Literature DB >> 29953507

Aortic root evaluation prior to transcatheter aortic valve implantation-Correlation of manual and semi-automatic measurements.

Barbora Horehledova1,2, Casper Mihl1,2, Chris Schwemmer3, Babs M F Hendriks1,2, Nienke G Eijsvoogel1,2, Bastiaan L J H Kietselaer1,2,4, Joachim E Wildberger1,2, Marco Das1,2.   

Abstract

BACKGROUND: Pre-procedural TAVI planning requires highly sophisticated and time-consuming manual measurements performed by experienced readers. Semi-automatic software may assist with partial automation of assessment of multiple parameters. The aim of this study was to evaluate differences between manual and semi-automatic measurements in terms of agreement and time.
METHODS: One hundred and twenty TAVI candidates referred for the retrospectively ECG-gated CTA (2nd and 3rd generation dual source CT) were evaluated. Fully manual and semi-automatic measurements of fourteen aortic root parameters were assessed in the 20% phase of the R-R interval. Reading time was compared using paired samples t-test. Inter-software agreement was calculated using the Intraclass correlation coefficient (ICC) in a 2-way mixed effects model. Differences between manual and semi-automatic measurements were evaluated using Bland-Altman analysis.
RESULTS: The time needed for evaluation using semi-automatic assessment (3 min 24 s ± 1 min 7 s) was significantly lower (p<0.001) compared to a fully manual approach (6 min 31 sec ± 1 min 1 sec). Excellent inter-software agreement was found (ICC = 0.93 ± 0.0; range:0.90-0.95). The same prosthesis size from manual and semi-automatic measurements was selected in 92% of cases, when sizing was based on annular area. Prosthesis sizing based on annular short diameter and perimeter agreed in 99% and 96% cases, respectively.
CONCLUSION: Use of semi-automatic software in pre-TAVI evaluation results in comparable results in respect of measurements and selected valve prosthesis size, while necessary reading time is significantly lower.

Entities:  

Mesh:

Year:  2018        PMID: 29953507      PMCID: PMC6023104          DOI: 10.1371/journal.pone.0199732

Source DB:  PubMed          Journal:  PLoS One        ISSN: 1932-6203            Impact factor:   3.240


Introduction

Transcatheter aortic valve implantation (TAVI) provides a minimally invasive therapeutical option to patients with severe aortic valve stenosis who are not eligible for conventional valve replacement [1-8]. Precise assessment of aortic root prior to the intervention is fundamental for patient’s outcome. It allows for recognition of suitable patients, selection of the correct valve for replacement, and therefore minimizes the risk of peri-procedural complications [9]. Need for comprehensive imaging with high spatial and temporal resolution, which is not impaired by dynamic movement of aortic root during cardiac cycle is the reason why multidetector row CT (MDCT) has been addressed as the method of choice in pre-TAVI evaluation [10-13]. MDCT assessment of aortic root was recently also considered a predictive factor for severity of paravalvular leakage, which further supports the role of MDCT in TAVI planning [11, 14]. However, tedious manual assessment of aortic root dimensions is highly sophisticated and therefore requires an experienced reader who is able to reliably extract necessary data from the scan. Semi-automatic software is designed to determine anatomical structures of aortic annulus on MDCT scan, allowing partial automation of measurements and in this respect simplifying the assessment, hereby making TAVI planning faster. Similar assisting software tools, with different levels of automation, have been previously described in the limited extent of four or five aortic root dimensions [15, 16], however, in our clinical practice fourteen aortic root parameters are routinely assessed in each TAVI candidate, following the expert consensus guidelines on MDCT imaging before TAVI [17, 18]. The aim of this study was to evaluate the agreement and time effectiveness of a prototype of a semi-automatic software with the standard manual assessment in aortic root evaluation prior to TAVI.

