Yili He1, Daqing Gu1, Xiangzhi Kong1, Zhiqiang Feng1, Weishang Lin1, Yunfeng Cai1.
Abstract
OBJECTIVE: To establish an applicable and highly sensitive patient-based real-time quality control (PBRTQC) program based on a data model constructed with patients' results of a procalcitonin point-of-care testing (POCT) analyzer.Entities:
Keywords: internal quality control; moving rate of positive results; point-of-care testing; procalcitonin
Mesh:
Substances:
Year: 2022 PMID: 35257410 PMCID: PMC8993656 DOI: 10.1002/jcla.24320
Source DB: PubMed Journal: J Clin Lab Anal ISSN: 0887-8013 Impact factor: 2.352
Probability of detection of different biases by the conventional multi‐rule internal quality control program
| Bias magnitude (ng/ml) | |||||||
|---|---|---|---|---|---|---|---|
| QC Rule | 0.01 | 0.05 | 0.1 | 0.24 | 0.5 | 1.0 | |
|
Level 1 QC 1.50 ng/ml | 13S | 0.23% | 1.5% | 8.2% | 70.8% | 100% | 100% |
| 22S | 0.11% | 1.3% | 11% | 84.8% | 100% | 100% | |
| 41S | 0.16% | 2.8% | 22% | 95.5% | 100% | 100% | |
|
Level 2 QC 19.88 ng/ml | 13S | 0.14% | 0.17% | 0.21% | 0.39% | 1.0% | 4.9% |
| 22S | 0.055% | 0.071% | 0.097% | 0.22% | 0.83% | 5.8% | |
| 41S | 0.070% | 0.094% | 0.14% | 0.34% | 1.7% | 13% | |
The conventional QC program uses 1.50 ng/ml quality control product to trigger the 41S rule first, which requires a bias of more than 0.24 ng/ml, with 95.5% probability.
Abbreviation: QC, quality control.
Parameters of MA and MR program with various block sizes
| MA procedure | MR procedure | ||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|
|
| Truncated data | Block Size | LCL | UCL | FR (%) | COV | Positive Rate | Block Size | LCL | UCL | FR (%) |
| 0.2 | 6.0% | 10 | <0.02 | 0.34 | 1.89 | 0.02 | 94.57% | 10 | 78.37 | >100 | 1.53 |
| 20 | <0.02 | 0.27 | 0.66 | 20 | 85.01 | >100 | 10.81 | ||||
| 30 | 0.03 | 0.24 | 0.18 | 30 | 84.24 | >100 | 4.28 | ||||
| 40 | 0.05 | 0.23 | 0.27 | 40 | 85.80 | >100 | 6.31 | ||||
| 50 | 0.06 | 0.22 | 0.27 | 50 | 87.03 | >100 | 7.26 | ||||
| 60 | 0.07 | 0.21 | 0.49 | 60 | 88.03 | >100 | 6.15 | ||||
| 70 | 0.07 | 0.20 | 1.04 | 70 | 88.26 | >100 | 10.07 | ||||
| 80 | 0.08 | 0.20 | 0.54 | 80 | 89.49 | >100 | 7.69 | ||||
| 90 | 0.08 | 0.20 | 1.00 | 90 | 90.15 | >100 | 11.05 | ||||
| 100 | 0.08 | 0.20 | 1.10 | 100 | 90.66 | >100 | 8.74 | ||||
| 0.3 | 4.8% | 10 | <0.02 | 0.38 | 2.56 | 0.03 | 81.46% | 10 | 46.73 | >100 | 0.37 |
| 20 | <0.02 | 0.31 | 3.48 | 20 | 58.73 | >100 | 0.21 | ||||
| 30 | 0.02 | 0.28 | 1.79 | 30 | 63.28 | >100 | 0.04 | ||||
| 40 | 0.04 | 0.26 | 1.36 | 40 | 66.29 | >100 | 0.13 | ||||
| 50 | 0.05 | 0.25 | 0.62 | 50 | 69.24 | 98.15 | 0.50 | ||||
| 60 | 0.06 | 0.24 | 0.62 | 60 | 71.35 | 96.13 | 0.55 | ||||
| 70 | 0.07 | 0.23 | 0.85 | 70 | 72.70 | 94.89 | 0.97 | ||||
| 80 | 0.07 | 0.23 | 0.76 | 80 | 74.13 | 93.56 | 6.75 | ||||
| 90 | 0.07 | 0.22 | 0.85 | 90 | 75.29 | 92.46 | 7.51 | ||||
| 100 | 0.08 | 0.22 | 1.04 | 100 | 76.24 | 91.52 | 12.85 | ||||
| 0.4 | 4.4% | 10 | <0.02 | 0.46 | 2.11 | 0.04 | 69.42% | 10 | 25.48 | >100 | 0.21 |
