In vivo dosimetry during conformal radiotherapy: requirements for and findings of a routine procedure.

J H Lanson1, M Essers, G J Meijer, A W Minken, G J Uiterwaal, B J Mijnheer.   

Abstract

PURPOSE: Conformal radiotherapy requires accurate knowledge of the actual dose delivered to a patient. The impact of routine in vivo dosimetry, including its special requirements, clinical findings and resources, has been analysed for three conformal treatment techniques to evaluate its usefulness in daily clinical practice.
MATERIALS AND METHODS: Based on pilot studies, routine in vivo dosimetry quality control (QC) protocols were implemented in the clinic. Entrance and exit diode dose measurements have been performed during two treatment sessions for 378 patients having prostate, bladder and parotid gland tumours. Dose calculations were performed with a CT-based three-dimensional treatment planning system. In our QC-protocol we applied action levels of 2.5% for the prostate and bladder tumour group and 4.0% for the parotid gland patients. When the difference between the measured dose at the dose specification point and the prescribed dose exceeded the action level the deviation was investigated and the number of monitor units (MUs) adjusted. Since an accurate dose measurement was necessary, some properties of the on-line high-precision diode measurement system and the long-term change in sensitivity of the diodes were investigated in detail.
RESULTS: The sensitivity of all diodes decreased by approximately 7% after receiving an integrated dose of 10 kGy, for 4 and 8 MV beams. For 34 (9%) patients the difference between the measured and calculated dose was larger than the action level. Systematic errors in the use of a new software release of the monitor unit calculation program, limitations of the dose calculation algorithms, errors in the planning procedure and instability in the performance of the accelerator have been detected.
CONCLUSIONS: Accurate in vivo dosimetry, using a diode measurement system, is a powerful tool to trace dosimetric errors during conformal radiotherapy in the range of 2.5-10%, provided that the system is carefully calibrated. The implementation of an intensive in vivo dosimetry programme requires additional staff for measurements and evaluation. The patient measurements add only a few minutes to the total treatment time per patient and guarantee an accurate dose delivery, which is a prerequisite for conformal radiotherapy.

Entities:  

Mesh:

Year:  1999        PMID: 10577686     DOI: 10.1016/s0167-8140(99)00074-2

Source DB:  PubMed          Journal:  Radiother Oncol        ISSN: 0167-8140            Impact factor:   6.280


  7 in total

1.  Pretreatment verification of dose calculation and delivery by means of measurements with PLEXITOM™ phantom.

Authors:  Paweł Wołowiec; Paweł Kukołowicz; Krzysztof Lis
Journal:  Rep Pract Oncol Radiother       Date:  2013-02-04

2.  In patient dose reconstruction using a cine acquisition for dynamic arc radiation therapy.

Authors:  Angelo Piermattei; Andrea Fidanzio; Luigi Azario; Francesca Greco; Alessandra Mameli; Savino Cilla; Luca Grimaldi; Guido D'Onofrio; Boris Giuseppe Augelli; Gerardina Stimato; Diego Gaudino; Sara Ramella; Rolando D'Angelillo; Francesco Cellini; Lucio Trodella
Journal:  Med Biol Eng Comput       Date:  2009-02-17       Impact factor: 2.602

3.  Breast in vivo dosimetry by EPID.

Authors:  Andrea Fidanzio; Francesca Greco; Alessandra Mameli; Luigi Azario; Mario Balducci; Maria Antonietta Gambacorta; Vincenzo Frascino; Savino Cilla; Domenico Sabatino; Angelo Piermattei
Journal:  J Appl Clin Med Phys       Date:  2010-09-02       Impact factor: 2.102

4.  Errors in radiation oncology: a study in pathways and dosimetric impact.

Authors:  Eric E Klein; Robert E Drzymala; James A Purdy; Jeff Michalski
Journal:  J Appl Clin Med Phys       Date:  2005-08-12       Impact factor: 2.102

5.  A Monte Carlo and experimental investigation of the dosimetric behavior of low- and medium-perturbation diodes used for entrance in vivo dosimetry in megavoltage photon beams.

Authors:  Mohammad Amin Mosleh-Shirazi; Sareh Karbasi; Daryoush Shahbazi-Gahrouei; Shahram Monadi
Journal:  J Appl Clin Med Phys       Date:  2012-11-08       Impact factor: 2.102

6.  Evaluation of automated pre-treatment and transit in-vivo dosimetry in radiotherapy using empirically determined parameters.

Authors:  Evy Bossuyt; Reinhilde Weytjens; Daan Nevens; Sarah De Vos; Dirk Verellen
Journal:  Phys Imaging Radiat Oncol       Date:  2020-11-09

7.  Pre-clinical and clinical evaluation of the HYPERSCINT plastic scintillation dosimetry research platform for in vivo dosimetry during radiotherapy.

Authors:  Imke Schoepper; Sonja Dieterich; Earl Alonzo Trestrail; Michael Sean Kent
Journal:  J Appl Clin Med Phys       Date:  2022-02-21       Impact factor: 2.102

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.