Maud Jaccard1, Ambroise Champion1, Angèle Dubouloz1, Cristina Picardi1, Jérôme Plojoux2, Paola Soccal2, Raymond Miralbell1,3, Giovanna Dipasquale1, Francesca Caparrotti1.
Abstract
BACKGROUND AND PURPOSES: Motion management is crucial for optimal stereotactic body radiotherapy (SBRT) of moving targets. We aimed to describe our clinical experience with real-time tracking of lung-specific electromagnetic transponders (EMTs) for SBRT of early stage non-small cell lung cancer in free-breathing (FB) or deep inspiration breath-hold (DIBH).Entities:
Keywords: Deep inspiration breath-hold; Electromagnetic transponder; Intra-fraction motion; Lung SBRT; Real-time tracking
Year: 2019 PMID: 33458292 PMCID: PMC7807938 DOI: 10.1016/j.phro.2019.11.002
Source DB: PubMed Journal: Phys Imaging Radiat Oncol ISSN: 2405-6316
Geometric, setup and treatment characteristics of the seven patients treated with EMT real-time tracking.
| P1 | P2 | P3 | P4 | P5 | P6 | P7 | |
|---|---|---|---|---|---|---|---|
| Tumor location | Apical segment of upper left lobe | Lingular segment of upper left lobe | Apical segment of lower left lobe | Overlapping anterior segment of upper left lobe and superior segment lower left lobe | Lower right lobe | Upper right lobe | Upper left lobe |
| PTV volume (cm3) | 11.5 | 9.1 | 24.7 | 77.4 | 11.5 | 7.1 | 13.2 |
| Treatment modality | FB | DIBH | FB | FB | DIBH | DIBH | DIBH |
| Days between EMT implantation and simulation CT | 15 | 9 | 32 | 8 | 7 | 7 | 41 |
| Days between simulation CT and 1st treatment | 7 | 11 | 15 | 13 | 28 | 14 | 21 |
| Prescribed dose | 8 × 7.5 Gy | 8 × 7.5 Gy | 5 × 12 Gy | 8 × 7.5 Gy | 3 × 20 Gy | 3 × 20 Gy | 5 × 12 Gy |
| Overall treatment time (days) | 17 | 19 | 10 | 21 | 7 | 8 | 14 |
| Tumor-EMT centroid distance (mm) | 4 | 13 | 18 | 10 | 11 | 15 | 17 |
| Inter-EMTs distances on planning CT (mm) | 12 | 39/45/38 | 9/60/57 | 49/54/58 | 31/26/20 | 46/46/37 | 33/32/61 |
| GTV centroid motion amplitude on 4DCT (LR/SI/AP) (mm) | <1/<1/<1 | 2/4/3 | 2/6/2 | 1/4/2 | 2/8/4 | 1/4/6 | 2/2/2 |
| Median setup time [range] (min) | 8.0 [4.6–9.9] | 11.0 [9.3–18.1] | 10.5 [5.3–14.6] | 8.7 [4.9–34.1] | 10.5 [9.4–13.0] | 19.5 [8.6–22.4] | 13.1 [11.5–13.5] |
| Median inter-EMTs distance differences [range] (mm) | 0.1 [0.0–0.2] | 1.4 [0.0–3.7] | 1.7 [0.7–4.3] | 1.8 [1.3–3.3] | 0.6 [0.3–1.4] | 2.7 [0.6–4.5] | 1.0 [0.1–1.9] |
| Median geometrical residual [range] (mm) | 0.1 [0.0–0.1] | 0.9 [0.6–1.7] | 1.3 [0.8–2.0] | 1.3 [0.8–1.7] | 0.4 [0.4–0.6] | 1.9 [1.7–2.0] | 0.8 [0.7–1.0] |
| Median session time [range] (min) | 10.3 [7.3–12.0] | 17.1 [12.4–30.0] | 16.4 [10.0–20.3] | 15.3 [7.6–36.5] | 13.0 [12.8–28.1] | 33.5 [25.5–39.3] | 22.1 [21.5–23.4] |
| Median number of beam holds per fraction [range] | 0 [0–1] | 4.5 [0–14] | 20.5 [2–36] | 2 [0–29] | 12 [9–18] | 6 [5–11] | 6 [4–8] |
Abbreviations: PTV: planning target volume, EMT: electromagnetic transponder, FB: free-breathing, DIBH: deep inspiration breath-hold, LR: left–right, SI: supero-inferior, AP: antero-posterior, P#: patient #.
