Anna H Brouwer1,2, Gerard C de Wit3, Joke H de Boer4, Maria M van Genderen3,4.
Abstract
PURPOSE: This study sought to investigate whether there is an optimal position of the Dawson, Trick, and Litzkow (DTL) electrodes when measuring the full-field electroretinogram (ERG) for monitoring purposes.Entities:
Keywords: Dawson, Trick, and Litzkow (DTL) electrodes; Electrophysiology; Electroretinogram (ERG)
Mesh:
Year: 2019 PMID: 31686287 PMCID: PMC7205847 DOI: 10.1007/s10633-019-09733-3
Source DB: PubMed Journal: Doc Ophthalmol ISSN: 0012-4486 Impact factor: 2.379
Fig. 1Example of the two DTL positions. Representative example of the two Dawson, Trick, and Litzkow (DTL) positions that were used: the lower lid position (LLP) (left) and the fornix position (FP) (right)
ERG results and differences of two DTL electrode positions
| Lower lid | Fornix | ||||
|---|---|---|---|---|---|
| Median | [IQR] | Median | [IQR] | ||
| a-wave amplitude* | |||||
| 0.3 | − 11.8 | [-15.5, − 8.1] | − 10.1 | [− 13.0, − 7.4] | < 0.001*** |
| 1.0 | − 19.4 | [− 25.1, − 14.3] | − 16.3 | [− 20.1, − 11.9] | < 0.001*** |
| 3.0 | − 29.7 | [− 37.8, − 22.2] | − 24.9 | [− 30.5, − 19.8] | < 0.001*** |
| 10.0 | − 46.1 | [− 57.2, − 35.4] | − 37.4 | [− 45.9, − 31.4] | < 0.001*** |
| a-wave implicit time* | |||||
| 0.3 | 19.0 | [18.2, 20.0] | 18.9 | [18.0, 20.1] | 0.445 |
| 1.0 | 17.1 | [16.4, 18.0] | 17.1 | [16.3, 17.8] | 0.115 |
| 3.0 | 15.7 | [15.1, 16.4] | 15.8 | [15.1, 16.5] | 0.816 |
| 10.0 | 14.8 | [14.3, 15.5] | 14.8 | [14.2, 15.5] | 0.119 |
| b-wave amplitude* | |||||
| 0.3 | 33.3 | [23.3, 42.0] | 26.2 | [20.4, 34.6] | < 0.001*** |
| 1.0 | 76.0 | [55.5, 97.6] | 63.1 | [49.9, 82.4] | < 0.001*** |
| 3.0 | 119.8 | [92.2, 151.7] | 103.2 | [81.7, 129.1] | < 0.001*** |
| 10.0 | 118.4 | [93.0, 144.6] | 105.0 | [82.6, 122.8] | < 0.001*** |
| b-wave implicit time* | |||||
| 0.3 | 27.0 | [25.8, 30.0] | 27.2 | [25.6, 29.9] | 0.718 |
| 1.0 | 28.9 | [27.8, 31.0] | 28.8 | [27.7, 30.8] | 0.105 |
| 3.0 | 31.6 | [30.7, 33.0] | 31.6 | [30.7, 33.0] | 0.129 |
| 10.0 | 35.9 | [34.8, 37.3] | 36.0 | [34.7, 37.2] | 0.055 |
| 30 Hz flicker peak | |||||
| Amplitude | 67.1 | [47.2, 86.3] | 58.0 | [43.9, 74.7] | < 0.001*** |
| Implicit time | 29.1 | [27.5, 31.3] | 29.1 | [27.6, 31.3] | 0.461 |
Descriptive statistics of the ERG responses of the two DTL positions (LLP and FP)
*Results of different flash strengths in candela. seconds/meters squared of the light-adapted ERG
***Statistically significant
ERG Electroretinogram, DTL Dawson, Trick, and Litzkow electrode, IQR interquartile range, LLP lower lid position, FP fornix position
Amplitude ratio and implicit time difference between the two DTL positions
| Amplitude ratio | Median | [IQR] | Percentiles | |
|---|---|---|---|---|
| 2.5th | 97.5th | |||
| a-wave | ||||
| 0.3* | 1.18 | [0.85, 1.56] | 0.35 | 3.73 |
| 1.0* | 1.18 | [0.93, 1.52] | 0.55 | 3.06 |
| 3.0 | 1.19 | [0.98, 1.43] | 0.55 | 2.34 |
| 10.0 | 1.19 | [1.00, 1.44] | 0.71 | 2.11 |
| b-wave | ||||
| 0.3* | 1.19 | [1.01, 1.39] | 0.73 | 2.23 |
| 1.0* | 1.13 | [1.00, 1.33] | 0.79 | 1.80 |
| 3.0 | 1.15 | [1.00, 1.33] | 0.78 | 1.75 |
| 10.0 | 1.12 | [1.00, 1.31] | 0.76 | 1.79 |
| 30 Hz flicker response | ||||
| 1.12 | [0.97, 1.31] | 0.73 | 1.80 | |
