Barbora Horehledova1,2, Casper Mihl1,2, Ellen Boswijk1,2, Genevieve A J C Crombag1,2, Estelle C Nijssen1, Patty J Nelemans3, Leo F Veenstra4, Joachim E Wildberger1,2, Marco Das1,2.
Abstract
BACKGROUND: Multidetector computed tomography (MDCT) plays a key role in patient assessment prior to transcatheter aortic valve implantation (TAVI). However, to date no consensus has been established on what is the optimal pre-procedural imaging protocol. Variability in pre-TAVI acquisition protocols may lead to discrepancies in aortic annulus measurements and may potentially influence prosthesis size selection.Entities:
Year: 2020 PMID: 32396570 PMCID: PMC7217477 DOI: 10.1371/journal.pone.0232673
Source DB: PubMed Journal: PLoS One ISSN: 1932-6203 Impact factor: 3.240
Scan and contrast media application parameters.
| Retrospective ECG-gated protocol | Non ECG-synchronized high pitch protocol | |
|---|---|---|
| Scan direction | Cranio-caudal | Cranio-caudal |
| Tube voltage [kV] | 100 | 100 |
| Tube current [mAs] | 320 | 150 |
| Dose modulation | CARE Dose4D | CARE Dose4D |
| CTDIvol [mGy] | 62 | 5 |
| Rotation time [s] | 0.28 | 0.28 |
| Temporal resolution [ms] | 75 | 75 |
| Pitch | 0.17 | 3 |
| Slice collimation | 2x2x64x0.6 | 2x2x64x0.6 |
| Slice width [mm] | 0.75/0.7 | 1.5/1 |
| Reconstruction kernel | B26f | I30f |
| Contrast media | Iopromide 300 (Ultravist) | |
| Test bolus | 20ml CM * 7.2ml/s followed by 15ml NaCl* 7.2ml/s | |
| Main bolus | 75ml CM * 7.2ml/s (100%) | |
| 50ml CM/NaCl * 7.2ml/s (50/50%) | ||
| 25ml NaCl* 7.2ml/s | ||
| Iodine delivery rate [gI/s] | 2.16 | |
CM = contrast medium; ECG = electrocardiogram; gI = grams of iodine; kV = kilovolt; mAs = milliamper-second; mGy = milligray; ml = milliliter; mm = millimeter; NaCl = saline; s = second
Baseline characteristics of the patient population.
| Mean±SD | Range | |
|---|---|---|
| Gender | --- | |
| Age (years) (n = 50) | 81±5 | 67–88 |
| • female (n = 26; 52%) | • 82±5 | • 67–88 |
| Height [cm] | 166±10 | 144–198 |
| Weight [kg] | 74 ±11 | 52–98 |
| BSA [m2] | 1.8±0.2 | 1.53–2.33 |
| Aortic stenosis characteristic derived from TTE: | ||
| Heart rate [bpm] | 75 ±18 | 50–128 |
| Ejection Fraction [%] | 54±12 | 24–73 |
| AVA [cm2] | 0.81±0.2 | 0.4–1.6 |
| Maximum pressure gradient [mmHg] | 71±25 | 25–138 |
| Mean pressure gradient [mmHg] | 44±16 | 12–88 |
(AVA = aortic valve area; cm = centimeter; m = meter; mmHg = millimeters of mercury; n = number; SD = standard deviation; TTE-transthoracic echocardiography; % = percent)
Aortic root dimensions used in prosthesis size selection.
| Blant-Altman analysis shows the systematic and random discrepancies in aortic annulus measurements between max-systole minus | ||||
|---|---|---|---|---|
| Short diameter (mm) | 22.5±1.7 | 20.9 ±1.8 | 1.6±1.2 | - 0.7–3.8 |
| Area (mm2) | 478.3±64 | 430.0±59 | 48.3±27.3 | - 5.2–101.8 |
| DA (mm) | 24.6±1.7 | 23.3±1.6 | 1.3±0.7 | - 0.1–2.6 |
| Perimeter (mm) | 79.6±5.7 | 76.7±5.6 | 3.0±2.4 | 1.7–7.6 |
| Short diameter (mm) | 22.5±1.7 | 20.9 ±1.7 | 1.5±1.4 | -1.2–4.2 |
| Area (mm2) | 478.3±64 | 441.4±63 | 36.9±28.1 | -18.1–91.9 |
| DA (mm) | 24.6±1.7 | 23.6±1.7 | 1.0±0.7 | -0.5–2.4 |
| Perimeter (mm) | 79.6±5.7 | 77.2±5.6 | 2.4±2.8 | -3.0–7.8 |
(DA = area derived diameter; mm = millimeter; mm2 = square millimeter; SD = standard deviation)
Pairwise comparison of selected prosthesis sizes between the max-systolic versus end-diastolic and non-ECG-synchronized measurements.
| • Green cells represent cases with agreement in prosthesis size between max-systole and end-diastole/non-ECG-synchronized measurement | |||||||||
| Max-systolic (20%) | End-diastolic (70%) | Non-ECG-synchronized | |||||||
| n.: | 20mm | 23mm | 26mm | 29mm | 20mm | 23mm | 26mm | 29mm | |
| 20mm | 1 | 0 | 0 | 0 | 1 | 0 | 0 | 0 | |
| 23mm | 3 | 8 | 0 | 0 | 3 | 8 | 0 | 0 | |
| 26mm | 0 | 7 | 23 | 0 | 0 | 9 | 21 | 0 | |
| 29mm | 0 | 0 | 7 | 1 | 0 | 0 | 5 | 3 | |
| n.: | 20mm | 23mm | 26mm | 29mm | 20mm | 23mm | 26mm | 29mm | |
| 20mm | 1 | 0 | 0 | 0 | 1 | 0 | 0 | 0 | |
| 23mm | 3 | 9 | 0 | 0 | 2 | 10 | 0 | 0 | |
| 26mm | 0 | 6 | 23 | 0 | 0 | 6 | 23 | 0 | |
| 29mm | 0 | 0 | 7 | 1 | 0 | 0 | 5 | 3 | |
| n.: | 23mm | 26mm | 29mm | Not-suitable | 23mm | 26mm | 29mm | ||
| 23mm | 4 | 0 | 0 | 1 | 3 | 0 | 0 | ||
| 26mm | 7 | 24 | 0 | 0 | 7 | 24 | 0 | ||
| 29mm | 0 | 8 | 7 | 0 | 0 | 9 | 6 | ||
| n.: | 23mm | 26mm | 29mm | Not-suitable | 23mm | 26mm | 29mm | Not-suitable | |
| 23mm | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | |
| 26mm | 0 | 5 | 0 | 0 | 0 | 5 | 0 | 0 | |
| 29mm | 0 | 6 | 21 | 0 | 0 | 5 | 22 | 0 | |
| Not-suitable | 0 | 0 | 9 | 9 | 0 | 0 | 6 | 12 | |
| n.: | 23mm | 26mm | 29mm | Not-suitable | 23mm | 26mm | 29mm | ||
| 23mm | 5 | 0 | 0 | 1 | 4 | 0 | 0 | ||
| 26mm | 9 | 22 | 0 | 0 | 7 | 24 | 0 | ||
| 29mm | 0 | 8 | 6 | 0 | 0 | 9 | 5 | ||
(DA = area derived diameter; ESV = Edwards Sapien Valve; MCV = Medtronic Core Valve)