Intra- and interfraction motion of EMT centroids positions during beam-on for the seven patients treated with EMT real-time tracking.
| P1 | P2 | P3 | P4 | P5 | P6 | P7 | ||
|---|---|---|---|---|---|---|---|---|
| Tolerance thresholds for EMT centroid beam gating (mm) | LR | 3.0 | 3.0 | 5.0 | 5.0 | 3.0 | 3.0 | 3.0 |
| SI | 3.0 | 3.0 | 8.0 | 7.0 | 3.0 | 3.0 | 3.0 | |
| AP | 3.0 | 3.0 | 8.0 | 5.0 | 3.0 | 3.0 | 3.0 | |
| Median [min, max] EMT centroid motion excursions during all sessions (mm) | LR | 0.7 [0.6–0.7] | 2.7 [1.7–4.7] | 1.0 [0.8–1.7] | 2.7 [2.2–6.9] | 2.3 [2.1–2.9] | 2.2 [1.9–3.5] | 2.4 [1.2–2.5] |
| SI | 2.2 [1.8–2.5] | 2.0 [0.9–4.5] | 8.7 [8.1–10.7] | 9.4 [4.9–11.1] | 4.5 [3.9–4.6] | 2.0 [1.5–3.0] | 2.9 [2.2–4.2] | |
| AP | 1.5 [1.3–1.6] | 2.5 [1.7–3.3] | 5.5 [5.0–6.5] | 3.5 [3.0–5.5] | 4.6 [3.1–5.0] | 1.9 [1.7–2.7] | 1.8 [1.4–2.3] | |
| SD of median EMT positions (mm) | LR | 0.2 | 0.8 | 1.0 | 1.4 | 0.8 | 1.3 | 1.0 |
| SI | 0.6 | 1.1 | 2.8 | 0.8 | 1.0 | 0.6 | 0.7 | |
| AP | 0.6 | 0.5 | 1.1 | 0.7 | 0.5 | 0.8 | 0.6 |
Abbreviations: EMT: electromagnetic transponder, SD: standard deviation, LR: left–right, SI: supero-inferior, AP: antero-posterior, P#: patient #.
Dosimetric comparison of DIBH vs. FB treatment planning for the four patients treated in DIBH. Only the OAR receiving a Dmean > 2 Gy are reported.
| P2 | P5 | P6 | P7 | ||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| DIBH | FB | DIBH | FB | DIBH | FB | DIBH | FB | ||||||
| PTV | Volume [cm3] | 9.1 | 13.2 | −31 | 11.5 | 20.9 | −45 | 7.1 | 10.1 | −30 | 13.2 | 18.7 | −29 |
| Ipsilateral lung | V5 Gy (%) | 15.6 | 27.1 | −42 | 18.7 | 23.4 | −20 | 11.8 | 17.4 | −30 | 16.2 | 19.0 | −15 |
| V20 Gy (%) | 3.6 | 7.5 | −52 | 3.8 | 5.8 | −34 | 3.1 | 5.4 | −42 | 3.2 | 4.5 | −29 | |
| V30 Gy (%) | 1.8 | 4.2 | −57 | 1.9 | 3.3 | −42 | 1.7 | 3.2 | −47 | 1.8 | 2.5 | −30 | |
| Main bronchus (P2)/heart (P5)/chest wall (P7) | Dmean (Gy) | 8.7 | 13.7 | −36 | 2.5 | 3.2 | −22 | – | – | – | 8.7 | 10.2 | −15 |
| D2% (Gy) | 22.4 | 28.2 | −21 | 7.5 | 9.2 | −18 | – | – | – | 47.5 | 66.1 | −28 | |
Abbreviations: PTV: planning target volume, FB: free-breathing, DIBH: deep inspiration breath-hold, OAR: organ at risk, Vx Gy: volume receiving x Gy, D2%: dose received by 2% of the volume, Dmean: mean dose, P#: patient #.
Fig. 1Patient 2 treated in deep inspiration breath-hold: a) example of cone-beam CT imaging at linac (framed in yellow) overlaid with planning CT, showing also the planning target volume (PTV) contour (red) and the electromagnetic transponder (EMT) (pink), b) EMT centroid positions during session 8. The shaded grey area show the moments when the treatment beam was ON. The red horizontal lines indicate the ±3 mm tolerance threshold for beam gating. (For interpretation of the references to colour in this figure legend, the reader is referred to the web version of this article.)
Fig. 2Kilo-Voltage imaging performed during deep inspiration breath-hold (DIBH) treatment of patient 2 in a single breath-hold. A selection of images acquired every 10° from 270° to 80° imaging beam angle is shown (top), as well as a zoom on the region were the electromagnetic transponders (EMTs) are located (bottom). The horizontal yellow lines show that the EMTs had negligible supero-inferior motion during DIBH, in accordance with the EMT tracking system signal. The electromagnetic array is visible on the images acquired at angle 330°, 360°, 30° and 60°. (For interpretation of the references to colour in this figure legend, the reader is referred to the web version of this article.)
Fig. 3Boxplots showing 1st percentile, 1st quartile, median, 3rd quartile and 99th percentile of the vertical, longitudinal and lateral electromagnetic transponders centroid positions registered during beam-on for each fractions of the seven patients. The red lines show the tolerance thresholds for beam gating. (For interpretation of the references to colour in this figure legend, the reader is referred to the web version of this article.)