Comparison of five automated hematology analyzers in a university hospital setting: Abbott Cell-Dyn Sapphire, Beckman Coulter DxH 800, Siemens Advia 2120i, Sysmex XE-5000, and Sysmex XN-2000.

Mathias Bruegel, Dorothea Nagel, Manuela Funk, Petra Fuhrmann, Johannes Zander, Daniel Teupser.   

Abstract

BACKGROUND: Various types of automated hematology analyzers are used in clinical laboratories. Here, we performed a side-by-side comparison of five current top of the range routine hematology analyzers in the setting of a university hospital central laboratory.
METHODS: Complete blood counts (CBC), differentials, reticulocyte and nucleated red blood cell (NRBC) counts of 349 patient samples, randomly taken out of routine diagnostics, were analyzed with Cell-Dyn Sapphire (Abbott), DxH 800 (Beckman Coulter), Advia 2120i (Siemens), XE-5000 and XN-2000 (Sysmex). Inter-instrument comparison of CBCs including reticulocyte and NRBC counts and investigation of flagging quality in relation to microscopy were performed with the complete set of samples. Inter-instrument comparison of five-part differential was performed using samples without atypical cells in blood smear (n=292). Automated five-part differentials and NRBCs were additionally compared with microscopy.
RESULTS: The five analyzers showed a good concordance for basic blood count parameters. Correlations between instruments were less well for reticulocyte counts, NRBCs, and differentials. The poorest concordance for NRBCs with microscopy was observed for Advia 2120i (Kendall's τb=0.37). The highest flagging sensitivity for blasts was observed for XN-2000 (97% compared to 65%-76% for other analyzers), whereas overall specificity was comparable between different instruments.
CONCLUSIONS: To the best of our knowledge, this is the most comprehensive side-by-side comparison of five current top of the range routine hematology analyzers. Variable analyzer quality and parameter specific limitations must be considered in defining laboratory algorithms in clinical practice.

Entities:  

Mesh:

Year:  2015        PMID: 25720071     DOI: 10.1515/cclm-2014-0945

Source DB:  PubMed          Journal:  Clin Chem Lab Med        ISSN: 1434-6621            Impact factor:   3.694


  13 in total

1.  Reliable assessment of bone marrow and bone marrow concentrates using automated hematology analyzer.

Authors:  Venkata P Mantripragada; Nicolas S Piuzzi; Jaiben George; Wesley Bova; Mitchell Ng; Cynthia Boehm; George F Muschler
Journal:  Regen Med       Date:  2019-07-19       Impact factor: 3.806

2.  A comparison of Sysmex-XN 2000 and Yumizen H2500 automated hematology analyzers.

Authors:  Milena Małecka; Olga Ciepiela
Journal:  Pract Lab Med       Date:  2020-10-29

3.  Association of white blood cell count with breast cancer burden varies according to menopausal status, body mass index, and hormone receptor status: a case-control study.

Authors:  Byoungjin Park; Hye Sun Lee; Ji Won Lee; Seho Park
Journal:  Sci Rep       Date:  2019-04-08       Impact factor: 4.379

4.  Letter to the editor: Prognostic significance of preoperative and follow-up neutrophil-to-lymphocyte ratio and platelet-to-lymphocyte ratio in patients with non-metastatic clear cell renal cell carcinoma.

Authors:  Beuy Joob; Viroj Wiwaniitkit
Journal:  Investig Clin Urol       Date:  2019-06-25

5.  Analytical comparison between two hematological analyzer systems: Mindray BC-5180 vs Sysmex XN-1000.

Authors:  Jing Wang; Shengmei Zhao; Zhenguo Su; Xiangdong Liu
Journal:  J Clin Lab Anal       Date:  2019-06-20       Impact factor: 2.352

6.  Association of the neutrophil to lymphocyte ratio and white blood cell count with response to pharmacotherapy in unipolar psychotic depression: An exploratory analysis.

Authors:  Cornelis F Vos; Tom K Birkenhäger; Willem A Nolen; Walter W van den Broek; Marieke J H Coenen; Sophie E Ter Hark; Robbert-Jan Verkes; Joost G E Janzing
Journal:  Brain Behav Immun Health       Date:  2021-08-05

7.  Evaluation of automated hematology analyzer DYMIND DH76 compared to SYSMEX XN 1000 system.

Authors:  Milena Velizarova; Teodora Yacheva; Mariana Genova; Dobrin Svinarov
Journal:  J Med Biochem       Date:  2021-09-03       Impact factor: 3.402

8.  Three new inflammatory markers C reactive protein to albumin ratio, neutrophil to lymphocyte ratio, and platelet to lymphocyte ratio correlated with relapsing polychondritis disease activity index.

Authors:  Xiaoyu Cao; Mengzhu Zhao; Huijuan Li; Dong Xu; Mengtao Li; Xuan Zhang; Fengchun Zhang; Yong Hou; Xiaofeng Zeng
Journal:  Clin Rheumatol       Date:  2021-06-23       Impact factor: 2.980

9.  Neutrophil-to-Lymphocyte Ratio, Platelet-to-Lymphocyte Ratio and Heart Failure.

Authors:  Viroj Wiwanitkit
Journal:  Arq Bras Cardiol       Date:  2016-03       Impact factor: 2.000

10.  Validation of a DNA methylation reference panel for the estimation of nucleated cells types in cord blood.

Authors:  Andres Cardenas; Catherine Allard; Myriam Doyon; E Andres Houseman; Kelly M Bakulski; Patrice Perron; Luigi Bouchard; Marie-France Hivert
Journal:  Epigenetics       Date:  2016-09-26       Impact factor: 4.528

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.