Rectangular cuffs may overestimate blood pressure in individuals with large conical arms.

Paolo Palatini1, Elisabetta Benetti, Claudio Fania, Giacomo Malipiero, Francesca Saladini.   

Abstract

OBJECTIVES: Although the upper arm has the shape of a truncated cone, cylindrical cuffs and bladders are currently used for blood pressure (BP) measurement. The aim of this study was to ascertain whether cylindrical and tronco-conical cuffs provide different readings according to arm size and shape.
DESIGN: We studied 220 individuals with arm circumference ranging from 22 to 42.5 cm. Four different cylindrical and four different tronco-conical bladders of appropriate size were used. Sequential same-arm measurements were performed in triplicate by two observers using the two cuffs in a random order. In 100 individuals, the actual pressure transmitted to the arm surface by the two cuffs at the central point was also measured.
RESULTS: Upper arm shape was tronco-conical in all of the individuals. In a multiple regression, conicity was related to arm circumference (P < 0.001) and length (P = 0.001). Arm conicity and size were independently related to the between-cuff SBP (P = 0.001 and 0.002, respectively) and DBP (P = 0.001 and <0.001, respectively) discrepancies. In the group with arm circumference of 37.5-42.5 cm, the cylindrical cuff overestimated BP measured with the tronco-conical cuff by 2.0 ± 0.4/1.8 ± 0.3 mmHg (P = 0.001 and <0.001). In this group, 15% of individuals found hypertensive with the cylindrical cuff were normotensive when assessed with the conical cuff. Differences as great as 9.7/7.8 mmHg were found in individuals with large arms and slant angle equal to or less than 83°.
CONCLUSION: In obese people, the upper arm may have a pronounced tronco-conical shape and cylindrical cuffs may overestimate BP. Tronco-conical cuffs should be used for BP measurement in individuals with large arms.

Entities:  

Mesh:

Year:  2012        PMID: 22278143     DOI: 10.1097/HJH.0b013e32834f98a0

Source DB:  PubMed          Journal:  J Hypertens        ISSN: 0263-6352            Impact factor:   4.844


  10 in total

1.  Validating prediction equations for mid-arm circumference measurements in adults: National Health and Nutrition Examination Survey, 2001-2012.

Authors:  Tatiana Nwankwo; Yechiam Ostchega; Guangyu Zhang; Jeffery P Hughes
Journal:  Blood Press Monit       Date:  2015-06       Impact factor: 1.444

2.  Hypothalamic-pituitary-adrenal-axis dysregulation and double product increases potentiate ischemic heart disease risk in a Black male cohort: the SABPA study.

Authors:  Leoné Malan; Christiaan E Schutte; Ala'a Alkerwi; Saverio Stranges; Nicolaas T Malan
Journal:  Hypertens Res       Date:  2017-02-09       Impact factor: 3.872

Review 3.  Techniques for self-measurement of blood pressure: limitations and needs for future research.

Authors:  Paolo Palatini; Gerhard N Frick
Journal:  J Clin Hypertens (Greenwich)       Date:  2012-01-23       Impact factor: 3.738

4.  Prognostic Value of Ambulatory Blood Pressure in the Obese: The Ambulatory Blood Pressure-International Study.

Authors:  Paolo Palatini; Gianpaolo Reboldi; Lawrence J Beilin; Edoardo Casiglia; Kazuo Eguchi; Yutaka Imai; Kazuomi Kario; Takayoshi Ohkubo; Sante D Pierdomenico; Joseph E Schwartz; Lindon Wing; Paolo Verdecchia
Journal:  J Clin Hypertens (Greenwich)       Date:  2015-10-05       Impact factor: 3.738

5.  Impact of cuff positioning on blood pressure measurement accuracy: may a specially designed cuff make a difference?

Authors:  Grzegorz Bilo; Oscar Sala; Carlotta Perego; Andrea Faini; Lan Gao; Anna Głuszewska; Juan Eugenio Ochoa; Dario Pellegrini; Laura Maria Lonati; Gianfranco Parati
Journal:  Hypertens Res       Date:  2017-01-12       Impact factor: 3.872

Review 6.  Current applications and limitations of European guidelines on blood pressure measurement: implications for clinical practice.

Authors:  Giuliano Tocci; Barbara Citoni; Giulia Nardoianni; Ilaria Figliuzzi; Massimo Volpe
Journal:  Intern Emerg Med       Date:  2022-03-31       Impact factor: 5.472

7.  Cuff challenges in blood pressure measurement.

Authors:  Paolo Palatini; Roland Asmar
Journal:  J Clin Hypertens (Greenwich)       Date:  2018-07       Impact factor: 3.738

8.  The Effect of Cuff Size on Blood Pressure Measurement in Obese Surgical Patients: A Prospective Crossover Clinical Trial.

Authors:  Serpil Yüksel; Gülay Altun-Uğraş; Nurhan Altınok; Necla Demir
Journal:  Florence Nightingale J Nurs       Date:  2020-07-03

9.  Upper limb dimensions in adults presenting for elective surgery - implications for blood pressure measurement.

Authors:  Christopher Chow; Peter Ceglowski; Katie Lehane; Anita Pelecanos; Kellie Wren; Victoria A Eley
Journal:  BMC Anesthesiol       Date:  2020-04-04       Impact factor: 2.217

10.  Correlation of patient characteristics with arm and finger measurements in Asian parturients: a preliminary study.

Authors:  Ming Jian Lim; Chin Wen Tan; Hon Sen Tan; Rehena Sultana; Victoria Eley; Ban Leong Sng
Journal:  BMC Anesthesiol       Date:  2020-08-31       Impact factor: 2.217

  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.