Deformation dynamics and mechanical properties of the aortic annulus by 4-dimensional computed tomography: insights into the functional anatomy of the aortic valve complex and implications for transcatheter aortic valve therapy.

Ashraf Hamdan1, Victor Guetta, Eli Konen, Orly Goitein, Amit Segev, Ehud Raanani, Dan Spiegelstein, Ilan Hay, Elio Di Segni, Michael Eldar, Ehud Schwammenthal.   

Abstract

OBJECTIVES: The purpose of this study was to assess deformation dynamics and in vivo mechanical properties of the aortic annulus throughout the cardiac cycle.
BACKGROUND: Understanding dynamic aspects of functional aortic valve anatomy is important for beating-heart transcatheter aortic valve implantation.
METHODS: Thirty-five patients with aortic stenosis and 11 normal subjects underwent 256-slice computed tomography. The aortic annulus plane was reconstructed in 10% increments over the cardiac cycle. For each phase, minimum diameter, ellipticity index, cross-sectional area (CSA), and perimeter (Perim) were measured. In a subset of 10 patients, Young's elastic module was calculated from the stress-strain relationship of the annulus.
RESULTS: In both subjects with normal and with calcified aortic valves, minimum diameter increased in systole (12.3 ± 7.3% and 9.8 ± 3.4%, respectively; p < 0.001), and ellipticity index decreased (12.7 ± 8.8% and 10.3 ± 2.7%, respectively; p < 0.001). The CSA increased by 11.2 ± 5.4% and 6.2 ± 4.8%, respectively (p < 0.001). Perim increase was negligible in patients with calcified valves (0.56 ± 0.85%; p < 0.001) and small even in normal subjects (2.2 ± 2.2%; p = 0.01). Accordingly, relative percentage differences between maximum and minimum values were significantly smallest for Perim compared with all other parameters. Young's modulus was calculated as 22.6 ± 9.2 MPa in patients and 13.8 ± 6.4 MPa in normal subjects.
CONCLUSIONS: The aortic annulus, generally elliptic, assumes a more round shape in systole, thus increasing CSA without substantial change in perimeter. Perimeter changes are negligible in patients with calcified valves, because tissue properties allow very little expansion. Aortic annulus perimeter appears therefore ideally suited for accurate sizing in transcatheter aortic valve implantation.
Copyright © 2012 American College of Cardiology Foundation. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2012        PMID: 22222074     DOI: 10.1016/j.jacc.2011.09.045

Source DB:  PubMed          Journal:  J Am Coll Cardiol        ISSN: 0735-1097            Impact factor:   24.094


  41 in total

1.  Assessment of annular distensibility in the aortic valve.

Authors:  John O'Dea; David J Nolan
Journal:  Interact Cardiovasc Thorac Surg       Date:  2012-06-13

2.  Cyclic changes in area- and perimeter-derived effective dimensions of the aortic annulus measured with multislice computed tomography and comparison with metric intraoperative sizing.

Authors:  Won-Keun Kim; Alexander Meyer; Helge Möllmann; Andreas Rolf; Susanne Möllmann; Johannes Blumenstein; Arnaud Van Linden; Christian W Hamm; Thomas Walther; Jörg Kempfert
Journal:  Clin Res Cardiol       Date:  2016-02-18       Impact factor: 5.460

Review 3.  Guidance of transcatheter aortic valve replacement by echocardiography.

Authors:  Rebecca T Hahn
Journal:  Curr Cardiol Rep       Date:  2014-01       Impact factor: 2.931

Review 4.  The time has come to extend the expiration limit of cryopreserved allograft heart valves.

Authors:  Jan Burkert; Petra Kochová; Zbyněk Tonar; Robert Cimrman; Tereza Blassová; Ramadan Jashari; Radovan Fiala; Jaroslav Špatenka
Journal:  Cell Tissue Bank       Date:  2020-06-24       Impact factor: 1.522

5.  Assessment of aortic valve tract dynamics using automatic tracking of 3D transesophageal echocardiographic images.

Authors:  Sandro Queirós; Pedro Morais; Wolfgang Fehske; Alexandros Papachristidis; Jens-Uwe Voigt; Jaime C Fonseca; Jan D'hooge; João L Vilaça
Journal:  Int J Cardiovasc Imaging       Date:  2019-01-30       Impact factor: 2.357

Review 6.  Highlights of the year in JACC 2012.

Authors:  Anthony N DeMaria; Jeroen J Bax; Gregory K Feld; Barry H Greenberg; Jennifer L Hall; Mark A Hlatky; Wilbur Y W Lew; João A C Lima; Ehtisham Mahmud; Alan S Maisel; Sanjiv M Narayan; Steven E Nissen; David J Sahn; Sotirios Tsimikas
Journal:  J Am Coll Cardiol       Date:  2013-01-22       Impact factor: 24.094

Review 7.  Paravalvular regurgitation following transcutaneous aortic valve replacement: predictors and clinical significance.

Authors:  Rebecca T Hahn; Susheel Kodali; Philippe Généreux; Martin Leon
Journal:  Curr Cardiol Rep       Date:  2014-05       Impact factor: 2.931

Review 8.  CT angiography after 20 years: a transformation in cardiovascular disease characterization continues to advance.

Authors:  Geoffrey D Rubin; Jonathon Leipsic; U Joseph Schoepf; Dominik Fleischmann; Sandy Napel
Journal:  Radiology       Date:  2014-06       Impact factor: 11.105

9.  The clinical anatomy and pathology of the human arterial valves: implications for repair or replacement.

Authors:  Michael G Bateman; Alexander J Hill; Jason L Quill; Paul A Iaizzo
Journal:  J Cardiovasc Transl Res       Date:  2013-01-17       Impact factor: 4.132

10.  Practical update on imaging and transcatheter aortic valve implantation.

Authors:  Gisela Feltes; Iván J Núñez-Gil
Journal:  World J Cardiol       Date:  2015-04-26
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.