|---|---|---|---|---|
| 1 | Tissue substitutes | CIRS | N/A | 10 | 6 | No | WET of tissue substitutes |
| 2 | Tissue substitutes | CIRS | N/A | 11 | 6 (3 for adults and 3 for children) | No | WET of tissue substitutes and animal tissues |
| 3 | Tissue substitutes | Gammex | N/A | 2 | 3 | No | WET of tissue substitutes and animal tissue, stoichiometric calibration |
| 4 | Stoichiometric | CIRS + Gammex | 7 | 12 (4 different anatomical regions; different CT scanners and different kVp settings) | No | WET of custom-made phantom of organic skull, fat and air | |
| 5 | Stoichiometric | CIRS | 11 | 2 (head and body) | No | No. Only relative inter-comparison of calibrations at different hospitals using radiological phantoms | |
| 6 | Stoichiometric | Gammex | 7 | 2 (head and body) | No | WET of animal tissues and human bone tissue. Also WET for half of an Alderson head phantom and a frozen pig head | |
| 7 | Stoichiometric | CIRS | 3 | 1 | No | WET of tissue substitutes and head phantom | |
| 8 | Stoichiometric | CIRS | 8 | 1 | Yes | WET of porcine and bovine tissues, and for two beam paths through a human cadaver in a cranium (bone + brain) and a thigh (muscle + adipose) region | |
| 9 | Stoichiometric | CIRS | 4 | 1 | No | WET of animal tissues | |
| 10 | Stoichiometric | CIRS | 7 | 1 | Yes | None | |
| 11 | Tissue substitutes + stoichiometric | CIRS + Gammex | 3 | 1 SECT + 1 DECT | Yes | WET of different phantoms (tissue substitutes) | |
| 12 | Stoichiometric | CIRS | NA | 3 (head, body and children) | No | Proton radiography of animal tissues |
Fig. 2Scored importance for SPR calculation in proton treatment planning. Importance score definition: 1: most important, 6: least important – more techniques could be scored equally. The median is shown by the red lines, and the blue boxes extend from the 25% to the 75% percentile. Abbreviation: NA: No answer. (For interpretation of the references to colour in this figure legend, the reader is referred to the web version of this article.)