A Palmer1, J Kearton, O Hayman.
Abstract
OBJECTIVES: The objective of this study was to determine current radiotherapy linear accelerator quality control (QC) practice in the UK, as a comparative benchmark and indicator of development needs, and to raise awareness of QC as a key performance indicator.Entities:
Mesh:
Year: 2012 PMID: 22674707 PMCID: PMC3500805 DOI: 10.1259/bjr/46195110
Source DB: PubMed Journal: Br J Radiol ISSN: 0007-1285 Impact factor: 3.039
Questionnaire topics
| Frequency of QC schedule review and perceived need for productivity/efficiency improvement |
| Use of quality system management tools in making decision on QC test schedules |
| Guidance and publications used to make decisions on QC test schedules |
| Ease of access to linear accelerators to perform QC and extent of evening and weekend working |
| Levels of completion of planned QC work |
| Time spent performing linear accelerator and patient-specific QC |
| Involvement of stakeholders; knowledge of tolerances and reporting of results |
| Number of linear accelerators, manufacturer and specification |
QC, quality control.
Figure 1“Word cloud” diagram of primary influences determining UK radiotherapy centres' QC schedules for linear accelerators (font size proportional to number of citations) [3,8,15-21].
Proportion of centres routinely using indicated combinations of normal day, evening, weekend and early morning sessions to undertake linear accelerator QC
| Time linear accelerator QC is undertaken | Proportion of centres | ||||||||
| 30% | 30% | 12% | 7% | 7% | 5% | 5% | 2% | 2% | |
| Early morning | Y | Y | Y | Y | |||||
| Normal day | Y | Y | Y | Y | Y | Y | Y | Y | |
| Evening | Y | Y | Y | Y | Y | ||||
| Weekend | Y | Y | Y | Y | |||||
QC, quality control; Y, yes.
Time (hours) spent undertaking linear accelerator QC testing
| Time category | Minimum value | First quartile | Mean | Third quartile | Maximum value |
| Total machine time ( hours per linear accelerator per month) | 3.0 | 10.0 | 15.0 | 20.0 | 35.0 |
| Total time including offline analysis ( hours per linear accelerator per month) | 5.0 | 13.1 | 19.5 | 26.2 | 56.0 |
| Total time for patient-specific IMRT QC per patient | 0.0 | 1.0 | 1.5 | 2.1 | 10.0 |
IMRT, intensity-modulated radiotherapy; QC, quality control.
Stated frequency of QC checks and percentage occurrence within submitted physics department QC schedules, for checks identified in IPEM Report 81 [7], with comparison with previously recommended frequency
| QC test | IPEM Report 81 frequency | Included in physics QC schedules | Daily | Weekly | Between weekly and monthly | Monthly | Quarterly or 6-monthly | Annual |
| Output constancy measurement | Daily | 48% | 9% | – | – | – | – | |
| Light field size | Daily | 52% | 9% | – | 8% | – | – | |
| Laser and crosswire alignment | Daily | 52% | 17% | – | 8% | – | – | |
| Door/maze interlock | Daily | 61% | 7% | 8% | 21% | – | – | |
| ODI | Daily | 48% | 9% | 9% | 0% | – | – | |
| Output “calibration” measurement | Weekly | 87% | – | 35% | 5% | – | – | |
| ODI at different distances | Weekly | 44% | 20% | 20% | 20% | – | – | |
| Pointers | Weekly | 57% | 8% | 23% | – | 8% | – | |
| Wedge factor | Weekly | 65% | 13% | 20% | 7% | 6% | – | |
| Flatness and symmetry at zero gantry angle | 2-weekly | 70% | – | 6% | 13% | – | – | |
| Emergency off switches | Monthly | 30% | 14% | 14% | 29% | – | – | |
| Shielding tray interlock | Monthly | 17% | – | – | – | 25% | – | |
| Back-up timer interlock | Monthly | 9% | – | – | – | – | – | |
| Gantry and collimator rotation scales | Monthly | 74% | 12% | 0% | 12% | 6% | 6% | |
| Light field dimension, various sizes | Monthly | 43% | 10% | 10% | 0% | 10% | 20% | |
| Isocentre “quick” check | Monthly | 43% | 10% | 10% | 20% | – | – | |
| Table lateral and longitudinal scale | Monthly | 57% | 8% | 8% | – | 8% | – | |
| Table height scale | Monthly | 52% | – | 8% | – | 8% | 16% | |
| Photon and electron energy, two-depth measure | Monthly | 83% | – | – | 11% | – | – | |
| Radiation and light field alignment, at one size | Monthly | 74% | – | – | 6% | 12% | – | |
| MLC leaf position, standard field | Monthly | 43% | 10% | 20% | 10% | 20% | 10% | |
| Radiation and light field alignment, other sizes | Quarterly | 52% | – | – | – | 34% | 8% | |
| Back-up dosemeter and timer | Quarterly | 23% | 17% | – | – | 33% | – | |
| Field size at extended distance | Quarterly | 30% | – | – | – | – | 14% | |
| Output variation with gantry angle | Quarterly | 65% | – | – | – | 7% | 13% | |
| Wedge factor with gantry angle | Quarterly | 65% | – | – | – | 20% | 7% | |
| Output for non-standard fields | Quarterly | 52% | – | – | – | 8% | 33% | |
| Flatness and symmetry non-zero gantry angle | Quarterly | 61% | – | – | – | 36% | 21% | |
| MLC stability with gantry rotation | Quarterly | 26% | – | – | – | 33% | – | |
| MLC leakage between leaves | Quarterly | 43% | – | – | – | 30% | 10% | |
| Radiation isocentre | 6-monthly | 35% | – | – | – | – | 12% | |
| Calibration of constancy check device | 6-monthly | 61% | – | – | – | 36% | 7% | |
| Linearity of dosimetry system | 6-monthly | 78% | – | – | – | 22% | 17% | |
| Table deflection under load | Annual | 39% | – | – | – | 11% | 11% | |
| Isocentre “definitive” check | Annual | 35% | – | – | – | – | 12% | |
| PDD and profiles in 3D scanning water tank | Annual | 48% | – | – | – | – | – | |
| Field size reproducibility | Annual | 9% | – | – | – | – |
3D, three-dimensional; IPEM, Institute of Physics and Engineering in Medicine; MLC, multileaf collimator; ODI, optical distance indicator; PDD, percentage depth dose; QC, quality control.
