| Literature DB >> 36092771 |
Elena N Vasina1, Natalie Kong2, Peter Greer1,2, Jose Baeza Ortega1, Tomas Kron3, Joanna J Ludbrook2, David Thwaites4, Joerg Lehmann1,2,4.
Abstract
Background and purpose: Real-time treatment monitoring with the electronic portal imaging device (EPID) can conceptually provide a more accurate assessment of the quality of deep inspiration breath-hold (DIBH) and patient movement during tangential breast radiotherapy (RT). A system was developed to measure two geometrical parameters, the lung depth (LD) and the irradiated width (named here skin distance, SD), along three user-selected lines in MV EPID images of breast tangents. The purpose of this study was to test the system during tangential breast RT with DIBH. Materials and methods: Measurements of LDs and SDs were carried out in real time. DIBH was guided with a commercial system using a marker block. Results from 17 patients were assessed. Mean midline LDs, <mLDs>, per tangent were compared to the planned mLDs; differences between the largest and smallest observed <mLDs> (<mSDs>) per tangent were calculated.Entities:
Keywords: DIBH; Deep inspiration breath-hold; EPID; Electronic portal imaging device; Internal anatomy; Intra-fraction motion; Lung depth; Real-time monitoring; Tangential breast radiotherapy
Year: 2022 PMID: 36092771 PMCID: PMC9450128 DOI: 10.1016/j.phro.2022.08.002
Source DB: PubMed Journal: Phys Imaging Radiat Oncol ISSN: 2405-6316
Fig. 1MV EPID image of a tangential breast field illustrating the lung depth (LD) in red and the skin distance (SD) in light blue. Both parameters are shown at the default locations (superior, midline, and inferior at 25%, 50% and 75% of the field height, respectively). These locations are user selectable and should be chosen to be clinically optimal based on the anatomy. (For interpretation of the references to colour in this figure legend, the reader is referred to the web version of this article.)
Fig. 2Workflow of real-time recording of MV frames with the C-DOG software and real-time measurements of LD (lung depth) and SD (skin distance) with the image processing software.
Fig. 3Lung depth (LD) and skin distance (SD) vs beam-on time during fraction 10 of Patient 1. Vertical lines separate the four beams. LDs and SDs are close to their planned values (shown by the horizontal lines of the same colour). Short beam-off instances are present in the second halves of beams 1 and 3 when the MLC leaves come in: this is seen as the gaps between the data points of LDs (beam-holds). When the MLC leaves block the patient’s skin, the measured SDs become 0 mm (data not shown).
Per tangent mean midline lung depth (
| Patient | <mLD> vs TW | Maximum of [<mLD> – planned mLD], mm | Number of fractions with <mLD> outside TW/Total number of fractions | Number of fractions with sEPID outside TW/Number of fractions with sEPID | ||
|---|---|---|---|---|---|---|
| Gantry angle 1* | Gantry angle 2 | Gantry angle 1* | Gantry angle 2 | |||
| 1 | Below & Above | −16.2 | 11/14 | 10/14 | 1/2 | 0/2 |
| 2 | Below | −6.8 | 7/10 | 4/12 | 7/10 [KV] | 4/12 |
| 3 | Below & Above | 10 | 5/9 | 4/10 | 0/1 | 2/2 |
| 4 | Below | −9.5 | 9/14 | 8/14 | 2/5 | 0/5 |
| 5 | Below | −10.3 | 9/10 | 9/10 | 0/1 | 0/1 |
| 6 | Above | 10.5 | 9/10 | 5/14 | NA | 0/4 |
| 7 | Below | −13.3 | 8/8 | 7/8 | ||
| 8 | Below | −7.9 | 9/11 | 6/11 | ||
| 9 | Below | −10.7 | 3/8 | 8/8 | 2/7 | 7/8 |
| 10 | Below & Above | −6.2 | 1/3 | 4/5 | 1/1 | 3/2 |
| 11 | Within | −2.3 | 0/12 | 0/12 | 0/12 | 0/12 |
| 12 | Below | −8.1 | 3/3 | 3/3 | ||
| 13 | Below | −7.1 | 5/12 | 4/12 | ||
| 14 | Below | −3.4 | 1/6 | 2/8 | NA | 1/2 |
| 15 | Above | 8.6 | 2/3 | 1/3 | ||
| 16 | Above | 3.4 | NA | 1/3 | NA | 1/3 |
| 17 | Below | −8.4 | 3/3 | 1/5 | 1/2 | 0/2 |
*Due to clinical imaging for position verification, some of the first beams were not available for observation for patients 1, 2, 3, 6, 10, 14, 16, and 17. NA: not available.
Fig. 4The differences between the measured and the planned LD and SD values (superior, midline, and inferior parameters are grouped together) for each beam of the tangential breast treatment of Patient 1. Plotted are the mean values with the error bars representing one standard deviation. The planned LDs and SDs are provided in Supplementary Table 1S.
Ranges of the mean midline LDs and SDs: Difference between largest and smallest mean (per tangent) measured throughout the treatment.
| Patient | Gantry angle 1 | Gantry angle 2 | ||
|---|---|---|---|---|
| <mLD>range | <mSD>range | <mLD>range | <mSD>range | |
| 1 | 20.5* | 18.1 | 18.7* | 20.4 |
| 2 | 4.7 | 3.9 | 3.0 | 3.3 |
| 3 | 12.5 | 10 | 10 | 8.5 |
| 4 | 11.8* | 14.2 | 10.5* | 17.2 |
| 5 | 8.4 | 10.3 | 8.7 | 10.2 |
| 6 | 7.7 | 8.5 | 9.9 | 9.7 |
| 7 | 8.3 | 6.8 | 11.8* | 8.4 |
| 8 | 5.3 | 5.3 | 5.5 | 5.6 |
| 9 | 6.6 | 7.9 | 4.7 | 6.2 |
| 10 | 5.9 | 7.6 | 10 | 8.4 |
| 11 | 2.7 | 6.6 | 3.1 | 10.8 |
| 12 | 2.4 | 2.3 | 3.4 | 3.8 |
| 13 | 5.1 | 3.9 | 8.9 | 7.4 |
| 14 | 6 | 5 | 5 | 5.5 |
| 15 | 6.1 | 5 | 3.1 | 5.3 |
| 16 | NA | NA | 6.4 | 8.1 |
| 17 | 4.8 | bolus | 9.5 | bolus |
*The asterisks indicate that values can be underestimated due to instances of midline LDs being recorded as 0 mm and hence possibly being negative (chest wall not fully in the beam). NA: not available.