Evaluation of automated pre-treatment and transit in-vivo dosimetry in radiotherapy using empirically determined parameters.

Evy Bossuyt1, Reinhilde Weytjens1, Daan Nevens1,2, Sarah De Vos1, Dirk Verellen1,2.   

Abstract

BACKGROUND AND
PURPOSE: First reports on clinical use of commercially automated systems for Electronic Portal Imaging Device (EPID)-based dosimetry in radiotherapy showed the capability to detect important changes in patient setup, anatomy and external device position. For this study, results for more than 3000 patients, for both pre-treatment verification and in-vivo transit dosimetry were analyzed.
MATERIALS AND METHODS: For all Volumetric Modulated Arc Therapy (VMAT) plans, pre-treatment quality assurance (QA) with EPID images was performed. In-vivo dosimetry using transit EPID images was analyzed, including causes and actions for failed fractions for all patients receiving photon treatment (2018-2019). In total 3136 and 32,632 fractions were analyzed with pre-treatment and transit images respectively. Parameters for gamma analysis were empirically determined, balancing the rate between detection of clinically relevant problems and the number of false positive results.
RESULTS: Pre-treatment and in-vivo results depended on machine type. Causes for failed in-vivo analysis included deviations in patient positioning (32%) and anatomy change (28%). In addition, errors in planning, imaging, treatment delivery, simulation, breath hold and with immobilization devices were detected. Actions for failed fractions were mostly to repeat the measurement while taking extra care in positioning (54%) and to intensify imaging procedures (14%). Four percent initiated plan adjustments, showing the potential of the system as a basis for adaptive planning.
CONCLUSIONS: EPID-based pre-treatment and in-vivo transit dosimetry using a commercially available automated system efficiently revealed a wide variety of deviations and showed potential to serve as a basis for adaptive planning.
© 2020 The Authors.

Entities:  

Keywords:  In-vivo; Transit dosimetry

Year:  2020        PMID: 33458354      PMCID: PMC7807610          DOI: 10.1016/j.phro.2020.09.011

Source DB:  PubMed          Journal:  Phys Imaging Radiat Oncol        ISSN: 2405-6316


Introduction

In-vivo dosimetry is recommended in radiotherapy to avoid major treatment errors and to improve accuracy [1], [2], [3], [4], [5], [6], [7], [8], [9], [10], [11], [12]. In clinical routine, placing the traditional point detectors on the patient’s skin is not always feasible and requires additional setup time. Moreover, it is often not possible to use these in combination with Intensity Modulated Radiation Therapy (IMRT) or Volumetric Modulated Arc Therapy (VMAT) for which verification is limited to pre-treatment solutions. Since Electronic Portal Imaging Device (EPID) images contain dose information, many groups have investigated their use for radiotherapy dose measurement [12], [13], [14], [15], [16], [17], [18], [19], [20], [21], [22]. Pre-treatment dose verification and in-vivo dosimetry using EPIDs has become routine practice in a growing number of clinics in recent years. Mainstream acceptance, however, has been hampered by a lack of commercially available solutions. Fully automated EPID-based systems are now emerging making it feasible to perform dosimetric quality assurance (QA) on every field for every patient. For a comprehensive literature review on electronic portal imaging for radiotherapy dosimetry the reader is referred to van Elmpt et al. [23]. Olch and colleagues [22] published the first report of the clinical use of a commercially available automated system for EPID-based dosimetry. It presents results for 855 fractions using the first treatment fraction as baseline to compare with later fractions, showing the system’s capability to detect important changes in patient setup, anatomy and external device position. We present results for more than 30,000 fractions, of which most are based on absolute dose prediction (as opposed to comparing to baseline), investigating if this broadens the variety of detected deviations. There is currently no consensus on which parameters for gamma analysis are most appropriate for specific treatment sites or equipment used [12], [18], [22], [23]. A common set of parameters is not feasible for clinical practice. This study investigates the possibility to determine parameters empirically, balancing the rate between detection of clinically relevant problems and the number of false positive results.