Materials and methods

Study population and data collection

Between April 2014 and April 2016, 120 patients with severe AS were retrospectively evaluated. All patients were referred from the cardiology outpatient department for pre-interventional assessment of aortic root dimensions and peripheral arteries. Patients with a history of valve replacement were excluded from the data analysis. All procedures performed in studies involving human participants were in accordance with the ethical standards of the institutional and/or national research committee and with the 1964 Helsinki declaration and its later amendments or comparable ethical standards. The approval for this study was obtained from the Institutional Review Board and the local medical ethical research committee (METC). Due to the retrospective nature of this study a waiver of written informed consent was issued by the Institutional Review Board. The data were coded and analyzed anonymously. The local METC (METC—Maastricht University Medical Center) reference number is 15-4-202.

Imaging protocol & post-processing

Retrospectively ECG-gated spiral MDCT of the aortic root was performed on a 2nd and 3rd generation Dual source CT scanner (Somatom Definition Flash & Somatom Force, Siemens Healthcare GmbH, Forchheim, Germany) following an established institutional protocol [19]. Images were reconstructed at every 10% of the R–R interval with individually adapted field of view (FOV) at 0.6 mm slice thickness with an increment of 0.4mm using iterative reconstruction (strength 3) and l26f kernel. Measurements were performed with dedicated post-processing software (Syngo.via™ Siemens Healthcare GmbH, Forchheim, Germany).

Objective CT image quality

CT image quality was quantified, in terms of attenuation in Hounsfield Units (HU), standard deviation (SD), signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR). Vascular attenuation was assessed with manually placed circular regions of interest (ROI) at the level of sinotubular junction. For noise and signal estimation another ROI was placed in the perivascular tissue in the left ventricular myocardium. CNR was calculated as vascular attenuation minus perivascular tissue attenuation, divided by the SD of the perivascular tissue attenuation. SNR was calculated as vascular attenuation divided by the SD of vascular attenuation [19]. Image quality was considered diagnostic with vascular attenuation values > 200 HU [20, 21] and CNR > 3 [22].

Subjective CT image quality

The subjective image quality, in terms of presence or absence of cardiac motion artifacts, was qualitatively evaluated by an experienced cardiovascular radiologist (CM) using previously published 4-point Likert scale [23]: grade 1 –non-diagnostic: impaired image quality that precluded appropriate evaluation of the aortic root due to severe motion artifacts; grade 2—diagnostic: reduced image quality due to motion artifacts, but sufficient to assess aortic root dimensions; grade 3—good: presence of motion artifacts, but fully preserved ability to reliably assess aortic annulus dimensions; grade 4—excellent: complete absence of motion artifacts. Cardiac motion artifacts were defined as beam-hardening, stair-stepping, blurring, ghosting, streaking, linear bands, areas of isolated or multiple discontinuity or dark shadows [24].

Aortic annulus and leaflet calcifications

The presence of aortic annulus calcification was qualitatively analyzed and graded in consensus by 2 observers (BH, CM) using a 5-point Likert scale: grade 0: absent annulus calcification; grade 1—minimal (< 25% of total annulus circumference); grade 2—mild (25–50% of total annulus circumference); grade 3—moderate (50–75% of total annulus circumference); grade 4—severe (75–100% of total annulus circumference) [25]. The aortic valve leaflet calcifications were graded as present or absent.

Time tracking

Case evaluation time was recorded within each assessment. A time tracking macro tool (programmed in Microsoft Excel, Microsoft corporation, Redmond, WA, USA) was built in the electronic worksheet. Therefore pre-TAVI measurements and reading time were recorded simultaneously to assure uniformity and accuracy of time measurements, regardless of method or reader performing the evaluation. The time tracking tool assessed the software loading time, describing the time before measurements could be carried out and time used for assessment of aortic root measurements as individual time points within evaluation. Mean necessary reading time was established for manual and semi-automatic approach. In manual assessment, the effective diameters derived from annular area and perimeter (eff. DA and eff. DP, respecitvely) had to be additionally calculated with dedicated formulas (Fig 1), as these values are not directly available during manual analysis. However, the time needed to calculate effective diameters is not significant and was therefore disregarded.
Fig 1

Formulas used for calculation of effective diameter.

eff. DA = effective diameter derived from aortic annulus area, eff. DP = effective diameter derived from aortic annulus perimeter.

