Literature DB >> 28317681

Comparison of biometry and intraocular lens power calculation performed by a new optical biometry device and a reference biometer.

Bruna V Ventura1, Marcelo C Ventura2, Li Wang2, Douglas D Koch2, Mitchell P Weikert2.   

Abstract

PURPOSE: To compare biometry measurements and intraocular lens (IOL) power calculations obtained by a new optical A-scan interferometer biometer (Galilei G6) (new biometer) and a partial coherence interferometer biometer (IOLMaster 500) (reference biometer).
SETTING: Private practice, Recife, Brazil.
DESIGN: Retrospective comparative study.
METHODS: Cataract patients who had biometric measurements with the new biometer and the reference biometer were included in the study. Comparisons were performed for axial length (AL), keratometry (K), anterior chamber depth (ACD), and IOL power calculation to reach emmetropia with an Acrysof SN60WF IOL using the Haigis formula. The Pearson correlation coefficient and the 95% limits of agreement (LoA) were calculated. The paired Student t test and the Wilcoxon test were used to assess differences between devices.
RESULTS: Eighty-eight eyes (88 patients) were studied. Both biometers provided statistically similar mean IOL power to reach emmetropia, AL, keratometry, and ACD measurements (P > .05). The differences in these variables did not vary as their mean values increased, and there was a strong positive correlation between the values obtained by both devices for each variable. The 95% LoA values for AL, mean keratometry, ACD, and IOL power were 0.27 mm, 1.08 diopters (D), 0.66 mm, and 1.56 D, respectively.
CONCLUSIONS: The 2 devices were comparable with regard to mean IOL power, mean AL, K, and ACD measurements. However, the wide range of differences between the devices suggests they should not be used interchangeably.
Copyright © 2016 ASCRS and ESCRS. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2017        PMID: 28317681     DOI: 10.1016/j.jcrs.2016.11.033

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  7 in total

1.  Comparison of Anterior Segment Parameters and Axial Length Using Two Scheimpflug Devices with Integrated Optical Biometers.

Authors:  Chantaka Supiyaphun; Sasivimol Rattanasiri; Passara Jongkhajornpong
Journal:  Clin Ophthalmol       Date:  2020-10-23

2.  The repeatability and agreement of biometric measurements by dual Scheimpflug device with integrated optical biometer.

Authors:  Hassan Hashemi; Sara Sardari; Abbasali Yekta; Mehdi Khabazkhoob
Journal:  Sci Rep       Date:  2022-05-11       Impact factor: 4.996

3.  Ocular biometry and refractive outcomes using two swept-source optical coherence tomography-based biometers with segmental or equivalent refractive indices.

Authors:  Miki Kamikawatoko Omoto; Hidemasa Torii; Sachiko Masui; Masahiko Ayaki; Kazuo Tsubota; Kazuno Negishi
Journal:  Sci Rep       Date:  2019-04-25       Impact factor: 4.379

4.  Intraocular Lens Calculation after Refractive Surgery: A Long-Term Retrospective Comparison of Eight Formulas.

Authors:  David B Rosen; Madeline B Heiland; Mitchell Tingey; Harry Y Liu; Paul Kang; Benjamin Buckner; Yasmyne C Ronquillo; Phillip C Hoopes; Majid Moshirfar
Journal:  Med Hypothesis Discov Innov Ophthalmol       Date:  2019

Review 5.  Cornea and anterior eye assessment with placido-disc keratoscopy, slit scanning evaluation topography and scheimpflug imaging tomography.

Authors:  Raul Martin
Journal:  Indian J Ophthalmol       Date:  2018-03       Impact factor: 1.848

6.  Repeatability, reproducibility, and comparison of ocular biometry using a new optical coherence tomography-based system and another device.

Authors:  Adam Wylęgała; Bartłomiej Bolek; Robert Mazur; Edward Wylęgała
Journal:  Sci Rep       Date:  2020-09-02       Impact factor: 4.379

7.  Comparison of Biometry Measurements Using Standard Partial Coherence Interferometry versus New Scheimpflug Tomography with Integrated Axial Length Capability.

Authors:  Jorge Selem Haddad; Eliza Barnwell; Karolinne Maia Rocha; Renato Ambrosio; George O Waring Iv
Journal:  Clin Ophthalmol       Date:  2020-02-04
  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.