Materials and methods

The study was conducted in the Iridium Kankernetwerk, Belgium with ten linear accelerators: three ‘old generation’ machines: Varian Clinacs; and seven ‘new generation’ machines: six TrueBeams and one TrueBeam STX (Varian Oncology Systems, Palo Alto, CA). Two treatment planning systems were used to create plans: RayStation 9A (RaySearch Laboratories, Stockholm, Sweden) for stereotactic plans (both intra and extra-cranial) and Eclipse v13.6 (Varian Medical Systems, Palo Alto, CA) for all other types of plans. An automated web-based system was installed early 2017, for both pre-treatment and in-vivo QA based on EPID measurements (PerFRACTIONTM, part of SunCHECKTM, Sun Nuclear Corporation). Clinical use of the software (version 1.7) started in October 2017 on two of the machines, gradually including all other machines until in February 2018 the software was used for all machines and all patients (version 2.0 in May 2018, version 2.2 in November 2019), pushing DICOM data to the server, actively retrieving images, and automatically calculating results in the background. Failing fractions were checked daily by the responsible physicists and physicians, and appropriate actions were taken. For this study, calculated results and comments on causes and actions for failed fractions were extracted from the system using structured query language (SQL) scripts. A clinical validation of the efficiency and performance on detection of errors was carried out, along with an analysis of causes and actions taken for failed fractions. Pre-treatment QA (‘Fraction 0′) was performed for all VMAT plans. Preparation consisted of exporting plans to PerFRACTION, assigning a ‘QA template’ containing (treatment site specific) tolerance levels, and scheduling integrated images. All images were automatically retrieved after measurement and compared with calculated predicted dose images using a Global Gamma analysis of 3%/2mm, a low dose threshold of 10% and a passing tolerance level of 98%. All treatment plans were supervised by a physicist, including the results of the ‘Fraction 0′ measurement. If the automatic verification failed, plan and results were checked in detail and re-calculated with 3%/3mm and/or 4%/2mm. If still failing, results were discussed with a physician. In total 3136 fractions from 2686 patients were analyzed. Transit EPID dosimetry with integrated images was performed in 43% of all treatment fractions: as standard, for the first three days of treatment and weekly thereafter (or more if fraction results failed); daily in the case of short treatment schedules. All causes of failed fractions were analyzed. If failure was due to more than one reason, the cause with the greatest contribution to the failure was assigned. In the start-up period between October 2017 and August 2018, in-vivo results were calculated using ‘relative’ (image-to-image) comparison, using one of the first acquired images as a reference (baseline). After this period, a first analysis was performed in which 8621 fractions with transit EPID measurements were included [24]. In September 2018 ‘absolute’ verification was introduced allowing comparison of images to calculated predicted dose. Breasts with nodal irradiation remained being calculated relative to a baseline image, because a longitudinal panel shift was needed for analysis of the entire breast and predicted dose could not be calculated with these panel shifts. A second analysis was performed for patients treated between September 2018 and August 2019 with in total 24,011 fractions from 3671 patients using transit EPID measurements. The latter will be the main subject of this report. During the start-up period we tried to determine ‘appropriate’ parameters for gamma analysis, balancing the rate between detection of clinically relevant problems and the number of false positive results. Using the ‘QA templates’ in the software, ‘standard’ parameters for gamma analysis were assigned. We chose to create templates depending on the treatment site, as the treatment site largely determines the used immobilization and treatment technique, and acceptable tolerance. We ended up with seven different sets of analysis parameters for non-stereotactic plans and three for stereotactic plans, summarized in Table 1.
Table 1

Summary of empirically determined parameters for gamma analysis of in-vivo transit dosimetry results.