Formulas used for calculation of effective diameter.

eff. DA = effective diameter derived from aortic annulus area, eff. DP = effective diameter derived from aortic annulus perimeter. Software loading time was tracked, but excluded from further evaluation because semi-automatic measurements were performed in preclinical interface.

Measurements

Pre-TAVI measurements were performed in fully manual and semi-automatic approach in each patient in the 20% phase of cardiac cycle, according to previous literature [26]. Results of earlier evaluation (manual or semi-automatic; inter-observer) were blinded during the subsequent data analysis with the remaining method. The following dimensions of the aortic root were assessed, in accordance to the expert consensus guidelines of the Society of Cardiovascular Computed Tomography (SCCT) [17]: short and long annulus diameter cross-sectional area of aortic annulus (defined as an area within an oval or ring formed by linking the most basal portions of the leaflet attachments) aortic annulus perimeter distance from the aortic annulus to the center of the ostium of the left and right coronary artery widest diameter at the sinotubular junction aortic root diameter at the level of the left and right coronary artery ostium widest portion of the coronary sinuses diameter length of left and right aortic valve leaflet eff. DP and eff. DA Agreement between manual and semi-automatic method and inter-reader agreement in semi-automatic approach were estimated in all measured parameters. In patients suitable for TAVI procedure the TAVI prosthesis size (Edwards Sapien Valve, Edwards Lifesciences Corp, Irvine, USA) was theoretically selected according to published recommendations [17, 27]. The agreement in selected prosthesis size derived from manual and semi-automatic method was evaluated.

Manual measurements

The manual assessment was performed with a dedicated software workflow (CT TAVI planning, Syngo.via™ VB10A Siemens). The 20% phase of the R-R interval was selected. Image data were manipulated in order to create an oblique transversal plane corresponding to the aortic annulus, crossing the most basal attachments of the aortic leaflets (Fig 2A), where aortic annulus diameters, area and perimeter were assessed. Eff. DA and eff. DP were calculated manually. An oblique transversal plane perpendicular to the course of the aorta was used to measure the dimensions of the widest portion of aortic root, the diameter at the level of sinotubular junction and at the level of coronary artery ostia. The oblique sagittal or coronal plane was used for the measurement of the distances from aortic annulus to the coronary artery ostia, and the length of the aortic leaflets (Fig 2B). Manual assessment was performed by a research fellow specially trained for this analysis (BH), with one year of experience in TAVI planning.
Fig 2

CT TAVI planning, Syngo.via™ VB10A Siemens, Siemens Healthcare GmbH, Forchheim Germany.

(A) The oblique transversal plane crossing the most basal attachments of the aortic valve leaflets. (B) The oblique sagittal (coronal) plane of the aortic valve showing leaflets.

CT TAVI planning, Syngo.via™ VB10A Siemens, Siemens Healthcare GmbH, Forchheim Germany.

(A) The oblique transversal plane crossing the most basal attachments of the aortic valve leaflets. (B) The oblique sagittal (coronal) plane of the aortic valve showing leaflets.

Semi-automatic measurements

Semi-automatic measurements were performed with a prototype version of syngo.CT Cardiac Planning software (Syngo.via™ VB20A, Siemens). The semi-automatic software automatically detects the three most basal attachment points of aortic valve leaflets, therefore opens the user interface directly in the annulus plane in the endsystolic phase. Measurements of short and long annulus diameter, annular area and perimeter are fully automatic. Likewise, eff. DA and eff. DP are calculated by the software and presented instantly. A centerline is computed through the aortic root and ascending aorta perpendicularly to the estimated aortic annulus plane. A dedicated tool (diameter ruler) displays automatically minimum and maximum diameters, perimeter, cross-sectional area and effective diameters at respective levels of the aortic root and ascending aorta while scrolling along the centerline. This allows for semi-automatic assessment of aortic root diameters (widest portion of aortic sinuses, diameter at sinotubular junction, diameter at the level of coronary ostia). Another dedicated tool assists only with image manipulation in order to visualize both left and right coronary ostia in one oblique sagittal plane (see Fig 3). Measurements of the distance from the aortic annulus to the coronary ostia as well as the length of coronary leaflets are then assessed manually.
Fig 3

Automatic aortic annulus plane of prototype version of syngo.CT Cardiac Planning.

