| Literature DB >> 31681865 |
Arthur J Olch1,2, Kyle O'Meara3, Kenneth K Wong1,2.
Abstract
PURPOSE: To characterize the clinical utility of a new commercially available system for daily patient treatment quality assurance using electronic portal imaging detector (EPID) exit dose images. METHODS AND MATERIALS: The PerFRACTION automated quality assurance system was used to acquire integrated EPID images for every field every day for 60 treatment courses for 57 patients. Four thousand seventy-nine field values from 855 fractions were analyzed. Gamma passing rates were computed by the system for each field daily. Passing rates and pass-fail status were recorded by treatment modality (intensity modulated radiation therapy or 3-dimensional conformal radiotherapy) and location. When failures occurred, an attempt was made to determine the reason.Entities:
Year: 2019 PMID: 31681865 PMCID: PMC6817722 DOI: 10.1016/j.adro.2019.04.001
Source DB: PubMed Journal: Adv Radiat Oncol ISSN: 2452-1094
Overview of the analysis of all DFVs, fields, fractions, and courses at both the 2%/2 mm and the 3%/3 mm gamma tolerances
| 2%, 2 mm | 3%, 3 mm | |
|---|---|---|
| Total DFV | 4079 | 4079 |
| Total DFV failed | 834 | 340 |
| % DFV failed (M1) | 20.5 | 8.3 |
| Total fields | 260 | 260 |
| Total fields failed | 61 | 21 |
| Field failed (M2), % | 23.5 | 8.1 |
| No. of fields with ≥1 DFV failure | 125 | 63 |
| Fields with ≥1 DFV failure (M3), % | 48.1 | 24.2 |
| Total fractions | 855 | 855 |
| Total fractions failed | 152 | 65 |
| Fraction failed (M4), % | 17.8 | 7.6 |
| Total courses | 60 | 60 |
| Total courses failed | 36 | 17 |
| Courses failed (M5), % | 60.0 | 28.3 |
Abbreviations: DFV = daily field value; M1-5 = the 5 metrics described in the Methods and Materials.
Data acquisition by location and modality
| Treatment location | No. of courses | No. of daily field values | No. of fields | No. of fractions |
|---|---|---|---|---|
| Brain | 33 | 3036 | 189 | 496 |
| Abdomen | 7 | 95 | 13 | 51 |
| Chest | 6 | 71 | 10 | 42 |
| Extremity | 2 | 51 | 5 | 25 |
| H&N | 7 | 519 | 30 | 117 |
| Pelvis | 2 | 91 | 5 | 40 |
| Spine | 3 | 216 | 8 | 84 |
| Treatment modality | ||||
| 3DCRT | 16 | 207 | 29 | 114 |
| IMRT | 44 | 3872 | 231 | 741 |
| Total | 60 | 4079 | 260 | 855 |
Abbreviations: 3DCRT = 3-dimensional conformal radiotherapy; H&N = head and neck; IMRT = intensity modulated radiation therapy.
Figure 1Field failure due to body position change. (a) Although there is an external device located within this field, the pattern of pixel failure best matches the shape and location of the shoulder. (b) The failing pixels in the region in which the patient's shoulder is located; slight adjustments in the positioning of the shoulder would alter the amount of radiation detected by the electronic portal imaging detector.
Figure 2Field failure due to internal anatomy change. The failing (orange) pixels (b) correspond with regions in the lungs in image. (a) The best explanation for gamma failures in this case would be internal changes, such as the density of the lungs or position of the diaphragm.
Treatment location and modality failure rate (%) at 2%/2 mm and 3%/3 mm by course, DFV, fraction, and field
| Treatment location | Courses failed, % | Total DFV fail rate | Fraction fail rate | Field fail rate | ||||
|---|---|---|---|---|---|---|---|---|
| 2,2 | 3,3 | 2,2 | 3,3 | 2,2 | 3,3 | 2,2 | 3,3 | |
| Brain | 57.6 | 21.2 | 18.1 | 5.8 | 13.2 | 4.0 | 18.0 | 5.1 |
| Abdomen | 42.9 | 42.9 | 44.2 | 32.1 | 50.0 | 42.9 | 50.0 | 25.0 |
| Chest | 83.3 | 66.7 | 53.9 | 18.5 | 53.4 | 18.8 | 75.0 | 16.7 |
| Extremity | 50.0 | 0.0 | 8.3 | 1.4 | 11.8 | 2.9 | 0.0 | 0.0 |
| H&N | 71.4 | 42.9 | 38.7 | 20.1 | 35.9 | 21.4 | 47.6 | 27.6 |
| Pelvis | 50.0 | 0.0 | 7.7 | 0.0 | 7.7 | 0.0 | 0.0 | 0.0 |
| Spine | 66.7 | 0.0 | 11.9 | 1.7 | 12.3 | 1.4 | 0.0 | 0.0 |
| Treatment modality | ||||||||
| 3DCRT | 81.3 | 56.3 | 56.7 | 24.7 | 55.2 | 29.2 | 65.6 | 25.0 |
| IMRT | 52.8 | 18.2 | 15.0 | 5.7 | 11.6 | 4.4 | 14.3 | 4.8 |
| Overall failure rates | 60.0 | 28.3 | 20.5 | 8.3 | 17.8 | 7.6 | 23.5 | 8.1 |
Abbreviations: 3DCRT = 3-dimensional conformal radiotherapy; DFV = daily field value; H&N = head and neck; IMRT = intensity modulated radiation therapy.
P values for 2%, 2 mm, 0.037 by location, <.0001 by modality.
Reason for DFV failures by treatment modality at 2%/2 mm tolerances
| Modality | Reason | No. of fields with ≥1 DFV failure | Fields with ≥1 DFV failure, % | No. of DFV failed | Total DFVs failed, % |
|---|---|---|---|---|---|
| 3DCRT | |||||
| BPC | 9 | 31.0 | 65 | 31.4 | |
| EDP | 0 | 0.0 | 0 | 0.0 | |
| U | 3 | 10.3 | 9 | 4.4 | |
| IAC | 10 | 34.5 | 38 | 18.4 | |
| IMRT | |||||
| BPC | 26 | 11.3 | 179 | 4.6 | |
| EDP | 28 | 12.2 | 134 | 3.5 | |
| U | 46 | 19.9 | 402 | 10.4 | |
| IAC | 3 | 1.3 | 7 | 0.2 |
Abbreviations: 3DCRT = 3-dimensional conformal radiation therapy; BPC = body position change; DFV = daily field value; EDP = external device position; IAC = internal anatomic change; IMRT = intensity modulated radiation therapy; U = unknown.