Repeatability and reproducibility for intraocular pressure measurement by dynamic contour, ocular response analyzer, and goldmann applanation tonometry.

Michael Sullivan-Mee1, Gretchen Gerhardt, Kathy D Halverson, Clifford Qualls.   

Abstract

PURPOSE: To evaluate and compare the intraocular pressure measurement variability between Goldmann applanation tonometry (GAT), Pascal dynamic contour tonometry (DCT), and ocular response analyzer (ORA) tonometry.
METHODS: Subjects were prospectively recruited from consecutive Albuquerque VA Medical Center eye clinic patients that were previously diagnosed with ocular hypertension, glaucoma suspect, primary open-angle, or normal pressure glaucoma. Two sets of intraocular pressure measurements (3-4 ORA, 2 DCT, and 2 GAT) were obtained approximately 15 minutes apart. Each set was obtained by 1 of 2 examiners using random examiner sequences. ORA was measured first in both eyes, followed by either DCT or GAT, which were obtained in random order. Repeatability was assessed by examining variability of consecutive measurements by the same examiner, and reproducibility was examined by assessing variability between the first and second measurement sets.
RESULTS: One hundred and twenty eyes of 60 subjects were included. Mean age was 64.1+/-9.6 years and 58/60 (97%) were male. Intraobserver repeatability was highest for GAT, followed closely by DCT, and then ORA. Intersession reproducibility was similar for all methods, although a tonographic effect may have corrupted GAT and DCT reproducibility results. We found no repeatability or reproducibility differences between eyes, between examiners, or between measurement sets.
CONCLUSIONS: Although some intermethod variability differences were identified, all 3 methods in this study demonstrated clinically acceptable measurement repeatability and reproducibility. This result, in conjunction with the finding that variability was not different between eyes, examiners, or measurement sets, suggests that DCT and ORA are reliable enough to be clinically useful.

Entities:  

Mesh:

Year:  2009        PMID: 20010245     DOI: 10.1097/IJG.0b013e31819c487d

Source DB:  PubMed          Journal:  J Glaucoma        ISSN: 1057-0829            Impact factor:   2.503


  12 in total

1.  Corneal Deformation Response and Ocular Geometry: A Noninvasive Diagnostic Strategy in Marfan Syndrome.

Authors:  Lauren C Beene; Elias I Traboulsi; Ibrahim Seven; Matthew R Ford; Abhijit Sinha Roy; Robert S Butler; William J Dupps
Journal:  Am J Ophthalmol       Date:  2015-10-24       Impact factor: 5.258

2.  Reproducibility of ocular response analyzer measurements and their correlation with central corneal thickness.

Authors:  Joanna Wasielica-Poslednik; Fatmire Berisha; Shakhsanam Aliyeva; Norbert Pfeiffer; Esther M Hoffmann
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2010-08-10       Impact factor: 3.117

3.  Repeatability and reproducibility of Goldmann applanation, dynamic contour, and ocular response analyzer tonometry.

Authors:  Allen Shawlun Wang; Luciana M Alencar; Robert N Weinreb; Ali Tafreshi; Sunil Deokule; Gianmarco Vizzeri; Felipe A Medeiros
Journal:  J Glaucoma       Date:  2013-02       Impact factor: 2.503

4.  Efficacy and tolerability of the fixed combination of brinzolamide 1% and timolol 0.5% in daily practice.

Authors:  Ines Lanzl; Thomas Raber
Journal:  Clin Ophthalmol       Date:  2011-03-02

5.  Measuring accurate IOPs: Does correction factor help or hurt?

Authors:  Pinakin Gunvant; Robert D Newcomb; Elliot M Kirstein; Victor E Malinovsky; Richard J Madonna; Richard E Meetz
Journal:  Clin Ophthalmol       Date:  2010-07-21

6.  Comparison of Goldmann and Pascal tonometry in relation to corneal hysteresis and central corneal thickness in nonglaucomatous eyes.

Authors:  G Mangouritsas; S Mourtzoukos; A Mantzounis; L Alexopoulos
Journal:  Clin Ophthalmol       Date:  2011-08-03

7.  Detecting IOP Fluctuations in Glaucoma Patients.

Authors:  Brenda Nuyen; Kaweh Mansouri
Journal:  Open Ophthalmol J       Date:  2016-02-29

8.  Reliable intraocular pressure measurement using automated radio-wave telemetry.

Authors:  Eleftherios I Paschalis; Fabiano Cade; Samir Melki; Louis R Pasquale; Claes H Dohlman; Joseph B Ciolino
Journal:  Clin Ophthalmol       Date:  2014-01-07

Review 9.  Ageing and ocular surface immunity.

Authors:  Alireza Mashaghi; Jiaxu Hong; Sunil K Chauhan; Reza Dana
Journal:  Br J Ophthalmol       Date:  2016-07-04       Impact factor: 4.638

10.  Citius, Altius, Fortius: Agreement between Perkins and Dynamic Contour Tonometry (Pascal) and the Impact of Altitude.

Authors:  Oscar Albis-Donado; Shibal Bhartiya; Marina Gil-Reyes; Giovanna Casale-Vargas; Nancy Arreguin-Rebollar; Malik Y Kahook
Journal:  J Curr Glaucoma Pract       Date:  2018-03-01
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.