Syngo.via™ VB20A, Siemens Healthcare GmbH, Forchheim Germany.

Automatic aortic annulus plane of prototype version of syngo.CT Cardiac Planning.

Syngo.via™ VB20A, Siemens Healthcare GmbH, Forchheim Germany. In addition, the reproducibility of semi-automatic measurements was evaluated. A research fellow specially trained in pre-TAVI evaluation with one year of experience (BH; reader 1) and a cardiovascular radiologist (CM; reader 2) performed the semi-automatic measurements independently and fully blinded to each other, in all 120 cases.

Statistical analysis

Dimensions of each anatomical structure and time needed for evaluations were evaluated using descriptive statistics (mean ± SD) calculated with independent samples t-test. Correlation and agreement between the manual and semi-automatic method were determined with the use of the paired samples t-test and Bland-Altman methods with 95% confidence interval (CI; mean difference ± 1.96 x SD). Inter-observer agreement was calculated using the Intraclass correlation coefficient (ICC) in a 2-way mixed effects model Bland-Altman methods with 95% CI. Statistical analysis was conducted using Statistical Package for Social Sciences version 23.0 (SPSS Inc., Chicago, IL, USA). All p-values are 2-sided, and a p-value < 0.05 was considered statistically significant.

Results

Baseline characteristics

Baseline characteristics are listed in Table 1. The study population consisted of 68 male and 52 female patients with an average age of 78 ± 8 years.
Table 1

Baseline characteristics.

Baseline characteristicsMeanRange
Age [years]78 ± 843–90
GenderMale / Female68 (57%) / 52 (43%)
Height [cm]167 ± 10135–193
Weight [kg]77 ± 1945–150
BMI [kg/m2]28 ± 616–46

cm = centimeter; kg = kilogram; m = meter

cm = centimeter; kg = kilogram; m = meter All examinations were of diagnostic quality, allowing aortic valve evaluation in all 120 included cases. The mean attenuation value at sinotubular junction was 426 ± 108 HU, with mean SNR 12 ± 6 and mean CRN 11 ± 5. Cardiac motion artifacts were completely absent (grade 4) in 101 images (84%), non-significant motion artifacts (grade 3) occurred in 19 scans (16%). No scans presented with non-diagnostic or diagnostic level of artifacts (grade 1 or 2). The annulus calcification was present in total of 118 cases (98%). In 105 patients (88%) was the extent of annular calcification graded as minimal and in 13 as mild (18%). In 5 cases (4%) the aortic annulus calcification extended caudally into the left ventricular outflow tract. The aortic valve leaflets were calcified in all assessed patients. All time tracking values in all established time points are stated in Table 2. Mean necessary reading time for manual measurements was 6 min 31 s ± 1 min 1 s. Mean necessary reading time for semi-automatic measurements were 3 min 24 s ± 1 min 7 s.
Table 2

Time tracking of manual and semi-automatic measurements.

