Jing Wang1, Shengmei Zhao2, Zhenguo Su1, Xiangdong Liu2.
Abstract
BACKGROUND: To compare the Mindray BC-5180 and Sysmex XN-1000 instruments by analyzing the results of complete blood count in the external quality assessment in Shandong Province in 2018.Entities:
Keywords: Bland-Altman; Mindray BC-5180; Passing-Bablok; Sysmex XN-1000; complete blood count; consistency
Mesh:
Year: 2019 PMID: 31218736 PMCID: PMC6805265 DOI: 10.1002/jcla.22955
Source DB: PubMed Journal: J Clin Lab Anal ISSN: 0887-8013 Impact factor: 2.352
Statistics of various brands of instruments used in the laboratories
| Brand | Place of origin | Model | Number of laboratories |
|---|---|---|---|
| Mindray | China | BC‐5180 | 36 |
| BC‐5390 | 31 | ||
| BC‐6800 | 28 | ||
| BC‐5380 | 26 | ||
| BC‐6900 | 16 | ||
| BC‐5300 | 15 | ||
| BC‐3000 | 15 | ||
| BC‐2600 | 10 | ||
| BC‐5500 | 9 | ||
| BC‐5310 | 7 | ||
| BC‐5100 | 4 | ||
| BC‐1800 | 4 | ||
| BC‐5800 | 3 | ||
| BC‐300 | 3 | ||
| BC‐2800 | 3 | ||
| Sysmex | Japan | XN‐1000 | 36 |
| XE‐2100 | 32 | ||
| XS‐1000i | 32 | ||
| XS‐800i | 30 | ||
| XT‐1800i | 30 | ||
| XN‐2000 | 29 | ||
| KX‐21 | 21 | ||
| XN‐9000 | 20 | ||
| XT‐2000i | 19 | ||
| Others | 56 | ||
| NIHON KOHDEN | Japan | MEK‐8222K | 31 |
| MEK‐7222K | 30 | ||
| MEK‐6318K | 25 | ||
| HORIBA | Japan | ABX‐Pentra | 26 |
| ABX‐Micros | 25 | ||
| Others | 79 | ||
| Total | 731 |
Figure 1The results of two instruments, Mindray BC‐5180 and Sysmex XN‐1000, were tested by t test. The P values of the five test items of WBC, RBC, PLT, Hb, and HCT were all >0.05
A passing–Bablok regression analysis for Mindray BC‐5180和Sysmex XN‐1000 comparison
| Equation | 95% CI for intercept | 95% CI for slope |
| |
|---|---|---|---|---|
| WBC |
| −0.220 to 0.020 | 1.000 to 1.022 | 0.986 |
| RBC |
| −0.434 to 0.010 | 1.000 to 1.091 | 0.997 |
| PLT |
| 0.000 to 0.796 | 0.963 to 1.006 | 0.989 |
| Hb |
| 0.000 to 0.000 | 1.000 to 1.000 | 0.993 |
| HCT |
| −1.230 to 0.300 | 1.000 to 1.050 | 0.984 |
r: Pearson test was used for obtain the correlation coefficient.
Figure 2Comparison of test results of WBC, RBC, PLT, Hb, and HCT of Mindray BC‐5180 vs Sysmex XN‐1000 (A) Passing and Bablok regression analysis for WBC on 360 EQA results was y = −0.020 + 1.000x. (B) Passing and Bablok regression analysis for RBC on 360 EQA results was y = −0.352 + 1.074x. (C) Passing and Bablok regression analysis for PLT on 360 EQA results was y = 0.356 + 0.972x. (D) Passing and Bablok regression analysis for Hb on 360 EQA results was y = 0.000 + 1.000x. (E) Passing and Bablok regression analysis for HCT on 360 EQA results was y = −0.412 + 1.023x
Figure 3Bland‐Altman plots: comparing the WBC (A), RBC (B), PLT (C), Hb (D), and HCT (E) of Mindray BC‐5180 vs Sysmex XN‐1000. The dotted lines represent 95% limits of agreement