Normalization (Local/Global)Dose Difference Tolerance (%)Distance Tolerance (mm)Low Dose Threshold (%)Passing Tolerance Level (%)
BreastLocal762090
Whole Brain RadioTherapyLocal732090
Palliative treatmentsLocal752093
H&N and BrainGlobal332095
RectumGlobal552093
Other treatment sites with maskGlobal532095
Other treatment sites without mask (including lung, pelvis, abdomen,…)Global552095
Stereotactic 1 mmLocal1012095
Stereotactic 2 mmLocal1022095
Stereotactic 3 mmLocal1032095
Summary of empirically determined parameters for gamma analysis of in-vivo transit dosimetry results. A Global Gamma Analysis with 3% dose difference tolerance, 3 mm distance tolerance, 10% low dose threshold and 95% passing tolerance level was used at the start, using one of the first acquired images as baseline. However, these levels turned out to be too strict for most plans with an excessive number of false positives. The challenge is to detect large discrepancies that could have clinical consequences as opposed to minor issues and therefore find a good balance between the detection of clinically relevant problems without increasing the rate of both false negatives, or false positives. It was decided to use the Planning Target Volume (PTV) margin as distance tolerance, a patient shift within the PTV margin being considered as clinically acceptable. The dose difference tolerance was increased to 5% for most treatments except Brain and H&N, because the variation in patient thickness produced failed fractions that were clinically irrelevant. It was increased to 7% for breast, Whole Brain Radiation Therapy (WBRT) and palliative treatments and to 10% for stereotactic treatments (the very small fields caused acceptable large dose differences in the high dose regions). The low dose threshold was increased to 20%, as the 10% threshold produced fractions failing in low dose regions close to field edges (considered clinically less relevant). A Global Gamma Analysis was applied for VMAT plans and a Local Gamma Analysis for 3D plans and stereotactic plans. This was based on reports showing that global analysis masks some problems, but also that local analysis magnifies irrelevant errors at low doses [25]. Since integrated images for VMAT plans consist of a summation of dose from all angles, they are less sensitive to geometric errors and local analysis tends to magnify irrelevant errors for large VMAT plans. The passing tolerance levels were set at 95% except for rectum and palliative treatments (93%) and for breast and WBRT treatments (90%). Due to the skin flash, a small shift in breast and WBRT treatments causes a lot of points to fail and 95% was considered too strict. In rectum treatments 93% provided a value that seemed to balance the detection of clinically relevant air bubbles against irrelevant ones.

Results

Pre-treatment QA proved to be time-efficient, requiring five to ten minutes per plan, including preparation, measurement and verification. Fig. 1 shows results for pre-treatment QA, with 88% of fractions passing the automatically calculated gamma analysis of 3%/2mm with a 98% tolerance level (Fig. 1a). In 8% of cases, results were scored as ‘acceptable’ by a physicist with a gamma analysis of 4%/2mm or 3%/3mm. In 2% of cases, results were determined to be acceptable after discussion with a physician. In less than 1% of cases, results were unacceptable and the plan was adjusted. A difference was observed between results for different machine types, Fig. 1b and 1c respectively. On the new generation machines, 97% of measurements passed with the standard 3%/2mm analysis, compared to 77% on the old generation machines.
Fig. 1

Pre-treatment QA results: a) All results, b) Only old generation machine type results, c) Only new generation machine type results.

Pre-treatment QA results: a) All results, b) Only old generation machine type results, c) Only new generation machine type results. Of the 8621 fractions of the first in-vivo analysis after the start-up period, 1523 failed (18%), of which 7% were false positives and 11% caused by patient related issues. Of the 24,011 fractions of the second analysis, 3766 failed (16%), of which 6% were false positives and 10% caused by patient related issues. Of these 6% false positives, more than 4% were measurements performed on old generation machines, less than 2% on new generation machines. Fig. 2 shows an analysis of these measured fractions, classified per treatment site and/or technique.
Fig. 2

Analysis of measured in-vivo fractions, classified per treatment site and/or technique, indicating the number of passed fractions, the number of false positives and the number of failed fractions due to planning problems, deviations in patient positioning and changes in patient anatomy.

Analysis of measured in-vivo fractions, classified per treatment site and/or technique, indicating the number of passed fractions, the number of false positives and the number of failed fractions due to planning problems, deviations in patient positioning and changes in patient anatomy. Fig. 3 shows all causes and actions of the 3766 failed fractions of the second analysis in detail: false positives were mostly caused by interrupted beams (20%) and imager calibration issues (10%). Patient related failed fractions were in 32% of cases caused by a deviation in patient positioning and in 28% of cases by anatomy changes such as weight loss, deviations in bladder or rectal filling and tumor shrinkage. Actions most commonly taken were: repeating the measurement while taking extra care in positioning, breath hold technique, shoulder or arm position (54%) or adding extra imaging to obtain more information and to help differentiating between true and false positives (14%). In 24% of cases no extra actions were taken (e.g. acceptable for palliative cases or last fractions of treatment). In 4% of cases, plan adjustments were made. Fig. 4 shows all actions for failed fractions classified per treatment site and/or technique.
Fig. 3

Causes and actions for failed fractions in the second analysis: a) Causes: software problems including software bugs and irradiation on another (matched) machine with a different imager causing a deviation when comparing to a baseline of another machine; wrong imager position; problems with imager calibration; interrupted beams causing missing dose in the image; planning problems; deviations in patient position; changes in patient anatomy. b) Actions: taking a new measurement; adding extra pre-treatment imaging; plan adjustment; taking measures regarding patient preparation (e.g. bladder and rectum protocol); taking no action; adjusting the used tolerances.