Manual assessment:Mean ± SDp—value
 • Total Reading time7 min 36 s ± 1 min 7 s<0.001
 • Loading time1 min 5 s ± 0 min 23 s<0.001
 • Measurements time6 min 31 s ± 1 min 1 s<0.001
Semi-automatic assessment:
 • Total Reading time3 min 55 s ± 1min 19 s<0.001
 • Loading time0 min 31 s ± 0 min 21 s<0.001
 • Measurements time3 min 24 s ± 1 min 7 s<0.001

Total reading time = loading time + measurements time; Loading time = time before measurements could be carried; Measurement time = time for assessing evaluated measurements; SD = standard deviation

Total reading time = loading time + measurements time; Loading time = time before measurements could be carried; Measurement time = time for assessing evaluated measurements; SD = standard deviation

Manual and semi-automatic measurements

All manual and semi-automatic measurements are stated in the S1 Table. Excellent inter-software agreement was found between manual and semi-automatic method (ICC = 0.93 ± 0.02 for parameters with possible semi-automatic assessment and ICC = 0.83 ± 0.17 for all accessed dimensions; Table 3) [28]. The lowest ICC between manual and semi-automatic assessment was established for distance from aortic annulus to left and right coronary ostium (ICC = 0.46 and 0.51, respectively). The highest agreement was found in measurement of area, perimeter and effective diameters (ICC = 0.95). The mean difference between manual and semi-automatic assessment of short diameter, annular area and perimeter, eff. DA and eff. DP were 0.31 mm (95% CI: -2.18 to 2.81), 14.38 mm2 (95% CI: -58.37 to 87.13), 0.39 mm (95% CI: -5.70 to 6.49), 0.35 mm (95%: -1.48 to 2.18), and 0.12 mm (95% CI: -1.83 to 2.06), respectively (Fig 4A–4D).
Table 3

Intraclass correlation agreement (ICC) between manual and semi-automatic approach and inter-observer agreement in semi-automatic approach.

ICC manual & semi-automatic measurementsICC semi-automatic measurements
Short annulus diameter0.9110.967Automated measurements
Long annulus diameter0.9160.965
Annulus area0.9450.98
Eff. DA0.9470.984
Perimeter0.9500.984
Eff. DP0.9500.983
Widest portion of aortic root0.920.988Semi-automatic measurements
Diameter—Left ostium0.9020.977
Diameter—Right ostium0.8980.963
Diameter—sinotubular junction0.9180.970
Dist. annulus to the left ostium0.4640.918Manual measurements
Dist. annulus to the right ostium0.5130.883
Left leaflet length0.6960.652
Right leaflet length0.7340.649

dist. = distance; eff. DA = diameter derived from aortic annulus area; eff. DP = diameter derived from aortic annulus perimeter; eff. = effective; ICC = Intraclass correlation coefficient

Fig 4

Bland-Altman plots demonstrating agreement between fully manual and semi-automatic MDCT measurements of annular dimensions commonly used in prosthesis size selection shown with 95% confidence interval.

The middle line presents the mean difference (expressed in mm for aortic annulus diameters and perimeter, mm2 for aortic annulus area) and the upper and lower lines represent 95% confidence interval. (A) Short Axis Diameter; (B) Aortic Annulus Area; (C) Effective Diameter (Area); (D) Aortic Annulus Perimeter; (E) Effective Diameter (Perimeter).

dist. = distance; eff. DA = diameter derived from aortic annulus area; eff. DP = diameter derived from aortic annulus perimeter; eff. = effective; ICC = Intraclass correlation coefficient

Bland-Altman plots demonstrating agreement between fully manual and semi-automatic MDCT measurements of annular dimensions commonly used in prosthesis size selection shown with 95% confidence interval.

The middle line presents the mean difference (expressed in mm for aortic annulus diameters and perimeter, mm2 for aortic annulus area) and the upper and lower lines represent 95% confidence interval. (A) Short Axis Diameter; (B) Aortic Annulus Area; (C) Effective Diameter (Area); (D) Aortic Annulus Perimeter; (E) Effective Diameter (Perimeter). Excellent inter-observer correlation in semi-automatic software was found in all assessed dimensions (ICC = 0.92 ± 0.12). The lowest observed mean ICC was found for the length of the cusps (ICC = 0.65) and the highest was found for the widest portion of the coronary sinuses diameter (ICC = 0.99). The mean inter-observer difference was ≤ -0.1 mm for annular diameters, -0.2 mm (95% CI: -4.02 to 3.62,) for annular perimeter and—3.3 mm2 (95% CI: -54.59 to 47.99) for annular area (Fig 5).
Fig 5

Aortic annulus diameter measured manually and with semi-automatic software.

mm = millimeter.

