Atsushi Ikeura1, Taketoshi Kushida1, Kenichi Oe1, Yoshihisa Kotani2, Muneharu Ando1, Takashi Adachi1, Takanori Saito1.
Abstract
STUDYEntities:
Keywords: Computed tomography; Lumbar vertebrae; Pedicle screws; Pig; Prosthesis failure
Year: 2020 PMID: 31906614 PMCID: PMC7280915 DOI: 10.31616/asj.2019.0170
Source DB: PubMed Journal: Asian Spine J ISSN: 1976-1902
Fig. 1.Photographs showing the screw types used in the present study. (A) 4.0-LEG (LEGACY; diameter, 4.0 mm; length, 30 mm), (B) 4.5-LEG (LEGACY; diameter, 4.5 mm; length, 30 mm), and (C) 4.5-SOL (SOLERA; diameter, 4.5 mm; length, 30 mm).
Fig. 2.Images of the porcine vertebrae. (A) Photograph of a 30-mmlong screw inserted through a 2-mm-thick washer. (B) Computed tomographic image showing the 10×30-mm rectangular region of interest (enclosed by the black lines); cortical margins were excluded to prevent volume averaging.
Fig. 3.Biomechanical testing. (A) The autograph used to determine pull-out strength (N) at a displacement rate of 1.0 mm/min. (B) A screw head attached to the testing machine. (C) The bone–screw complex fixed inside the metal box. (D) Typical load–displacement curve for screw axial pull-out.
Computed tomography values and pull-out strengths for each screw type
| Pedicle screw | CT valuea) (HU) | Pull-out strength[ |
|---|---|---|
| 4.0-LEG | 412.9±107.5 | 757.7±379.7 |
| 4.5-LEG | 470.1±82.0 | 1157.0±275.0[ |
| 4.5-SOL | 432.5±96.8 | 1085.9±532.8[ |
Values are presented as mean±standard deviation. 4.0-LEG: LEGACY; diameter, 4.0 mm; length, 30 mm; thread pitch, 2.75 mm; 4.5-LEG: LEGACY; diameter, 4.5 mm; length, 30 mm; thread pitch, 2.75 mm; 4.5-SOL: SOLERA; diameter, 4.5 mm; length, 30 mm; thread pitch, 2.0 mm in the cortical zone and 4 mm in the cancellous zone.
HU, Hounsfield units.
Groups were compared using one-way analysis of variance followed by the Tukey multiple comparison test.
p<0.05 for 4.0-LEG vs. 4.5-LEG and 4.5-SOL.
Fig. 4.Relationships between the CT values and pull-out strengths (N) of (A) 4.0-LEG (LEGACY; diameter, 4.0 mm; length, 30 mm), (B) 4.5-LEG (LEGACY; diameter, 4.5 mm; length, 30 mm), and (C) 4.5-SOL (SOLERA; diameter, 4.5 mm; length, 30 mm) screws. The CT values were significantly positively correlated with the pull-out strengths (N) of all pedicle screws (p<0.001). CT, computed tomography; HU, Hounsfield units.