Fig. 4

Analysis of the actions for failed fractions in the second analysis classified per treatment site and/or technique: taking a new measurement; adding extra pre-treatment imaging; plan adjustment; taking measures regarding patient preparation (e.g. bladder and rectum protocol); taking no action; adjusting used tolerances.

Causes and actions for failed fractions in the second analysis: a) Causes: software problems including software bugs and irradiation on another (matched) machine with a different imager causing a deviation when comparing to a baseline of another machine; wrong imager position; problems with imager calibration; interrupted beams causing missing dose in the image; planning problems; deviations in patient position; changes in patient anatomy. b) Actions: taking a new measurement; adding extra pre-treatment imaging; plan adjustment; taking measures regarding patient preparation (e.g. bladder and rectum protocol); taking no action; adjusting the used tolerances. Analysis of the actions for failed fractions in the second analysis classified per treatment site and/or technique: taking a new measurement; adding extra pre-treatment imaging; plan adjustment; taking measures regarding patient preparation (e.g. bladder and rectum protocol); taking no action; adjusting used tolerances. Examples of a wide variety of detected errors and deviations, together with the appropriate corrective actions, are illustrated in Appendix A, including decision charts for taking actions when a measurement is out of tolerance (Fig. A.1). Several of the discovered errors would have led to a dose difference of the total treatment of more than 5% if uncorrected. Some of them are also discussed in the Appendix. Examples of what could be detected when comparing to a baseline include: patient position (Fig. A.2), breast swelling (Fig. A.3), pneumonia (Fig. A.4), planning problems where the skin flash tool was not well used (Fig. A.5), unintended table shift (Fig. A.6) and tumor reduction (Fig. A7, Fig. A8). More errors could be caught when comparing to predicted dose, including planning issues (Fig. A.9), problems at simulation (Fig. A10, Fig. A11, Fig. A12), problems with patient preparation (Fig. A.13), errors during imaging and treatment (Fig. A.14), weight loss at the start of treatment (Fig. A.15), issues related to the breath hold technique (Fig. A.16) and problems with immobilization devices (Fig. A.17).
Fig. A1

Decision charts for when a measurement is out of tolerance. The first decision chart is meant for physicists including detecting false positives, the second decision chart is meant for physicians to take actions for patient related errors. Results are always compared with the available imaging. Taking Cone Beam Computed Tomography (CBCT) is often one of the first actions.

Fig. A2

Deviation in breast position. On the left, in-vivo software results. On the right, offline review images where the integrated images could be matched with the Digitally Reconstructed Radiograph (DRR).

Fig. A3

Breast swelling. On the left, in-vivo software results. On the right, offline review images where the integrated images could be matched with the DRR.

Fig. A4

Pneumonia. On the left, in-vivo software results. On the right, transversal views of planning CT and CBCT’s with the structures projected from the planning CT.

Fig. A5

Planning problem: skin flash tool not well used. On the left, in-vivo software results. On the right, Beam’s Eye View (BEV) of the planned dose in the planning system.

Fig. A6

Wrong table shift. On the left, in-vivo software results. On the right, offline review results showing a table shift of almost 7 cm before the last field.

Fig. A7

Tumor reduction. On the left, in-vivo software results. On the right, frontal and transversal views of the CBCT’s, with the structures projected from the planning CT.

Fig. A8

Tumor reduction and weight loss. On the top left, in-vivo software results. On the top right, sagittal and transversal views of the original plan. On the bottom left, sagittal view of the CBCT and transversal view of the original plan on the evaluation CT. On the bottom right, the change in Dose Volume Histogram for several structures (squares representing the original plan on the original CT and triangles the original plan on the evaluation CT).

Fig. A9

Planning problem not avoiding hip implant. On the top, transversal and frontal view of the plan on the planning CT. On the bottom, in-vivo software results.

Fig. A10

Problem where slow breathing induced artefacts masking large tumor movement. On the top left, in-vivo software results. On the bottom right, sagittal and frontal view of the planning CT.

Fig. A11

Simulated with the arms down, but planned with the arms up without notification. On the left, in-vivo software results. On the right, offline review images of setup fields and transversal view of the plan on the planning CT.