Aortic annulus diameter measured manually and with semi-automatic software.

mm = millimeter.

TAVI prosthesis size selection

Ninety-five patients finally were eligible for TAVI procedure (Fig 6). Theoretical prosthesis sizing based on short annular diameter was possible in 94 cases, as appropriate prosthesis size is not available for one patient with annulus diameter of 29.1 mm. Prosthesis sizing derived from short annular diameter resulted in size agreement in 93 cases (99%). In one case the manual assessment suggested TAVI prosthesis one size larger compared to the semi-automatic assessment. The same prosthesis size was suggested in 86 cases (92%) when sizing was based on annular area and in 86 cases (96%) based on annular perimeter.
Fig 6

Theoretical TAVI prosthesis size selection (Edwards SapienValve).

n = number; TAVI = transcatheter aortic valve implantation. Theoretical prosthesis size selection according to recommendations for Edwards SapienValve (Edwards Lifesciences Corp, Irvine, USA).

Theoretical TAVI prosthesis size selection (Edwards SapienValve).

n = number; TAVI = transcatheter aortic valve implantation. Theoretical prosthesis size selection according to recommendations for Edwards SapienValve (Edwards Lifesciences Corp, Irvine, USA).

Discussion

This study demonstrates that the evaluated semi-automatic software provides reliable aortic root measurements with an excellent inter-observer agreement. The prototype of the semi-automatic software is able to automatically recognize aortic annulus dimensions with an excellent agreement to manual assessment (ICC = 0.91–0.95), while the semi-automatic software is significantly less time consuming. Precise measurements of aortic annulus dimensions are not only essential for selection of patients suitable for TAVI procedure, they are also crucial for particular selection of prosthesis size, which directly impacts patient’s outcome [9, 14]. Special attention was therefore paid to the annular dimensions suggested to be used in prosthesis size selection according to the industry guidelines. The mean difference between manual and semi-automatic measurements was ≤ 0.35 mm for annular diameters, 0.39 mm for annular perimeter and 14.38 mm2 for annular area with an excellent ICC of 0.91 or higher. Even though the difference was statistically significant for short aortic annulus, annular area and eff. DA, it should not be considered as clinically relevant according to previous literature [29]. This assumption is further supported with a 92–99% agreement between manual and semi-automatic assessment, when hypothetical valve sizing was applied in this study. Particularly, when agreement in suggested valve size of 59.4% was reported for manual measurements performed by different readers in study by Lou et al. in 2015 [15]. The moderate agreement in the annulus to coronary artery ostium distance (ICC = 0.46; 0.51) can be explained by the different data segmentation between methods. In a manual approach, measurements are carried out in a coronal plane, whereas the semi-automatic software arranges the aortic root along a calculated centerline, perpendicular to the aortic annulus plane, slightly affecting ostial position. However, comparable moderate agreement (ICC 0.48; inter-rater difference of up to 6.7 mm) was previously also reported for inter-rater agreement in manual assessment [30]. More importantly, the mean difference of 3.5 mm in annulus-ostium distance was previously considered acceptable in study by Watanabe et al. [16]. Nevertheless, the calculated centerline in semi-automatic software also offers a higher confidence while assessing the semi-manual measurements of aortic root diameters (ICC = 0.90–0.92). A dedicated tool “diameter ruler” works along the calculated centerline in the curved planar reformation (CPR) segment, directly presenting the minimum and maximum diameter at respective level of aortic root while scrolling. This feature allows the reader to distinguish the true maximum diameter faster and in more sophisticated way in comparison to manual assessment, where multiple manual measurements at different levels of aortic root have to be performed. Therefore, apart from the increased level of user convenience regarding measurements, semi-automatic software also demonstrated a greater efficiency in terms of time consumption. Evaluating pre-TAVI scans with semi-automatic software was approximately 3 minutes faster, allowing to process twice as many pre-TAVI datasets, while delivering comparable results. Manual assessment was slower mainly due to need for initial data manipulation, in order to recognize aortic annulus plane. Inter-observer agreement in this study was nearly perfect using the semi-automatic software, which offered reproducible results in terms of measurements, without necessarily reflecting the experience of the reader. Data segmentation, in matter of recognition or position of basal insertions of aortic valve cusps, may not satisfy the reader completely in some cases (e.g. due to poor image quality, etc.) [15], however, only in 4 out of the 120 cases in this study the reader decided to redefine the basal hinge points. A limitation of this study was that only two readers were able to perform semi-automatic measurements on the prototype version of software due to limited availability of workstations with the prototype software, and only one of these readers performed the manual measurements in all 120 cases. This study aimed to evaluate inter-software agreement and reproducibility rather than accuracy, because of the absence of true reference standard, such as direct anatomic measurement. Correlation of pre-TAVI measurements to post-interventional outcome was not possible in most cases, as our institution serves as a TAVI referral center and many patients undergo follow up in an external facilities.