Results of the amplitude ratio (lower lid position/fornix position) and implicit time difference (lower lid position–fornix position) of the different flash strengths in candela. seconds/meters squared of the light-adapted ERG
*Candela·seconds/meters squared
Fig. 2Example of ERG curves of the two DTL positions. Representative example of ERG curves obtained from the same patient with the DTL at the lower lid position (black) and the fornix position (gray). Abbreviations: DTL Dawson, Trick, and Litzkow electrode, ERG electroretinogram
Fig. 3Amplitude ratio and implicit time differences of the two DTL positions. Scatter plots showing the implicit time differences (implicit time of LLP–LP) and amplitude ratio (LLP/FP) of the electroretinogram results all eyes for the light-adapted b-wave 3.0 a cds/m2. Uveitis eyes are indicated in as circles and unaffected eyes as squares. The large diamond indicates the median amplitude ratio (1.15) and mean implicit time difference (0.07). Abbreviations: DTL Dawson, Trick, and Litzkow electrode, cds/m2 cd · seconds/squared meters, LLP lower lid position, FP fornix position
Intraclass correlation coefficient between the two DTL positions
| ICC | CI | ||
|---|---|---|---|
| a-wave amplitude* | |||
| 0.3* | 0.584 | 0.494 | 0.658 |
| 1.0 | 0.654 | 0.578 | 0.715 |
| 3.0 | 0.678 | 0.608 | 0.736 |
| 10.0 | 0.716 | 0.655 | 0.767 |
| a-wave implicit time* | |||
| 0.3 | 0.788 | 0.742 | 0.826 |
| 1.0 | 0.830 | 0.793 | 0.860 |
| 3.0 | 0.856 | 0.825 | 0.882 |
| 10.0 | 0.857 | 0.826 | 0.883 |
| b-wave amplitude* | |||
| 0.3 | 0.865 | 0.836 | 0.889 |
| 1.0 | 0.900 | 0.878 | 0.918 |
| 3.0 | 0.885 | 0.860 | 0.905 |
| 10.0 | 0.842 | 0.807 | 0.870 |
| b-wave implicit time* | |||
| 0.3 | 0.946 | 0.934 | 0.955 |
| 1.0 | 0.973 | 0.967 | 0.978 |
| 3.0 | 0.954 | 0.944 | 0.962 |
| 10.0 | 0.982 | 0.978 | 0.985 |
| 30 Hz flicker Peak | |||
| Amplitude | 0.897 | 0.874 | 0.915 |
| Implicit time | 0.979 | 0.975 | 0.983 |
ICCs showing reliability between the ERG with two DTL positions: lower lid position and fornix position
*Results of different flash strengths in candela seconds/meters squared of the light-adapted ERG
ICC Intraclass correlation coefficient, ERG electroretinogram, DTL Dawson, Trick, and Litzkow electrode, CI confidence interval
Coefficients of variation of amplitudes of the two DTL positions
| Lower lid position | Fornix position | ||
|---|---|---|---|
| a-wave* | |||
| 0.3 | 24.2 | 21.8 | 0.054 |
| 1.0 | 19.6 | 19.3 | 0.904 |
| 3.0 | 17.7 | 15.7 | 0.014 |
| 10.0 | 15.5 | 14.5 | 0.202 |
| b-wave* | |||
| 0.3 | 21.2 | 19.7 | 0.187 |
| 1.0 | 20.5 | 19.3 | 0.267 |
| 3.0 | 36.0 | 33.7 | 0.239 |
| 10.0 | 32.7 | 30.8 | 0.259 |
| 30 Hz flicker* | |||
| 21.8 | 20.5 | 0.256 | |
CoV of amplitudes were calculated after data were transformed to fit a normal distribution of two DTL positions (lower lid position and fornix position). The R package cvequality (version 0.1.3; Marwick and Krishnamoorthy 2018) was used to test for significant differences in CoV of the two DTL positions. Based on Bonferroni’s correction (more tests than shown in this table), P values below < 0.006 were considered as statistically significant
*Results of different flash strengths in candela seconds/meters squared of the light-adapted ERG
ERG Electroretinogram, DTL Dawson, Trick, and Litzkow electrode, CI confidence interval, CoV coefficients of variation