Fig. A12

Patient swallowed during simulation. On the left, in-vivo software results. On the right, offline review images of the CBCT, with the structures projected from the planning CT.

Fig. A13

Air in the rectum. On the left, in-vivo software results. On the right, transversal view of the CBCT, with the structures projected from the planning CT.

Fig. A14

Large table shift not applied after online match. On the left, in-vivo software results. On the right, offline review images of setup fields.

Fig. A15

Weight loss at first fraction. On the left, in-vivo software results. On the right, structures of the planning CT projected on a transversal view of the CBCT.

Fig. A16

Bad breathing technique. On the left, in-vivo software results. On the right, offline review setup image matched with the DRR.

Fig. A17

Belly board problem. On the left, in-vivo software results. On the top right, structures of the planning CT projected on a sagittal view of the CBCT.

Discussion

This study presents clinical results for more than 3000 patients, for EPID-based pre-treatment and in-vivo transit dosimetry. Analysis with an empirically determined set of parameters shows a wide variety of detected errors and deviations. Our measurements showed pre-treatment QA to be time-efficient (five to ten minutes) and revealing some deviations, with results differing between machine types. Extra verifications with Arc CHECK and point measurements showed most of the deviations to be acceptable and causes for failure mostly related to technical problems with both machines and imagers (e.g. the imager position of the aS1000 imagers not being corrected mechanically during VMAT and hence less stable than the aS1200 imagers). In addition, the more complex plans showed better results at new generation machines. It is known from log file analysis that MLC positioning errors substantially differ between machine types [26], [27]. Omitting pre-treatment verification for non-hypo-fractionated plans seems feasible, since in-vivo transit dosimetry comparing to predicted dose is performed, which would also indicate problems with patient plans or machine failures. However, if the patient has failing in-vivo fractions, it would require an extra ‘Fraction 0′ measurement to eliminate plan issues as the cause of failure. This would introduce an extra workflow and more manual work compared to performing the measurement for every VMAT plan before treatment as a standard procedure. In the future it might be possible to predict ‘difficult’ plans using complexity thresholds or Machine Learning. Properly generating and cleaning the large amount of current data will open the door for Machine Learning tools to filter out machine and specific issues before the actual QA is needed. Tolerance levels for in-vivo dosimetry were empirically determined in the start-up period, using an AMARA-principle: detecting errors, but only As Many As Reasonably Achievable, taking into account economic and societal factors. Economic factors include costs of in-vivo systems and time spent on measuring and analyzing results, including false positives. Societal factors include patient comfort, e.g. implanting detectors or placing detectors on the skin reduce patient comfort compared to deploying an imager during treatment. Finding a good balance will be different for every in-vivo system and every department. Comparing clinical results from multi-center collaborations might help to determine a set of generally accepted recommendations. An AMARA-principle can be based on a few pillars. The first is knowing the sensitivity, strengths and weaknesses of the system. This can influence decisions on the tolerance levels to be used, but also on decisions such as the level of image guidance to be used; e.g. in-vivo EPID dosimetry is quite sensitive to a change in patient thickness [28], but the sensitivity to patient position variations is dependent on treatment site [18]: in anatomical regions with few density differences like abdominal regions, a shift in patient’s position will have less effect on the transmitted radiation than in Head & Neck (H&N) regions. Also, sensitivity can depend on technique: e.g. VMAT could be less sensitive to detect patient position variations compared to IMRT, which in turn could be less sensitive than 3D conformal radiotherapy [19]. Secondly, reducing the number of false positives and introducing an automated probability score for being false positive, might make it less likely a real error will be classified as a false positive due to the observer becoming ‘error fatigued’. As a first step, a script has been introduced to identify false positives related to interrupted beams or incorrect positioning of the imager: a report is generated showing the number of detected Monitor Units in the image and the position of the imager compared with the planned values. The latter will be extended to produce suggestions for possible causes and actions in the near future. Finally, re-evaluating tolerance levels on a regular base, especially after introducing software or hardware changes or introducing new immobilization devices, imaging or treatment strategies, is necessary to re-assess if parameters can be made tighter to diminish the number of false negatives without adding false positives or clinically irrelevant errors. Software upgrades in which shifted imager positions can be taken into account or replacement of equipment, might also influence decisions on used tolerance levels. The latter is illustrated in the adjustments made to tolerance levels during the startup period to decrease the number of false positives (Table 1). Using these tolerance levels, a level of 16% failed fractions was reached, of which 6% were false positives and 10% were caused by patient related issues. This was judged to be an acceptable balance between the detection of clinically relevant problems and the number of false positive results. In the report of Olch et al. [22], tighter tolerance levels were used and the 3%/3mm analysis only produced 8% failed fractions. However, the study reports only 57 patients, all being treated on a new generation machine (of which we also reported less failed fractions) and strongly immobilized, probably producing a higher degree of positioning certainty (close to 3 mm). In our study, results of 3671 patients were analyzed, receiving treatments using various immobilization devices and imaging strategies, supporting the suggestion of determining tolerance levels according to an AMARA-principle allowing variation based on the changes in treatment strategies, desired sensitivity levels and/or machine type. Evaluating results on a regular basis offers important insights in the quality of treatments and indicates possible items for improvement, or helps departments to decide on future projects. An illustration is given in Fig. 2 where the largest number of positioning issues was related to extremities (12%), breast boost (10%), esophagus (9%) and Head & Neck (H&N) patients (9%). As a result, extra pre-treatment imaging was introduced for breast boosts and more attention was given to proper positioning and avoiding shoulders in treatment planning for H&N. The largest number of anatomy changes were seen for pelvis (10%), rectum (10%) and abdomen (9%). Earlier reports have shown that these regions might have a lower sensitivity to patient position [18], [28]. From Fig. 4 it can be observed that most plan adjustments occurred in lung (11%), abdominal (10%) and H&N treatments (9%), showing the software’s potential to serve as a basis for adaptive planning. However, further investigation is necessary to determine the number of false negatives and the limitations of the system. Currently, a study is being conducted to investigate the sensitivity in detecting H&N patients that require an adaptive plan during the course of treatment and to estimate the number of false negatives using current tolerances. In conclusion, a commercially available automated pre-treatment and in-vivo transit dosimetry system has been clinically implemented for all patients, efficiently revealing a wide variety of deviations using an empirically determined set of parameters for gamma analysis. Results show its potential to serve as a basis for adaptive planning and the number of false positive results dependent on machine type. Tolerance levels should be wisely chosen according to an AMARA-principle, balancing the rate between detection of clinically relevant problems and the number of false positive results.