Conclusion

The use of semi-automatic software in pre-TAVI evaluation results in comparable outcome in respect of measurements and selected valve prosthesis size in comparison to a manual approach, while the necessary reading time is significantly lower. Inter-observer agreement of semi-automatic measurements is excellent, proposing a possibility to standardize aortic annulus measurements, with a high confidence level regardless of reader experience.

Pre-TAVI evaluation assessment with manual and semi-automatic approach expressed as mean values ± SD and mean difference with 95% confidence interval from Bland-Altman analysis.

CI = confidence interval; man = manual; mm = millimeter; SA = semi-automatic method; SA1 = semi-automatic method performed by readered 1; SA2 = semi-automatic method performed by readered 2; sign. = significance. (DOCX) Click here for additional data file.
  29 in total

1.  Aortic annulus sizing: echocardiographic versus computed tomography derived measurements in comparison with direct surgical sizing.

Authors:  Jörg Kempfert; Arnaud Van Linden; Lukas Lehmkuhl; Ardawan J Rastan; David Holzhey; Johannes Blumenstein; Friedrich W Mohr; Thomas Walther
Journal:  Eur J Cardiothorac Surg       Date:  2012-03-07       Impact factor: 4.191

Review 2.  Cardiothoracic CT angiography: current contrast medium delivery strategies.

Authors:  Markus Weininger; J Michael Barraza; Corey A Kemper; John F Kalafut; Philip Costello; U Joseph Schoepf
Journal:  AJR Am J Roentgenol       Date:  2011-03       Impact factor: 3.959

3.  3-dimensional aortic annular assessment by multidetector computed tomography predicts moderate or severe paravalvular regurgitation after transcatheter aortic valve replacement: a multicenter retrospective analysis.

Authors:  Alexander B Willson; John G Webb; Troy M Labounty; Stephan Achenbach; Robert Moss; Miriam Wheeler; Christopher Thompson; James K Min; Ronen Gurvitch; Bjarne L Norgaard; Cameron J Hague; Stefan Toggweiler; Ronald Binder; Melanie Freeman; Rohan Poulter; Steen Poulsen; David A Wood; Jonathon Leipsic
Journal:  J Am Coll Cardiol       Date:  2012-02-22       Impact factor: 24.094

4.  Percutaneous transcatheter implantation of an aortic valve prosthesis for calcific aortic stenosis: first human case description.

Authors:  Alain Cribier; Helene Eltchaninoff; Assaf Bash; Nicolas Borenstein; Christophe Tron; Fabrice Bauer; Genevieve Derumeaux; Frederic Anselme; François Laborde; Martin B Leon
Journal:  Circulation       Date:  2002-12-10       Impact factor: 29.690

5.  Manual versus automatic detection of aortic annulus plane in a computed tomography scan for transcatheter aortic valve implantation screening.