| 20 | <0.02 | 0.37 | 1.30 | 20 | 38.63 | >100 | 0.33 | ||||
| 30 | <0.02 | 0.34 | 1.13 | 30 | 44.79 | 95.47 | 0 | ||||
| 40 | 0.02 | 0.32 | 1.09 | 40 | 48.69 | 91.83 | 0.08 | ||||
| 50 | 0.02 | 0.31 | 0.66 | 50 | 51.84 | 88.79 | 0.08 | ||||
| 60 | 0.03 | 0.29 | 0.35 | 60 | 54.57 | 86.15 | 0.63 | ||||
| 70 | 0.04 | 0.28 | 0.75 | 70 | 56.57 | 84.22 | 0.93 | ||||
| 80 | 0.04 | 0.28 | 0.53 | 80 | 58.10 | 82.79 | 1.06 | ||||
| 90 | 0.05 | 0.27 | 0.49 | 90 | 56.31 | 81.61 | 0.90 | ||||
| 100 | 0.05 | 0.27 | 0.76 | 100 | 60.04 | 80.81 | 0.73 | ||||
| 0.5 | 3.8% | 10 | <0.02 | 0.52 | 2.36 | 0.05 | 60.58% | 10 | 12.94 | 107.39 | 0.08 |
| 20 | <0.02 | 0.42 | 1.81 | 20 | 26.04 | 94.31 | 0 | ||||
| 30 | <0.02 | 0.38 | 1.38 | 30 | 32.01 | 88.41 | 0 | ||||
| 40 | <0.02 | 0.36 | 1.04 | 40 | 36.28 | 84.34 | 0 | ||||
| 50 | 0.01 | 0.34 | 0.57 | 50 | 40.19 | 80.59 | 0.13 | ||||
| 60 | 0.02 | 0.33 | 0 | 60 | 43.26 | 77.60 | 0.42 | ||||
| 70 | 0.02 | 0.32 | 0.22 | 70 | 45.39 | 75.56 | 1.23 | ||||
| 80 | 0.03 | 0.32 | 0.40 | 80 | 46,71 | 74.33 | 0.98 | ||||
| 90 | 0.03 | 0.32 | 0.93 | 90 | 47.54 | 73.53 | 0.77 | ||||
| 100 | 0.03 | 0.31 | 0.53 | 100 | 48.13 | 72.92 | 0.73 | ||||
| 0.6 | 3.1% | 10 | <0.02 | 0.61 | 1.79 | 0.06 | 53.84% | 10 | 3.30 | >100 | 0 |
| 20 | <0.02 | 0.48 | 0.51 | 20 | 16.80 | 88.77 | 0.04 | ||||
| 30 | <0.02 | 0.43 | 0.21 | 30 | 22.56 | 84.11 | 0.08 | ||||
| 40 | <0.02 | 0.39 | 0.30 | 40 | 26.89 | 79.96 | 0 | ||||
| 50 | <0.02 | 0.37 | 0 | 50 | 30.69 | 76.33 | 0 | ||||
| 60 | 0.03 | 0.36 | 0 | 60 | 33.96 | 73.15 | 0.08 | ||||
| 70 | 0.04 | 0.34 | 0.17 | 70 | 36.20 | 70.98 | 0.17 | ||||
| 80 | 0.05 | 0.34 | 0.13 | 80 | 37.71 | 69.55 | 0.34 | ||||
| 90 | 0.05 | 0.33 | 0.09 | 90 | 38.46 | 68.81 | 0.68 | ||||
| 100 | 0.06 | 0.33 | 0.04 | 100 | 38.94 | 68.28 | 0.30 | ||||
| 0.7 | 2.91% | 10 | <0.02 | 0.61 | 2.68 | 0.07 | 48.62% | 10 | <0 | 97.41 | 0.17 |
| 20 | <0.02 | 0.48 | 0.60 | 20 | 11.28 | 84.91 | 0.04 | ||||
| 30 | <0.02 | 0.43 | 0.21 | 30 | 17.18 | 79.01 | 0 | ||||
| 40 | <0.02 | 0.39 | 0.30 | 40 | 21.81 | 74.51 | 0 | ||||
| 50 | 0.03 | 0.37 | 0.09 | 50 | 25.74 | 70.73 | 0 | ||||
| 60 | 0.03 | 0.36 | 0.35 | 60 | 28.82 | 67.74 | 0 | ||||
| 70 | 0.04 | 0.34 | 0.17 | 70 | 30.74 | 65.86 | 0.04 | ||||
| 80 | 0.05 | 0.34 | 0.18 | 80 | 31.76 | 64.90 | 0.34 | ||||
| 90 | 0.05 | 0.33 | 0.22 | 90 | 32.21 | 64.46 | 0 | ||||
| 100 | 0.06 | 0.33 | 0.04 | 100 | 32.43 | 64.19 | 0 | ||||
| 0.8 | 2.59% | 10 | <0.02 | 0.69 | 1.82 | 0.08 | 43.86% | 10 | <0 | 88.42 | 0.74 |
| 20 | <0.02 | 0.54 | 0.38 | 20 | 9.38 | 75.35 | 0.21 | ||||
| 30 | <0.02 | 0.47 | 0.38 | 30 | 14.57 | 70.14 | 0.17 | ||||
| 40 | <0.02 | 0.44 | 0.26 | 40 | 18.38 | 66.43 | 0.33 | ||||
| 50 | 0.03 | 0.42 | 0.09 | 50 | 21.45 | 63.51 | 0.08 | ||||
| 60 | 0.02 | 0.40 | 0.09 | 60 | 23.73 | 61.32 | 0.34 | ||||
| 70 | 0.03 | 0.39 | 0.13 | 70 | 25.12 | 59.94 | 0.72 | ||||
| 80 | 0.04 | 0.38 | 0.13 | 80 | 25.90 | 59.17 | 0.42 | ||||
| 90 | 0.05 | 0.38 | 0.09 | 90 | 26.30 | 58.74 | 0.60 | ||||
| 100 | 0.05 | 0.37 | 0 | 100 | 26.61 | 58.33 | 0.13 | ||||