Declaration of Competing Interest

The authors declare the following financial interests/personal relationships which may be considered as potential competing interests: Iridium Kankernetwerk is a member of the SunCHECK Customer Advisory Board and a Reference center for Sun Nuclear Corporation, but there has been no significant financial support for this work that could have influenced its outcome.
  21 in total

1.  Overview of 3-year experience with large-scale electronic portal imaging device-based 3-dimensional transit dosimetry.

Authors:  Ben J Mijnheer; Patrick González; Igor Olaciregui-Ruiz; Roel A Rozendaal; Marcel van Herk; Anton Mans
Journal:  Pract Radiat Oncol       Date:  2015-07-09

2.  Quantifying the performance of in vivo portal dosimetry in detecting four types of treatment parameter variations.

Authors:  C Bojechko; E C Ford
Journal:  Med Phys       Date:  2015-12       Impact factor: 4.071

Review 3.  A literature review of electronic portal imaging for radiotherapy dosimetry.

Authors:  Wouter van Elmpt; Leah McDermott; Sebastiaan Nijsten; Markus Wendling; Philippe Lambin; Ben Mijnheer
Journal:  Radiother Oncol       Date:  2008-08-14       Impact factor: 6.280

4.  Error detection during VMAT delivery using EPID-based 3D transit dosimetry.

Authors:  B Mijnheer; A Jomehzadeh; P González; I Olaciregui-Ruiz; R Rozendaal; P Shokrani; H Spreeuw; R Tielenburg; A Mans
Journal:  Phys Med       Date:  2018-10-11       Impact factor: 2.685

5.  Interfractional trend analysis of dose differences based on 2D transit portal dosimetry.

Authors:  L C G G Persoon; S M J J G Nijsten; F J Wilbrink; M Podesta; J A D Snaith; T Lustberg; W J C van Elmpt; F van Gils; F Verhaegen
Journal:  Phys Med Biol       Date:  2012-09-21       Impact factor: 3.609

6.  In vivo dosimetry during conformal radiotherapy: requirements for and findings of a routine procedure.

Authors:  J H Lanson; M Essers; G J Meijer; A W Minken; G J Uiterwaal; B J Mijnheer
Journal:  Radiother Oncol       Date:  1999-07       Impact factor: 6.280

Review 7.  In vivo dosimetry in external beam radiotherapy.