Authors:  Arnaud Van Linden; Jörg Kempfert; Johannes Blumenstein; Helge Möllmann; Won-Keun Kim; Serap Alkaya; Christian Hamm; Thomas Walther
Journal:  Eur J Cardiothorac Surg       Date:  2014-01-14       Impact factor: 4.191

6.  Computed tomography-based evaluation of aortic annulus, prosthesis size and impact on early residual aortic regurgitation after transcatheter aortic valve implantation.

Authors:  Nicola Buzzatti; Francesco Maisano; Azeem Latib; Micaela Cioni; Maurizio Taramasso; Marco Mussardo; Antonio Colombo; Ottavio Alfieri
Journal:  Eur J Cardiothorac Surg       Date:  2012-05-02       Impact factor: 4.191

7.  Aortic annulus and ascending aorta: comparison of preoperative and periooperative measurement in patients with aortic stenosis.

Authors:  Michal Smíd; Jirí Ferda; Jan Baxa; Jakub Cech; Tomás Hájek; Boris Kreuzberg; Richard Rokyta
Journal:  Eur J Radiol       Date:  2009-02-23       Impact factor: 3.528

8.  Evaluation of aortic root and valve calcifications by multi-detector computed tomography.

Authors:  Andrew L Rivard; Thomas Bartel; Richard W Bianco; Kasia Simura O'Donnell; Johannes Bonatti; Wolfgang Dichtl; Ricardo C Cury; Gudrun M Feuchtner
Journal:  J Heart Valve Dis       Date:  2009-11

9.  Percutaneous aortic valve replacement for severe aortic stenosis in high-risk patients using the second- and current third-generation self-expanding CoreValve prosthesis: device success and 30-day clinical outcome.

Authors:  Eberhard Grube; Gerhard Schuler; Lutz Buellesfeld; Ulrich Gerckens; Axel Linke; Peter Wenaweser; Barthel Sauren; Friedrich-Wilhelm Mohr; Thomas Walther; Bernfried Zickmann; Stein Iversen; Thomas Felderhoff; Raymond Cartier; Raoul Bonan
Journal:  J Am Coll Cardiol       Date:  2007-06-06       Impact factor: 24.094

10.  MDCT evaluation of aortic root and aortic valve prior to TAVI. What is the optimal imaging time point in the cardiac cycle?

Authors:  Tomas Jurencak; Jakub Turek; Bastiaan L J H Kietselaer; Casper Mihl; Madeleine Kok; Vincent G V A van Ommen; Leen A F M van Garsse; Estelle C Nijssen; Joachim E Wildberger; Marco Das
Journal:  Eur Radiol       Date:  2015-02-24       Impact factor: 5.315

View more
  3 in total

Review 1.  Preoperative TAVR Planning: How to Do It.

Authors:  Rodrigo Petersen Saadi; Ana Paula Tagliari; Eduardo Keller Saadi; Marcelo Haertel Miglioranza; Carisi Anne Polanczyck
Journal:  J Clin Med       Date:  2022-05-05       Impact factor: 4.964

2.  Retrospectively ECG-gated helical vs. non-ECG-synchronized high-pitch CTA of the aortic root for TAVI planning.

Authors:  Barbora Horehledova; Casper Mihl; Ellen Boswijk; Genevieve A J C Crombag; Estelle C Nijssen; Patty J Nelemans; Leo F Veenstra; Joachim E Wildberger; Marco Das
Journal:  PLoS One       Date:  2020-05-12       Impact factor: 3.240

3.  Aortography Keypoint Tracking for Transcatheter Aortic Valve Implantation Based on Multi-Task Learning.

Authors:  Viacheslav V Danilov; Kirill Yu Klyshnikov; Olga M Gerget; Igor P Skirnevsky; Anton G Kutikhin; Aleksandr A Shilov; Vladimir I Ganyukov; Evgeny A Ovcharenko
Journal:  Front Cardiovasc Med       Date:  2021-07-19
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.