The i was used in the MA procedure to eliminate a certain number of outliers, for example, i = 0.2 means that the population mean ± 0.2 SD was the truncated limit, this eliminated data accounting for 6% of the population data. When N = 30, the lower and upper control limits were 0.03 and 0.24, respectively, and the false rejection rate was 0.18%. The cut‐off value was used in the MR method to assign binary values to patient data, for example, COV = 0.03, if the patient result was greater than the COV, then it was regarded as “1”; otherwise, it was regarded as “0.” When N = 10, the lower control limit was 46.73%, and the upper control limit was invalid. The false rejection rate was 0.37%.
Abbreviations: COV, cut‐off value; FR, false rejection rate; LCL, lower control limit; MA, moving average; MR, moving rate; UCL, upper control limit.
FIGURE 1The minimum MNPed of different parameters and biases in MA and MR programs. (A) The minimum MNPed curves obtained by the MA program after simulating different COVs combined with different Ns to detect different errors and eliminating the parameters of FR ≧1.0%. For example, the curve of i = 0.3 is curve of the minimum MNPed obtained when running N = 50~90 with FR < 1.0%. Since the minimum MNPed value was concentrated on the curve with N = 0.2, N = 0.2 was selected. (B) is similar to the above, COV = 0.03 and COV = 0.07 were selected as parameters of MR program
FIGURE 2Bias detection curves for MA and MR programs with different block sizes. (A) MA program with i = 0.2. (B) MR program with COV = 0.03. (C) MR program with COV = 0.07. Different curves represent MNPeds measured by QC programs with different block sizes (N)
FIGURE 3Validation charts for optimal MA or MR program for the bias detection. (A) MA program with i = 0.2 and N = 20. (B) MA program with i = 0.2 and N = 30. (C) MR program with COV = 0.03 and N = 10. (D) MR program with COV = 0.07 and N = 10. The blue strips represent MNPeds, the error lines represent the maximum number of results required for bias detection, and the red lines represent the daily run size of specimens
FIGURE 4β Curves of bias detection for optimal MA and MR programs. β represents the false negative rate detected by the program after introducing bias, also known as type II error. (A), (B) The MA programs with i = 0.2, N = 20, and i = 0.2, N = 30, when bias ≥ 0.2 ng/ml, β < 1%. (C) The MR program with COV = 0.03, N = 10, when negative bias ≥ 0.4 ng/ml, β < 1%. (D) The MR program with COV = 0.07, N = 10, when bias ≥ 0.06 ng/ml, β < 1%
FIGURE 5Simulation diagram of the MR program to detect systematic errors. The green horizontal line represents the average MR of the positive results. The red horizontal line indicates the upper control limit. The dashed lines represent the original data sets. The blue line represents the data set after introducing bias to the simulation at a random point. For the MR program with COV = 0.07 and N = 10, the lower limit of the control was < 0%, so there was no lower control limit in the sample diagram