Bold indicates the most common value, and hence the column that is recommended to the reader.
Stated frequency of QC checks and percentage occurrence within submitted physics department QC schedules, for checks not included in IPEM Report 81 [7]
| QC test | Included in physics QC schedules | Per patient | Daily | Weekly | Between weekly and monthly | Monthly | Quarterly or 6-monthly | Annual |
| MV/kV imaging quality | 43% | – | 22% | 11% | – | 11% | 11% | |
| kV imaging isocheck and known shifts | 22% | – | 20% | – | – | – | ||
| MV/kV imager collision interlocks | 9% | – | 50% | – | – | 50% | – | – |
| kV/MV image registration | 9% | – | – | – | – | 50% | – | 50% |
| kV imager dose | 9% | – | – | – | 50% | 50% | – | – |
| kV/MV imager movement calibration | 39% | – | – | – | – | 44% | 12% | |
| MLC “picket fence” film | 52% | – | – | 16% | – | – | ||
| MLC positions in complex test fields | 13% | – | – | – | 34% | 66% | – | – |
| MLC carriage skew | 13% | – | – | – | – | – | – | |
| Output at different dose rates | 13% | – | – | – | – | – | 33% | |
| Dynamic wedge profile | 35% | – | – | – | – | 25% | 25% | |
| 26% | – | – | – | 17% | 17% | – | ||
| Table isocentre rotation | 44% | – | – | – | 10% | 20% | – | |
| Collimator position with electron applicators | 44% | – | – | 10% | 10% | 20% | 10% | |
| Film of matching fields | 26% | – | – | – | – | 33% | – | |
| Radiation and light alignment, asymmetric | 30% | – | – | – | 14% | 29% | 14% | |
| IMRT point dose | 30% | – | – | – | – | – | – | |
| IMRT fluence map | 30% | – | – | – | – | – | – | |
| MLC sweep gap | 40% | – | 25% | 25% | – | 25% | 25% | – |
| MLC travel range | 10% | – | – | – | – | – | – | |
| MLC travel speed | 19% | – | – | – | 20% | 20% | 20% | |
| IMRT dose rate constancy | 10% | – | – | – | – | – | – | |
| Radiation protection survey | 9% | – | – | – | – | – | – |
IMRT, intensity-modulated radiotherapy; IPEM, Institute of Physics and Engineering in Medicine; MLC, multileaf collimator; ODI, optical distance indicator; QC, quality control.
Modal value and range of stated tolerance or action value for selected QC checks, where sufficient responses were provided for significance (n>5)
| QC test | Range minimum tolerance | Modal tolerance | Range maximum tolerance |
| Output constancy measurement | ±1.0% | ± | ±5.0% |
| Output “calibration” measurement | ±1.5% | ± | ±3.0% |
| Light field sizes | ±1.0 mm | ± | ±2.0 mm |
| Lasers and crosswire alignments | ±1.0 mm | ± | ±2.0 mm |
| Optical distance indicator | ±1.0 mm | ± | ±5.0 mm |
| Wedge factor | ±1.0% | ± | ±2.0% |
| Gantry and collimator scales | ±0.2° | ± | ±1.0° |
| Table lateral and longitudinal scale | ±0.5 mm | ± | ±2.0 mm |
| Table height scale | ±1.0 mm | ± | ±5.0 mm |
| Radiation and light field alignment | ±1.0 mm | ± | ±2.0 mm |
| MLC leaf position | ±1.0 mm | ± | ±2.0 mm |
| MLC leakage between leaves | ±2.0% | ± | ±5.0% |
| Linearity of dosimetry system | ±1.0% | ± | ±2.0% |
| Couch deflection under load | ±3.0 mm | ± | ±5.0 mm |
MLC, multileaf collimator; QC, quality control.