PURPOSE: The aim of this retrospective cohort study was to evaluate intraocular pressure (IOP) changes during femtosecond laser-assisted cataract surgery (FLACS) using two different patient interface systems.Entities:
Year: 2019 PMID: 31662896 PMCID: PMC6778884 DOI: 10.1155/2019/5986895
Source DB: PubMed Journal: J Ophthalmol ISSN: 2090-004X Impact factor: 1.909
Demographic and clinical characteristics at the baseline.
| Patient demographics and preoperative clinical status | |||||
|---|---|---|---|---|---|
| Parameter | Group 1 | Group 2 |
| ||
| Mean ± SD | Range | Mean ± SD | Range | ||
| Age (y) | 72.4 ± 7.6 | 47, 89 | 71.5 ± 13.9 | 23, 89 | 0.69 |
| Male : female (n) | 33 : 28 | — | 27 : 28 | — | — |
| Left eye : right eye (n) | 37 : 24 | — | 32 : 23 | — | — |
| ACD (mm) | 3.06 ± 0.24 | 2.61, 3.6 | 3.17 ± 0.38 | 2.58, 4.06 | 0.42 |
| CCT ( | 560.9 ± 35.6 | 525, 630 | 555.2 ± 28.5 | 509, 634 | 0.38 |
| AL (mm) | 23.84 ± 1.49 | 21.56, 28.9 | 23.92 ± 1.36 | 21.93, 27.3 | 0.76 |
| Preop IOP (mmHg) | 14.1 ± 0.4 | 10, 22 | 13.8 ± 0.4 | 10, 20 | 0.15 |
ACD: anterior chamber depth; CCT: central cornea thickness; AL: axial length.
Mean IOP data at the different study time points of the two groups.
| Mean IOP data at the different study time points | |||||||
|---|---|---|---|---|---|---|---|
| Study time points | Group 1 | Group 2 | |||||
| Mean ± SD (mmHg) |
|
| Mean ± SD (mmHg) |
|
|
| |
| Before suction | 14.1 ± 0.4 | 1.0 | — | 13.8 ± 0.4 | 1.0 | — | 0.68 |
| After vacuum | 33.2 ± 1.1 | <0.001 | — | 24.2 ± 1.4 | <0.001 | — | <0.001 |
| After phaco | 21.4 ± 0.9 | <0.001 | <0.001 | 20.2 ± 1.2 | <0.001 | 0.67 | 0.395 |
| 1 day after | 14.0 ± 0.4 | 1.0 | <0.001 | 12.7 ± 0.3 | 1.0 | <0.001 | 0.01 |
| 7 days after | 13.9 ± 0.3 | 1.0 | <0.001 | 13.2 ± 0.3 | 1.0 | <0.001 | 0.09 |
Before suction: IOP measured before laser treatment; after vacuum: IOP measured just after vacuum turned off and suction ring removed on lying patients; after phaco: IOP measured after cataract surgery; 1 day after: IOP measured at 1 day after surgery; 7 days after: IOP measured at 7 days after surgery. p value1: statistical significance of IOP values compared to preoperative values; p value2: statistical significance of IOP values compared to values obtained after suction phase; p value3: statistical significance of the IOP difference between the two groups at each study time points. Group 1: liquid optics interface. Group 2: SoftFit™ Patient Interface.
Figure 1Intraocular pressure values measured at different time points during FLACS performed with two different devices: Catalys and LenSx. The box is determined by the central mean as well as the 25th and 75th percentiles. The whiskers represent the minimum value (0th percentile) and the maximum value (100th percentile).