| Literature DB >> 32867679 |
Ming Jian Lim1, Chin Wen Tan1,2, Hon Sen Tan1, Rehena Sultana3, Victoria Eley4, Ban Leong Sng5,6.
Abstract
BACKGROUND: Accurate blood pressure (BP) measurement depends on appropriate cuff size and shape in relation to the arm. Arm dimensions outside the recommended range of cuff sizes or trunco-conical arms may result in inaccurate BP measurements. Measuring BP using finger cuffs is a potential solution. Arm cuff size is based on mid-arm circumference (MAC), and trunco-conicity is quantified by conicity index. We aimed to determine the correlation of MAC, body mass index (BMI), and weight with conicity index.Entities:
Keywords: Blood pressure; Body weight; Obstetrics
Year: 2020 PMID: 32867679 PMCID: PMC7457531 DOI: 10.1186/s12871-020-01131-6
Source DB: PubMed Journal: BMC Anesthesiol ISSN: 1471-2253 Impact factor: 2.217
Parturient and gestational characteristics
| Non-obesea | Obesea | Total | |
|---|---|---|---|
| Age (years) | 33.9 ± 4.4 | 33.7 ± 3.7 | 33.8 ± 4.2 |
| Gestational age (weeks) | 38.2 ± 1.1 | 38.1 ± 1.0 | 38.2 ± 1.1 |
| Height (m) | 1.6 ± 0.1 | 1.6 ± 0.1 | 1.6 ± 0.1 |
| Weight (kg) | 66.1 ± 8.0 | 86.5 ± 11.4 | 73.3 ± 13.5 |
| BMI (kg/m2) | 26.2 ± 2.5 | 34.5 ± 4.0 | 29.1 ± 5.1 |
| Race / ethnicity | |||
| Chinese | 119 (61.3) | 48 (45.3) | 167 (55.7) |
| Malay | 36 (18.6) | 34 (32.1) | 70 (23.3) |
| Indian | 15 (7.7) | 14 (13.2) | 29 (9.7) |
| Others | 24 (12.4) | 10 (9.4) | 34 (11.3) |
| ASA physical status | |||
| I | 98 (50.5) | 32 (30.2) | 130 (43.3) |
| II | 95 (49.0) | 62 (58.5) | 157 (52.3) |
| III | 1 (0.5) | 12 (11.3) | 13 (4.3) |
| Chronic hypertension | 3 (1.5) | 2 (1.9) | 5 (1.7) |
| Gestational hypertensionb | 6 (3.1) | 5 (4.7) | 11 (3.7) |
| Gestational diabetes | 18 (9.3) | 14 (13.2) | 32 (10.7) |
Values are represented as mean ± standard deviation (SD) or number (%)
ASA American Society of Anesthesiologists, BMI body mass index
a Obese and non-obese groups were defined as BMI ≥ 30 kg/m2 and BMI < 30 kg/m2 respectively
b Gestational hypertension defined as having a blood pressure ≥ 140/90 on two separate occasions more than 6 h apart, without proteinuria and diagnosed after 20 weeks of gestation
Arm and finger measurements
| Non-obesea | Obesea | Total | |
|---|---|---|---|
| Right arm measurements (cm) | |||
| Length | 33.9 ± 2.2 [28, 40] | 34.4 ± 2.5 [26, 41] | 34.1 ± 2.3 [26, 41] |
| Proximal circumference | 29.1 ± 3.3 [20, 38] | 35.0 ± 5.0 [14, 49] | 31.4 ± 4.9 [14, 49] |
| MAC | 25.5 ± 2.6 [20, 36] | 31.4 ± 4.0 [15, 43] | 27.6 ± 4.2 [15, 43] |
| Distal circumference | 23.7 ± 2.3 [19, 34] | 28.7 ± 3.6 [13, 40] | 25.4 ± 3.7 [13, 40] |
| Left arm measurements (cm) | |||
| Length | 33.9 ± 2.2 [27, 41] | 34.4 ± 2.4 [26, 40] | 34.1 ± 2.3 [26, 41] |
| Proximal circumference | 29.1 ± 3.3 [19, 38] | 35.5 ± 4.5 [14, 49] | 31.2 ± 4.9 [14, 49] |
| MAC | 25.5 ± 2.7 [19, 37] | 31.4 ± 4.0 [15, 43] | 27.6 ± 4.3 [15, 43] |
| Distal circumference | 23.7 ± 2.4 [19, 35] | 28.7 ± 3.6 [13, 41] | 25.5 ± 3.7 [13, 41] |
| Right arm conicity index (%) | 5.1 ± 2.0 [1, 12] | 6.4 ± 2.3 [1, 14] | 5.6 ± 2.2 [1, 14] |
| Left arm conicity index (%) | 4.8 ± 1.9 [0, 10] | 6.2 ± 2.0 [1, 12] | 5.3 ± 2.1 [0, 12] |
| Right finger circumference (cm) | 5.2 ± 0.3 [5, 6] | 5.5 ± 0.4 [5, 7] | 5.3 ± 0.4 [4, 7] |
| Left finger circumference (cm) | 5.1 ± 0.3 [4, 6] | 5.5 ± 0.3 [5, 7] | 5.2 ± 0.4 [4, 7] |
Values are represented as mean ± standard deviation (SD) [min, max]
BMI body mass index; MAC mid-arm circumference
a Obese and non-obese groups were defined as BMI ≥ 30 kg/m2 and BMI < 30 kg/m2 respectively
Univariate linear regression analysis describing the associations between BMI, weight, and MAC with conicity indices
| ß (95% CI) | r (95% CI) a | ||
|---|---|---|---|
| Right conicity index | |||
| MAC | 0.20 (0.25–0.26) | 0.39 (0.29–0.48) | < 0.01 |
| Weight | 0.06 (0.04–0.07) | 0.35 (0.25–0.45) | < 0.01 |
| BMI | 0.18 (0.13–0.22) | 0.41 (0.31–0.50) | < 0.01 |
| Left conicity index | |||
| MAC | 0.20 (0.15–0.25) | 0.41 (0.32–0.51) | < 0.01 |
| Weight | 0.06 (0.04–0.07) | 0.39 (0.29–0.48) | < 0.01 |
| BMI | 0.18 (0.13–0.22) | 0.43 (0.34–0.52) | < 0.01 |
BMI body mass index, CI confidence interval, MAC mid-arm circumference
a Pearson correlation coefficient
Fig. 1Frequency distribution of left and right arm cuff sizes (n = 300). Data presented as % of study population
Fig. 2Frequency distribution of left and right finger cuff sizes for Nexfin™ or CNAP™ devices (n = 300). Data presented as % of study population
Parturient satisfaction survey regarding blood pressure measurements
| Question | Non-obesea | Obesea | Unadjusted | |
|---|---|---|---|---|
| Long time taken for blood pressure reading | 19 (9.8) | 9 (8.5) | 0.86 (0.37–1.96) | 0.71 |
| Cuff pops off | 6 (3.1) | 26 (24.5) | 10.18 (4.04–25.69) | < 0.01 |
| Need to put cuff on lower arm or leg | 1 (0.5) | 3 (2.8) | 5.62 (0.58–54.70) | 0.14 |
| Need to change to different cuff | 4 (2.1) | 18 (17.0) | 9.72 (3.19–29.56) | < 0.01 |
Values are represented as number (%)
BMI body mass index, CI confidence interval
a Obese and non-obese groups were defined as BMI ≥ 30 kg/m2 and BMI < 30 kg/m2 respectively