Authors:  Ben Mijnheer; Sam Beddar; Joanna Izewska; Chester Reft
Journal:  Med Phys       Date:  2013-07       Impact factor: 4.071

8.  Sensitivity study of an automated system for daily patient QA using EPID exit dose images.

Authors:  Audrey H Zhuang; Arthur J Olch
Journal:  J Appl Clin Med Phys       Date:  2018-03-06       Impact factor: 2.102

9.  EPID based in vivo dosimetry system: clinical experience and results.

Authors:  Sofia Celi; Emilie Costa; Claas Wessels; Alejandro Mazal; Alain Fourquet; Pascal Francois
Journal:  J Appl Clin Med Phys       Date:  2016-05-08       Impact factor: 2.102

10.  First Report of the Clinical Use of a Commercial Automated System for Daily Patient QA Using EPID Exit Images.

Authors:  Arthur J Olch; Kyle O'Meara; Kenneth K Wong
Journal:  Adv Radiat Oncol       Date:  2019-04-12
View more
  8 in total

1.  Comparing treatment uncertainty for ultra- vs. standard-hypofractionated breast radiation therapy based on in-vivo dosimetry.

Authors:  Yawo A C Fiagan; Evy Bossuyt; Melanie Machiels; Daan Nevens; Charlotte Billiet; Philip Poortmans; Thierry Gevaert; Dirk Verellen
Journal:  Phys Imaging Radiat Oncol       Date:  2022-05-13

Review 2.  X-change symposium: status and future of modern radiation oncology-from technology to biology.

Authors:  Stefanie Corradini; Maximilian Niyazi; Dirk Verellen; Vincenzo Valentini; Seán Walsh; Anca-L Grosu; Kirsten Lauber; Amato Giaccia; Kristian Unger; Jürgen Debus; Bradley R Pieters; Matthias Guckenberger; Suresh Senan; Wilfried Budach; Roland Rad; Julia Mayerle; Claus Belka
Journal:  Radiat Oncol       Date:  2021-02-04       Impact factor: 4.309

3.  Extending in aqua portal dosimetry with dose inhomogeneity conversion maps for accurate patient dose reconstruction in external beam radiotherapy.

Authors:  Igor Olaciregui-Ruiz; Julia-Maria Osinga-Blaettermann; Karen Ortega-Marin; Ben Mijnheer; Anton Mans
Journal:  Phys Imaging Radiat Oncol       Date:  2022-04-14

4.  Using eclipse scripting to fully automate in-vivo image analysis to improve treatment quality and safety.

Authors:  Ananta Raj Chalise; Casey Bojechko
Journal:  J Appl Clin Med Phys       Date:  2022-03-22       Impact factor: 2.243

5.  What is the optimal input information for deep learning-based pre-treatment error identification in radiotherapy?

Authors:  Cecile J A Wolfs; Frank Verhaegen
Journal:  Phys Imaging Radiat Oncol       Date:  2022-08-27

6.  First clinical experience with real-time portal imaging-based breath-hold monitoring in tangential breast radiotherapy.

Authors:  Elena N Vasina; Natalie Kong; Peter Greer; Jose Baeza Ortega; Tomas Kron; Joanna J Ludbrook; David Thwaites; Joerg Lehmann
Journal:  Phys Imaging Radiat Oncol       Date:  2022-08-13

7.  Verification of stereotactic radiosurgery plans for multiple brain metastases using a virtual phantom-based procedure.

Authors:  Juan-Francisco Calvo-Ortega; Peter B Greer; Sandra Moragues-Femenía; Miguel Pozo-Massó; Joan Casals-Farran
Journal:  Rep Pract Oncol Radiother       Date:  2022-07-29

8.  Reduction of systematic dosimetric uncertainties in volumetric modulated arc therapy triggered by patient-specific quality assurance.

Authors:  Anton Mans; Roel Rozendaal; Tomas Janssen; Eugène Damen; Jochem Kaas; Anke van Mourik; Ben Mijnheer
Journal:  Phys Imaging Radiat Oncol       Date:  2022-01-20
  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.